K211432 · Phantomics, Inc. · LLZ · Nov 23, 2021 · Radiology
Device Facts
Record ID
K211432
Device Name
Myomics Q
Applicant
Phantomics, Inc.
Product Code
LLZ · Radiology
Decision Date
Nov 23, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Myomics Q is intended to be used for viewing, post-processing and analysis of cardiac magnetic resonance (MR) images in a Digital Imaging and Communications in Medicine (DICOM) Standard format. It enables: - Importing cardiac MR images in DICOM format. - Supporting clinical diagnostics by analysis of cardiac MR images using display functionality such as panning, windowing, zooming through series/slices of the images. - Supporting clinical diagnostics analysis of the heart in cardiac MR images and signal intensity. - Software package is designed to support the physician compliance assessment, document and follow up heart disease by cardiac MRI. It shall be used by qualified medical professionals, experienced in examining cardiovascular MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. This device is a software application that can be used as a stand-alone product or in a network environment.
Device Story
Myomics Q is a software application for qualitative and quantitative analysis of cardiac MR images. It imports DICOM-formatted cardiac MR images; provides tools for viewing (panning, windowing, zooming) and manual contouring (endocardium/epicardium) for T1, T2, and Late Gadolinium Enhancement (LGE) analysis. Operated by qualified medical professionals in hospital or private practice settings, the device functions as a stand-alone or networked application. Users perform manual contouring to generate polar map reports for T1 and T2 analysis. The output assists physicians in assessing cardiac function and heart disease. By providing standardized visualization and quantitative measurement tools, the device supports clinical decision-making and patient follow-up.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was validated by comparing Myomics Q against reference device cvi42. Testing included verification of DICOM import/export, patient data display, and contouring tools. Quantitative analysis of Native T1, Post T1, and T2 polar maps showed results within ±5% of cvi42, attributed to user-dependent contouring variability. Software validation followed IEC 62304 and FDA cybersecurity guidance.
Technological Characteristics
Software-only medical device (SaMD). Operates on Windows OS. Supports DICOM standard for image input/output. Features include manual contouring tools (endocardium, epicardium, move, pinch, nudge, freehand), windowing, panning, and zooming. Connectivity via network or stand-alone. No hardware components. Software validation conducted per IEC 62304:2006+A1:2015.
Indications for Use
Indicated for viewing, post-processing, and analysis of cardiac MR images in DICOM format. Intended for use by qualified medical professionals experienced in cardiovascular MR image evaluation to support clinical diagnostics, compliance assessment, documentation, and follow-up of heart disease. No specific age or gender restrictions; use may be limited by cardiac MR scanner acquisition capabilities.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
Phantomics Inc. % Milly Milly Regulatory Affairs Consultant KMC, Inc. Room no. 1709, 123, Digital-ro 26-gil, Guro-gu Seoul, 08390 REPUBLIC OF KOREA
Re: K211432
Trade/Device Name: Myomics Q Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: September 13, 2021 Received: October 4, 2021
Dear Milly Milly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211432
Device Name Mynomics Q
Myomics O is intended to be used for viewing, post-processing and analysis of cardiac magnetic resonance (MR) images in a Digital Imaging and Communications in Medi-cine (DICOM) Standard format. It enables:
- Importing cardiac MR images in DICOM format.
- Supporting clinical diagnostics by analysis of cardiac MR images using display functionality such as panning, windowing, zooming through series/slices of the images. *
- Supporting clinical diagnostics analysis of the heart in cardiac MR images and signal intensity.
- Software package is designed to support the physician compliance assessment, document and follow up heart disease by cardiac MRI.
It shall be used by qualified medical professionals, experienced in examining cardiovascular MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. This device is a soft-ware application that can be used as a stand-alone product or in a network environment.
The target population for the device is not restricted, however the image acquisition by a cardiac MR scanner may limit the use of the device for certain sectors or the public.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Phantomics. The logo consists of a stylized head with a closed eye above the word "PHANTOMICS" in blue, sans-serif font. The logo is simple and modern.
# 510(k) SUMMARY
This summary of 510(k) for K211432 –safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: Sept 13, 2021
### 1. INFORMATION
1.1 Submitter Information
- = Submitter Name: Phantomics Inc.
