Training: 3723 cases (80% of total dataset). Validation: 378 cases (10% of total dataset).
—
AI performance testing: 728 cases total (including 77 cases for this module).
—
Indications for Use
Myomics is intended to be used for viewing, post-processing, qualitative evaluation of cardiovascular magnetic resonance (MR) images in a Digital Imaging and Communications in Medicine (DICOM) standard format. It enables a set of tools to assist physicians in qualitative assessment of cardiac images and quantitative measurements of the heart and adjacent vessels; and to view the presence or absence of physician-identified lesion in blood vessels. The target population for manual workflows of Myomics is not restricted; however, semi-automated machine learning algorithms of Myomics are intended for an adult population. The software comprises various analysis modules, including AI-powered algorithms, for a comprehensive evaluation of MR images. Myomics is used for cardiac images acquired from a 3.0 T MR scanner. Myomics shall be used only for cardiac images acquired from an MR scanner. It shall be used by qualified medical professionals, experienced in examining cardiovascular MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process.
Device Story
Myomics is a software application for analyzing cardiovascular MR images in DICOM format; functions as a standalone product or integrated into hospital/private practice environments. Inputs: cardiovascular MR images (3.0 T). Processing: software provides tools for qualitative assessment and quantitative measurements of heart/vessels; includes AI-powered semi-automated segmentation modules for myocardium (Native T1, Post T1, T2, CINE, LGE PSIR, CINE RV, LGE Magnitude). Output: processed images, segmentations, and measurements displayed via GUI. Used by qualified medical professionals to assist in diagnostic decision-making; enables visualization of lesions and cardiac function. Benefits: provides automated/semi-automated analysis to support clinical evaluation of heart disease.
Clinical Evidence
Bench testing only. No clinical data. Performance validated using 728 anonymized patient cases across multiple MR vendors. AI modules (Native T1, Post T1, T2, CINE, LGE PSIR, CINE RV, LGE Magnitude) evaluated using DICE score metric; all modules achieved average DICE score > 0.7. Software verification and validation conducted per ISO 13485, IEC 62304, and ISO 14971.
Technological Characteristics
Software-based medical image management and processing system. Operates on Microsoft Windows. Connectivity: DICOM standard networking. Standards: IEC 62304:2006+A1:2015 (software development), ISO 14971:2019 (risk management). Features semi-automated machine learning algorithms for myocardium segmentation. Supports file formats: dcm, json, nii, csv.
Indications for Use
Indicated for viewing, post-processing, and qualitative/quantitative evaluation of cardiovascular MR images. Intended for use by qualified medical professionals for diagnostic decision-making. Semi-automated machine learning algorithms are intended for an adult population. Used for cardiac images acquired from 3.0 T MR scanners.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 28, 2025
Phantomics Inc. Soeun Baek Manager 152, Magokseo-ro, Gangseo-gu A-609 Seoul, 07788 Korea, South
Re: K241922
Trade/Device Name: Myomics Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH, LLZ Dated: January 21, 2025 Received: January 21, 2025
Dear Soeun Baek:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Samuel for
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241922
Device Name Myomics
### Indications for Use (Describe)
Myomics is intended to be used for viewing, post-processing, qualitative evaluation of cardiovascular magnetic resonance (MR) images in a Digital Imaging and Communications in Medicine (DICOM) standard format. It enables a set of tools to assist physicians in qualitative assessment of cardiac images and quantitative measurements of the heart and adjacent vessels; and to view the presence or absence of physician-identified lesion in blood vessels. The target population for manual workflows of Myomics is not restricted; however, semi-automated machine learning algorithms of Myomics are intended for an adult population.
The software comprises various analysis modules, including AI-powered algorithms, for a comprehensive evaluation of MR images.
Myomics is used for cardiac images acquired from a 3.0 T MR scanner.
Myomics shall be used only for cardiac images acquired from an MR scanner. It shall be used by qualified medical professionals, experienced in examining cardiovascular MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "PHANTOMICS". The logo consists of a circular icon with a stylized human face inside, positioned to the left of the company name. The text "PHANTOMICS" is written in a bold, sans-serif font, and the entire logo is colored in a bright blue hue.
# K241922
Image /page/4/Picture/3 description: The image shows a logo for "PHANTOMICS". The logo consists of the word "PHANTOMICS" in blue, arranged vertically along the left side of a blue circle. Inside the circle is a stylized face with closed eyes, a nose, and a smiling mouth, all rendered in the same blue color as the circle and text. The overall design is simple and clean, with a focus on the company name and a minimalist representation of a face.
