K241543 · Vinbrain Joint Stock Company · QIH · Dec 6, 2024 · Radiology
Device Facts
Record ID
K241543
Device Name
DrAid™ for Liver Segmentation
Applicant
Vinbrain Joint Stock Company
Product Code
QIH · Radiology
Decision Date
Dec 6, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Liver segmentation
—
Mean Dice >= 0.95; 95% CI lower bound of Dice scores >= 0.90; 95% CI upper bound of HD95 score <= 4.0
Mean Dice: 0.9649; 95% CI Dice: [0.9631, 0.9667]; Mean HD95: 1.7061; 95% CI HD95: [1.5595, 1.8526]
—
—
450 contrast-enhanced CT scans of 150 patients from US medical institutions.
3 (US board-certified radiologists)
Liver volume measurement
—
95% CI upper bound of Volume Error <= 5%
Mean Volume Error: 2.7269%; 95% CI: [2.4308%, 3.0230%]
—
—
450 contrast-enhanced CT scans of 150 patients from US medical institutions.
3 (US board-certified radiologists)
Indications for Use
DrAid™ for Liver Segmentation is a web-based software, non-invasive image analysis application designed for the visualization, evaluation, and reporting of liver and physician identified lesions using multiphase images (with slice thickness <= 3.0mm) of patients aged and older than 22 years old obtained from CT scanners. The software provides a range of tools for image viewing, processing, and reporting. The software enables professionals, including physicians and technicians, to review and analyze multiphase volume datasets of the liver. DrAid™ for Liver Segmentation operates in a semiautomated quantitative imaging function, utilizing an Al algorithm to generate liver segmentation that is then editable by the physician if necessary. Additionally, the device provides tools for manual segmentation within user input of seed points and boundary editing for physician-identified lesions within the liver. Professionals can assess liver volume (mm³), liver lesion volume (mm³), and maximum lesion diameter (mm), position, thereby aiding in evaluation and treatment planning. It is important to note that the software is intended for use by trained professionals, including physicians and technicians. The image source for analysis is DICOM, allowing compatibility with standard medical imaging data formats.
Device Story
DrAid™ for Liver Segmentation is a web-based software application for clinical use by physicians and technicians. It processes multiphase DICOM CT images to assist in liver and lesion analysis. The device utilizes an AI algorithm for semi-automated liver segmentation, which users can manually refine. It also provides tools for manual segmentation of lesions via seed points and boundary editing. The system calculates liver volume, lesion volume, and maximum lesion diameter. Outputs are displayed for clinical review to aid in evaluation and treatment planning. The device does not generate diagnoses; it serves as a quantitative imaging tool to support clinical decision-making by qualified professionals.
Clinical Evidence
Bench testing only. Performance validated on 450 contrast-enhanced CT scans from 150 patients (age 60 ±12). Ground truth established by 3 US board-certified radiologists. Primary endpoints: Liver segmentation mask Dice score (Mean 0.9649, 95% CI [0.9631, 0.9667]) and HD95 (Mean 1.7061, 95% CI [1.5595, 1.8526]). Liver volume measurement error (NVE) mean 2.7269% (95% CI [2.4308%, 3.0230%]). Generalization verified across multiple CT manufacturers (Siemens, GE, Philips, Toshiba).
Technological Characteristics
Web-based software application; DICOM-compatible; utilizes AI algorithm for semi-automated segmentation; manual editing tools for seed points and boundaries; supports MPR visualization; runs on standard hospital/clinic workstations.
Indications for Use
Indicated for patients aged 22 years and older requiring visualization, evaluation, and reporting of liver and physician-identified lesions using multiphase CT images (slice thickness <= 3.0mm).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
VinBrain Joint Stock Company Steven Quoc Hung Truong Chief Executive Officer No 7 Bang Lang 1 Street, Vinhomes Riverside Ecological Urban Area. Long Bien District Hanoi, 100000 Vietnam
December 6, 2024
Re: K241543
Trade/Device Name: DrAid™ for Liver Segmentation Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: November 7, 2024 Received: November 7, 2024
Dear Steven Quoc Hung Truong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation reguires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241543
Device Name DrAid™ for Liver Segmentation
#### Indications for Use (Describe)
DrAid™ for Liver Segmentation is a web-based software, non-invasive image analysis application designed for the visualization, evaluation, and reporting of liver and physician identified lesions using multiphase images (with slice thickness <= 3.0mm) of patients aged and older than 22 years old obtained from CT scanners. The software provides a range of tools for image viewing, processing, and reporting.
The software enables professionals, including physicians and technicians, to review and analyze multiphase volume datasets of the liver. DrAid™ for Liver Segmentation operates in a semautomated quantitative imaging function, utilizing an Al algorithm to generate liver segmentation that is then editable by the physician if necessary. Additionally, the device provides tools for manual segmentation within user input of seed points and boundary editing for physician-identified lesions within the liver. Professionals can assess liver volume (mm³), liver lesion volume (mm³), and maximum lesion diameter (mm), position, thereby aiding in evaluation and treatment planning.
