Retrospective analysis of vital sign data from 794 patients enrolled in prospective clinical safety studies (2020-2024)
Retrospective analysis of clinical data was used to evaluate and demonstrate the effectiveness of the WARD-CSS alert reduction algorithms (time duration and alert filters) compared to standard threshold-based monitoring.
Retrospective Alert Reduction Analysis; Retrospective analysis of clinical data; Study Period: 2020-2024
Hospitalized medical and surgical patients; Sample Size: 794 patients
Standard patient monitors (threshold-only alerting)
Number of alerts generated by different methodologies (thresholds only vs. thresholds with time durations vs. WARD-CSS filters)
Indications for Use
WARD-CSS is a clinical decision support system that remotely integrates, analyzes and displays continuous vital sign data (via a mobile or web application) from medical devices for non-pediatric hospitalized patients within non-critical care units. WARD-CSS uses a set of standardized rules based on scientific and clinical evidence to detect and alert on clinically relevant vital sign deviations when used by trained health care professionals in hospitals. WARD-CSS is not intended to replace current monitoring practices or replace health care professionals' judgment. WARD-CSS is a tool intended to help health care professionals manage monitored patients and make clinical care decisions.
Device Story
WARD-CSS is a standalone clinical decision support software for non-critical care hospital units. It integrates continuous vital sign data (HR, SpO2, RR, Temp, BP) from medical devices. The system uses knowledge-based, threshold-driven algorithms to detect deviations (e.g., desaturation, tachycardia, hypotension). It features alert reduction methodologies: time duration requirements for thresholds and alert filters (reissue time and prioritization). The system consists of a mobile app for HCPs to receive notifications and document vitals, a web app for administrative management, and a backend server for data processing. It provides a real-time monitoring overview and alerts to assist HCPs in clinical management. It does not replace professional judgment or current monitoring practices; it aims to reduce alert fatigue compared to standard monitors.
Clinical Evidence
Retrospective analysis of 794 subjects from prospective clinical safety studies (2020-2024) across medical and surgical cohorts. Primary endpoint: alert reduction efficacy. WARD-CSS methodology (thresholds + time duration + alert filters) compared to threshold-only systems. Results: 97.8% total reduction in alerts (median 9.0 vs 417.0). Bench testing included software verification (IEC 62304) and human factors (IEC 62366-1).
Technological Characteristics
Standalone software; mobile/web application platform; backend server processing. Threshold-based knowledge-based algorithms. Connectivity: integrates data from medical devices. Standards: IEC 62304 (software), IEC 62366-1 (human factors).
Indications for Use
Indicated for non-pediatric hospitalized patients in non-critical care units. Used by trained healthcare professionals to monitor vital signs (heart rate, oxygen saturation, respiration rate, temperature, blood pressure) and receive alerts for clinically relevant deviations.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 14, 2025
WARD 24/7 ApS % Melissa DeHass Principal Consultant ROM+ 2790 Mosside Boulevard Suite 800 Monroeville, Pennsylvania 15146
Re: K241958
Trade/Device Name: WARD-CSS (v1.2.x) Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MWI Dated: July 3, 2024 Received: July 3, 2024
Dear Melissa Dehass:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer W. Shih -S
Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241958
Device Name WARD-CSS
#### Indications for Use (Describe)
WARD-CSS is a climical decision support system that remotely integrates, analyzes continuous vital sign data (via a mobile or web application) from medical devices for non-pediatric hospitalized patients within non-critical care units.
WARD-CSS uses a set of standardized rules based on scientific and clinical evidence to detect and alert on clinically relevant vital sign deviations when used by trained health care professionals in hospitals.
WARD-CSS is not intended to replace current monitoring practices or replace health care professionals' judgment. WARD-CSS is a tool intended to help health care professionals manage monitored patients and make clinical care decisions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
#### DATE PREPARED
December 20, 2024
## MANUFACTURER AND 510(k) OWNER
Ward 24/7ApS HealthTechHub Copenhagen Danneskiold-Samsøes Allé 41 1434 København K Denmark Official Contact: Madolina Christina Telephone: +45 53845636 Email: madolina.christian@ward247.com
### REPRESENTATIVE/CONSULTANT
Melissa DeHass Principal RQM+ Telephone: 717-476-0702 Email: mdehass(@rqmplus.com
## DEVICE INFORMATION
| Proprietary Name/Trade Name | WARD-Clinical Software Support (WARD-CSS) |
|-----------------------------|--------------------------------------------------------------------------|
| Common Name | Monitor, physiological, patient (with arrhythmia detection of<br>alarms) |
| Regulation Number | 870.2300 |
| Class | 2 |
| Product Code | MWI |
| Premarket Review | Traditional 510(k) |
| Review Panel | Cardiovascular |
### PREDICATE DEVICE IDENTIFICATION
WARD-CSS is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary<br>Predicate |
|---------------|---------------------------------------------------------------------------------------------|----------------------|
| K213335 | Capsule Surveillance System /<br>Capsule Surveillance Technologies, SAS /Capsule Tech, Inc. | ✓ |
{5}------------------------------------------------
## DEVICE DESCRIPTION
WARD-CSS is a stand-alone software intended for use in continuous monitoring of patients and near real-time analysis of vital signs for the purpose of notifying healthcare professionals in case of clinically relevant vital sign deviations.
