Stela Primer: - Dentin and enamel bonding Stela Automix: - Direct Class I, II, III and V cavities. Ideally Class I and II. - Base or liner - Core build-ups.
Device Story
Stela Automix System is a bulk-fill radiopaque self-cure dental restorative system for professional use. Components include Stela Primer (adhesive liquid) and Stela Automix (double-barrel syringe restorative paste). Primer applied to dentin/enamel via brush; Automix dispensed via mixing tip. System bonds to tooth structure via ionic bonds; polymerization occurs in situ upon contact. Used in dental clinics by dentists for direct restorations, base/liner, and core build-ups. Eliminates need for scrubbing or light curing. Benefits include simplified application and restorative procedure.
Clinical Evidence
No clinical data. Evidence consists of bench testing and biocompatibility assessments. Bench testing evaluated working time, setting time, flexural strength, shade, color stability, radio-opacity, and shear bond strength to dentin/enamel per ISO 4049 and ISO 29022:2013. Biocompatibility testing included cytotoxicity (ISO 7405, ISO 10993-5), delayed-type hypersensitivity (ISO 10993-10), irritation/intracutaneous reactivity (ISO 10993-10), systemic toxicity (ISO 10993-11), and genotoxicity (ISO 10993-3). Results indicate the system meets ISO requirements and performs comparably to predicate devices.
Technological Characteristics
Self-cure resin-based dental restorative system. Components: Primer (adhesive) and Automix (bulk-fill radiopaque composite). Materials: Resin-based. Sensing/Actuation: Chemical polymerization in situ. Connectivity: None. Sterilization: Not specified. Algorithm: None. Standards: ISO 4049 (performance), ISO 29022 (bond strength), ISO 10993 (biocompatibility).
Indications for Use
Indicated for dental professionals performing direct restorations of Class I, II, III, and V cavities, base or liner placement, and core build-ups, as well as dentin and enamel bonding.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 28, 2022
SDI Limited Quynh Jewell Regulatory Affairs Manager 3-15 Brunsdon Street, Bayswater Melbourne, Victoria 3153 Australia
### Re: K222581
Trade/Device Name: Stela Automix System Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF, KLE Dated: August 26, 2022 Received: August 26, 2022
### Dear Quynh Jewell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M. ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222581
Device Name Stela Automix System
Indications for Use (Describe) Stela Primer: - Dentin and enamel bonding
Stela Automix:
- Direct Class I, II, III and V cavities. Ideally Class I and II.
- Base or liner
- Core build-ups.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the letters "SDI" in a black square, followed by the letter "s" in a larger font size. The letters "SDI" are white and bolded. The letter "s" is also bolded and appears to be part of a word that is cut off.
# SDI Limite
Establishment Registration No: 3004140838
PIN: MD6132486
510(k) NOTIFICATION (Traditional): Stela Automix System – Summary – August 2022
# K222581
#### Submitter Information 1.
SDI Limited 3-15 Brunsdon Street Bayswater, Victoria 3153 Australia
| Contact Person: | Quynh Jewell |
|-----------------|----------------------------|
| Position: | Regulatory Affairs Manager |
| Company: | SDI Limited |
| Phone: | +61 3 8727 7111 |
| Email: | Quynh.Jewell@sdi.com.au |
Date Prepared: 9 September 2022
#### Device Details 2.
The Stela Automix System comprises the following components:
| Proprietary Name: | Stela Automix |
|--------------------|----------------------------------|
| Common Name: | Dental restorative material |
| Regulation Name: | Tooth shade resin material |
| Regulation Number: | 872.3690 |
| Product Code: | EBF |
| Regulatory Class: | II |
| Proprietary Name: | Stela Primer |
| Common Name: | Dental adhesive |
| Regulation Name: | Resin tooth bonding agent |
| Regulation Number: | 872.3200 |
| Product Code: | KLE |
| Regulatory Class: | II |
| Proprietary Name: | Mixing tip and application brush |
| Common Name: | Mixing tip and application brush |
| Regulation Name: | Resin applicator |
| Regulation Number: | 872.3140 |
| Product Code: | KXR |
| Regulatory Class: | I |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for SDI Limited. The logo consists of the letters "SDI" in a bold, sans-serif font, set against a black square. To the right of the square, the words "SDI Limited" are written in a smaller, regular font.
510(k) NOTIFICATION (Traditional): Stela Automix System – Summary – August 2022
#### ന് Predicate Device
| Submitter's Device | Primary Predicate Device | Secondary Predicate Device |
|--------------------|--------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Stela Primer | Name: ParaBond1<br>Manufacturer: Coltene<br>510(k) Number: K053040 | Name: Scotchbond Universal Adhesive2<br>Manufacturer: 3M ESPE<br>510(k) Number: K110302 |
| Stela Automix | Name: Fill-Up!<br>Manufacturer: Coltene<br>510(k) Number: K150218 | Name: Filtek Supreme Ultra Universal<br>Restorative<br>Manufacturer: 3M ESPE<br>510(k) Number: K083610 |
1 ParaBond was originally cleared in ParaCem Universal DC 510(k) based on the National Library of Medicine GUDID database noting ParaBond's 510(k) number as K053040. 2 Scotchbond Universal Adhesive was originally cleared as Adhesive EXL 759
No reference devices were used in this submission.
It is noted the mixing tip and application brush are exempt from the premarket notification procedure under Regulation 872.3140.