- 트 Address
- : B-1224, 247, Gonghang-daero, Gangseo-gu, Seoul, 07803, Republic of Korea
- . ■ Telephone Number: +82-2-6347-2019 Fax: +82-504-329-2510
### 1.2 Contact Person
■
- . Name: Milly (Consultant / KMC, Inc.)
- Address: Rm. No. 1709, 123, Digital-ro 26-gil, Guro-gu, 08390, Republic of Korea
- Telephone Number: +82-70-8965-5554 " Fax: +82-2-2672-0579
- . E-mail: milly@kmcerti.com
### 2. DEVICE INFORMATION
- 2.1 Trade Name / Proprietary Name: Myomics Q
- 2.2 Common Name: Image Processing System
- 2.3 Classification Name: Medical Image Management and Processing System
- 2.4 Product Code: LLZ
- 2.5 Classification Regulation: 21CFR 892.2050
- 2.6 Device Class: Class II
(special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern)
- 2.7 Classification Panel: Radiology
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Phantomics. The logo consists of a circular icon with a stylized figure inside, above the word "PHANTOMICS" in a sans-serif font. The color of the logo is a light blue.
### 3. PREDICATE DEVICE
| Item | Predicate Device | Reference Device |
|--------------------------|------------------------------------|------------------------------------|
| Manufacturer | Circle Cardiovascular Imaging Inc. | Circle Cardiovascular Imaging Inc. |
| Device Name (Trade Name) | cmr42 | cvi42 |
| 510(k) Number | K082628 | K141480 |
### 4. SUBJECT DEVICE DESCRIPTION
Myomics Q is software application for evaluating cardiovascular images in a DICOM Standard format. The software can be used as a stand-alone product that can be integrated into a hospital or private practice environment. This device has a graphical user interface which allows users to analyze cardiac MR Images qualitatively and quantitatively.
### 5. INDICATIONS FOR USE
Myomics Q is intended to be used for viewing, post-processing and analysis of cardiac magnetic resonance (MR) images in a Digital Imaging and Communications in Medicine (DICOM) Standard format. It enables:
- Importing cardiac MR images in DICOM format. .
- . Supporting clinical diagnostics by analysis of cardiac MR images using display functionality such as panning, windowing, zooming through series/slices of the images. *
- Supporting clinical diagnostics analysis of the heart in cardiac MR images and signal . intensity.
- Software package is designed to support the physician compliance assessment, . document and follow up heart disease by cardiac MRI.
It shall be used by qualified medical professionals, experienced in examining and evaluating cardiovascular MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. This device is a software application that can be used as a stand-alone product or in a network environment.
The target population for the device is not restricted, however the image acquisition by a cardiac MR scanner may limit the use of the device for certain sectors or the public.
{5}------------------------------------------------
# SUBSTANTIAL EQUIVALENCI ್ರೆ
| | Subject Device | Primary Device | Reference Device | Equivalence<br>Result |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Manufacturer | Phantomics Inc. | Circle Cardiovascular Imaging Inc. | Circle Cardiovascular Imaging Inc. | |
| Device Name | Myomics Q | cmr42 | cvi42 | Unique<br>Information |
| 510(k) Number | K211432 | K082628 | K141480 | |
| Regulation Number | 21CFR892.2050 | 21CFR892.2050 | 21CFR892.2050 | <span style="text-align: center;">☑ Same</span><br>☐ Similar<br>☐ Different |
| Regulation Name | Medical Image Management and<br>Processing System | Medical Image Management and<br>Processing System | Medical Image Management and<br>Processing System | <span style="text-align: center;">☑ Same</span><br>☐ Similar<br>☐ Different |
| Classification | Class II | Class II | Class II | <span style="text-align: center;">☑ Same</span><br>☐ Similar<br>☐ Different |
| Product Code | LLZ | LLZ | LLZ | <span style="text-align: center;">☑ Same</span><br>☐ Similar<br>☐ Different |
