The Excelsior XT-27 Microcatheter and Excelsior XT-27 Flex Microcatheter are intended to assist in the delivery of embolization particles, diagnostic agents (such as contrast media), and interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter. The Excelsior XT-27 Pre-Shaped Microcatheter and Excelsior XT-27 Flex Pre-Shaped Microcatheter are intended to assist in the delivery of embolization particles, diagnostic agents (such as contrast media), and interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter.
Device Story
Sterile, single-lumen, 0.027-inch ID microcatheter; used to access distal neurovasculature. Features graded shaft stiffness, radiopaque platinum/iridium tip marker for fluoroscopic visualization, and Hydrolene hydrophilic coating to reduce friction. Operated by physicians in clinical settings; used with guide catheter and steerable guidewire. Provides pathway for delivery of embolization particles, contrast media, and stents. Modification involves updated labeling for compatibility with guide catheters having a minimum 0.046-inch ID. Benefits include improved trackability and device delivery in neurovascular procedures.
Clinical Evidence
Bench testing only. Studies included guide catheter compatibility (track and pull-back forces), tensile strength of joints and marker bands after simulated use, and evaluation of hydrophilic coating integrity and particulate generation. All tests passed.
Technological Characteristics
Materials: PTFE, Pebax, 304 stainless steel wire, Nylon, Santoprene. Tip marker: 90% Platinum/10% Iridium. Coating: Hydrolene (hydrophilic polymer). Dimensions: 0.027 inch ID; 135 cm or 150 cm effective length; 6 cm or 18 cm distal shaft lengths. Sterilization: Ethylene Oxide (EtO) gas. Single-lumen, steerable design.
Indications for Use
Indicated for use in the neurovasculature to assist in the delivery of embolization particles, diagnostic agents (contrast media), and interventional devices (stents) using a catheter with a 0.027-inch inner diameter.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
March 21, 2025
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in a smaller font size.
Stryker Neurovascular Manmeet Chawla Senior Project Manager Regulatory Affairs 47900 Bayside Parkway Fremont, California 94538
Re: K242289
Trade/Device Name: Excelsior XT-27 Microcatheter, Excelsior XT-27 Flex Microcatheter, Excelsior XT-27 Pre-Shaped Microcatheter, Excelsior XT-27 Flex Pre-Shaped Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, KRA Dated: February 21, 2025 Received: February 21, 2025
Dear Manmeet Chawla:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242289
#### Device Name
Excelsior XT-27 Microcatheter, Excelsior XT-27 Flex Microcatheter,
Excelsior XT-27 Pre-Shaped Microcatheter, Excelsior XT-27 Flex Pre-Shaped Microcatheter
#### Indications for Use (Describe)
The Excelsior XT-27 Microcatheter and Excelsior XT-27 Flex Microcatheter are intended to assist in the delivery of embolization particles, diagnostic agents (such as contrast media), and interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter.
The Excelsior XT-27 Pre-Shaped Microcatheter and Excelsior XT-27 Flex Pre-Shaped Microcatheter are intended to assist in the delivery of embolization particles, diagnostic agents (such as contrast media), and interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Stryker Neurovascular. The word "stryker" is in bold, black, sans-serif font. Below it, the word "Neurovascular" is in a smaller, black, sans-serif font. There is a registered trademark symbol to the right of the word "stryker".
K242289
## 510(k) Summary
| Submitter: | Stryker Neurovascular<br>47900 Bayside Parkway<br>Fremont, CA 94538<br>Phone: 510-423-2500 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Manmeet Chawla<br>E-mail: manmeet.chawla@stryker.com |
| Date Prepared: | March 20, 2025 |
| Device Name: | Excelsior® XT-27® Microcatheter, Excelsior® XT-27® Flex Microcatheter,<br>Excelsior® XT-27® Pre-Shaped Microcatheter, Excelsior® XT-27® Flex Pre-<br>Shaped Microcatheter |
| Common Name: | Percutaneous catheter |
| Classification Name: | Catheter, Percutaneous, Neurovasculature (21 CFR 870.1250, QJP) and<br>Catheter, Continuous Flush (21 CFR 870.1210, KRA) |
| Device Classification: | Class II |
| Product Code: | QJP and KRA |
| Predicate Device: | Excelsior® XT-27® Microcatheter (K113778, cleared 20-April-2012) |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Stryker Neurovascular. The word "stryker" is in bold, sans-serif font, with the "s" in lowercase and the rest of the letters in uppercase. To the right of the "r" is the registered trademark symbol. Below "stryker" is the word "Neurovascular" in a smaller, sans-serif font.
## DEVICE DESCRIPTION:
The subject devices hereafter referred to as Excelsior® XT-27® Microcatheters are sterile, single lumen, 0.027 inch inner diameter (ID) microcatheters with one tip marker designed to aid the physician in accessing distal neurovasculature when used with a guide catheter and steerable guidewire. Graded shaft stiffness ranging from a highly flexible tip to a semi-rigid proximal section aids the physician in tracking over selectively placed guidewires. A luer fitting located on the microcatheter hub is used for the attachment of accessories. One radiopaque tip marker is positioned at the distal tip of the device to facilitate fluoroscopic visualization® XT-27® Microcatheters are coated on the outer surface with Hydrolene™ coating which reduces friction during manipulation in the vessel.