The following 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Device Act 1990 and 21 CFR 807.92(c).
Date: May 21, 2024
#### l. SUBMITTER
- Submitter Name: Phantomics Inc.
- Address: A-609, 152, Magokseo-ro, Gangseo-gu, Seoul, Republic of Korea
- Data Prepared: May 21 2024
- I Telephone Number: +82-10-5755-2840
- I Contact Person: Soeun Baek
- Email: soeunbaek@phantomics.io
#### II. DEVICE
- Device's Trade Name: Myomics
- Device's Common Name: Automated Radiological Image Processing Software
- I Regulation Number: 21CFR 892.2050
- Primary Product Code: QIH
- Secondary Product Code: LLZ
- Device Class: Class II (special controls; voluntary standards-Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern)
- I Classification Panel: Radiology
#### lll. PREDICATE DEVICE
The primary predicate is Myomics Q (K211432) manufactured by Phantomics Inc. The cvi42 Auto Imaging Software Application (K213998), manufactured by Circle Cardiovascular Imaging Inc. is used as a secondary predicate device. The predicate devices have not been subject to design-related recall.
# IV. SUBJECT DEVICE DESCRIPTION
Myomics is a software application for analysis cardiovascular MR images in DICOM Standard format. The software can be used as a stand-alone product that can be integrated into a hospital or private practice environment. This device has a graphical user interface which allows users to analyze cardiovascular MR images qualitatively and quantitatively.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for "PHANTOMICS". The logo consists of a stylized head with three curved lines above it, followed by the word "PHANTOMICS" in blue, sans-serif font. The logo is simple and modern.
# V. INTENDED USE
Viewing, post-processing, qualitative and quantitative evaluation of blood vessels and cardiovascular MR images in DICOM format.
# VI. INDICATIONS FOR USE
The Indications for Use statement for the Myomical to the predicate device; the differences do not alter the intended use of the device, nor do they affect the safety and effectiveness of the device relative to the predicate.
| Table 1. Comparison to the predicate devices | | | |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Primary Device | Reference Device |
| | <i>Myomics</i> | <i>Myomics Q</i> (K211432) | <i>cvi42 Auto</i> (K213998) |
| | Manufactured by Phantomics | Manufactured by Phantomics | Manufactured by Circle |
| Indication<br>for use | <i>Myomics</i> is intended to be used for viewing, post-<br>processing, qualitative and quantitative evaluation<br>of cardiovascular magnetic resonance (MR)<br>images in a Digital Imaging and Communications<br>in Medicine (DICOM) standard format. | <i>Myomics Q</i> is intended to be used for viewing,<br>post-processing and analysis of cardiac magnetic<br>resonance (MR) images in a Digital Imaging and<br>Communications in Medicine (DICOM) Standard<br>format. It enables: | <i>cvi42 Auto</i> is intended to be used for<br>viewing, post-processing, qualitative and<br>quantitative evaluation of cardiovascular<br>magnetic resonance (MR) and computed<br>tomography (CT) images in a Digital Imaging<br>and Communications in Medicine (DICOM)<br>Standard format. |
| | It enables a set of tools to assist physicians in<br>qualitative assessment of cardiac images and<br>quantitative measurements of the heart and<br>adjacent vessels; and to view the presence or<br>absence of physician-identified lesion in blood<br>vessels. | - Importing cardiac MR images in DICOM format.<br>- Supporting clinical diagnostics by analysis of<br>cardiac MR images using display functionality such<br>as panning, windowing, zooming through<br>series/slices of the images.<br>- Supporting clinical diagnostics analysis of the<br>heart in cardiac MR images and signal intensity.<br>- Software package is designed to support the<br>physician-to-physician compliance assessment,<br>document and follow up heart disease by cardiac<br>MRI. | It enables a set of tools to assist physicians in<br>qualitative assessment of cardiac images and<br>quantitative measurements of the heart and<br>adjacent vessels; perform calcium scoring;<br>and to confirm the presence or absence of<br>physician-identified lesion in blood vessels. |
| | The target population for manual workflows of<br><i>Myomics</i> is not restricted; however, semi-<br>automated machine learning algorithms of<br><i>Myomics</i> are intended for an adult population. | It shall be used by qualified medical professionals,<br>experienced in examining and evaluating<br>cardiovascular MR images, for the purpose of<br>obtaining diagnostic information as part of a<br>comprehensive diagnostic decision-making process. | The target population for <i>cvi42 Auto's</i><br>manual workflows is not restricted; however,<br><i>cvi42 Auto's</i> semi-automated machine<br>learning algorithms are intended for an adult<br>population. |
| | The software comprises various analysis modules,<br>including Al-powered algorithms, for a<br>comprehensive evaluation of MR images. | This device is a software application that can be<br>used as a stand-alone product or in a network<br>environment. | <i>cvi42 Auto</i> shall be used only for cardiac<br>images acquired from an MR or CT scanner.<br>It shall be used by qualified medical<br>professionals, experienced in examining and<br>evaluating cardiovascular MR or CT images, |
| | <i>Myomics</i> is used for cardiac images acquired<br>from a 3.0 T MR scanner. | | |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for "PHANTOMICS". The logo consists of a circular icon with three stylized figures inside, positioned above the text "PHANTOMICS". The text is in a sans-serif font and is colored in blue. The logo appears to be for a company or organization named "PHANTOMICS".
