K242603 · Medacta International S.A. · PBF · Nov 19, 2025 · Orthopedic
Device Facts
Record ID
K242603
Device Name
MyPAO SA guides
Applicant
Medacta International S.A.
Product Code
PBF · Orthopedic
Decision Date
Nov 19, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K242603 · Nov 19, 2025
MyPAO SA guides
Medacta International S.A.
Real-world clinical treatment data (patient cases treated by surgeons in Europe)
Real-world clinical data from patients treated in routine clinical practice in Europe (France and Switzerland) were used to validate the design and demonstrate the safety and effectiveness of the MyPAO SA guides by comparing performance metrics and intra-operative feedback against the predicate device.
Design Validation of MyPAO SA System; Retrospective/Observational analysis of real-world clinical treatment cases; Follow-up/Duration: Intra-operative and post-operative follow-up period
Patients requiring periacetabular osteotomy; Number of Sites: Multiple sites across Europe (France and Switzerland)
Predicate device (K220706)
Ability to replicate pre-operative planning (mean absolute errors of angles), intra-operative surgeon feedback, and occurrence of device-related complications/adverse events.
Indications for Use
MyPAO SA patient-specific guides are devices intended to be used as anatomical guides. The guides are specifically designed based on the CT images of the patient (i.e., pelvis, proximal femur, distal femur). The guides are intended to assist the surgeon in the positioning of the acetabular fragment during periacetabular osteotomies to treat patients who require to undergo periacetabular osteotomy. MyPAO SA guides are intended for single use only.
Device Story
Patient-specific 3D-printed surgical guides; derived from patient CT scans (pelvis, proximal/distal femur). System includes Preop SA guide for drilling/marking and Realignment SA Guide for intraoperative acetabular fragment positioning; accompanied by 3D-printed bone models. Used in clinical settings by surgeons during periacetabular osteotomy. Guides facilitate less invasive surgical access by preserving Rectus Femoris insertion on Anterior Inferior Iliac Spine. Surgeon uses guides to replicate preoperative plan; output is physical anatomical alignment. Benefits include improved surgical precision, reduced invasiveness, and verified correction of acetabular fragment position.
Clinical Evidence
Clinical design validation using real-world data from patients treated in Europe. Effectiveness assessed by comparing mean absolute errors in correction angles against predicate (K220706); results showed subject device mean absolute errors were within acceptability margins and lower than predicate. Intra-operative surgeon feedback indicated full satisfaction. Safety evaluated by monitoring complications/adverse events; zero device-related complications or adverse events reported intra-operatively or during post-operative follow-up.
Technological Characteristics
Patient-specific 3D-printed guides; non-sterile, single-use. Design based on patient CT imaging. Anchoring via two non-parallel pins. Materials, biocompatibility, and production process flow are identical to predicate K220706. No electronic components or software algorithms described; mechanical anatomical guidance system.
Indications for Use
Indicated for patients requiring periacetabular osteotomy (PAO) to assist surgeons in positioning the acetabular fragment.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 19, 2025
Medacta International S.A.
% Christopher Lussier
Senior Director, Quality, Regulatory and Clinical Research
Medacta USA
6386 Global Drive, Suite 101
Memphis, Tennessee 38141
Re: K242603
Trade/Device Name: MyPAO SA guides
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: PBF
Dated: March 12, 2025
Received: October 16, 2025
Dear Christopher Lussier:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242603 - Christopher Lussier
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K242603 - Christopher Lussier
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242603 | |
| Device Name MyPAO SA Guides | |
| Indications for Use (Describe) MyPAO SA patient-specific guides are devices intended to be used as anatomical guides. The guides are specifically designed based on the CT images of the patient (i.e., pelvis, proximal femur, distal femur). The guides are intended to assist the surgeon in the positioning of the acetabular fragment during periacetabular osteotomies to treat patients who require to undergo periacetabular osteotomy. MyPAO SA guides are intended for single use only. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
MyPAO SA guides
Traditional 510(k)
Medacta International SA
# 510(k) Summary
## I. Submitter
Medacta International SA
Strada Regina
6874 Castel San Pietro (CH)
Switzerland
Phone (+41) 91 696 60 60
Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA
Applicant Correspondent: Chris Lussier, Senior Director, Quality, Regulatory, and Clinical Research, Medacta USA
Date Prepared: 30 August, 2024
Date Revised: 13 November, 2025
## II. Device
| Device Proprietary Name: | MyPAO SA guides |
| --- | --- |
| Common or Usual Name: | Orthopaedic surgical planning and instrument guides |
| Classification Name: | Single/multiple component metallic bone fixation appliances and accessories. |
| Primary Product Code | PBF |
| Regulation Number: | 21 CFR 888.3030 |
| Device Classification | II |
## III. Predicate Device
Substantial equivalence is claimed to the following primary predicate device:
> MyPAO guides cleared within K220706
## IV. Device Description
The MyPAO SA guides are a set of patient specific 3D printed guides produced using patients' CT scans, intended to assist the surgeon in reaching the desired level of correction during periacetabular osteotomies.