WARD-CSS utilizes knowledge-based algorithms to evaluate clinically relevant vital signs deviations to help drive clinical management.
The system is intended to be used as an adjunct to current monitoring practice in the general med/surg floors of the hospitals
The system assists healthcare professionals when monitoring patients on their wards by:
- Providing a real-time monitoring overview of vital signs for all patients. ●
- Alerting the healthcare professionals when a patient deteriorates. ●
The following types of alerts are detected by WARD-CSS in the vital sign data:
- Desaturation ●
- Hypertension
- Hypotension ●
- Bradypnea ●
- Tachypnea ●
- Tachycardia
- Bradycardia
- Hypotension and Bradycardia
- Hypotension and Tachycardia ●
- Bradypnea and Desaturation
- Fever
The WARD-CSS consists of a Mobile App, Web App and Backend Server. The Mobile App is used by healthcare professionals (HCPs) to monitor patients. The HCP will receive notifications of the alerts to their mobile phones. Within this app, the HCP can also document vital signs into an electronic health record system. The Web App is used by administrative users to manage hospitals, wards, users, and monitors. The Backend Server is used to receive and process all incoming data and manage all data used in the apps.
## Intended use/Indications for use
WARD-CSS is a clinical decision support system that remotely integrates, analyzes and displays continuous vital sign data (via a mobile or web application) from medical devices for nonpediatric hospitalized patients within non-critical care units.
WARD-CSS uses a set of standardized rules based on scientific and clinical evidence to detect and alert on clinically relevant vital sign deviations when used by trained health care professionals in hospitals.
WARD-CSS is not intended to replace current monitoring practices or replace health care professionals' judgment. WARD-CSS is a tool intended to help health care professionals manage monitored patients and make clinical care decisions.
{6}------------------------------------------------
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
WARD 24/7 ApS believes that the WARD-CSS is substantially equivalent to the predicate device based on the information summarized here:
The subject device has a similar design and dimensions, and uses similar or identical materials, as the device cleared in K21335. The subject device has the same intended use and similar technological characteristics to the devices cleared in K213335, and the subject device has undergone testing to ensure the device is as safe and effective as the predicate.
| WARD-CSS<br>(Subject Device) | Capsule Surveillance System (K213335)<br>(Predicate Device) | Comments | |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | MWI | MWI | Same |
| Regulation | 870.2330 | 870.2330 | Same |
| Indications for Use | WARD-CSS is a clinical<br>decision support system<br>that remotely integrates,<br>analyzes and displays<br>continuous vital sign data<br>(via a mobile and web<br>application) from medical<br>devices for non-pediatric<br>hospitalized patients<br>within non-critical care<br>units.<br><br>WARD-CSS uses a set of<br>standardized rules based<br>on scientific and clinical<br>evidence to detect<br>clinically relevant vital<br>sign deviations (alerts)<br>when used by trained<br>health care professionals<br>in hospitals.<br><br>WARD-CSS is not<br>intended to replace health<br>care professionals'<br>judgment, but rather to<br>assist health care<br>professionals in making<br>timely, informed, higher<br>quality decisions.<br><br>WARD-CSS is not<br>intended to replace<br>current monitoring<br>practices but is intended<br>to help drive clinical | Capsule Surveillance<br>System is a clinical<br>decision support device<br>that integrates, analyzes,<br>and displays data from<br>multiple sources<br>including medical devices<br>and healthcare<br>information systems. It<br>uses standardized rules<br>that are based on<br>customer approved<br>clinical practices,<br>protocols and policies to<br>create clinically relevant<br>alerts in health care<br>facilities when used by<br>clinician physicians or<br>appropriate medical staff<br>under the direction of<br>physicians. It is not<br>intended to replace<br>clinicians' judgement, but<br>rather to assist clinicians<br>in making timely,<br>informed, higher, quality<br>decisions. Capsule<br>Surveillance may be<br>configured for secondary<br>monitoring and alerting<br>intended to be relied upon<br>in deciding to take<br>immediate clinical action.<br>Capsule Surveillance may<br>also be configured for | Similar |
| | WARD-CSS<br>(Subject Device) | Capsule Surveillance<br>System (K213335)<br>(Predicate Device) | Comments |
| | management. | remote display of<br>physiological data and<br>alerts not intended to be<br>relied upon in deciding to<br>take immediate clinical<br>action. | |
| Intended User | Healthcare professionals | | Same |
| Intended<br>Patient<br>Population | Hospitalized adult<br>patients (non-pediatric<br>population) hospitalized<br>patients within non-<br>critical departments. | Hospital patients | Same,<br>WARD-<br>CSS and<br>Capsule<br>Surveillance<br>System are<br>utilized by<br>adult<br>patients not<br>in the<br>intensive<br>care unit<br>(ICU). |
| Patient<br>Monitoring | Adjunctive | Adjunctive | Same |
| Vital Signs | Heart rate Oxygen saturation Respiration rate Temperature Blood pressure | Heart rate Oxygen saturation Respiration rate Temperature Blood pressure | Same |
| Platforms | Mobile App<br>Web App | Mobile App<br>Web App | Same |
| Algorithms | Threshold-based | Threshold-based | Same |
| Vital Sign<br>Monitoring<br>Resolution | 1-minute | | |
| Alerts | Direct detection from<br>vital sign data | Direct detection from<br>vital sign data | Same |
{7}------------------------------------------------
{8}------------------------------------------------
## SUMMARY OF NON-CLINICAL TESTING
The following tests were performed to demonstrate safety based on current industry standards:
- . Software Testing (IEC 62304)
- . Human Factors (IEC 62366-1)
The results of these tests indicate that WARD-CSS is substantially equivalent to the predicate device, the Capsule Surveillance System (K213335).
