K230088 · 4Web Medical, Inc. · SAI · Mar 21, 2024 · Orthopedic
Device Facts
Record ID
K230088
Device Name
Ankle Truss System (ATS)
Applicant
4Web Medical, Inc.
Product Code
SAI · Orthopedic
Decision Date
Mar 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K230088 · Mar 21, 2024
Ankle Truss System (ATS)
4Web Medical, Inc.
Retrospective multi-center clinical evaluation
A retrospective, multi-center study was conducted to evaluate the clinical performance (fusion, pain, functional outcomes) and safety (adverse events, secondary surgical interventions) of the ATS device compared to standard IM nail treatments for TTC fusion in patients with failed ankle arthrodesis or arthroplasty.
Retrospective, single arm, multi-center evaluation
Patients undergoing TTC fusion as a salvage procedure following failed ankle arthrodesis or failed ankle arthroplasty; Number of Sites: Multi-center
IM nail for TTC fusion
Fusion, pain, functional outcomes, device and procedure related adverse events, and rates of subsequent secondary surgical intervention (SSSI)
Indications for Use
The 4WEB Medical Ankle Truss System (ATS) is for use as an accessory to the Stryker T2 Ankle Arthrodesis Nail or the Stryker Valor Hindfoot Fusion Nail as part of a tibiotalocalcaneal fusion construct in a salvage procedure following failed ankle arthrodesis or failed ankle arthroplasty for patients at risk for loss of limb. The ATS is not intended for standalone use. The anatomical landmarks necessary for the design and creation of ATS devices that are patient matched must be present and identifiable on appropriate radiography scans. The ATS is intended for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
Device Story
Permanently implanted titanium alloy (Ti6Al4V) structures in spherical or cuboidal configurations; open architecture truss design. Used as accessory to Stryker T2 Ankle Arthrodesis Nail or Stryker Valor Hindfoot Fusion Nail for tibiotalocalcaneal fusion. Implanted during salvage procedures for failed ankle arthrodesis or arthroplasty in patients at risk for limb loss. Provides mechanical support; open space allows bone graft integration. Patient-matched design based on radiography scans. Used in clinical settings by orthopedic surgeons. Benefits include structural support for fusion in complex salvage cases.
Clinical Evidence
Retrospective, single-arm, multi-center evaluation. Compared ATS device to control treatments (IM nail alone) for TTC fusion in failed total ankle arthrodesis patients. Endpoints: fusion, pain, functional outcomes, safety (adverse events), and secondary surgical intervention (SSSI) rates.
Technological Characteristics
Titanium alloy (Ti6Al4V) construction; additive manufacturing technique. Spherical or cuboidal configurations with open architecture truss. Mechanical testing per ASTM F2077 and ASTM F1264. Corrosion and wear assessment performed.
Indications for Use
Indicated for patients at risk for limb loss requiring tibiotalocalcaneal fusion as a salvage procedure following failed ankle arthrodesis or failed ankle arthroplasty. Used as an accessory to Stryker T2 Ankle Arthrodesis Nail or Stryker Valor Hindfoot Fusion Nail. Requires identifiable anatomical landmarks on radiography for patient-matched design. Not for standalone use.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
4WEB Medical Lateral Spine Truss System - Interbody Fusion Device (K211388)
Submission Summary (Full Text)
{0}------------------------------------------------
March 21, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
4WEB Medical, Inc. % Kathy Herzog Regulatory Consultant DuVal & Associates, P.A. 1820 Medical Arts Building 825 Nicollet Mall Minneapolis. Minnesota 55402
Re: K230088
Trade/Device Name: Ankle Truss System (ATS) Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: SAI, HSB Dated: March 19, 2024 Received: March 20, 2024
Dear Kathy Herzog:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming
{1}------------------------------------------------
product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Farzana Farzana Sharmin -S Sharmin -S Date: 2024.03.21
15:03:45 -04'00'
Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230088
Device Name Ankle Truss System (ATS)
#### Indications for Use (Describe)
The 4WEB Medical Ankle Truss System (ATS) is for use as an accessory to the Stryker T2 Ankle Arthrodesis Nail or the Stryker Valor Hindfoot Fusion Nail as part of a tibiotalocalcaneal fusion construct in a salvage procedure following failed ankle arthrodesis or failed ankle arthroplasty for patients at risk for loss of limb. The ATS is not intended for standalone use.
