K171174 · E.M.S Electro Medical Systems S.A · EFB · Dec 18, 2017 · Dental
Device Facts
Record ID
K171174
Device Name
PERIO-FLOW nozzle
Applicant
E.M.S Electro Medical Systems S.A
Product Code
EFB · Dental
Decision Date
Dec 18, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The PERIO-FLOW nozzle, used with EMS PERIO-FLOW dental handpieces, is intended for patients suffering from periodontal disease. The PERIO-FLOW nozzle, used with EMS PERIO-FLOW dental handpieces, is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.
Device Story
Flexible thermoplastic nozzle accessory for EMS PERIO-FLOW dental handpieces; connects to dental operative unit turbine. Transforms air/powder/water mixture into directed spray for subgingival plaque removal. Features printed markings at 3mm and 5mm to guide insertion depth. Used by dental professionals in clinical settings. Output is physical removal of plaque; aids in periodontal disease management. Single-use, non-sterile device.
Clinical Evidence
Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-1. Comparison testing verified connection force, pressure resistance, and nozzle fixation. Functional testing confirmed powder exit hole placement, exit direction, and durability of markings. All acceptance criteria met.
Technological Characteristics
Thermoplastic polyester elastomer nozzle with printed markings. Dimensions: 20.3mm length, 2.96mm width at taper, 2.03mm at tip. Operates via air-polishing (water/air/powder mixture). Compatible with PERIO (Glycine) powder. Service pressure 1-2.2 bar; static air pressure 2.7-3.5 bar. Single-use, non-sterile.
Indications for Use
Indicated for patients with periodontal disease requiring non-surgical removal of subgingival plaque in pockets up to 5 mm following initial periodontal treatment.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
E.M.S. Electro Medical Systems S.A. c/o Christina Henza Regulatory Can-Do-Medical 31 CH. de la Vuarpillière CH - 1260 Nyon Switzerland
December 18, 2017
Re: K171174
Trade/Device Name: PERIO-FLOW nozzle Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I Product Code: EFB Dated: November 17, 2017 Received: November 17, 2017
Dear Christina Henza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K171174
Device Name PERIO-FLOW nozzle
Indications for Use (Describe)
The PERIO-FLOW nozzle, used with EMS PERIO-FLOW dental handpieces, is intended for patients suffering from periodontal disease.
The PERIO-FLOW nozzle, used with EMS PERIO-FLOW dental handpieces, is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
× | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510 (K) SUMMARY FOR PERIO-FLOW nozzle
- SUBMITTTER/ 510(K) HOLDER l. E.M.S. Electro Medical Systems S.A. Ch. de la Vuarpillière 31 CH - 1260 Nyon Switzerland
Primary Contact:
Suzanne FASSIO-HARDY, Regulatory Affairs Manager – Dental Direct: +41 22 99 44 771 Email: sfassiohardy@ems-ch.com
Submission Contact:
Christina Henza chenza@can-do-medical.com
- II. Date Prepared: December 14, 2017
- III. DEVICE NAME
Proprietary Name: PERIO-FLOW nozzle Common/Usual Name: Accessories Classification Name: Dental Handpiece and Accessories (872.4200) Device Class: I Product Code: EFB
- IV. PREDICATE DEVICES
Primary Predicate: E.M.S. Electro Medical Systems S.A., AIR-FLOW handy 3.0 PERIO (K132480 cleared on 02/24/2014
- V. DEVICE DESCRIPTION
The PERIO-FLOW nozzle is a flexible, thermoplastic accessory to the PERIO-FLOW handpiece that is used with the EMS AIR-FLOW devices such as AIR-FLOW MASTER, AIR-FLOW MASTER PIEZON and AIR-FLOW handy PERIO and AIR-FLOW handy 3.0 Perio. The PERIO-FLOW nozzle includes printed markings and is molded from an alternate version of the cleared thermoplastic polyester elastomer. The marks are arranged to ensure that the user is aware of the location of the tip with respect to the pocket. There is a
{4}------------------------------------------------
mark at 3 mm and 5 mm so that the user can evaluate the location of the tip during a procedure. The 5mm mark is bolder to indicate the depth limit. The line and the square located above the 5 mm line serve to ensure that the user does not insert the nozzle deeper than 5mm in the pocket. The PERIO-FLOW nozzle is provided non-sterile. The nozzle is single use only and is intended for use under clean conditions.
