K160952 · Alma, Ltd. · GEX · May 5, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160952
Device Name
Alma Diode Tabletop Laser
Applicant
Alma, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 5, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Alma Diode Tabletop Laser is intended for use in dermatologic and general surgical procedures.
Device Story
The Alma Diode Tabletop Laser is a surgical laser system used in dermatologic and general surgical procedures. It utilizes diode laser modules (810 nm, 980 nm, or 1470 nm) to deliver laser energy to soft tissue via optical fibers. The system includes a main console, footswitch, and an automatic pull-back accessory. The pull-back accessory controls the rate of fiber movement along the vein wall during endovenous procedures, ensuring consistent energy delivery and minimizing treatment time. The device is operated by clinicians in a clinical setting. The clinician uses an LCD touch screen to configure parameters and monitor the procedure. By automating the fiber withdrawal, the device provides consistent exposure, which can improve procedural efficiency and potentially benefit the patient by ensuring uniform vein occlusion.
Clinical Evidence
Bench testing only. Verification testing confirmed mechanical pull-back speeds matched requested settings. Software was verified and validated per IEC 62304 and FDA guidance. No clinical data presented.
Technological Characteristics
Diode laser source; wavelengths 810nm, 980nm, 1470nm; optical fiber delivery; LCD touch screen interface; 100-240V AC power. Pull-back accessory provides automatic fiber movement at 1, 2, and 4 mm/sec. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62304 (software).
Indications for Use
Indicated for endovenous laser surgery for saphenous incompetent veins (810 nm); endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux (980 nm, 1470 nm); and laser assisted lipolysis (980 nm, 1470 nm).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized graphic of three human profiles facing right. The profiles are stacked on top of each other, creating a sense of depth and unity. The overall design is simple and professional, reflecting the department's role in public health and human services.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 5, 2016
Alma Ltd % Ms. Kathy Maynor Regulatory Consultant 26 Rebecca Ct Homosassa, Florida 34446
Re: K160952
Trade/Device Name: Alma Diode Tabletop Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 1, 2016 Received: April 5, 2016
Dear Ms. Maynor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160952 "
Device Name Alma Diode Tabletop Laser
Indications for Use (Describe)
Intended Use The Alma Diode Tabletop Laser is intended for use in dermatologic and general surgical procedures.
Indications for Use
laser assisted lipolysis.
The Alma Diode Tabletop Laser includes three possible diode laser modules depending on the customer order. Diode Laser Modules:
The indications for use for the 810 nm Alma Diode Tabletop Laser include: - The Alma 810 nm diode tabletop laser is indicated for endovenous laser surgery for saphenous incompetent veins.
The indications for use for the 980 nm Alma Diode Tabletop Laser include: -The Alma 980 nm diode tabletop laser is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The Alma 980 nm diode tabletop laser is further indicated for laser assisted lipolysis.
The indications for use for the 1470 nm Alma Diode Tabletop Laser include: -The Alma 1470 nm diode tabletop laser is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The Alma 1470 nm diode tabletop laser is further indicated for
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
## Section 8 – 510(k) Summary or 510(k) Statement
#### I. General Information
| Submitter: | Alma Lasers, Ltd,<br>Halamish St. POB 30<br>Industrial Park<br>Caesarea Israel 38900 |
|---------------------------|--------------------------------------------------------------------------------------|
| Contact Person: | Kathy Maynor<br>Consultant<br>352-586-3113 (cell) |
| Summary Preparation Date: | May 4, 2016 |
#### II. Names
| Device Names: | Alma Diode Tabletop Laser |
|-------------------------------|-----------------------------------|
| Primary Classification Names: | Surgical Powered Light Instrument |
#### III. Predicate Devices
| K # | Predicate Device |
|---------|---------------------------|
| K140005 | ALMA DIODE TABLETOP LASER |
| K133774 | DIOTECH CO LTD - ATOVEN |
#### IV. Product Description
The Alma Diode Tabletop Laser is comprised of the following major components:
- The main console unit 1.
- 2. Pull-back
- 3. Footswitch
- Accessories 4.
{4}------------------------------------------------
#### V. Indications for Use
### Intended Use
The Alma Diode Tabletop Laser is intended for use in dermatologic and general surgical procedures.
Indications for Use
The Alma Diode Tabletop Laser includes three possible diode laser modules depending on the customer order.
Diode Laser Modules:
The indications for use for the 810 nm Alma Diode Tabletop Laser include: - The Alma 810 nm diode tabletop laser is indicated for endovenous laser surgery for saphenous incompetent veins.
The indications for use for the 980 nm Alma Diode Tabletop Laser include:
-The Alma 980 nm diode tabletop laser is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The Alma 980 nm diode tabletop laser is further indicated for laser assisted lipolysis.
The indications for use for the 147 0nm Alma Diode Tabletop Laser include:
-The Alma 1470 mm diode tabletop laser is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The Alma 1470 nm diode tabletop laser is further indicated for laser assisted lipolysis.