It is important to note that the software is intended for use by trained professionals. including physicians and technicians. The image source for analysis is DICOM, allowing compatibility with standard medical imaging data formats.
Note: DrAid™ for Liver Segmentation does not generate diagnoses or potential findings directly.
The interpretation of the image data and the clinical decision-making process should be performed by qualified healthicare professionals. The installation and deployment of the software medical device should be carried out by VinBrain and trained technicians.
Caution: Federal law restricts this device to sale by or on the order of a physician.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a sequence of alphanumeric characters. The sequence starts with the letter 'K', followed by the numbers '241543'. The characters are displayed in a clear, sans-serif font, and they are all aligned horizontally.
## 510(k) Summary
# DrAid™ for Liver Segmentation
| Name and Address of Applicant: | VinBrain Joint Stock Company<br>No 7 Bang Lang 1 Street,<br>Vinhomes Riverside Ecological Urban Area,<br>Viet Hung Ward, Long Bien District,<br>Hanoi, 100000, Vietnam |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Steven Quoc Hung Truong , Chief Executive Officer |
| Telephone No.: | 84 (981) 927-516 |
| Email Address: | anh3.pham@vinbrain.net |
| Date of Submission: | November 7, 2024 |
| Device Name: | DrAidTM for Liver Segmentation |
| Product Code: | QIH |
| Regulation Name: | Medical image management and processing system |
| Regulation Number: | 892.2050 |
| Classification: | Class II |
| Classification Name: | System, Image Processing, Radiological |
#### Identification of Predicate Device:
510(k) Number: K131498 Device Name: IQQA-LIVER MULTIMODALITY SOFTWARE Manufacturer: EDDA TECHNOLOGY, INC.
## 1. Device Description Summary
Device Identification:
Key device components:
- AI algorithm for liver segmentation ।
- Measurement algorithm -
- DICOM Processing Module for CT images -
- Liver Segmentation viewer -
- Results Export Module -
- Device Characteristics: -
- Software -
Environment of Use:
- Healthcare facility/hospital -
Brief Written Description:
- Explanation of how the device works/principle of operation: DrAid™ for Liver Segmentation is a web-based software that processes and analyzes multiphase CT images in DICOM format. The software utilizes AI algorithms for semi-automated liver segmentation, combined with manual editing capabilities. Additionally, the device provides tools for manual segmentation
{5}------------------------------------------------
with user input of seed points and boundary editing for physician-identified lesions within the liver.
- Key Features for SE/Performance: -
- Visualization modes: i
- Original DICOM 2D image viewing ।
- -MPR visualization
- -Manual correction tools:
- Seed point placement -
- Boundary editing for lesions -
- । Segmentation refinement
- Reporting tool. ।
- Energy Source: -
- -Web-based application running on standard hospital/clinic workstations
### 2. Indications for use
DrAid™ for Liver Segmentation is a web-based software, non-invasive image analysis application designed for the visualization, evaluation, and reporting of liver and physician identified lesions using multiphase images (with slice thickness <= 3.0mm) of patients aged and older than 22 years old obtained from CT scanners. The software provides a range of tools for image viewing, processing, and reporting.
The software enables professionals, including physicians and technicians, to review and analyze multiphase volume datasets of the liver. DrAid™ for Liver Segmentation operates in a semiautomated quantitative imaging function, utilizing an Al algorithm to generate liver segmentation that is then editable by the physician if necessary. Additionally, the device provides tools for manual segmentation within user input of seed points and boundary editing for physician-identified lesions within the liver. Professionals can assess liver volume (mm3), liver lesion volume (mm²), and maximum lesion diameter (mm), position, thereby aiding in evaluation and treatment planning.
It is important to note that the software is intended for use by trained professionals, including physicians and technicians. The image source for analysis is DICOM, allowing compatibility with standard medical imaging data formats.
Note: DrAid™ for Liver Segmentation does not generate diagnoses or potential findings directly.
{6}------------------------------------------------
The interpretation of the image data and the clinical decision-making process should be
performed by qualified healthcare professionals. The installation and deployment of the software
medical device should be carried out by VinBrain and trained technicians.
Caution: Federal law restricts this device to sale by or on the order of a physician.
## 3. Indication for use comparison
Both the subject and predicate devices are intended for segmentation of liver and physician identified lesions for use by trained healthcare professionals and are compatible with DICOM image data, which is essential for medical imaging applications.