| | Subject Device | Primary Device | Reference Device |
|---|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| - | <i>Myomics</i><br>Manufactured by Phantomics | <i>Myomics Q</i> (K211432)<br>Manufactured by Phantomics | <i>cvi42 Auto</i> (K213998)<br>Manufactured by Circle |
| | Myomics shall be used only for cardiac images<br>acquired from an MR scanner. It shall be used by<br>qualified medical professionals, experienced in<br>examining and evaluating cardiovascular MR<br>images, for the purpose of obtaining diagnostic<br>information as part of a comprehensive diagnostic<br>decision-making process. | The target population for the device is not<br>restricted, however the image acquisition by a<br>cardiac MR scanner may limit the use of the device<br>for certain sectors or the public. | for the purpose of obtaining diagnostic<br>information as part of a comprehensive<br>diagnostic decision-making process. |
# VII. SUBSTANTIAL EQUIVALENCE
| - | Subject Device | Primary Device | Reference Device |
|-------------------|--------------------------------------------|--------------------------------------------|--------------------------------------------|
| Device Name | Myomics | Myomics Q | cvi42 Auto Imaging Software Application |
| 510(k) Number | - | K211432 | K213998 |
| Manufacturer | Phantomics Inc. | Phantomics Inc. | Circle Cardiovascular Imaging Inc. |
| Regulation Number | 21CFR892.2050 | 21CFR892.2050 | 21CFR892.2050 |
| Regulation Name | Picture Archiving and Communication System | Picture Archiving and Communication System | Picture Archiving and Communication System |
| Classification | Class II | Class II | Class II |
| Product Code | QIH | LLZ | QIH |
# Table 2. General comparison to the predicate devices
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for the company "PHANTOMICS". The logo consists of a stylized face above the company name. The text is in a sans-serif font and is colored blue.
| Table 3. Feature comparison table of Myomics with the predicate devices | | | |
|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| - | Subject Device | Primary Device | Reference Device |
| | Myomics<br>Manufactured by Phantomics | Myomics Q (K211432)<br>Manufactured by Phantomics | cvi42 Auto (K213998)<br>Manufactured by Circle |
| Imaging Modalities | MR | MR | MR and CT |
| Imaging Type | Cardiovascular | Cardiovascular | Cardiovascular |
| DICOM Compliant<br>Networking | Yes | Yes | Yes |
| Import and Display<br>MR images | Yes | Yes | Yes |
| Images can be<br>displayed by study<br>and series | Yes | Yes | Yes |
| Segmentation of<br>regions of interest | Yes (Endocardium and epicardium contour<br>segmentation) | Yes (Endocardium and epicardium contour<br>segmentation) | Yes (Endocardium and epicardium contour<br>segmentation) |
| Store images | Yes (File Format: dcm, json, nii, csv) | Yes (File Format: dcm, json, nii, csv) | Yes (File Format: PDF, XML) |
| Machine Learning<br>Based Algorithm | Yes (Semi-automatic segmentation) | No | Yes (Semi-automatic segmentation) |
| Quantitative<br>Assessment of<br>Cardiac Function | Yes (Manual segmentation, and semi-<br>automatic segmentation using Machine<br>Learning technique of four heart chambers in<br>short-axis views) | Yes (Manual segmentation) | Yes (Manual segmentation, and semi-<br>automatic segmentation using Machine<br>Learning technique of four heart chambers in<br>long and short-axis views) |
| Operating System | Microsoft Windows | Microsoft Windows | Mac OS and Microsoft Windows |
| Software<br>Development<br>Standard | IEC 62304:2006+A1:2015 | IEC 62304:2006+A1:2015 | IEC 62304:2006+A1:2015 |
| Risk Management<br>Standard | ISO 14971:2019 | ISO 14971:2007 | ISO 14971:2019 |
| | | | | Table 3. Feature comparison table of Myomics with the predicate devices | |
|--|--|--|--|-------------------------------------------------------------------------|--|
| | | | | | |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for "PHANTOMICS". The logo consists of a stylized head with three curved lines above the word "PHANTOMICS". The word "PHANTOMICS" is written in a sans-serif font and is in blue. The logo is simple and modern.