The principle of operation of the proposed PERIO-FLOW nozzle is identical to predicate device nozzle (DT-113) which was cleared with the predicate system AIR-FLOW handy 3.0 PERIO (K132480 cleared on 02/24/2014).
EMS dental handpieces connect to a standard turbine connection on a dental operative unit and deliver a mixture of water, air, and the prophylaxis powder to a treatment site. The air /powder mixture exits the distal end of the PERIO-FLOW handpiece through the PERIO-FLOW nozzle where it meets with a water spray in the pocket.
#### VI. INDICATIONS FOR USE
The PERIO-FLOW nozzle, used with EMS PERIO-FLOW dental handpieces, is intended for patients suffering from periodontal disease.
The PERIO-FLOW nozzle, used with EMS PERIO-FLOW dental handpieces, is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.
### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS OF PROPOSED COMPARED TO THE PREDICATE DEVICE
The proposed PERIO-FLOW nozzle is an additional accessory intended to be used with all currently available EMS PERIO-FLOW dental handpieces. The substantial equivalence claim is specific to the nozzle accessory within the predicate submission, all other system components are unaffected by this submission.
At a high level, the subject and predicate devices are based on the following same technological elements:
- . Nozzle Geometry (general shape and hole placement)
- . Projection of water/air/powder mixture
- Use with PERIO (Glycine) powder
The following technological differences exist between the subject and predicate devices:
- . The material is a different thermoplastic polyester elastomer
- . The subject device includes printed markings
The questions regarding the material change and addition of markings are whether it is biocompatible and whether the use is adequately described within the instructions. These questions apply to both the new device and the predicate. Therefore, the proposed device, PERIO-FLOW nozzle, is substantially equivalent with regards to the nozzle portion of the legally marketed predicate AIR-FLOW handy 3.0 PERIO (K132480 cleared on 02/24/2014).
{5}------------------------------------------------
| Substantial Equivalence Table | | | | |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Item for Comparison | Proposed Device<br>(PERIO-FLOW Nozzle DT-476) | Predicate Device<br>(PERIO-FLOW Nozzle DT-113) | Explanation of<br>Variation | |
| Regulatory Information | Name | PERIO-FLOW nozzle | EMS AIR-FLOW handy 3.0 PERIO | N/A |
| | 510(k)# | K171174 | K132480 | N/A |
| | Predicates | K132480 | K082791 | N/A |
| | Product Code | EFB | EFB | Same. |
| | Class | 1 | 1 | Same. |
| | Combination Product | No | No | Same. |
| | Regulation Number | 872.4200 | 872.4200 | Same. |
| | Regulation Generic<br>Name | Dental Handpiece and<br>Accessories | Dental Handpiece and Accessories | Same. |
| Intended use | Regulation Intended<br>Use | "to prepare dental cavities for<br>restorations, such as fillings, and<br>for cleaning teeth." | "to prepare dental cavities for<br>restorations, such as fillings, and for<br>cleaning teeth." | Same. |
| | Indications | The PERIO-FLOW nozzle, used<br>with EMS PERIO-FLOW dental<br>handpieces, is intended for<br>patients suffering from<br>periodontal disease.<br><br>The PERIO-FLOW nozzle, used<br>with EMS PERIO-FLOW dental<br>handpieces, is indicated for the<br>non-surgical removal of<br>subgingival plaque in pockets up<br>to 5 mm after initial periodontal<br>treatment. | The AIR-FLOW handy 3.0 PERIO is<br>intended for patients suffering from<br>periodontal disease.<br><br>The AIR-FLOW handy 3.0 PERIO is<br>indicated for the non-surgical<br>removal of subgingival plaque in<br>pockets up to 5 mm after initial<br>periodontal treatment. | Same. |