| Indication for use | Myomics Q is intended to be used for<br>viewing, post-processing and analysis<br>of cardiac magnetic resonance (MR)<br>images in a Digital Imaging and<br>Communications in Medicine<br>(DICOM) Standard format. It enables:<br>• Importing cardiac MR<br>images in DICOM format. | cmr 42 is intended to be used for<br>viewing, post-processing and analysis<br>of cardiac magnetic resonance (MR)<br>images in a Digital Imaging and<br>Communications in Medicine<br>(DICOM) Standard format.<br>It enables; | Cvi42 vascular analysis add-on is an<br>image analysis software package add-<br>on for evaluating CT and MR images<br>of blood vessels. Combining digital<br>image processing and visualization<br>tools such as multiplaner<br>reconstruction (MPR), thin/think<br>maximum intensity projection (MIP) | <span style="text-align: center;">☑ Same</span><br>☐ Similar<br>☐ Different |
{6}------------------------------------------------
| • Supporting clinical<br>diagnostics by analysis of<br>cardiac MR images using<br>display functionality such as<br>panning, windowing,<br>zooming through<br>series/slices of the images.* | • Importing Cardiac MR Images in<br>DICOM format | thin and think, inverted MIP thin and<br>think, volume rendering technique<br>(VRT), curved planner reformation,<br>processing tools such as bone removal<br>(based on both single energy and dual<br>energy) table removal and evaluation<br>tools (vessel centerline calculation,<br>lumen calculation, stenosis<br>calculation) and reporting tools (lesion<br>location, lesion characteristics) and<br>key images), the software package is<br>designed to support the physician in<br>conforming the presence or absence of<br>physician identified lesion in blood | | | | | | OS<br>(Operating System) | | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|------------------------------|-------------------------------|----------------|----------------------------|----------------------------|------------------------------|
| • Supporting clinical<br>diagnostics analysis of the<br>heart in cardiac MR images<br>and signal intensity. | • Supporting clinical diagnostics by<br>qualitative analysis of the cardiac<br>MR images using display<br>functionality such as panning,<br>windowing, zooming, navigation<br>through series/slices and phases. | vessels and evaluation, documentation<br>and follow up of any such lesions. | | information as part of a comprehensive<br>diagnostic decision-making process.<br>This device is a software application<br>that can be used as a stand-alone<br>product or in a network environment.<br>The target population for the device is<br>not restricted, however the image<br>acquisition by a cardiac MR scanner<br>may limit the use of the device for<br>certain sectors or the public. | professionals, experienced in<br>examining and evaluating<br>cardiovascular MR images, for the<br>purpose of obtaining diagnostic<br>information as part of a comprehensive<br>diagnostic decision-making process.<br>cmr 42 is a software application that<br>can be used as a standalone product or<br>in a networked environment.<br>The target population for the cmr42 is<br>not restricted, however the image<br>acquisition by a cardiac magnetic<br>resonance scanner may limit the use of<br>the device for certain sectors of the<br>general public. | be used as a stand-alone product or in<br>a networked environment.<br>The target population for the cvi42 is<br>not restricted. | | | Window | Mac OS, Microsoft, Windows | Mac OS, Microsoft, Windows | |
| • Software package is<br>designed to support the<br>physician-to-physician<br>compliance assessment,<br>document and follow up<br>heart disease by cardiac<br>MRI. | • Supporting clinical diagnostics by<br>quantitative measurement of the<br>heart and adjacent vessels in<br>cardiac MR images, specifically<br>distance, area, volume and mass | It shall be used by qualified medical<br>professionals, experienced in<br>examining and evaluating<br>cardiovascular CT or MR images, for<br>the purpose of obtaining diagnostic<br>information as part of a comprehensive<br>diagnostic decision-making process. | | | cmr 42 shall not be used to view or<br>analyze images of any part of the body<br>except the cardiac magnetic resonance<br>images acquired from a cardiovascular<br>magnetic resonance scanner. | | | DICOM Compliant<br>networking | Yes | Yes | Yes | Same<br>Similar<br>Different |