The Excelsior® XT-27® Microcatheters are available in effective lengths of both 135 cm (53.1 inch) and 150 cm (59.1 inch), with two distal shaft configurations achieved through distal shaft lengths of 6 cm (XT-27 model) and 18 cm (XT-27 Flex model). Both straight tip and pre-shaped versions are available. Refer to Table 1.
| Product Description | UPN | Effective Length (cm) /<br>Tip Length (cm) |
|-------------------------------------|---------------|--------------------------------------------|
| Excelsior® XT-27®, Straight | M003XT2735810 | 135 / 6 |
| Excelsior® XT-27®, Pre-Shaped | M003XT2735910 | 135 / 6 |
| Excelsior® XT-27®, Straight | M003XT2750810 | 150 / 6 |
| Excelsior® XT-27®, Pre-Shaped | M003XT2750910 | 150 / 6 |
| Excelsior® XT-27®, Flex, Straight | M003XT2735820 | 135 / 18 |
| Excelsior® XT-27®, Flex, Pre-Shaped | M003XT2735920 | 135 / 18 |
| Excelsior® XT-27®, Flex, Straight | M003XT2750820 | 150 / 18 |
| Excelsior® XT-27®, Flex, Pre-Shaped | M003XT2750920 | 150 / 18 |
| | | Table 1. Excelsior® XT-27® Microcatheter Types | |
|--|--|------------------------------------------------|--|
|--|--|------------------------------------------------|--|
Please note that all Excelsior® XT-27® Microcatheters referenced in this submission are considered the subject device.
The subject Excelsior® XT-27® Microcatheters are a modification to the predicate device, Excelsior® XT-27 Microcatheter (K113778), for the labeling change to recommend the use of the Excelsion® XT-27 % Microcatheters with guide catheters with a minimum ID of 0.046 inch (1.17 mm). The technological characteristics and principles of operation remain unchanged except for the Excelsior® XT-27% Microcatheters with guide catheters with a minimum ID of 0.046 inch (1.17 mm). The device design, materials, manufacturing, packaging, and sterilization methods, biocompatibility data, bench-top data, and stability are directly applicable from the predicate device (K113778) and from other changes made since that did not require a 510(k) submission.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Stryker Neurovascular. The word "stryker" is written in a bold, sans-serif font, with the letters closely spaced together. Below the word "stryker" is the word "Neurovascular", written in a smaller, sans-serif font. There is a registered trademark symbol to the right of the word "stryker".
# Accessories
Each Excelsion® XT-27® Microcatheter is provided with a shaping mandrel and peel away introducer sheath within a separate inner Tyvek™ pouch (accessory pouch) for straight configurations and inside a pre-shaped tray for the pre-shaped configurations. The Excelsior® XT- 27® Microcatheters are available in a single pack (one unit per package) only. The device pouch, Instruction for Use (IFU), and pointer page are provided inside a shelf carton.
## INTENDED USE /INDICATIONS FOR USE:
The indications for use are similar between the predicate Excelsior® XT-27® Microcatheter (K113778) and subject Excelsion® XT-27® Microcatheters. Minor modifications for clarity were made to the indications for use (refer to comparison in Table 2). Changes do not impact the intended use of the device.
Indications for Use for Excelsior® XT-27® Microcatheter and Excelsion® XT-27® Flex Microcatheter:
The Excelsior® XT-27® Microcatheter and Excelsior® XT-27® Flex Microcatheter are intended to assist in the delivery of embolization particles, diagnostic agents (such as contrast media), and interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter.
# Indications for Use for Excelsior® XT-27® Pre-Shaped Microcatheter and Excelsior® XT-27® Flex (Pre-Shaped Microcatheter:
The Excelsion® XT-27® Pre-Shaped Microcatheter and Excelsior® XT-27® Flex Pre-Shaped Microcatheter are intended to assist in the delivery of embolization particles, diagnostic agents (such as contrast media), and interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter.
# TECHNOLOGICAL CHARACTERISTICS AND COMPARISON TO THE PREDICATE DEVICE:
Stryker Neurovascular has demonstrated that the subject Excelsior® XT-27® Microcatheters are substantially equivalent to the predicate Excelsior® XT-27® Microcatheter (K113778) based on the same intended use, same or similar materials, similar design, and the same fundamental operating principles.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Stryker Neurovascular. The word "stryker" is in bold, sans-serif font, with the "s" slightly overlapping the "t". Below "stryker" is the word "Neurovascular" in a smaller, sans-serif font. A registered trademark symbol is located to the right of the "r" in "stryker".