#### 4. Device Description
The Stela Automix System is an easy to use bulk-fill radiopaque self-cure dental restorative system for dental professional use. The system comprises:
- Stela Automix a double-barrel syringe containing two pastes which are mixed and . dispensed with a single-use mixing tip and
- Stela Primer - an adhesive liquid packaged in a bottle. A bendable disposable brush applicator is used to apply Stela Primer.
The combination of Stela Primer and Stela Automix will bond to both enamel and dentin without scrubbing or light curing.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for SDI Limited. The logo consists of the letters "SDI" in a bold, sans-serif font, set against a black square. To the right of the black square, the words "SDI Limited" are written in a smaller, less bold font.
510(k) NOTIFICATION (Traditional): Stela Automix System – Summary – August 2022
#### Indications of Use ട.
The indications for each component of the Stela Automix System are below.
Stela Primer
1) Dentin and enamel bonding
1) Direct restoration of Class I, II, III and V cavities. Ideally Class I and II,
2) Base or liner,
Stela Automix
- 3) Core build-ups
For Stela Primer, the indications are identical to the primary predicate. In comparison to the secondary predicate, Stela Primer's indications are phrased in a broad sense whereas the predicate specifies each use. This difference in wording is not expected to alter the intended use nor affect the safety or effectiveness of Stela Primer compared to the secondary
For Stela Automix, the indications are nearly identical to the primary predicate with the exception of the direct restoration of Class III and IV cavities which are indicated by the secondary predicate according to its Instructions for Use.
### Comparison of Technological Characteristics with the Predicate 6. Device
The Stela Automix System is based on well-established existing self-cure resin technology. The System's two components consist of a 1) combined primer plus bonding agent (Stela Primer) and 2) restorative bulk-fill resin-based material (Stela Automix).
Stela Primer binds to tooth structure through ionic bonds. Bonding of Stela Primer to Stela Automix occurs in situ upon contact via polymerisation.
At a high level, the Stela Automix System has the same technological characteristics as the predicate devices. Where differences exist these are as follows:
- . Stela Primer number of steps
- Similar to Scotchbond Universal Adhesive in that it consists of one bottle O containing the necessary elements to prime and bond: whereas ParaBond consists of three bottles: a conditioner, adhesive A and adhesive B.
- Stela Primer curing ●
- Self-cure like ParaBond; whereas Scotchbond Universal Adhesive is light O cured.
- Stela Automix material and curing
- radiopaque bulk-fill composite like both predicates. Stela Automix is self-O curing while Fill-Up and Filtek Supreme Ultra Universal Restorative are dualcure and light-cure, respectively.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for SDI Limited. The logo consists of the letters "SDI" in white on a black square, followed by the words "SDI Limited" in a smaller, sans-serif font. The text is aligned horizontally, with the black square on the left and the company name on the right.
510(k) NOTIFICATION (Traditional): Stela Automix System – Summary – August 2022
#### 7. Performance Data
## Biocompatibility testing
The Stela Automix System is considered an external communicating device in long term contact with tissue/bone/dentin. A biological risk assessment and testing were conducted to assess the biocompatibility of the Stela Automix System:
- Cytotoxicity tests (ISO 7405:2018, ISO 10993-5:2009) ●
- Delayed-type hypersensitivity (ISO 10993-10:2010)
- Irritation/intracutaneous reactivity (ISO 10993-10:2010) ●
- Systemic toxicity (ISO 10993-11:2017) ●
- Genotoxicity (ISO 10993-3:2014).
The results of the biocompatibility testing and risk assessment demonstrates a low potential for an unacceptable adverse biological response from contact of the component materials of the device with the body.
## Non-clinical performance data
Bench testing was conducted to determine the performance of the Stela Automix System compared to the primary and secondary predicate devices. Testing was conducted on the system where Stela Primer was used in conjunction with Stela Automix and compared to the equivalent system for the primary and/or secondary predicate device(s). The tests performed were:
- Working time (ISO 4049) ●
- . Setting time (ISO 4049)
- Flexural strength (ISO 4049) ●
- Shade (ISO 4049) ●
- Colour stability (ISO 4049) ●
- Radio-opacity (ISO 4049) ●
- Shear bond strength to dentin (ISO 29022:2013) .
- . Shear bond strength to enamel (ISO 29022:2013).
Bench testing demonstrated the Stela Automix System met the relevant ISO standard requirement, where applicable, and performed comparably to the predicate devices for the parameters tested.
## Clinical performance data
This section is not applicable.
## Electrical safety and electromagnetic compatibility (EMC)
This section is not applicable.
## Software verification and validation testing
This section is not applicable.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for SDI Limited. The logo consists of the letters "SDI" in a bold, sans-serif font on a black background. To the right of the black box, the words "SDI Limited" are written in a smaller, less bold font on a white background. The overall design is simple and professional.
510(k) NOTIFICATION (Traditional): Stela Automix System – Summary – August 2022
## Mechanical and acoustic testing
This section is not applicable.
#### Conclusion Regarding Substantial Equivalence 8.
A suite of bench and biocompatibility tests shows Stela Automix System performs comparably to the predicate devices (when used on a system basis) and possesses a low likelihood of an unacceptable adverse biological response from contact of the component materials of the device with the body. The Stela Automix System's indications for use are identical or very similar to the predicate devices and they are all based on the same restorative materials technology. On this basis, the Stela Automix System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.