| It shall be used by qualified medical<br>professionals, experienced in<br>examining and evaluating<br>cardiovascular MR images, for the<br>purpose of obtaining diagnostic | • Supporting clinical diagnostics by<br>using area and volume<br>measurements for measuring LV<br>function and derived parameters<br>cardiac output and cardiac index<br>in long axis and short axis cardiac<br>MR images | cvi42 is a software application that can | image from all MRI | Yes | Yes | Yes | Same<br>Similar<br>Different | Display of Images | Study & Series | Study & Series | Study & Series | Same<br>Similar<br>Different |
| | • Flow quantifications based on<br>velocity encodes images | It shall be used by qualified medical | Store images | File Format: dcm, json, csv, nii | File Format: PDF, XML | File Format: PDF, XML | Same<br>Similar<br>Different | | | | | |
| Quantities<br>assessment of<br>cardiac function | Yes<br>(Myocardiac) | Yes<br>(Myocardiac and Blood Flow) | Yes<br>(Myocardiac and Blood Flow) | Same<br>Similar<br>Different | | | | | | | | |
| Corresponding Tool | Yes<br>(Contour) | Yes<br>(Contour) | Yes<br>(Contour) | Same<br>Similar<br>Different | | | | | | | | |
| Reports containing<br>visualization of<br>images and<br>quantitative<br>parameters | T1, T2<br>T1 & T2 Mapping<br>Late Gd Enhancement (LGE) | T1, T2<br>T1 & T2 Mapping<br>Late Gd Enhancement (LGE)<br>Perfusion<br>Flow Analysis | T1, T2<br>T1 & T2 Mapping<br>Late Gd Enhancement (LGE)<br>Perfusion<br>Flow Analysis | Same<br>Similar<br>Different | | | | | | | | |
Page 4
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
page 4
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for "PHANTOMICS". The logo features a stylized head with closed eyes above the word "PHANTOMICS" in blue, sans-serif font. The head is drawn with simple lines, giving it a minimalist and modern appearance.
| Same Items | Description |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number | The proposed regulation number is “21CFR892.2050”. The<br>Regulation is related to the Medical Image Management and<br>Processing System.<br>It is the same point between the subject device (Myomics Q)<br>and the predicate device (cmr42) including the reference<br>device (cvi42). |
| Regulation Name | The proposed regulation name is Medical Image<br>Management and Processing System in accordance with<br>21CFR892.2050.<br>The subject device and predicate device are an image<br>analysis software for evaluating cardiac MR images.<br>It is the same point between the subject device (Myomics Q)<br>and the predicate device (cmr42) including the reference<br>device (cvi42). |
| Classification | In accordance with same regulation number<br>(21CFR892.2050), FDA provide the regulatory<br>classification. It is Class II and same point between the<br>subject device (Myomics Q) and the predicate device<br>(cmr42) including the reference device (cvi42). |
| Product Code | The proposed product code of the subject device is “LLZ”.<br>It is the same product code with the predicate device (cmr42)<br>including the reference device (cvi42). |
| Image from all MRI | All devices are derived and saved to MR DICOM Image file<br>only. It is same point |
| DICOM Compliant<br>networking | All devices are derived and saved to MR DICOM Image file<br>only. It is same point. |
| Display of Images | All devices make to display clinical study and series related<br>to cardiac MR DICOM Image. It is same point |
| Corresponding Tool | All devices control the contour tool for post-processing. It is<br>same point. |
1) Same points between the subject device and the predicate device
2) Similar points between the subject device and the predicate device
| Same Items | Description |
|------------|-------------|
|------------|-------------|
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for "PHANTOMICS". The logo consists of a stylized graphic above the word "PHANTOMICS" in all capital letters. The graphic appears to be a stylized representation of a face or head, possibly suggesting a futuristic or technological theme. The text is in a sans-serif font and a blue color.
