Direct restoration of Class I, II, III and V cavities (ideally Class I and II), base or liner, core build-ups.
Device Story
DenMat Bulk Fill is a tooth-shaded, radiopaque, dual-cured (auto-cure with light-cure acceleration) composite resin. It is used by dental professionals in clinical settings for direct dental restorations and core build-ups. The device is delivered via a dual-barrel syringe with auto-mixing tips. Its low-viscosity, flowable consistency allows for adaptation to prepared cavity walls. The dual-cure mechanism provides an unlimited depth of cure in a single application, eliminating the need for layering and allowing for use in areas where light transmission is inadequate. By simplifying the restorative process and ensuring structural integrity through established composite chemistry, the device facilitates efficient dental procedures and provides durable restorations for patients.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing, including flexural strength, diametral strength, work time, set time, water solubility, water absorption, and color stability. Biocompatibility assessed per ISO 10993-1, including cytotoxicity, sensitization, reactivity, genotoxicity (Ames/Mouse Lymphoma), pyrogenicity, and implantation.
Technological Characteristics
Dual-cured (auto-cure/light-cure) composite resin. Radiopaque. Low-viscosity flowable form factor. Delivered via dual-barrel syringe with auto-mixing tips. Curing light wavelength 400-500 nm. Complies with ISO 4049, ISO 10993 (1, 3, 5, 10, 11, 18, 23), ISO 7405, ISO 14971, and ISO 15223-1.
Indications for Use
Indicated for direct restoration of Class I, II, III, and V cavities (ideally Class I and II), as a base or liner, and for core build-ups.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 15, 2024
Den-Mat Holdings, LLC Kim Maynes Sr. Supervisor, Regulatory Affairs 1017 W. Central Lompoc, California 93436
Re: K242097
Trade/Device Name: DenMat Bulk Fill Composite Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: July 17, 2024 Received: July 18, 2024
Dear Kim Maynes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K242097
Device Name
DenMat Bulk Fill Composite
Indications for Use (Describe)
Direct restoration of Class I, II, III and V cavities (ideally Class I and II), base or liner, core build-ups.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Submission Summary - K242097
| Submitter's Name: | Den-Mat Holdings, LLC |
|------------------------------|--------------------------------------------|
| Address: | 1017 W. Central Avenue<br>Lompoc, CA 93436 |
| Phone Number: | 805-346-3700 |
| Contact Person: | Kim Maynes |
| Date of Summary Preparation: | 7/17/24 |
# 1. Submitter Information
# 2. Device Details
| Device Name: | DenMat Bulk Fill |
|------------------------------|------------------------------|
| Device Classification Name: | Material, Tooth Shade, Resin |
| Common Name: | Composite Resin |
| Classification: | II |
| Regulation number: | 21 CFR 872.3690 |
| Classification Product Code: | EBF |
# 3. Predicate Device
| Submitter's device | Predicate Device | Reference Device |
|--------------------|---------------------------------------|--------------------------------------|
| DenMat Bulk Fill | Name: Stela Automix | Name: DMRC Bulk Fill |
| | Manufacturer: SDI Limited (Australia) | Manufacturer: Danville Materials LLC |
| | 510(k) number: K222581 | 510(k) number: K151088 |
# 4. Device Description
DenMat Bulk Fill is a tooth shaded, radiopaque, dual-cured (auto-cure with light-cure acceleration) composite that offers an unlimited depth of cure in a single application without multiple layers and lightcuring when allowed to self-cure. DenMat Bulk Fill has the handling properties of a low viscosity or "flowable" composite that allows for adaptation to the prepared cavity walls. DenMat Bulk Fill is also ideal for core buildups, and any application where light transmission is inadequate. It is provided in multiple shades.
# 5. Indications for Use
Direct restoration of Class I, II, III and V cavities (ideally Class I and II), base or liner, core build-ups.
