K213335 · Capsule Surveillance Technologies, Sas /Capsule Tech, Inc. · MWI · Jan 14, 2022 · Cardiovascular
Device Facts
Record ID
K213335
Device Name
Capsule Surveillance System
Applicant
Capsule Surveillance Technologies, Sas /Capsule Tech, Inc.
Product Code
MWI · Cardiovascular
Decision Date
Jan 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Capsule Surveillance is a clinical decision support device that integrates, analyzes, and displays data from multiple sources including medical devices and healthcare information systems. It uses standardized rules that are based on customers approved clinical practices, protocols, and policies to create clinically relevant alerts in health care facilities when used by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended to replace clinicians' judgment, but rather to assist clinicians in making timely, informed, higher quality decisions. Capsule Surveillance may be configured for secondary monitoring and alerting intended to be relied upon in deciding to take immediate clinical action. Capsule Surveillance may also be configured for remote display of physiological data and alerts not intended to be relied upon in deciding to take immediate clinical action.
Device Story
Clinical decision support system; integrates data from medical devices and healthcare information systems (EMR). Operates via standardized, user-defined clinical rules/protocols to generate alerts. Used in hospital settings by physicians and medical staff. Accessed via dedicated workstation (kiosk mode) or web browser (remote display). Provides multi-patient matrix views and single-patient detailed views. Assists clinicians in timely, informed decision-making; supports secondary monitoring for immediate clinical action or remote display of physiological data. Benefits include improved clinical oversight and workflow efficiency.
Clinical Evidence
No clinical data. Bench testing only. Evidence includes human factors/usability testing, software verification/validation per IEC 62304, and alarm system testing per IEC 60601-1-8.
Technological Characteristics
Clinical decision support software; integrates data via HL7 feeds. Operates on workstations or web-based platforms. Complies with ANSI AAMI IEC 62304 (software lifecycle) and ANSI AAMI IEC 62366-1 (usability). Rule-based alert generation based on customer-defined clinical protocols.
Indications for Use
Indicated for use by clinical physicians or appropriate medical staff under the direction of physicians in healthcare facilities to integrate, analyze, and display physiological data (e.g., pulse, respiratory rate, SpO2, temperature, blood pressure) for clinical decision support and secondary monitoring.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
Decisio Health, Inc.'s Decision Health Patient Dashboard (K142106)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 14, 2022
Capsule Surveillance Technologies, SAS /Capsule Tech, Inc. Peter Kelley Director Quality & Regulatory 300 Brickstone Square Suite 203 Andover, Massachusetts 01810
Re: K213335
Trade/Device Name: Capsule Surveillance System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: December 13, 2021 Received: December 15, 2021
Dear Peter Kelley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Capsule Surveillance System
Indications for Use (Describe)
Capsule Surveillance is a clinical decision support device that integrates, analyzes, and displays data from multiple sources including medical devices and healthcare information systems. It uses standardized rules that are based on customers approved clinical protocols, and policies to create clinically relevant alerts in health care facilities when used by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended to replace clinicians' judgment, but rather to assist clinicians in making timely, informed, higher quality decisions. Capsule Surveillance may be configured for secondary monitoring and alerting intended to be relied upon in deciding to take immediate clinical action.
Capsule Surveillance may also be configured for remote display of physiological data and alerts not intended to be relied upon in deciding to take immediate clinical action.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | | | |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|---------------------|
| I. SUBMITTER | | | |
| Date Prepared | October 1st, 2021 | | |
| Submitter/Owner | Capsule Technologies, SAS / Capsule Tech, Inc.<br>FDA Establishment Number: 3003630387<br>300 Brickstone Square<br>Suite 203<br>Andover, MA 01810 USA<br>Phone: 978-697-4364 | | |
| Key Contact | Peter Kelley<br>Director Quality & Regulatory<br>peter.kelley@philips.com | | |
| 510(k) Submission Type | This is a traditional 510(k). | | |
| II. DEVICE | | | |
| Trade Name | Capsule Surveillance System | | |
| Common Name | Physiological Monitor | | |
| Classification Name | Panel & Name: Cardiovascular (OHT2)<br>Subpart & Division: Cardiac Electrophysiology, Diagnostics, and<br>Monitoring Devices (DHT2A)<br>21 CFR 870.2300<br>Regulatory Class: Class II<br>Product Code: MWI | | |
| III. PREDICATE DEVICE | | | |
| | 510(k) No. | Company Name<br>Device Name | Product<br>Code |
| Predicate Device | K142106 | Decisio Health, Inc.'s Decision Health<br>Patient Dashboard | MWI |
| Reference Device | K130208 | Cardiopulmonary Corp.'s Bernoulli<br>Enterprise Software | MHX,<br>MSX,<br>PFY |
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
## Capsule Surveillance System - description of the device per 21 CFR 807.92(a) (4)
The Capsule Surveillance System is a clinical decision support device that integrates data from multiple sources including medical devices and healthcare information systems.
The Capsule Surveillance Systems offers two methods of accessing the clinical application:
- . A workstation which is a dedicated desktop used exclusively to run the Capsule Surveillance System clinical application in kiosk mode.
- . The remote display which is any machine that accesses the Capsule Surveillance System via a web browser.
