The PMD Clean Redvolution (All 4005 models) is a hand-held device for over-the counter aesthetic purposes. The red light is intended for the use in treating wrinkles on the face.
Device Story
Hand-held device; utilizes red light emitting diodes (LEDs) at 630nm±5nm wavelength to irradiate facial skin for wrinkle treatment; includes reverse-side soft silicone bristles for facial cleansing prior to light therapy; powered by internal rechargeable Li-ion battery; operated by user in home environment; includes protective goggles to prevent ocular exposure; device output is red light energy; intended to improve facial appearance by treating wrinkles.
Clinical Evidence
No clinical data. Bench testing only. Device performance verified through electrical safety (IEC 60601-1, 60601-1-11, 60601-2-57), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), battery safety (IEC 62133-2), and biocompatibility (ISO 10993-5, -10, -23).
Indicated for over-the-counter aesthetic use on the face to treat wrinkles.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 25, 2024
Dongguan Hunter Electronic Technology Co., Ltd.
% Riley Chen
Official Correspondent
Feiying Drug & Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center,
No. 3101-90, Qianhai Road
Shenzhen, Guangdong 518052
China
Re: K241203
Trade/Device Name: PMD Clean Redvolution (4005-moab-NA, 4005-moab-INT, 4005-pepper-NA, 4005-pepper-INT, 4005-cream-NA, 4005-cream-INT, 4005-twilight-NA, 4005-twilight-INT, 4005-grey-NA, 4005-grey- INT)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS
Dated: April 30, 2024
Received: April 30, 2024
Dear Riley Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K241203 - Riley Chen
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}
K241203 - Riley Chen
Page 3
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
Digitally signed by
TANISHA L. HITHE -S
Date: 2024.07.25
21:25:28 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical
and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K241203
Device Name
PMD Clean Redvolution (4005-moab-NA, 4005-moab-INT, 4005-pepper-NA, 4005-pepper-INT, 4005-cream-NA, 4005-cream-INT, 4005-twilight-NA, 4005-twilight-INT, 4005-grey-NA, 4005-grey-INT)
Indications for Use (Describe)
The PMD Clean Redvolution (All 4005 models) is a hand-held device for over-the counter aesthetic purposes. The red light is intended for the use in treating wrinkles on the face.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
Dongguan Hunter Electronic Technology Co., Ltd.
510(k)s -510 (k) Summary
# 510 (k) Summary K241203
This “510(k) Summary” of 510(k) safety and effectiveness information is submitted in accordance with requirements of Title 21, CFR Section 807.92.
## (1) Applicant information:
510(k) owner's name: Dongguan Hunter Electronic Technology Co., Ltd.
Address: Room 301, Building 6, 74 Qiaodong Road, Tangxia Town, Dongguan City, Guangdong, P.R. China
Contact person: Hunter Ye
Phone number: +86 13554760590
Email: 13554760590@139.com
Date of summary prepared: 2024-7-25
## (2) Proprietary name of the device
Trade name/model: PMD Clean Redvolution (4005-moab-NA, 4005-moab-INT, 4005-pepper-NA, 4005-pepper-INT, 4005-cream-NA, 4005-cream-INT, 4005-twilight-NA, 4005-twilight-INT, 4005-grey-NA, 4005-grey-INT)
Common name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Classification: Light Based Over The Counter Wrinkle Reduction
Regulation number: 21 CFR 878.4810
Product code: OHS
Review panel: General & Plastic Surgery
Regulation class: Class II
## (3) Predicate & Reference device
| Predicate | Predicate Trade Name | Produce code |
| --- | --- | --- |
| K202055 | Looper (Model: ZX-579S) | OHS , OLP |
| Reference | Reference Trade Name | Produce code |
| K162098 | LED Phototherapy Device. Model:PL-120 | OHS , OLP |
## (4) Description/ Design of device:
PMD Clean Redvolution adopts light emitting diodes (LED) in the red (630nm±5nm) spectrum to irradiate on the face for the treatment of facial wrinkles. PMD Clean Redvolution adopts handheld design, LEDs are contained in the red light therapy treatment window, and on its reverse side, the soft silicone bristles can be used to clean the face before the red light treatment. The product is
Page 1 of 5
{5}
Dongguan Hunter Electronic Technology Co., Ltd. 510(k)s -510 (k) Summary
provided with internal battery and it can be charged by external adaptor. To prevent irradiation of LED lights to eyes during the treatment, PMD Clean Redvolution is equipped with goggles which blocks light energy from LEDs.
PMD Clean Redvolution includes 10 models, the only difference between these models is the enclosure color.
# (5) Indications for use:
The PMD Clean Redvolution (All 4005 models) is a hand-held device for over-the counter aesthetic purposes. The red light is intended for the use in treating wrinkles on the face.