| Indication for use | Although specific indication for use is similar between the |
|--------------------|-------------------------------------------------------------|
| | subject device and the predicate device including the |
| | reference device, all devices (subject device, predicate |
| | device and reference device) are intended to be used for |
| | viewing, post-processing and analysis of cardiac magnetic |
| | resonance (MR) images in a Digital Imaging and |
| | Communications in Medicine (DICOM) Standard format. It |
| | is same point. |
# 3) Different points between the subject device and the predicate device
| Same Items | Description |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| OS<br>(Operating System) | The subject device is available with window O.S. However,<br>predicated device and reference device are available with<br>Mac OS, Microsoft including Window O.S.<br>Although the OS (Operating System) is different between<br>subject device and predicate device including reference<br>device, Window O.S. is common O.S. under both devices.<br>In addition, Software Validation of the subject device by<br>utilizing Window O.S. had been conducted.<br>Thus, this different point is not effected to safety and<br>effectiveness of the subject device. |
| Store images | The subject device is intended to save the image file as the<br>following file format name; dcm, json, csv, nii<br>However, predicated device is intended to save the image<br>file as following file format name; PDF, DML<br>Although the file format name is different between subject<br>device and predicate device including reference device,<br>Software Validation of the subject device by utilizing the file<br>format such as dcm, json, csb, nii, had been conducted.<br>Thus, this different point is not effected to safety and<br>effectiveness of the subject device. |
| Quantities assessment of<br>cardiac function | The subject device is intended to analyze and assess the<br>myocardiac data. However, predicated device and reference<br>device are intended to analyze and assess the blood flow data<br>including myocardiac data. |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for "PHANTOMICS". The logo consists of a stylized graphic above the word "PHANTOMICS", which is written in a sans-serif font. The graphic appears to be a stylized representation of a person or figure. The color of the logo is a light blue.
| Although the assessment data are different between subject<br>device and predicate device including reference device,<br>Software Validation for the analysis of the subject device,<br>had been conducted.<br>Thus, this different point is not effected to safety and<br>effectiveness of the subject device. | The subject device is intended to display the report<br>containing T1, T2, T1 & T2 Mapping and LGE Data.<br>However, predicated device and reference device are<br>intended to display the report containing perfusion and flow<br>analysis including T1, T2, T1 & T2 Mapping and LGE Data<br>Although the display parameter is different between subject<br>device and predicate device including reference device,<br>Software Validation for displaying parameter of the subject<br>device, had been conducted.<br>Thus, this different point is not effected to safety and<br>effectiveness of the subject device. |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Reports containing<br>visualization of images and<br>quantitative parameters | |
{12}------------------------------------------------
# 7. NON-CLINICAL DATA
## 7.1 Safety Test
The safety requirement is not required because it is stand-alone software.
### 7.2 Software
The following tests were performed to assess effectiveness of software of the device. The test was performed in accordance with following standards.
| No. | Test Items | Standards |
|-----|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | General requirement for safety – Programmable<br>electrical medical systems (PEMS) | IEC 62304:2006+A1:2015 FDA Guidance (“Guidance for the<br>Content of Premarket Submissions for<br>Software Contained in Medical<br>Devices”) FDA Guidance (“Content of Premarket<br>Submissions for Management of<br>Cybersecurity in Medical Devices”) |
### 7.3 Performance Test
The following tests were performed to assess effectiveness of the device to support equivalence between subject device and reference device related to predicate device. The test was performed in accordance with following standards.