# 6. Comparison of Technological and Regulatory Characteristics with the Predicate Device
The similarities between DenMat Bulk Fill and the primary predicate Stela Automix and the reference device DMRC Bulk Fill, include the Regulatory Classification, Product Code, Intended Use of the device, Indications for Use, and Clinical condition of use. Furthermore, Technological Characteristics and Performance testing including Biocompatibility testing are similar.
{5}------------------------------------------------
There are no major differences but only minor differences between DenMat Bulk Fill and the primary predicate Stela Automix and the reference device DMRC Bulk Fill. Performance test results have slightly different values. In addition, the reference device, DMRC Bulk Fill, is a dual-cure bulk fill composite and uses identical curing light wavelength and curing time as DenMat Bulk Fill. The curing light intensity indicated for DMRC Bulk Fill is similar to DenMat Bulk Fill. These minor differences have no impact on the safety or performance of the device which demonstrates the subject device is substantially equivalent to the predicate device and the reference device.
Table 1 below compares DenMat Bulk Fill, the primary predicate Stela Automix and the reference device DMRC Bulk Fill with respect to intended use, technological characteristics and performance testing.
Non-clinical performance evaluations demonstrate that DenMat Bulk Fill is substantially equivalent to the primary predicate Stela Automix and the reference device DMRC Bulk Fill passes all standards, in terms of critical properties for restorative composites, including flexural and diametral strength, work time, set time, water solubility, water absorption and color stability.
| | Submitter's Device | Primary Predicate<br>Device | Reference Device | Comparisons |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Name | DenMat Bulk Fill | Stela Automix | DMRC Bulk Fill | |
| 510(k) number | K242097 | K222581 | K151088 | |
| Device<br>Classification Name | Tooth Shade Resin<br>Material | Tooth Shade Resin<br>Material | Tooth shade resin<br>material | Same |
| Regulation number | 21 CFR 872.3690 | 21 CFR 872.3690 | 21 CFR 872.3690 | Same |
| Classification<br>Product Code | EBF | EBF | EBF | Same |
| Indications for use | Direct restoration of<br>Class I, II, III and V<br>cavities (ideally<br>Class I and II), base<br>or liner, core build-<br>ups. | 1) Direct restoration<br>of Class I, II, III and V<br>cavities. Ideally<br>Class I and II,<br>2) Base or liner,<br>3) Core build-ups | DMRC Bulk Fill is<br>a dual-cure (auto-<br>cure) with light-<br>cure acceleration)<br>polymer-based<br>dental<br>restoratives that<br>when applied to<br>dental surfaces<br>pretreated with<br>suitable primers<br>or adhesives are<br>indicated for use<br>as the first<br>increment under<br>posterior<br>composites; for<br>core build-ups; | Same as the<br>primary<br>predicate |
| | | | and for luting<br>posts, crowns,<br>veneers and any<br>application where<br>light transmission<br>may be<br>inadequate | |
| Delivery System | Dual barrel syringe,<br>with auto mixing<br>tips | Dual barrel syringe,<br>with auto mixing<br>tips | Dual barrel<br>syringe, with auto<br>mixing tips | Same |
| Sterile | No | No | No | Same |
| Work time<br>(seconds)<br>(self cure) | 105 | 130 | 90 | Similar to<br>predicate<br>and<br>reference |
| Set time (seconds)<br>(self cure) | 145 | 170 | 125 | Similar to<br>predicate<br>and<br>reference |
| Curing mechanism | Dual cure<br>(self-cure and light-<br>cure) | Self-cure | Dual cure<br>(self-cure and<br>light-cure) | Same as<br>reference<br>device |
| Curing light<br>wavelength (nm) | 400 - 500 | N/A | 400 - 500 | Same as<br>reference<br>device |
| Curing time<br>(seconds)<br>(light cure) | 10 | N/A | 10 | Same as<br>reference<br>device |
| Curing light<br>intensity (mW/cm²) | > 1000 | N/A | > 600 for halogen<br>light and per<br>manufacturing<br>suggestions for<br>LED lights | Similar to<br>reference<br>device |
| Flexural strength<br>(MPa) (> 80) | 119 | 137 | 118 | Similar to<br>predicate<br>and<br>reference |
| Diametral strength<br>(MPa) (> 28) | 52 | 56 | 54 | Similar to<br>predicate<br>and<br>reference |