## SUMMARY OF CLINICAL TESTING
A literature review to support software algorithm development and determine the alert thresholds.
WARD-CSS reduces the number of alerts in comparison to standard patient monitors which alert immediately and repeatedly when a vital sign threshold is surpassed. (Standard patient monitors do not include a time duration component or any other alert filter).
WARD-CSS reduces the number of alerts in comparison to these standard patient monitors (indicating a basic threshold) by:
- 1) the introduction of a 'time duration' for a threshold to be surpassed within the alerting condition
- 2) additional 'alert filters' after the detection of the alerting conditions:
- 1. Reissue time filter: an alert of the same type is not allowed to trigger again within a specific time duration (in order to reduce the amount of alerts when a patient is continuously fulfilling an alerting condition).
- 2. Prioritization filter: an alert of lower priority is not allowed to trigger when an alert of higher priority is already active (this logic is only valid within the same alert groups like desaturation, bradycardia, tachycardia, hypertension, hypotension).
The combination of the '(alert)time duration' and 'alert filters', in addition to basic thresholds, theoretically lowers the number of alerts.
A retrospective analysis has been performed on four cohorts from prospective clinical safety studies conducted from 2020-2024 to measure the impact of these alert reduction methodologies.
The vital sign data of 1115 subjects (the number of subjects included in the WARD-CSS safety clinical studies) was primarily used to to confirm safety and not all data have been extracted for alert analysis.
Of the 1115 patients included in the safety studies, 794 patients were used to analyze the number of alerts. These patients are representative of both hospitalized medical and surgical cohorts.
Alert rates were analyzed for the three following cases:
- Thresholds only ●
- Thresholds with time durations .
- Thresholds with time durations and alert filters (WARD-CSS) .
In Table 1 below, the median and interquartile range for these cases are shown for all alert groups and each individual (bucketed) alerting group. A mean value is added to the results when the median is zero.
{9}------------------------------------------------
## Table 1. Alert Reduction Analysis
| Methodology | All alert groups | Hypertension + Hypotension | [Bradypnea + Tachypnea] | [Tachycardia + Bradycardia] | [Desaturation + Desaturation/ Bradypnea] | [Hypotension/Tachycardia + Hypotension/Bradycardia] |
|---------------------------------------------------------|----------------------------|-----------------------------|-------------------------|--------------------------------|------------------------------------------|-----------------------------------------------------|
| Only thresholds | 417.0<br>[177.9-<br>831.5] | 0.0 [0.0-1.0]<br>(mean 1.0) | 33.7<br>[6.8-157.9] | 11.9 [0.8-61.2]<br>(mean 76.3) | 256<br>[86.5-560.3] | 0.0 [0.0-0.0]<br>(mean 0.1) |
| Thresholds and time durations | 111<br>[26.5-338.3] | 0.0 [0.0-0.0]<br>(mean .04) | 4.7<br>[0.3-37.4] | 0.0 [0.0-1.9]<br>(mean 29.0) | 63.2<br>[9.2-227.6] | 0.0 [0.0-0.0]<br>(mean 0.0) |
| Thresholds, time durations and alert filters (WARD-CSS) | 9.0<br>[3.8-18.1] | 0.0 [0.0-0.0]<br>(mean 0.3) | 1.6<br>[0.3-4.9] | 0.0 [0.0-0.3]<br>(mean 1.0) | 4.7<br>[1.3 - 11.5] | 0.0 [0.0-0.0]<br>mean (0.0) |
The introduction of time durations gives a 74% reduction in alerts (over all alert types).
The alert filters reduce the alerts even further resulting in a total reduction of alerts by 97.8% by W ARD-CSS in comparison to patient monitors that alert only upon thresholds.
*WARD does not have any data within these studies on temperature, so it is not possible to calculate alert reduction numbers for fever (but the alert reduction follows the same methodologies).
### Summary
WARD-CSS reduces alert numbers compared to alert systems without- or with artefact removal.
{10}------------------------------------------------
## CONCLUSION
Based on the testing performed, including software verification, bench performance testing, and clinical testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The proposed WARD-CSS software and the Capsule Surveillance System, cleared under K213335 have the same intended use and indications, similar technological characteristics, and the same principles of operation. Thus, the proposed WARD-CSS software is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.