K242356 · Restor3D, Inc. · SAI · Mar 24, 2025 · Orthopedic
Device Facts
Record ID
K242356
Device Name
TIDAL Fusion Cage System
Applicant
Restor3D, Inc.
Product Code
SAI · Orthopedic
Decision Date
Mar 24, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K242356 · Mar 24, 2025
TIDAL Fusion Cage System
Restor3D, Inc.
Retrospective clinical evaluation of patients treated with the TIDAL Fusion Cage
A retrospective clinical evaluation was conducted to assess the safety and effectiveness of the TIDAL Fusion Cage System when used as an accessory to an intramedullary nail for tibiotalocalcaneal fusion. Endpoints included joint salvage, physical function, deformity correction, and absence of secondary surgical interventions or device-related serious adverse events.
Retrospective clinical evaluation of the TIDAL Fusion Cage; Retrospective, single arm clinical evaluation; Follow-up/Duration: Last follow-up
Patients undergoing tibiotalocalcaneal (TTC) fusion using an intramedullary nail and the TIDAL Fusion Cage
Not applicable for this study
Joint salvage, perceived physical function, deformity correction, absence of secondary surgical intervention, and absence of device-related serious adverse events
Indications for Use
The TIDAL Fusion Cage System is intended to be used as part of a tibiotalocalcaneal fusion construct in a salvage procedure following failed ankle arthrodesis or failed ankle arthroplasty for patients at risk of limb loss. The TIDAL Fusion Cage System is intended for use as an accessory to the DynaNail TTC Fusion System. The TIDAL Fusion Cage System is not intended for standalone use. The TIDAL Fusion Cage System is intended for use with autograft and/or allogenic bone graft.
Device Story
TIDAL Fusion Cage System is a single-use, titanium alloy (Ti-6Al-4V) implant designed as an accessory to the DynaNail TTC Fusion System for tibiotalocalcaneal fusion. The device features a central clearance hole to accommodate the intramedullary nail and a porous structure to facilitate packing of autograft or allogenic bone graft. It is used in salvage procedures for patients with failed ankle arthrodesis or arthroplasty who are at risk of limb loss. The device is implanted by a surgeon during a fusion procedure. The cage provides structural support and promotes fusion within the tibiotalocalcaneal construct, aiming to salvage the limb and improve patient function.
Clinical Evidence
Retrospective, single-arm clinical evaluation of the TIDAL Fusion Cage used with an intramedullary nail for TTC fusion. Effectiveness endpoints: joint salvage, physical function, and deformity correction. Safety endpoints: absence of secondary surgical intervention and device-related serious adverse events.
Technological Characteristics
Material: Titanium alloy (Ti-6Al-4V). Manufacturing: Additive manufacturing. Design: Porous structure with central clearance hole for intramedullary nail. Form factor: Varying shapes, heights, and diameters. Single-use implant.
Indications for Use
Indicated for patients at risk of limb loss requiring tibiotalocalcaneal fusion as a salvage procedure following failed ankle arthrodesis or failed ankle arthroplasty. Must be used as an accessory to the DynaNail TTC Fusion System with autograft and/or allogenic bone graft. Not for standalone use.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 24, 2025
restor3d, Inc. Brianna Prindle Director of Regulatory Affairs 4001 NC-54 Hwy, Suite 3160 Durham, North Carolina 27709
Re: K242356
Trade/Device Name: TIDAL Fusion Cage System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: SAI, HSB Dated: August 8, 2024 Received: February 20, 2025
Dear Brianna Prindle:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows the name "Farzana Sharmin -S" on the left side of the image. On the right side of the image, it says "Digitally signed by Farzana Sharmin -S" followed by the date "2025.03.24 18:58:58 -04'00'". The text on the right side of the image appears to be a digital signature.
Farzana Sharmin, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242356
Device Name TIDAL Fusion Cage System
#### Indications for Use (Describe)
The TIDAL Fusion Cage System is intended to be used as part of a tibiotalocalcaneal fusion construct in a salvage procedure following failed ankle arthrodesis or failed ankle arthroplasty for patients at risk of limb loss. The TIDAL Fusion Cage System is intended for use as an accessory to the DynaNail TTC Fusion System. The TIDAL Fusion Cage System is not intended for standalone use. The TIDAL Fusion Cage System is intended for use with autograft and/or allogenic bone graft.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "restor3d" in a bold, sans-serif font. The first six letters, "restor" are solid black, while the "3d" is outlined in white. The letters are closely spaced, giving the word a compact appearance. The overall design is simple and modern.