| Water sorption<br>(µg/mm³) (< 40) | 13 | 28 | 7 | Similar to<br>predicate<br>and<br>reference |
| Water solubility<br>(µg/mm³) (< 7.5) | 0.1 | -2.5 | 0.8 | Similar to<br>predicate |
| | | | | and<br>reference |
| Color stability | pass | pass | pass | Similar to<br>predicate<br>and<br>reference |
| FDA Recognized<br>Standards | ISO 4049<br>ISO 10993-1<br>ISO 10993-3<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11<br>ISO 10993-18<br>ISO 10993-23<br>ISO 7405<br>ISO 14971<br>ISO 15223-1 | ISO 4049<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11<br>ISO 10993-3<br>ISO 7405 | ISO 4049 | |
### Table 1. Comparison of DenMat Bulk Fill, Primary Predicate Device and Reference Device
{6}------------------------------------------------
{7}------------------------------------------------
# 7. Performance Testing
### Biocompatibility assessment
According to ISO 10993-1 Biological evaluation of medical devices —Part 1: Evaluation and testing within a risk management process, the subject device is categorized as an external communicating device with contact to tissue/bone/dentin and is permanent contact device. DenMat Bulk Fill's chemical composition, as well as materials and technological properties, are substantially equivalent to those of FDA cleared commercial devices as defined in our assessment. Chemicals in DenMat Bulk Fill have been well documented and used extensively in the dental industry for many years. Clinical experience research reveals acceptable biocompatibility in devices in the same category (externally communicating devices for dental restorations) as DenMat Bulk Fill. Therefore, the combination of all materials, chemicals, and processes has an established history of safe use in the intended applications.
DenMat Bulk Fill has been submitted to the following biocompatibility tests and no significant adverse effects have been identified:
- Cytotoxicity
- Maximization Test for Delayed-Type Hypersensitivity
- Intracutaneous (Intradermal) Reactivity
- Bacterial Reverse Mutation Assay (Ames Test)
- In Vitro Mouse Lymphoma
- Pyrogen
- Implantation effects
Comparable dental devices with FDA clearance have not been a focus of any advisory notice or recalls from FDA that are related to biocompatibility issues.
In conclusion, DenMat Bulk Fill demonstrates acceptable biocompatibility for the intended use.
{8}------------------------------------------------
### Non-clinical Performance Data:
The standards applicable to the submitter's device and the comparison of technological characteristics referenced include: ISO 4049:2019 Polymer-Based Restorative Material; ANSI/ADA Standard No. 27-2016 Polymer-based Restorative Materials (Modified adoption of ISO 4049:2009, Dentistry — Polymerbased restorative materials); ISO 7491:2000 Dental materials — Determination of color stability; and DenMat internal testing methods.
Non-clinical performance evaluations demonstrate that DenMat Bulk Fill is substantially equivalent to the primary predicate Stela Automix and the reference device DMRC Bulk Fill passes all standards in terms of critical properties for restorative composites, including flexural and diametral strength, work time, set time, water solubility, water absorption and color stability (reference Table 1).
### Clinical Performance Data:
Clinical data is not needed to characterize performance and establish substantial equivalence. The nonclinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on the subject devices.
### Electrical Safety and electromagnetic compatibility (EMC)
This section is not applicable.
### Software verification and validation testing:
This section is not applicable.
#### Mechanical and Acoustic Testing:
This section is not applicable.
### 8. Conclusion Regarding Substantial Equivalence:
The performance and biocompatibility tests show DenMat Bulk Fill performs comparably to the predicate device and possesses a low likelihood of an unacceptable adverse biological response from contact of the component materials of the device with the body. The Denmat Bulk Fill indications for use are identical to the predicate device and are based on the same restorative materials technology. On this basis, DenMat Bulk Fill is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.