A comparison of the subject and predicate devices is summarized in Table 2 below.
| Characteristic | Predicate Device<br>Excelsior® XT-27® Microcatheter<br>(K113778) | Subject Device<br>Excelsior® XT-27® Microcatheters<br>(K242289) |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Stryker | Same |
| Device Type | Percutaneous Catheter | Same |
| Regulation/Class | 870.1250, Class II<br>870.1210, Class II | Same |
| Product Code | DQY, KRA | QJP, KRA |
| Intended<br>Use/Indications<br>for Use | The Excelsior XT-27 Microcatheter is<br>intended to assist in the delivery of<br>diagnostic agents (such as contrast<br>media), therapeutic agents, and non-<br>liquid interventional devices (such as<br>stents) that are indicated for use in the<br>neurovasculature and with a catheter of<br>0.027 inches in inner diameter. | The Excelsior XT-27 Microcatheter and<br>Excelsior XT-27 Flex Microcatheter are<br>intended to assist in the delivery of<br>embolization particles, diagnostic agents<br>(such as contrast media), and<br>interventional devices (such as stents)<br>that are indicated for use in the<br>neurovasculature and with a catheter of<br>0.027 inches in inner diameter.<br><br>The Excelsior XT-27 Pre-Shaped<br>Microcatheter and Excelsior XT-27 Flex<br>Pre-Shaped Microcatheter are intended<br>to assist in the delivery of embolization<br>particles, diagnostic agents (such as<br>contrast media), and interventional<br>devices (such as stents) that are indicated<br>for use in the neurovasculature and with<br>a catheter of 0.027 inches in inner<br>diameter. |
# Table 2. Comparison of the Subject Devices with the Predicate Device
{8}------------------------------------------------
# Stryker Neurovascular Premarket Notification, Special 510(k)
| Excelsior® XT-27® and Excelsior® XT-27® Flex (Straight Microcatheters) |
|--------------------------------------------------------------------------|
| Excelsior® XT-27® and Excelsior® XT-27® Flex (Pre-Shaped Microcatheters) |
| Characteristic | Predicate Device<br>Excelsior® XT-27® Microcatheter<br>(K113778) | Subject Device<br>Excelsior® XT-27® Microcatheters<br>(K242289) |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| How Supplied | Single use, sterile, non-pyrogenic | Same |
| Method of<br>Sterilization | Ethylene Oxide (EtO) gas | Same |
| Materials | PTFE, Pebax, Stainless Steel wire,<br>Nylon, Santoprene | Same |
| Features | | |
| Tip Configuration | Straight tip Pre-Shaped tip Option for secondary steam shaping<br>by physician prior to use. | Same |
| Effective Lengths | 135 cm, 150 cm | Same |
| Proximal OD/<br>Distal OD | Proximal OD (outer diameter): 2.9F<br>Distal OD: 2.7F | Same |
| ID | 0.027 inch | Same |
| Hydrophilic<br>Coating Length | 80 cm | Same |
| Tip Length/Distal<br>Shaft Length | 6 cm, 18 cm | Same |
| Tip Marker | Platinum (90%) / Iridium (10%) alloy | Same |
| Metal<br>Reinforcement | 304 Stainless Steel | Same |
| Coating | Hydrophilic Polymer | Same |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Stryker Neurovascular. The word "stryker" is written in a bold, sans-serif font, with the "r" having a registered trademark symbol next to it. Below "stryker" is the word "Neurovascular" in a smaller, sans-serif font.
#### RISK ASSESSMENT:
Risk assessment of the subject modified Excelsior® XT-27® Microcatheters has been conducted in accordance with EN ISO 14971 and Stryker Neurovascular risk management procedures. Stryker Neurovascular has determined that the change in the minimum recommended guide catheter ID as outlined in the compatibility table of the Excelsior® XT-27® Microcatheters IFU raises no new questions of safety or effectiveness. The risk assessment conducted on the subject device did not result in any new hazards or harms.
#### PERFORMANCE DATA:
Non-clinical tests were performed to demonstrate the subject device performs as intended and substantial equivalence. Bench testing that was performed is summarized in Table 3.
| Test | Test Method Summary | Results |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Guide Catheter Compatibility<br>with 1.17 mm (0.046") ID | A track and Pull Back study was conducted to<br>evaluate the maximum forces required to completely<br>deliver and retrieve an Excelsior® XT-27®<br>Microcatheter inside a 0.046" ID guide catheter with<br>ancillary devices that represent worst-case sizes. | Pass |
| Tensile Strength for Joints and<br>Marker Band | Tensile strength tested after preconditioning by<br>simulated use with worst-case sized ancillary<br>devices and interventional devices. | Pass |
| Particulate and Coating<br>Integrity | Hydrophilic coating integrity and particulate<br>generation were evaluated under simulated use<br>conditions with Excelsior® XT-27® Microcatheters<br>used with 0.046" ID guide catheter, with comparison<br>to cleared comparator devices. | Pass |
## Table 3. Bench Testing Summary
# CONCLUSION:
The subject device is substantially equivalent to the predicate device based on similar intended use/indications for use, same or similar materials, same fundamental design, and the same operating principles. The non-clinical testing data support the subject device performs as intended.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.