4001 NC-54 Hwy Suite 3160 Durham, NC 27709 984.888.0593
# 510(k) Summary
## Date Prepared: March 24, 2025
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
## A.510(k) Owner:
restor3d, inc. 4001 NC-54 Hwy, Suite 3160 Durham, NC 27709
### B. Primary Correspondent:
Brianna Prindle Director of Regulatory Affairs brianna@restor3d.com 786-521-0501
### C. Premarket Notification:
| Submission Type: | Traditional 510(k) |
|----------------------|---------------------------------------------------------------------------------|
| Trade Name: | TIDAL Fusion Cage System |
| Classification Name: | Single/multiple components metallic bone fixation<br>appliances and accessories |
| Common Name: | Intramedullary fixation rod |
| Regulation Number: | 21 CFR 888.3020 |
| Product Code: | SAI, HSB |
| Classification: | II |
### D. Indications for Use:
The TIDAL Fusion Cage System is intended to be used as part of a tibiotalocalcaneal fusion construct in a salvage procedure following failed ankle arthrodesis or failed ankle arthroplasty for patients at risk of limb loss. The TIDAL Fusion Cage System is intended for use as an accessory to the DynaNail TTC Fusion System. The TIDAL Fusion Cage System is not intended for standalone use. The TIDAL Fusion Cage System is intended for use with autograft and/or allogenic bone graft.
# E. Predicate Devices:
The TIDAL Fusion Cage System is substantially equivalent to the following devices:
| 510(k) | Trade Name | Manufacturer |
|----------------------------|---------------------------------------|--------------|
| Primary Predicate Device | | |
| K171376 | DynaNail TTC Fusion System | MedShape |
| Secondary Predicate Device | | |
| K230088 | 4WEB Medical Ankle Truss System (ATS) | 4WEB |
| Reference Predicate Device | | |
| K201314 | restor3d Utility Wedge | restor3d |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "restor3d" in a bold, sans-serif font. The first six letters, "restor" are solid black, while the "3d" is outlined in black. The word is horizontally oriented and centered.
4001 NC-54 Hwy Suite 3160 Durham, NC 27709 984.888.0593
# F. Device Description:
The subject TIDAL Fusion Cage System is a single-use, implant grade titanium alloy (Ti-6Al-4V) device, available in varying shapes, heights and diameters, designed to be used as an accessory to the DynaNail TTC System for ankle fusion.
# G. Comparison of Technology
The subject TIDAL Fusion Cage System and predicate DynaNail TTC Fusion System have the same intended use for tibiotalocalcaneal (TTC) fusion. The indications for use of the TIDAL Fusion Cage System are a subset of the indications for use for the predicate DynaNail TTC Fusion System (K171376). The subject TIDAL Fusion Cage System is intended to be an accessory to the primary predicate DynaNail TTC Fusion System. The technological characteristics of the TIDAL Fusion Cage System differ from the predicate device, specifically, the material and manufacturing methods. The intended use, technological characteristics, materials and manufacturing of the subject TIDAL Fusion Cage System are identical to the secondary predicate device, the 4WEB Medical Ankle Truss System (ATS) (K230088). Both the subject device and reference device are made of titanium alloy (Ti-6Al-4V) and are manufactured using an additive manufacturing method.
# H. Comparison of Non-Clinical Performance
The TIDAL Fusion Cage System is substantially equivalent to the predicate device (DynaNail TTC Fusion System, K171376) in intended use, indications for use and performance specifications. The subject TIDAL Fusion Cage System is used as an accessory to the DynaNail TTC Fusion System, with a central clearance hole to accommodate the nail and porosity for packing of graft material. The subject TIDAL Fusion Cage System was subject to the following benchtop and clinical performance tests to support the assertion of substantial equivalence:
- Static Compressive Strength .
- Dynamic Compressive Strength
- Static and Dynamic Compression-Shear Strength ●
- Torsional Strength ●
- Dynamic Construct Fatigue with Subsequent Fretting-Corrosion Analysis ●
- . Nitinol Corrosion Test
- Wear Particle Analysis
# I. Clinical Performance Evaluation
restor3d completed a retrospective, single arm, clinical evaluation of the TIDAL Fusion Cage using an intramedullary (IM) nail for tibiotalocalcaneal (TTC) fusion. Clinical performance was evaluated against an IM nail for TTC fusion. Effectiveness endpoints included joint salvage with the TIDAL Fusion Cage in place at last follow-up, perceived, physical function, and deformity correction. Safety outcomes included absence of a subsequent secondary surgical intervention and device-related serious adverse events on the affected joints.
# J. Conclusions:
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "restor3d" in a bold, sans-serif font. The letters "restor" are filled in with black, while the "3d" is outlined in black. The word is horizontally oriented and centered in the image. The background is white.
4001 NC-54 Hwy Suite 3160 Durham, NC 27709 984.888.0593
Based on the data submitted, any differences of the TIDAL Fusion Cage System do not raise any different questions of safety or effectiveness, therefore, demonstrates substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.