K220706 · Medacta International S.A. · PBF · May 30, 2023 · Orthopedic
Device Facts
Record ID
K220706
Device Name
MyPAO Guides
Applicant
Medacta International S.A.
Product Code
PBF · Orthopedic
Decision Date
May 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Indications for Use
MyPAO patient-specific guides are devices intended to be used as anatomical guides are specifically designed based on the CT images of the patient (i.e., pelvis, proximal femur). The guides are intended to assist the surgeon in the positioning of the acetabular fragment during periacetabular osteotomies to treat patients who require to undergo periacetabular osteotomy. MyPAO guides are intended for single use only.
Device Story
MyPAO Guides are patient-specific, single-use anatomical guides designed from preoperative CT scans; used to assist surgeons in positioning the acetabular fragment during periacetabular osteotomies. System includes repositioning PSI guides (pre-op drill guide, ilium/acetabulum guides, realignment arc), preoperative/postoperative bone models, and reusable instruments. Surgeon uses MyPAO software to perform 3D preoperative planning, now including 3D femoral head coverage and acetabular anteversion parameters. Guides replicate desired correction before final fixation of acetabular fragment. Device used in clinical setting; aids surgeon in achieving planned realignment; benefits patient through improved surgical accuracy in complex pelvic osteotomies.
Clinical Evidence
Single-arm study (n=10; 8 female, 2 male; ages 18-53) conducted outside the U.S. compared preoperative planned correction to postoperative results using 2D planar x-rays. Primary endpoints: lateral center edge angle and acetabular index. Acceptance criteria: mean error ≤ 4 degrees. Results met criteria; no adverse effects or complications observed.
Technological Characteristics
Patient-specific, single-use, non-sterile guides and bone models. Designed via CT-based 3D planning. Software-driven design. Biocompatibility leveraged from K200792. Reusable instruments used in conjunction with guides.
Indications for Use
Indicated for patients requiring periacetabular osteotomy. Guides assist surgeons in positioning the acetabular fragment during the procedure.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 30, 2023
Medacta International S.A. % Chris Lussier Senior Director, Quality, Regulatory and Clinical Research Medacta USA 3973 Delp Street Memphis, Tennessee 38118
Re: K220706
Trade/Device Name: MyPAO Guides Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: PBF Dated: March 9, 2022 Received: March 10, 2022
Dear Chris Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Limin Sun -S
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220706
Device Name MyPAO Guides
### Indications for Use (Describe)
MyPAO patient-specific guides are devices intended to be used as anatomical guides are specifically designed based on the CT images of the patient (i.e., pelvis, proximal femur). The guides are intended to assist the surgeon in the positioning of the acetabular fragment during periacetabular osteotomies to treat patients who require to undergo periacetabular osteotomy. MyPAO guides are intended for single use only.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Sr. Director of Quality, Regulatory, and Clinical, Medacta USA Date Prepared: March 09, 2022 Date Revised: May 30, 2023
#### II. Device
| Device Proprietary Name: | MyPAO Guides |
|--------------------------|--------------------------------------------------------------------------------|
| Common or Usual Name: | Orthopaedic Surgical Planning And Instrument Guides |
| Classification Name: | Single/multiple component metallic bone fixation appliances<br>and accessories |
| Primary Product Code: | PBF |
| Regulation Number: | 21 CFR 888.3030 |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following primary predicate devices:
- Primary predicate device:
- >> VSP Orthopedics System, K190044, 3D Systems, Inc.
Reference device :
- A MySpine MySpine S2-Alar/Alar-Iliac Pedicle Screw Placement Guides, K200792, Medacta International SA
- A MyPAO Planning Report, K200589, Medacta International SA
#### Device Description IV.
The MyPAO Guides enlarge Medacta's MyPAO realignment system, already including MyPAO Planning report, in order to provide the surgeons a further help in the realignment phase after a Periacetabular Osteotomy.