# VIII. PERFORMANCE DATA AND TESTING
### 1) Performance Test
A bench performance test was conducted to verify the substantial equivalence in performance between the subject device (Myomics) and the predicate device (Myomics Q). These devices analyze cardiovascular MR images and have general worklist functions such as importing, exporting images, manually drawing contours, and reporting. It was also conducted to compare the segmentation function (endocardium and epicardium contour) between the subject device (Myomics) and the secondary predicate device (cvi42 auto). The bench performance test confirmed the substantial equivalence of the subject device (Myomics) to both predicate devices (cvi42 auto and Myomics Q).
### 2) Verification and validation
Software verification and validation of Subject device (Myomics) has been tested according to "ISO 13485:2016', "IEC 62304:2015" and "ISO 14971:2019" .and all documents have been made according to the Software Development Plan of Myomics.
### 3) Validation of AI Modules
The machine learning algorithms of Myomics were trained and tested using images from various major MR imaging device vendors. The data used for AI performance testing was not utilized during the algorithm training process. The training involved a dataset of 3723 anonymized cases, separate from the test set. This dataset was divided into training, validation, and test sets at a ratio of 80%, 10%, and 10% respectively, yielding 3723 cases for training and 378 cases each for validation and AI performance testing.
The AI performance testing dataset originally contained 378 test cases. To improve generalizability and evaluate performance across different MRI manufacturers, 50 additional cases were included per AI module, increasing the total number of test cases from 378 to 728.
| AI Module No. | Data Group (AI Model Name) | Training | Validation | Test |
|---------------|---------------------------------------|------------|------------|-----------|
| AIM-01 | Native T1 Map Myocardium Segmentation | 594 cases | 42 cases | 92 cases |
| AIM-02 | Post T1 Map Myocardium Segmentation | 498 cases | 41 cases | 91 cases |
| AIM-03 | T2 Map Myocardium Segmentation | 586 cases | 59 cases | 109 cases |
| AIM-04 | CINE Myocardium Segmentation | 640 cases | 40 cases | 90 cases |
| AIM-05 | LGE PSIR Myocardium Segmentation | 491 cases | 27 cases | 77 cases |
| AIM-06 | CINE RV Myocardium Segmentation | 437 cases | 142 cases | 192 cases |
| AIM-07 | LGE Magnitude Myocardium Segmentation | 477 cases | 27 cases | 77 cases |
| SUM | | 3723 cases | 378 cases | 728 cases |
#### Table 4. AI Module Description in Myomics
Across all MR machine manufacturers n= 728 anonymized patient images were used for the AI performance test of Myomics. This translates into 92 cases for Native T1 Map Myocardium Segmentation, 91 cases for Post T1 Map Myocardium Segmentation, 109 cases for T2 Map Myocardium Segmentation, 90 cases for CINE Map Myocardium Segmentation, 77 cases for LGE PSIR Myocardium Segmentation, 192 cases for CINE RV Myocardium Segmentation, and 77 cases for LGE Magnitude Myocardium Segmentation. All AI performance testing results met pre-defined acceptance criteria in Phantomics.
The AI performance acceptance criteria, defined using the DICE Score, were applied to evaluate the ML model's effectiveness in segmenting the Myocardium. The DICE Score is a common metric used in computer vision and medical imaging to measure the similarity or overlap between two sets. It is typically used in image segmentation and object detection. Upon evaluating the performance of each AI module used in Myomics, all modules attained an average DICE Score of over 0.7. All data and results from the AI performance test have been documented in accordance with quality system of Phantomics.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Phantomics. The logo consists of a stylized head in a circle above the word "PHANTOMICS" in blue, sans-serif font. The head is facing left and has a few lines indicating facial features.
# IX.CONCLUSION
In comparing substantial equivalence between the subject device and the predicate device, similarities are found in general, technical and material information. Despite some differences, the safety and performance test reports affirm the safety and effectiveness of the subject device. Therefore, we conclude that the subject device is substantially equivalent to the predicate device. The information provided in this premarket notification, which includes performance testing and comparisons with the predicate devices, attests to the safety and effectiveness of Myomics for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.