Use of the Capsule Surveillance System allows for simultaneous monitoring of different patients in a clinical setting, providing information from integrated data sources that monitor physiological parameters such as, pulse, respiratory rate, SpO2, temperature, and blood pressure.
#### V. INDICATIONS FOR USE
## Intended Use as required per 21 CFR 807.92(a)(5)
The Capsule Surveillance is a clinical decision support device that integrates, and displays data from multiple sources including medical devices and healthcare information systems. It uses standardized rules that are based on customers approved clinical practices, protocols, and policies to create clinically relevant alerts in health care facilities when used by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended to replace clinicians' judgment, but rather to assist clinicians in making timely, informed, higher quality decisions.
Capsule Surveillance may be configured for secondary monitoring and alerting intended to be relied upon in deciding to take immediate clinical action.
Capsule Surveillance may also be configured for remote display of physiological data and alerts not intended to be relied upon in deciding to take immediate clinical action.
### Comparison of Intended Uses for Subject Device and Predicate
| Name | Indications for Use/Intended Use |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Capsule<br>Surveillance<br>System<br>Subject Device | The Capsule Surveillance is a clinical decision support device that integrates,<br>analyzes, and displays data from multiple sources including medical devices<br>and healthcare information systems. It uses standardized rules that are based<br>on customers approved clinical practices, protocols, and policies to create<br>clinically relevant alerts in health care facilities when used by clinical<br>physicians or appropriate medical staff under the direction of physicians. It is |
{5}------------------------------------------------
| | not intended to replace clinicians' judgment, but rather to assist clinicians in<br>making timely, informed, higher quality decisions. |
|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Capsule Surveillance may be configured for secondary monitoring and<br>alerting intended to be relied upon in deciding to take immediate clinical<br>action. |
| | Capsule Surveillance may also be configured for remote display of<br>physiological data and alerts not intended to be relied upon in deciding to take<br>immediate clinical action. |
| K142106<br>Decisio Health<br>Patient<br>Dashboard<br>Predicate | The Decisio Health Patient Dashboard is a decision support device indicated<br>for aggregating, displaying, and managing physiologic and other patient<br>information. This information is generated by third party medical devices and<br>patient information systems. The device performs automated calculations on<br>patient data collected by third party devices based on approved clinical<br>protocols at patient care facilities. The Decisio Health Patient Dashboard is<br>intended for use by clinicians in healthcare facilities. |
| <b>VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE</b> | |
|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Similarities</b> | |
| Item of Comparison | Description/Rationale |
| Target Populations | Both devices are intended to be used by trained clinicians. |
| Environment of Use | Both devices are intended to be used in a hospital setting. |
| Matrix and Detailed Views | Both devices allow for matrix or multi-patient simultaneous monitoring, as well as detailed views/monitoring of single patients. |
| Data Input | Both devices use input from EMR systems as well as medical devices via HL7 feed for monitoring |
| <b>Differences</b> | |
| Item of Comparison | Description/Rationale |
| Methods for Accessing the Application | The subject Capsule Surveillance System allows access via a workstation or remotely via a web browser. The predicate Decisio Health Patient Dashboard allows access through a computer, while also permitting access through a patient monitor or mobile device. |
# Substantial Equivalence Summary
Operational and technological characteristics form the basis for the determination of substantial equivalence of the Capsule Surveillance System with the legally marketed predicate devices (K142106). The Capsule Surveillance System is substantially equivalent to the predicate devices.
{6}------------------------------------------------
#### PERFORMANCE DATA VII.
### Non-Clinical Tests - Harmonized Standards
The Capsule Surveillance System has passed all safety tests for demonstrated compliance with the harmonized standards below.
| Standard | FDA Recognition # | Title # |
|--------------------------------------------------------------------|-------------------|------------------------------------------------------------------------------------------------------------|
| ANSI AAMI IEC<br>62304:2006/A1:2016 | 13-79 | Medical device software – Software life cycle processes [Including Amendment 1<br>(2016)] |
| ANSI AAMI IEC<br>62366-<br>1:2015+AMDI:2020<br>(Consolidated Text) | 5-129 | Medical devices - Part 1 Application of usability engineering to medical devices,<br>including Amendment 1 |
# Non-clinical Bench Tests
Human factors and usability testing has been completed. Furthermore, testing in compliance with IEC 60601-1-8 for requirements for alarm systems in medical electrical equipment and medical electrical systems was completed, as well as certification to IEC 62304 for software lifecycle and IEC 62366.
No new issues of safety or effectiveness are introduced as a result of using this device.
# Clinical Studies
The Capsule Surveillance System, like the predicate device, did not require clinical trials.
FDA recognized standards, FDA guidance documents, harmonized standards, verification and validation, software validation, usability validation, and risk management activities have taken place for the Capsule Surveillance System.
Based upon the design, intended use, indications for use, classification, usability and safety testing the Capsule Surveillance System is substantially equivalent to the listed predicate device.
No new issues of safety or effectiveness are introduced as a result of using this device.
#### CONCLUSIONS VIII.
The results of the substantial equivalence assessment, taken together with non-clinical bench testing, electrical safety and electromagnetic compatibility, software verification and validation,
{7}------------------------------------------------
human factors and usability demonstrate that the Capsule Surveillance System does not raise additional questions of safety and effectiveness when compared to the predicate, performs as intended, and has performance characteristics that are substantially equivalent to the Decision Health Patient Dashboard predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.