Specifically, the current submission aims at obtaining the clearance for the changes made to MyPAO Planning report and for the following new devices:
{4}------------------------------------------------
- Reposition PSI guides (Pre-op drill guide, Ilium and Acetabulum Rep Guides, Realignment ● Arc):
- Preoperative and postoperative bone models; ●
- Reusable instruments to be used with the guides. ●
MyPAO Guides are patient specific guides designed on the basis of the patient's preoperative CT scan and subsequent preoperative planning and they aim at replicating the desired correction before the final fixation of the acetabular fragment.
The MyPAO Guides as well as their bone models are single-use and they are provided non-sterile.
The 3D pre-operative planning based on the patient's CT scans is performed through MyPAO software that is the same software already used to generate MyPAO Planning Report. The only change made to the already cleared MyPAO Planning report (K200589) regards the addition of two parameters in the report, i.e. 3D femoral head coverage and acetabular anteversion.
#### V. Indications for Use
MyPAO patient-specific guides are devices intended to be used as anatomical guides. The guides are specifically designed based on the CT images of the patient (i.e., pelvis, proximal femur). The guides are intended to assist the surgeon in the positioning of the acetabular fragment during periacetabular osteotomies to treat patients who require to undergo periacetabular osteotomy. MyPAO guides are intended for single use only.
The indications for use differ because the predicate for use in non-joint replacing osteotomies in the non-sacrum pelvis area, whereas the subject device has specific indications for use in periacetabular osteotomy (a type of non-joint replacement, non-sacrum pelvis indication). This specificity does not change the intended use of the instruments for guided orthopedic corrections of the pelvis.
#### VI. Comparison of Technological Characteristics
The subject and primary predicate device (K190044) indications for use are substantially equivalent: both are intended to be used as a surgical instrument to assist the surgeon during non-joint replacing osteotomies.
Both the subject and the predicate device include single use patient matched guides. The subject device differs from the predicate because it allows orientation of one fragment to the remaining pelvic bone, whereas the predicate allows for orientation of instruments to bone.
{5}------------------------------------------------
## Discussion
The subject and primary predicate device (K190044) indications for use are substantially equivalent: both are intended to be used as a surgical instrument to assist the surgeon during non-joint replacing osteotomies.
The comparison of technological characteristics and performance data provided within this submission, shows that there are no new risks associated with the subject devices design, and supports the substantial equivalence of the MyPAO Guides to the identified predicate devices.
The only change made to the already cleared MyPAO Planning report (K200589) has not any impact on device's safety and effectiveness since it simply regards the addition of two parameters in the report, i.e. 3D femoral head coverage and acetabular anteversion.
#### VII. Performance Data
Based on the risk analysis, testing activities were conducted to written protocols. The following validation and tests are being provided in support of the substantial equivalence determination:
## Non-Clinical Studies
- Software validation ●
- MyPAO Design validation to validate functionality and usability ●
- MyPAO Design validation to demonstrate the ability of MyPAO guides to replicate ● preoperative planning
- MyPAO guides positioning variability Cadaveric test validation ●
Biocompatibility data submitted in support of the reference device (K200792) was leveraged.
## Clinical Studies:
- 10 patients were treated by three surgeons using the MyPAO surgical planning software and . guides. The primary safety and effectiveness data included a surgeon formative evaluation and comparison of the preoperative planned correction as compared to the post-operative correction results. 2D planar x-rays were used to measure the final lateral center edge and acetabular index.
- This was a single-arm study with a literature control. The study was conducted outside O of the United States.
- The acceptance criteria was a mean error of 4 degrees or less for the lateral center edge o angle and acetabular index. This criteria was identified based on literature for conventional surgical technique accuracy.
- The clinical data set met the acceptance criteria. o
- The 10 patients included: ●
- 0 8 female
- 2 male o
- Ages: from 18 to 53 o
{6}------------------------------------------------
- . There were no observed adverse effects or complications.
## VIII. Conclusion
The information provided above supports that the MyPAO Guides are substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.