The MyPAO SA guides represent a range extension as well as a simplified approach with respect to the already existing MyPAO guides cleared within K220706. The MyPAO SA guides, and their related preoperative and postoperative 3D printed bone models, are single-use surgical instruments provided non-sterile to the end-user.
K242603 - Page 1 of 4
{5}
MyPAO SA guides
Traditional 510(k)
Medacta International SA
The MyPAO SA Realignment System is composed of:
- One Preop SA guide used to mark the path of the planned line and to drill holes that facilitate placement and fixation of the Realignment SA Guide during the realignment phase.
- One Realignment SA Guide placed on the acetabular and iliac fragments after osteotomy, allowing for intraoperatively assessing the positioning of the acetabular bone fragment.
- 3D printed bone models of the preoperative and postoperative condition of the patient.
The MyPAO SA guides are available in two configurations, allowing to perform both right and left hip surgeries.
Given their design, the MyPao SA guides require a less invasive surgical access with respect to the predicates MyPao guides (K220706). On the other hand, the new MyPao SA guides are not indicated in case of reverse PAO corrections.
## V. Indications for Use
MyPAO SA patient-specific guides are devices intended to be used as anatomical guides. The guides are specifically designed based on the CT images of the patient (i.e., pelvis, proximal femur, distal femur). The guides are intended to assist the surgeon in the positioning of the acetabular fragment during periacetabular osteotomies to treat patients who require to undergo periacetabular osteotomy. MyPAO SA guides are intended for single use only.
## VI. Comparison of Technological Characteristics
The MyPAO SA guides are identical to the MyPAO guides cleared within K220706 in terms of intended use and general technological characteristics. In particular, subject and predicate device share the following:
- functional features;
- anchoring points;
- production process flow;
- material;
- biocompatibility;
- device usage;
- packaging;
- shelf-life;
- sterilization (non-sterile);
The only differences between subject and predicate devices (K220706) are related to slight design changes implemented on both pre and post-osteotomy guides, in order to allow a less invasive surgical approach, preserving the insertion of the Rectus Femoris on the Anterior Inferior Iliac Spine (AIIS) during Periacetabular Osteotomy surgeries.
K242603 - Page 2 of 4
{6}
MyPAO SA guides
Traditional 510(k)
Medacta International SA
## Discussion
These changes do not pose any additional risks in terms of safety and performance as the unique positioning of the MyPAO SA guides on the bone is still guaranteed, since the guides are designed according to patient’s bone anatomy. Also, being the Post-op Realignment System a single component, the stiffness of the construct is not affected by external factors such as connecting elements, while its stability is still ensured by fixation through the two pins inserted into the bone with non-parallel directions.
## VII. Performance Data
Based on the risk analysis, testing activities were conducted to written protocols.
## Non-Clinical Studies
- Substantial Equivalence Assessment between the subject and predicate devices to demonstrate that no new worst cases are introduced in terms of technical characteristics, thus pre-clinical and clinical evaluations performed on the predicate devices can be leveraged for the subject devices;
- Guides positioning variability assessment demonstrating the substantial equivalence between the subject and predicate devices in terms of clinical, technological and biological characteristics as well as usability of the system;
- Designers variability assessment demonstrating minimal inter- and intra-designer variability.
## Clinical Studies:
- Design Validation by using real-world clinical data collected for the subject devices and compared to the ones of the predicate devices has been used.
In detail, patients representative of the intended user population were treated according to real clinical treatment by operating surgeons across Europe (France and Switzerland).
### Primary Effectiveness Endpoint
The subject devices’ effectiveness (ability to replicate pre-operative planning) was assessed by demonstrating substantial equivalent performance of the subject devices with respect to the predicate devices (K220706) and by collecting intra-operative surgeons’ feedback.
### Effectiveness Results
- Performance Equivalence Assessment: Different angles of interest have been measured. The mean absolute errors were within the respective predefined acceptability margin and lower for the MyPAO SA System compared to the predicate device. Variability was comparable between the two systems.
- Intra-operative surgeon feedback: Formative evaluation feedback from surgeons consistently expressed full intra-operative satisfaction with the correction achieved using the system.
### Primary Safety Endpoint
The safety of the subject system was evaluated by assessing the occurrence of complications or adverse events related to device use, both during surgery and throughout the patient’s post-
K242603 - Page 3 of 4
{7}
MyPAO SA guides
Traditional 510(k)
Medacta International SA
operative follow-up period. The analysis was deemed successful if no device-related complications or adverse events were reported.
**Safety Results**
No device-related complications or adverse events reported intra-operatively nor during the post-operative follow-up period.
**Summary**
The clinical performance demonstrated that the subject MyPAO SA System is a safe and effective device for its intended clinical application.
**VIII. Conclusion**
The information provided above supports that the subject device is substantially equivalent to the predicate device.
K242603 - Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.