{7}------------------------------------------------
# 4. Technological comparison
## Technical comparison with predicate device:
Both the subject and predicate device are software applications, supporting the evaluation of liver lesions from multiphase images and provide essential tools for image viewing, segmentation, and reporting. Both are designed for use by trained healthcare professionals and intended for use in a hospital healthcare environment. The differences in deployment platforms, liver segmentation technology, operating systems and some features do not compromise the safety and effectiveness of either device in their intended clinical applications.
The subject device provides semi-automatic segmentation of the liver using an AI algorithm with editable tool, whereas the predicate device provides manual segmentation for liver. The difference in liver segmentation techniques highlights the strength of DrAid™ for Liver Segmentation in leveraging an AI algorithm for liver segmentation. The utilization of advanced algorithms can enhance the efficiency and accuracy of liver segmentation. The difference in segmentation techniques does not raise different questions regarding the safety and effectiveness of DrAid™ for Liver Segmentation.
# 5. Non - Clinical and/or Clinical Test summary & conclusions
DrAid™ for Liver Segmentation has been rigorously evaluated and verified to align with user needs and intended applications through successful performance, functionality, and algorithmic tests. Testing outcomes confirm that DrAid™ for Liver Segmentation fulfills the device's user needs and requirements, demonstrating substantial equivalence in performance to the listed predicate and reference devices.
Performance testing was performed using the representative and independent US dataset and, to ensure that the performance and accuracy met the requirements and standards. The ground truth was established by the annotations provided by 3 US board-certified radiologists.
Overall Liver Segmentation Algorithm:
- Patients:
- o 450 contrast-enhanced CT scans of 150 patients
- o Age distribution: 60 ±12
- Sex distribution: 49.3% female, 50.7% male o
- Location of clinical sites: US medical institutions O
- o Image procedure: Contrast-enhanced CT images taken for diagnostic reading
The following table provides a summary of the validation results:
{8}------------------------------------------------
| Test<br>performed | Sample size<br>(number of<br>CT scans) | Acceptance criteria | Performance results |
|-------------------------------|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Liver<br>segmentation<br>mask | 450 | 1) Mean Dice $\ge$ 0.95<br>2) 95% CI lower bound<br>of Dice scores $\ge$ 0.90<br>3) 95% CI upper bound<br>of HD95 score $\le$ 4.0 | 1) Dice score:<br>+ Mean 1std: 0.9649 0.0195<br>+ 95% CI Dice: 0. 9649 [0.9631,<br>0.9667]<br>2) HD95:<br>+ Mean 1std: 1.70611.5800<br>+ 95% CI: 1.7061 [1.5595,<br>1.8526] |
It is evident that the performance of the DrAid™ liver segmentation algorithm evaluated on the representative test set meets the aforementioned acceptance requirements. The performance observed on the US-patient dataset, encompassing diverse tumors, steatosis cases, and hepatomegaly demonstrates the algorithm's robustness when confronted with challenging data.
The results of the liver volume measurement are provided below.
| Test performed | Sample size(number<br>of CT scans) | Acceptance criteria | Performance results |
|-----------------------------|------------------------------------|--------------------------------------------|-----------------------------------------------------------------------------|
| liver volume<br>measurement | 450 | 95% CI upper bound of<br>Volume Error ≤ 5% | NVE:<br>Mean 1std: 2.7269 %<br>3.1928 %<br>95% CI: [2.4308 %<br>, 3.0230 %] |
Furthermore, we evaluated the AI algorithm on main subgroups (i.e., different types of abnormalities, age groups, gender groups, phase groups, and ethnicity groups) to prove the model generalization capability on diversity data distributions. Detailed subgroup analysis is reported in the labeling.
The generalization capability of the DrAid™ liver segmentation algorithm is also verified using datasets collected from a wide range of CT scan manufacturers used in practical clinics. Details on the scanner types, data distribution, and performance results given in the table below demonstrate the robustness of the subject device to the different data sources acquired from the diverse manufacturers.
| CT brand | CT model name | Number<br>of scans | Data source | Mean Dice<br>[95% CI] |
|-----------------------|--------------------------------------------------|--------------------|-------------|-------------------------|
| Siemens | SOMATOM go.Up | 450 | USA | 0.9649 [0.9631, 0.9667] |
| GE Medical<br>Systems | LightSpeed 16, LightSpeed VCT,<br>Revolution ACT | 30 | USA | 0.9786 [0.9759, 0.9812] |
| Philips | Brilliance 64 | 30 | USA | 0.9747 [0.9726, 0.9768] |
{9}------------------------------------------------
| Toshiba | 00 | Vietnam | 0.9760 [0.9712, |
|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|-----------------|
| Aquilion ONE | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | 0.98097 |
### Conclusion:
In conclusion, all verification and validation activities demonstrated that the design specifications and technological characteristics of DrAid™ for Liver Segmentation meet the applicable requirements and standards and DrAid™ for Liver Segmentation is substantially equivalent to the currently marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.