The anatomical landmarks necessary for the design and creation of ATS devices that are patient matched must be present and identifiable on appropriate radiography scans. The ATS is intended for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date Prepared: | March 21, 2024 |
|-------------------------|----------------------------------------------------------------|
| Contact: | Jarod Oldham, VP of Quality and Regulatory |
| | 4WEB Medical, Inc. |
| | 2801 Network Blvd., Suite 620 |
| | Frisco, TX 75034 |
| | Phone: (800) 285-7090 |
| | Fax: 972-488-1816 |
| Regulatory Contact: | Kathy Herzog, Sr. Regulatory, Quality, & Compliance Consultant |
| | DuVal & Associates, PA |
| | herzog@duvalfdalaw.com |
| Trade Name: | Ankle Truss System (ATS) |
| Product Class: | Class II |
| Classification: | 21 CFR §888.3020 |
| Common Name: | Rod, Fixation, Intramedullary and Accessories |
| Product Code: | SAI |
| Secondary Product Code: | HSB |
#### Predicate Device(s):
The primary predicate device is the Stryker Valor Hindfoot Fusion Nail (K090857). An additional predicate is the Stryker T2 Ankle Arthrodesis Nail (K200880). A reference device is the 4WEB Medical Lateral Spine Truss System - Interbody Fusion Device (K211388).
#### Device Description:
The 4WEB Medical Ankle Truss System (ATS) is a permanently implanted device intended to provide mechanical support as an accessory to a tibiotalocalcaneal nail intended to fuse the tibiotalocalcaneal joint in a salvage procedure following a failed ankle arthrodesis or a failed ankle arthroplasty for patients at risk for loss of limb. ATS devices are titanium alloy (Ti6Al4V) structures in spherical or cuboidal configurations. The ATS device has an open architecture truss designed to provide structural support with open space throughout the implant to allow fusion. ATS devices are intended for use with the Stryker T2 Ankle Arthrodesis Nail or the Stryker Valor Hindfoot Fusion Nail as part of a tibiotalocalcaneal (TTC) fusion system.
#### Indications for Use:
The 4WEB Medical Ankle Truss System (ATS) is for use as an accessory to the Stryker T2 Ankle Arthrodesis Nail or the Stryker Valor Hindfoot Fusion Nail as part of a tibiotalocalcaneal fusion construct in a salvage procedure following failed ankle arthrodesis or failed ankle arthroplasty for patients at risk for loss of limb. The ATS is not intended for standalone use.
The anatomical landmarks necessary for the design and creation of ATS devices that are patient matched must be present and identifiable on appropriate radiography scans. The ATS is intended for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
{4}------------------------------------------------
## Subject and Predicate Device Comparison
Both the subject and predicate devices have the same intended use for tibiotalocalcaneal (TTC) fusion. The indications for use of the ATS are a subset of the indications for use for the predicate Stryker Valor Hindfoot Fusion Nail, K090857.
The technological characteristics of the ATS device differ from the predicate device. The materials, truss design, and manufacturing of the ATS are similar to the reference device (K211388). Both the subject device and reference device are made of titanium alloy (Ti6Al4V) and are manufactured using an identical additive manufacturing technique.
## Non-Clinical Performance Evaluation
The following non-clinical performance testing has been completed on the subject device:
- Static and dynamic testing per ASTM F2077 (With and Without an IM Nail) followed by testing per ASTM F1264
- Corrosion Assessment
- Wear Assessment
#### Clinical Performance Evaluation
4WEB Medical completed a retrospective, single arm, multi-center evaluation of the ATS device. Control treatments were an IM nail for TTC fusion for failed total ankle arthrodesis patients. Effectiveness endpoints included fusion, pain, and functional outcomes for the subject and control treatments. Safety outcomes included analysis of device and procedure related adverse events and rates of subsequent secondary surgical intervention (SSSI) for the subject and control treatments.
## Conclusion
The nonclinical and clinical performance testing demonstrate that the subject device is substantially equivalent to the predicate device, Stryker Valor Hindfoot Fusion Nail (K090857), when used as intended.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.