# (6) Technological characteristics and substantial equivalence:
| Item | Subject device | Predicate device | Reference device | Remark |
| --- | --- | --- | --- | --- |
| Trade name and model | PMD Clean Redvolution Model: 4005-moab-NA, 4005-moab-INT, 4005-pepper-NA, 4005-pepper-INT, 4005-cream-NA, 4005-cream-INT, 4005-twilight-NA, 4005-twilight-INT, 4005-grey-NA, 4005-grey-INT, | Looper (Model: ZX-579S) | LED Phototherapy Device. Model:PL-120 | / |
| 510 (k) number | Pending | K202055 | K162098 | / |
| Manufacturer | Dongguan Hunter Electronic Technology Co., Ltd. | Heat In A Click | Li-Tek Electronics Technologies | / |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Classification name | Light Based Over The Counter Wrinkle Reduction | Light Based Over The Counter Wrinkle Reduction, Over-The-Counter Powered Light Based Laser For Acne | Light Based Over the Counter Wrinkle Reduction, Over-The-Counter Powered Light Based Laser For Acne | Same |
| Product code | OHS | OLP, OHS | OLP, OHS | Same |
| Class | II | II | II | Same |
| Indications for use/ Intended use | The PMD Clean Redvolution (All 4005 models) is a hand-held | Looper (Model: ZX-579S) is a hand-held device for over-the counter aesthetic | The red light is intended for the treatment of periorbital | Same |
{6}
Dongguan Hunter Electronic Technology Co., Ltd.
510(k)s -510 (k) Summary
| | device for over-the counter aesthetic purposes. The red light is intended for the use in treating wrinkles on the face. | purposes. The Photon mode red light is indicated for the use in treating wrinkles on the face, and the blue light is indicated for the treatment of the mild to moderate inflammatory acne. | wrinkles, and the blue light is intended for the treatment of the mild to moderate inflammatory acne. | |
| --- | --- | --- | --- | --- |
| Location for use | Face | Entire Face | Face | Same |
| OTC or prescription | OTC | OTC | OTC | Same |
| Basic unit characteristics | | | | |
| Design | Hand-held device | Hand-held device | Hand-held device | Same |
| Power supply | 3.7V 1000mAh Li-ion Battery | DC 3.7V 1000mA Li battery | 3.7V 1050mAh Li battery | Same |
| Weight | 135g | 230g | 150g | Different |
| Dimensions | 175.2 * 56 * 37mm | 234.5mm x 30mm x 46mm | 187*65*51mm | Different |
| Irradiation area | 11cm² (±5%) | 12cm²±10% | 30cm²±5% | Similar |
| Software/Firmware/Microprocessor Control? | Yes | Yes | Yes | Same |
| Sterilization | Not required | No publicity | Not required | Same |
| Output specifications | | | | |
| Light source | Light Emitting Diodes (LED) | Light Emitting Diodes (LED) | Light Emitting Diodes (LED) | Same |
| Wavelength | Red: 630nm±5nm | Red: 630±10nm Blue: 415±10nm | Red: 630 ± 3nm Blue: 415 ± 3nm | Same |
| Irradiance | Red: 70mW/cm²±10% | Red: 55mW/cm²±10% Blue: 48mW/cm²±10% | Red: 80mW/cm²±10% Blue: 65mW/cm²±10% | Similar |
| Additional features | | | | |
| Treatment duration | 3 minutes for each part | 3 minutes per target area | 3 minutes per target area | Same |
| Materials of skin-contacting components | Silicone, plastic, glass | ABS Plastic & Stainless Steel | ABS plastic | Different |
| Environment for Operating | Temperature: 5-40°C Humidity: 10-80%RH Atmospheric pressure: 700hPa-1060hPa | Temperature: 5-35°C Humidity: 10-80%RH Atmospheric pressure: 700hPa-1060hPa | Temperature: 5 ~ 40°C Humidity: 10%-80% Atmospheric pressure: 700hPa-1060hPa | Same |
| Biocompatibility feature | Comply with ISO 10993-1, ISO 10993-5, ISO 10993-10 and ISO 10993-23 | Comply with ISO 10993-5:2009 and ISO 10993-10: 2010 | ABS plastic in hand hold part can be considered safety | Same |
{7}
Dongguan Hunter Electronic Technology Co., Ltd.
510(k)s -510 (k) Summary
| Electrical safety | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 | Same |
| --- | --- | --- | --- | --- |
| | IEC 60601-1-11 | IEC 60601-1-11 | IEC 60601-1-11 | |
| | IEC 60601-2-57 | IEC 60601-2-57 | IEC 60601-2-57 | |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | 60601-1-2 | Same |
| Photobiological safety | IEC 62471 | Unknown | IEC 62471 | Same |
## (7) Non-clinical and/or Clinical Tests Summary & Conclusions
Non-clinical testings have been conducted to verify that the PMD Clean Redvolution meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device. The testing results demonstrate that the subject device complies with the following standards:
- IEC 60601-1, Edition 3.2 2020-08 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2, Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-11, Edition 2.1 2020-07 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-57, Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- IEC 62471, First edition 2006-07 Photobiological safety of lamps and lamp systems
- IEC 62133-2, Edition 1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
The device has been tested for biocompatibility, it complies with the following standards:
- ISO 10993-5, Third edition 2009-06-01 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity
- ISO 10993-10, Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23, First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation
We have also conducted:
Software verification and validation test according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices"
Clinical test is not applicable, so there is no clinical data.
Page 4 of 5
{8}
Dongguan Hunter Electronic Technology Co., Ltd.
510(k)s –510 (k) Summary
Based on the above analysis and non-clinical tests performed, the subject device and predicate device have the same indications for use and technological characteristics, it can be concluded that the subject device is as safe, as effective and performs as well as the predicate device.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.