The Excelsior XT-27 Microcatheter is intended to assist in the delivery of diagnostic agents (such as contrast media), therapeutic agents, and non-liquid interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter.
Device Story
Sterile, single-lumen microcatheter; 0.027-inch ID; used by physicians to access distal neurovasculature. Operates via tracking over steerable guidewires within a guide catheter. Features graded shaft stiffness and Hydrolene outer coating to reduce friction. Includes radiopaque tip marker for fluoroscopic visualization. Available in 135cm/150cm lengths with straight or pre-shaped tips. Physician uses luer fitting to attach accessories; performs steam shaping as needed. Facilitates delivery of contrast media, therapeutic agents, and stents. Benefits patient by enabling precise navigation to distal neurovascular sites for diagnostic or interventional procedures.
Clinical Evidence
Bench testing only. Testing included visual inspection, dimensional measurements, corrosion resistance, simulated use (introduction, tracking, deployment, detachment), advancement/retraction force, and comprehensive biocompatibility (cytotoxicity, irritation, systemic toxicity, sensitization, hemocompatibility, pyrogenicity). All results met pre-defined acceptance criteria equivalent to predicate devices.
Indicated for patients requiring delivery of diagnostic or therapeutic agents and non-liquid interventional devices (e.g., stents) into the neurovasculature using a 0.027-inch inner diameter catheter.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
APR 2 0 2012
Stryker Neurovascular Traditional 510(k) Submission Excelsior® XT-27™ Microcatheter
i
. '•
K113778
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
.
| Trade Name: | Excelsior® XT-27™ Microcatheter |
|-------------------|------------------------------------------------------------------------------------------------|
| Generic Name: | Percutaneous Catheter, Microcatheter |
| Classification: | CLASS II, 21 CFR 870.1250 |
| Submitted By: | Stryker Neurovascular<br>47900 Bayside Parkway<br>Fremont, CA 94538-6515 |
| Contact: | Yoko Y Enrile |
| Predicate Device: | K000177, Renegade™ Hi-Flo™ Microcatheter<br>K042568, Excelsior® SL-10 Pre-Shaped Microcatheter |
Indications for Use: The Excelsior XT-27 Microcatheter is intended to assist in the delivery of
diagnostic agents (such as contrast media), therapeutic agents, and non-
liquid interventional devices (such as stents) that are indicated for use in
the neurovasculature and with a catheter of 0.027 inches in inner diameter.
Device Description: The Excelsior® XT-27™ Microcatheter is a sterile, single lumen 0.027 in
ID device with one tip marker designed to aid the physician in accessing
distal vasculature when used with a guide catheter and steerable
guidewire. Graded shaft stiffness ranging from a highly flexible tip to a
semi-rigid proximal section aids the physician in tracking over selectively
placed guidewires. A luer fitting located on the microcatheter hub is used
for the attachment of accessories. One radiopaque tip marker is positioned
at the distal tip of the device to facilitate fluoroscopic visualization. The
Excelsior XT-27 Microcatheter is coated on the outer surface with
Hydrolene™ coating which reduces friction during manipulation in the
vessel.
The Excelsior XT-27 Microcatheters are available in effective lengths of
both 135cm (53.1 inch) and 150 cm (59.1 inch), with two distal shaft
configurations achieved through distal shaft lengths of 6 cm (XT-27
model) and 18 cm (XT-27 Flex model). Both straight tip and Pre-Shaped
versions are available.
{1}------------------------------------------------
Accessories: Each Excelsior XT-27 Microcatheter is provided with accessories (a shaping mandrel, and peel away introducer) within a separate inner Tyvek™ pouch (Accessory pouch). The Excelsior XT-27 Microcatheter is available in a single pack (one unit per package) only. The device pouch and Directions for Use (DFU) are both provided inside a shelf carton.
Summary Differences in Technological Characteristics: The Excelsior XT-27 Microcatheters introduce several enhanced design features:
- a) The OD profile of the Excelsior XT-27 Microcatheter is smaller than the Renegade Hi-Flo Microcatheter.
- b) The Excelsior XT-27 Microcatheter has up to five zones of polymer stiffness which vary from the distal to proximal shaft and help to enhance the trackability and pushability of the catheter.
- c) The Excelsior XT-27 Microcatheter has a more supportive double helix winding reinforcement compared to the Renegade Hi-Flo Microcatheter.
### Bench Test Summary:
| Test | Result |
|-------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Visual Inspection:<br>Durable Hydrophilic Coating<br>Surface Defects<br>Surface-Extraneous matter | Met same criteria as<br>predicate |
| Dimensional Measurement:<br>Kink Radius of Curvature / Proximal<br>Shaft Kink<br>Distal OD Reduction<br>Tip Configuration<br>Catheter Hub | Met same criteria as<br>predicate |
| Corrosion Resistance | Met same criteria as<br>predicate |
{2}------------------------------------------------
. . . . . .
and the control of the county of
.