| | Treatment Site | Subgingival | Subgingival | Same. |
| | Anatomical sites | Teeth and soft tissues in the<br>mouth. | Teeth and soft tissues in the mouth | Same. |
| | Specific Treatment<br>site | Subgingival | Subgingival | Same. |
| | Contact duration | Limited ≤ 24 hours | Limited ≤ 24 hours | Same. |
| | Biocompatibility | Biocompatible | Biocompatible | Same. |
| | Material | Thermoplastic Polyester<br>Elastomer<br>Ink (printed markings) | Thermoplastic Polyester Elastomer | Equivalent. |
| | Sterility | Provided non-sterile | Provided non-sterile | Same. |
| | Shelf life | unrestricted | unrestricted | Same. |
| Technological Characteristics | General purpose | Cleaning and surface preparation | Cleaning and surface preparation | Same. |
| | Treatment | Air-polishing | Air-polishing | Same. |
| | Mechanism of<br>treatment | Projection of water/air/powder<br>mixture | Projection of water/air/powder<br>mixture | Same. |
| | Compatible<br>Prophylaxis<br>Powders | PERIO (Glycine) | PERIO (Glycine) | Same. |
| | Mode | continuous operation | continuous operation | Same. |
| | Service pressure to<br>the turbine | 1 to 2.2 bar (1000-2200 hPa) with<br>a service flow of 50-80 ml/min.) | 1 to 2.2 bar (1000-2200 hPa) with a<br>service flow of 50-80 ml/min.) | Same. |
| Substantial Equivalence Table | | | | |
| Item for Comparison | Proposed Device<br>(PERIO-FLOW Nozzle DT-476) | Predicate Device<br>(PERIO-FLOW Nozzle DT-113) | Explanation of<br>Variation | |
| Service pressure to<br>the turbine<br>connection: Air | Static pressure 2.7 to 3.5 bar<br>(2700-3500 hPa) | Static pressure 2.7 to 3.5 bar (2700-<br>3500 hPa) | Same. | |
| Nozzle Geometry | 20.3mm in length, the width of<br>the device at the taper is 2.96<br>mm thick where the taper begins<br>near the shoulder and 2.03mm<br>thick near the tip | 20.3mm in length, the width of the<br>device at the taper is 2.96 mm thick<br>where the taper begins near the<br>shoulder and 2.03mm thick near the<br>tip | Same. | |
{6}------------------------------------------------
#### VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing:
The biocompatibility evaluation for the PERIO-FLOW nozzle was conducted in accordance with the FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (Attachment A) published June 16, 2016. This device is categorized in ISO 10993-1:2009 as "Surface device breached or compromised surface" per section 5.2.1(c). The device will have limited contact of less than or equal to 24 hours. Testing completed on this device includes:
- . Cytotoxicity
- . Sensitization
- . Irritation
### Comparison Testing:
To verify that the nozzles will work with the specified dental handpieces, comparison testing was completed on the current material (predicate) and the new material (this submission) for the nozzle. Comparison testing consisted of measuring connection force, pressure resistance, and nozzle fixation.
To verify that the change of material does not impact the performance on the nozzle, functional testing was completed on the current material (predicate) and the new material (this submission) for the nozzle. Functional testing consisted of confirming the placement of powder exit holes on the nozzle, the direction of powder exits, and the durability of the markings.
All tests were successfully performed and all acceptance criteria were met.
#### IX. CONCLUSIONS
Based on the information and supporting documentation provided in the premarket notification, the PERIO-FLOW nozzle is substantially equivalent to the cited predicate device. Testing demonstrates that the PERIO-FLOW nozzle fulfills prospectively defined design and performance specifications.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.