| Test ID | Test Item | Standard | |
|---------|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| SPPT001 | Pre-requisite Verification | This test checks whether Test Software (Myomics Q and<br>cvi42) is installed on the appropriate system requirement<br>(OS: window) So, To verify that the pre-requisite is<br>completed for the execution of qualifi-cation | |
| SPPT002 | Import cardiac MR Images | To verify that import function is working. | |
| SPPT003 | Export cardiac MR Images | To verify that export MR images function is working. | |
| SPPT004 | Patient Data | To verify that patient information function is working. | |
| SPPT005 | Series Overview | To verify that series overview function is working<br>properly. | |
| SPPT006 | Contour<br>Drawing,<br>Image<br>Tools | Endocardium<br>Contour<br>Epicardium<br>Contour<br>Move<br>Pinch<br>Nudge<br>Curved Line<br>Free Hand | To verify that contour drawing functions are working<br>properly. |
| | Smoothing<br>Undo<br>Redo<br>Restart<br>Delete<br>Confirm<br>Zooming<br>Panning<br>Windowing | | |
| SPPT007 | T1 Analysis | • To verify that T1 analysis function is working properly.<br>* T1 Analysis is just T1 Image or T1 Map, but is not the analysis function of T1 Map automatically. | |
| SPPT008 | T2 Analysis | • To verify that T2 analysis function is working properly.<br>* T2 Analysis is just T2 Image or T2 Map, but is not the analysis function of T2 Map automatically. | |
| SPPT009 | LGE Analysis | • To verify that LGE analysis function is working properly.<br>* LGE Analysis is just LGE Image, but is not the analysis function of LGE Image automatically. | |
| SPPT010 | Quantitative compare of polar<br>map report in Native T1<br>analysis. | • To verify that the results of Myomics Q are very similar to the results of cvi42 in polar map report in Native T1 analysis.<br>* The quantitative is depending on the user contour accuracy. So, few deviation could be happened until ±5% between both s/w | |
| SPPT011 | Quantitative compare of polar<br>map report in Post T1<br>analysis. | • To verify that the results of Myomics Q are very similar to the results of cvi42 in polar map report in Post T1 analysis.<br>* The quantitative is depending on the user contour accuracy. So, few deviation could be happened until ±5% between both s/w | |
| SPPT012 | Quantitative compare of polar<br>map report in T2 analysis. | • To verify that the results of Myomics Q are very similar to the results of cvi42 in polar map report in T2 analysis<br>* The quantitative is depending on the user contour accuracy. So, few deviation could be happened until ±5% between both s/w | |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image contains the word "PHANTOMICS" in a sans-serif font. Above the word is a logo that appears to be a stylized face inside of a circle. The text and logo are in blue.
{14}------------------------------------------------
### 7.4 Discussion
According to IEC 62304:2006+A1:2015 including "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", Software Validation applied of cybersecurity risk management is passed to ensure safety and performance for subject device (Myomics O).
In addition, under bench test, control group is cvi42 because predicated device (cmr42) is end of production anymore. Although cvi42 is improved production from cmr42, cvi 42 includes same intended used and same function for Cardiac MR Image. So, cvi42 is control group because of reference product.
For the bench test between subject device (Myomics Q) and reference device (cvi42), both s/w demonstrate substantial equivalent of O.S. (Reference Test ID: SPPT001), DICOM Image Importing Function (Reference Test ID: SPPT002), DICOM Image Exporting Function (Reference Test ID: SPPT003), display of patient information (Reference Test ID: SPPT004), display of patient data (Cardiac MR Image; Series/Study Image) (Reference Test ID: SPPT005), contour control function (Reference Test ID: SPPT006) as a post-processing, display for T1 Map (Reference Test ID: SPPT007), for T2 Map (Reference Test ID: SPPT008) and for LGE Viewing (Reference Test ID: SPPT009),
Especially, both S/W are operated with almost quantitative evaluation Native T1 Map (Reference Test ID: SPPT010), Post T1 Map (Reference Test ID: SPPT011) and T2 Map (Reference Test ID: SPPT012) in the Polar Map. Quantitative evaluation are performed in cvi42 and Myomics Q for the analysis functions of Native T1 Map, Post T1 Map and T2 Map. The results of Myomics O did not show a difference of more than ±5% compared to the results of cvi42 (95% of cvi42 results < Myomics Q Results < 105% of cvi42 results). Its deviation could be happened. The reason that the test result of Myomics Q and the result of cvi42 do not match 100% is because the value of the Polar Map Report may vary slightly depending on how the contour is drawn. It leads the result values are slightly different depending on how the contour is drawn, but the result is almost the same. And, its data is just referenced to check image file, but it is not affected to directly clinical parameter for diagnostic justification of intended user. So, ±5% deviation data between subject device and reference device are not affected to clinical performance to achieve intended use.
According to above discussion, we claim the both S/W could be substantial equivalent.
{15}------------------------------------------------
# 8. CONCLUSION
Under the comparing substantial equivalence between the subject device and the predicate device including reference device, there are the same points such as general information, some technical and material information.
Although there is some deviation for comparing point between subject device and predicate device including reference device, the safety and performance test reports are supported to the safety and effectiveness by discussion of software validation and bench test.
In this regard, we conclude that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.