Air Powered Tooth Polishing System (Perio-Mate); Air Powered Tooth Polishing System (Perio Mate Nozzle Tip)
Applicant
Nakanishi, Inc.
Product Code
EFB · Dental
Decision Date
Sep 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
Perio-Mate: Perio-Mate is intended for the following application(s): Removal of supragival/subgingival soft deposits and soft deposits on prostheses (including implant abutments). Perio Mate Nozzle Tip: Perio Mate Nozzle Tip is intended for the following application(s): Removal of subgingival soft deposits and soft deposits on prostheses (including implant abutments). The Perio Mate Nozzle Tip and Perio-Mate is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.
Device Story
Air-powered dental polishing system; consists of handpiece and powder case; connects to dental chair units via couplings/hoses. Operates by mixing polishing powder and water with compressed air; sprays mixture onto treatment area. Perio-Mate used for supragingival/subgingival deposits; Perio Mate Nozzle Tip (single-use, EOG sterile) used for subgingival pockets. Operated by dental professionals in clinical settings. Output is pressurized air/water/powder stream; enables mechanical removal of plaque/stains from tooth surfaces and prostheses. Benefits include effective non-surgical cleaning of periodontal pockets and implant abutments.
Clinical Evidence
Bench testing only. Verification and validation testing performed per ISO 20608 (powder jet handpieces), ISO 17665-1 (moist heat sterilization), ISO 17664-1/2 (reprocessing), and ISO 10993-1 (biocompatibility). No clinical data presented.
Technological Characteristics
Air-powered dental handpiece system. Materials: biocompatible per ISO 10993-1. Energy source: compressed air from dental chair unit. Connectivity: hose/coupling interface per ISO 9186. Sterilization: moist heat for reusable handpiece; EOG for single-use nozzle tip. Mechanical principle: powder-air-water jet.
Indications for Use
Indicated for non-surgical removal of supragingival and subgingival soft deposits, plaque, and stains from teeth and prostheses (including implant abutments) in periodontal pockets up to 5 mm deep following initial periodontal treatment.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
September 20, 2024
Nakanishi Inc. Ikeda Satoru Regulatory Affairs Dept. Tix Tower Ueno 9F 4-8-1, Higashiueno Taito-ku, Tokyo 110-0015 Japan
Re: K241880
Trade/Device Name: Air Powered Tooth Polishing System (Perio-Mate); Air Powered Tooth Polishing System (Perio Mate Nozzle Tip) Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EFB Dated: June 28, 2024 Received: June 28, 2024
Dear Ikeda Satoru:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Bobak Shirmohammadi -S
For Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K241880
Device Name
Air Powered Tooth Polishing System (Perio-Mate);
Air Powered Tooth Polishing System (Perio Mate Nozzle Tip)
Indications for Use (Describe)
Perio-Mate:
Perio-Mate is intended for the following application(s):
Removal of supragingival soft deposits and soft deposits on prostheses (including implant abutments).
Perio Mate Nozzle Tip:
Perio Mate Nozzle Tip is intended for the following application(s):
Removal of subgingival soft deposits and soft deposits on prostheses (including implant abutments).
The Perio Mate Nozzle Tip and Perio-Mate is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) #: K241880
510(k) Summary
Prepared on: 2024-09-18
### Contact Details
21 CER 807.92(a)(1)
| Applicant Name | NAKANISHI INC. | 21 CFR 807.92(a)(1) |
|------------------------------------|----------------------------------------------------------------------------------------------------------------|---------------------|
| Applicant Address | 700 Shimohinata Kanuma Tochigi 322-8666 Japan | |
| Applicant Contact Telephone | +81-289-64-7277 | |
| Applicant Contact | Mr. Kikuchi Masaaki | |
| Applicant Contact Email | m-kikuchi@nsk-nakanishi.co.jp | |
| Correspondent Name | NAKANISHI INC. | |
| Correspondent Address | TIX TOWER UENO 9F 4-8-1, Higashiueno Taito-ku Tokyo 110-0015 Japan | |
| Correspondent Contact Telephone | +81-358-28-8590 | |
| Correspondent Contact | Mr. Ikeda Satoru | |
| Correspondent Contact Email | nskra-contact@nsk-nakanishi.co.jp | |
| Device Name | 21 CFR 807.92(a)(2) | |
| Device Trade Name | Air Powered Tooth Polishing System (Perio-Mate);<br>Air Powered Tooth Polishing System (Perio Mate Nozzle Tip) | |
| Common Name | Dental handpiece and accessories | |
| Classification Name | Handpiece, Air-Powered, Dental | |
| Regulation Number | 872.4200 | |
| Product Code(s) | EFB | |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K151912 | AIR-FLOW handy 3.0 PLUS | EFB |
| K171174 | PERIO-FLOW nozzle | EFB |
Perio-Mate
This device is used to remove stains and deposits by spraying the polishing powder and water mixed by compressed air when connected with chair units through couplings and hoses to the treatment area. This device consists of a handpiece and a powder case. Couplings connected with powder case of this device have different structures, This device is developed, considering the connection with each manufacture's type. Some of them have the direct connection with hose that is with accordance with ISO 9186. This device is delivered in a non-sterile state and sterilized by end-users as a reusable device. This device to be used for sub and supraginal purpose. During subgingival treatment, Perio Mate Nozzle Tip is of
{5}------------------------------------------------
the handpiece.