{5}------------------------------------------------
#### VI. Summary of Technical Characteristics
Table 1: Salient Characteristics of the 1470nm module and the Predicate Devices
| | K13<br>Alma diode tabletop laser | K100558<br>Quanta Diode Laser Family |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Parameter | | |
| Product<br>Code &<br>Regulation<br>No. | GEX<br>21 CFR 878.4810 | GEX<br>21 CFR 878.4810 |
| Laser<br>Wavelength<br>[nm] | 1470 | 1470 |
| Max power | 15W | 15W |
| Light/Laser<br>Source | Diode | Diode |
| Laser<br>Delivery | Optical Fiber | Optical Fiber |
| Operation<br>Mode | Continuous wave, single<br>pulse, pulsed | Continuous wave, single pulse,<br>pulsed |
| Pulse<br>Duration | 10-990ms | 3ms - 2.5s |
| Bare fiber<br>size | 200, 300, 320, 400, 600,<br>800, 1000 | 200, 300, 320, 400, 600, 800,<br>1000 |
| User<br>Interface | LCD touch screen | LCD touch screen |
| Aiming beam | 635nm | 650nm |
| Electrical<br>Requirements | 100-240, V AC 50-60 Hz,<br>6.3 A, | 100-240, V AC 50-60 Hz, 6.3<br>A, single phase, |
| Indications<br>for Use | The Alma 1470nm diode<br>tabletop laser is indicated<br>for use in endovenous<br>occlusion of the greater<br>saphenous vein in<br>patients with superficial<br>vein reflux. The Alma<br>1470nm diode tabletop<br>laser is further indicated<br>for laser assisted<br>lipolysis, | The QUANTA Diode Laser<br>System is intended for delivery<br>of laser light to soft tissue in the<br>contact and non contact mode<br>during surgical procedures<br>including<br>via endoscopes. The QUANTA<br>Diode Laser System is generally<br>indicated for use in endovenous<br>occlusion of the greater<br>saphenous vein in Patients with<br>Superficial Vein Reflux. The<br>QUANTA1470 Diode<br>Laser is further indicated for<br>laser assisted lipolysis |
{6}------------------------------------------------
| Manufacturer | Alma Lasers LTD | Alma Lasers LTD | DIOTECH CO LTD |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cleared Device | KXXXXXX | K140005 | K133774 |
| System Platform | Alma Diode Tabletop Laser | Alma Diode Tabletop Laser | ATOVEN |
| Accessory Description | Alma Lasers Pull-Back<br>Device | | Automatic fiber pull-<br>back driver ATOVEN |
| Wavelength<br>Supported | 1470nm | 810nm, 980nm, and<br>1470nm | 1470nm |
| Intended Use /<br>Indications For Use | The Alma Lasers Pull-Back<br>Device is an accessory<br>intended for use in<br>conjunction with the<br>previously cleared Alma<br>Diode Tabletop Laser<br>1470nm (K140005).<br>The Alma Diode Tabletop<br>Laser 1470 nm (K140005) is<br>intended for use in<br>dermatologic and general<br>surgical procedures:<br>endovenous occlusion of<br>the greater saphenous vein<br>in patients with superficial<br>vein reflux, and for laser<br>assisted lipolysis. | The Alma Diode Tabletop<br>Laser 1470 nm is intended<br>for use in dermatologic and<br>general surgical<br>procedures: endovenous<br>occlusion of the greater<br>saphenous vein in patients<br>with superficial vein reflux,<br>and for laser assisted<br>lipolysis. | ATOVEN is a diode<br>laser system<br>designed to delivery<br>of laser light to soft<br>tissue during general<br>surgery procedures.<br>This device intended<br>for treatment of reflux<br>of the saphenous<br>veins of thigh<br>associated with<br>varicose veins and<br>varicosities. |
| Pullback Accessory | | | |
| Dimensions (cm)<br>(H x W x D) | 5x20x8 | NA | Unk. |
| Weight (kg) | ~0.5 | NA | Unk. |
| Pullback Speed<br>(mm/sec) | 4 Steps:<br>1, 2, and 4 mm/sec | NA | 6 Steps:<br>0.5, 0.7, 1.0, 1.2, 1.5,<br>and 2.0 mm/sec |
| Pullback mode | Automatic | NA | Automatic |
| Acceptable fiber sizes | 400µm fiber<br>600µm fiber | NA | 400µm fiber<br>600µm fiber |
| Mode of Action | Controls the rate of fiber<br>movement along the vein<br>wall providing consistent<br>exposure and minimizing<br>treatment time. | NA | Controls the rate of<br>fiber movement along<br>the vein wall<br>providing consistent<br>exposure and<br>minimizing treatment<br>time. |
#### Safety and Effectiveness Information VII.
The review of the indications for use and technical characteristics provided demonstrates that the Alma diode tabletop laser is substantially equivalent to the predicate devices. Additional product electrical safety testing and EMC testing was successfully completed in accordance with the following standards:
{7}------------------------------------------------
- IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for Safety ●
- . IEC 60601-1-2 Medical Electrical Equipment 1-2 General Requirements for basic safety and essential performance Collateral Standard: Electromagnetic disturbances -Requirements and tests
## VIII. D Design Verification Testing
Design verification testing was performed for the Alma 1470mm diode tabletop laser with the pullback accessory. The testing consisted of performing mechanical tests to verify the actual pull back speeds matched the requested speeds.
Additionally, the software controlling the Alma 1470nm diode tabletop laser with the pull-back accessory was fully verified and validated in accordance with IEC 62304 and the FDA guidance document entitled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
#### IX. Conclusion
The Alma diode tabletop laser with the pull-back accessory was found to be substantially equivalent to the predicate device.
The Alma diode tabletop laser with the pull-back accessory shares the same or similar indications for use, similar design features, and functional features with, and thus is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.