.
:
:
.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
:
:
.
:
:
| Simulated use: | Met same criteria as predicate |
|---------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| * Introduction | |
| * Tracking | |
| * Reposition / Deployment | |
| * Detachment | |
| * Overall Performance | |
| Advancement / Retraction Force | Met same criteria as predicate |
| In-vitro Cytotoxicity, MEM Elution | Non-cytotoxic |
| Intracutaneous reactivity | Non-irritating |
| Acute systemic toxicity, Injection | Non-toxic |
| Sensitization, Guinea Pig<br>Maximization | Non- sensitizing |
| Hemocompatibility, Direct Contact;<br>Complement Activation C3a and<br>SC5b-9; PTT; blood cell counts and<br>hemoglobin / hematocrit levels | Non-hemolytic |
| Materials Mediated Pyrogen levels,<br>Rabbit test | Non-pyrogenic |
| USP impurities, <661> | None |
| Latex test, ELISA Inhibition | None |
·
:
{3}------------------------------------------------
| Predicate / Subject | Product Feature Comparison for Excelsior® SL-10 Pre-Shaped |
|---------------------|--------------------------------------------------------------------------|
| Device | Microcatheter (K042568) and Renegade™ Hi-Flo™ Microcatheter<br>(K000177) |
| Comparison: | (K000177) |
| Feature | K042568<br>Excelsior® SL-10<br>Pre-Shaped<br>Microcatheter | K000177<br>Renegade™ Hi-Flo™<br>Microcatheter | Subject Device |
|----------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Materials | PTFE, Pebax,<br>Stainless Steel<br>wire, Nylon,<br>Santoprene | PTFE, Pebax,<br>Vectran Fiber and<br>Pt/Ir wire, Nylon,<br>Santoprene | PTFE, Pebax,<br>Stainless Steel<br>wire, Nylon,<br>Santoprene |
| Tip shape | Preshaped | Straight | Offers both shapes<br>(Straight and<br>Preshaped) |
| | Option of steam<br>shaping by<br>physician for<br>proper adjustment<br>to the anatomy<br>prior to use | Option of steam<br>shaping by<br>physician for<br>proper adjustment<br>to the anatomy<br>prior to use | Option of steam<br>shaping by<br>physician for<br>proper adjustment<br>to the anatomy<br>prior to use |
| Effective<br>Lengths | 150 cm | 135 cm, 150 cm | 135 cm, 150 cm |
| Proximal<br>/Distal OD | Proximal OD 2.4F<br>Distal OD 1.7F | Proximal OD 3F<br>Distal OD 2.8F | Proximal OD: 2.9F<br>Distal OD: 2.7F |
| ID | 0.0165 inch | 0.027inch | 0.027inch |
| Hydrophilic<br>Coating<br>Length | 100 cm | 60 cm | 80 cm |
| Tip Length | 6cm | 10cm and 20cm | 6cm and 18cm |
| Tip Markers | 90%Pt-10%Ir | 90%Pt-10%Ir | 90%Pt-10%Ir |
| Coating | Polyvinylpyrrolidone Polyacrylamide | Hydropass<br>Hydrophilic<br>Coating | Polyvinylpyrrolidone Polyacrylamide |
| Method of<br>supply | Sterile, single-use,<br>non-pyrogenic | Sterile, single-use,<br>non-pyrogenic | Sterile, single-use,<br>non-pyrogenic |
Summary of Substantial Equivalence: The Excelsior XT-27 Microcatheter subject of this submission is substantially equivalent to the predicate devices with regard to design, materials, sterilization, principle of operation, performance, and indications for use.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to promoting health, well-being, and human services. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Stryker Neurovascular c/o Ms. Tina Lochner DEKRA Certification. Inc. 4377 County Line Road Chalfont, PA 18914
APR 2 0 2012
Re: K113778
Trade/Device Name: Excelsior XT-27 Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, KRA Dated: March 16, 2012 Received: March 22, 2012
Dear Ms. Lochner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, thereforey issues of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be fray be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Tina Lochner
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Ea. Hemm
Malvina B. Eydelman, MD Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K113778
Device Name: . . . . . _ _ _ Excelsior® XT-27™ Microcatheter
Indications For Use:
The Excelsior XT-27 Microcatheter is intended to assist in the delivery of diagnostic agents (such as contrast media), therapeutic agents, and non-liquid interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Quyuh Hoang
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number_
K113778
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.