Perio Mate Nozzle Tip
This device is connected to the Perio-Mate handpiece. The compressed air makes the tooth polishing agent in the Perio-Mate mixed and transfers it to the tip. The product is inseriodontal pocket, sprays the tooth polishing agent/water to the treatment area in the subgingival area, and enables to remove the deposits from teeth surface or prostheses. The product is EOG sterile single-use device.
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
Perio-Mate:
Perio-Mate is intended for the following application(s):
Removal of supragival/subgingival soft deposits and soft deposits on prostheses (including implant abutments).
Perio Mate Nozzle Tip:
Perio Mate Nozzle Tip is intended for the following application(s):
Removal of subgingival soft deposits and soft deposits on prostheses (including implant abutments).
The Perio Mate Nozzle Tip and Perio-Mate is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.
# Indications for Use Comparison
Both the Indications for Use for the subject device and the Indications for Use for the predicate device are intended for dental applications, including the removal of supragingival soft deposits on prostheses (including implant abutments).
The subject, predicate and reference device are indical for the non-surgical removal of subgingival plague in pockets up to 5 mm. The only difference between the Indictments for Use is the Device Name and product description.
# Technological Comparison
## 21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
The Air Powered Tooth Polishing System includes the Perio-Mate, which consists of a handpiece and the Perio Mate Nozzle Tip, which is connected to the end of the Perio-Mate handpiece.
Perio-Mate is used to remove stains and deposits by spraying the polishing powder and water mixed by compressed air when connected with chair units through couplings or hoses to the treatment area. This device is a non-sterile state and intended to be cleaned and sterilized by end-users as a reusable device. This device to be used for supragingival purpose.
During subgingival treatment, Perio Mate Nozzle Tip is connected to the handpiece. The compressed air makes the tooth polishing agent in the Perio-Mate mixed and transfers it to the tip. The product is inserted in a periodontal pocket, sprays the tooth polishing agent/water to the treatment area in the subgingival area, and enables to remove the deposits from teeth surface or prostheses.
The subject device shares these technical characteristics with the predicate device. These differences are only minor, such as the Perio Mate Nozzle Tip being an EOG sterilized disposable device, and dimensions. The Perio Mate Nozzle Tip being EOG sterilized, disposable, features that are not present in the predicate device but are present in the references reflect user preferences and market strategies and do not affect the substantial equivalence of the subject device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CER 807.92(b)
Non-Clinical Tests
For Air Powered Tooth Polishing System, the following non-clinical tests were performed to support substantial equivalence of the subject device.
Performance Test:
The Air Powered Tooth Polishing System subjected to verification and validation testing for product performance and reprocessing to support substantial equivalence. The rests demonstrate compliance with the requirements of the following standards and guidance.
· ISO 20608:2018 "Dentistry - Powder jet handpieces and powders"
• ISO 17665-1:2006 "Sterilization of health care products - Moist heat - Part 1: Requirements for the development validation and routine control of a sterilization process for medical devices"
· ISO 17664-1:2021 "Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 1: Critical and semi-critical medical devices"
· ISO 17664-2:2021 "Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 2: Non-critical medical devices."
{6}------------------------------------------------
· FDA guidance document "Dental Handpieces - Premarket Notification [510(k)] Submissions"
· FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" Biocompatibility Test:
Based on its Indications for Use, the Air Powered Tooth Polishing System, is classified as a device that has limited contact (less than 24 hours) with surface medical devices and breached or compromised surfaces.
To support substantial equivalence, biocompatibility testing was conducted. The test results demonstrate compliance with the requirements of the following standards and guidance:
- · ISO 10993-1:2018 "Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process"
- · FDA guidance document "Biological evaluation of medical devices Part 1: Evaluation and testing within a risk process"
The Non-Clinical Tests result support the substantial equivalence of the subject Air Polishing System to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.