K162098 · Li-Tek Electronic Technology Corporation · OLP · Jan 30, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162098
Device Name
LED Phototherapy Device
Applicant
Li-Tek Electronic Technology Corporation
Product Code
OLP · General, Plastic Surgery
Decision Date
Jan 30, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The red light is intended for the treatment of periorbital wrinkles, and the blue light is intended for the treatment of the mild to moderate inflammatory acne.
Device Story
LED Phototherapy Device (Model PL-120) is a handheld, battery-powered light therapy system for OTC use. Device utilizes two interchangeable LED lamp heads: blue light (415nm) for mild-to-moderate inflammatory acne and red light (630nm) for periorbital wrinkles. User selects lamp head and treatment duration (1-4 minutes or continuous) via device controls. Device is held 2-3cm from skin surface; no direct skin contact occurs. Output consists of photon light energy at specified wavelengths and power densities. Device provides non-invasive treatment; intended to improve skin appearance and manage acne symptoms. Safety features include auto-off timers and compliance with photobiological safety standards.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and performance verified via electrical safety (IEC 60601-1, IEC 60601-1-11), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and software verification/validation.
Technological Characteristics
Handheld LED device; ABS plastic housing; 3.7V 1050mAh rechargeable Li-ion battery. Emits red (630nm±3nm) or blue (415nm±3nm) light. Power density: 80mW/cm² (red), 65mW/cm² (blue). Irradiation area: 30cm². Operates at 2-3cm distance from skin. Microprocessor-controlled with selectable auto-off timers. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, and IEC 62471.
Indications for Use
Indicated for treatment of periorbital wrinkles (red light) and mild to moderate inflammatory acne (blue light) in patients using the device over-the-counter.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 30, 2017
Li-Tek Electronic Technology Corporation % Jet Li Regulation Manager Guangzhou Leta Testing Technology Co.,Ltd 6F, No.1 Tian Tairoad Science City, LuoGang District Guang Zhou City, China
Re: K162098
Trade/Device Name: LED Phototherapy Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: OLP and OHS Dated: December 31, 2016 Received: January 10, 2017
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
#### 510k preparation date: 2017-01-25
# 1. Submitter's Information
#### Sponsor:
- � Company Name: Li-Tek Electronics Technologies
- � Address: No.8~13,the industrial park of Jinshagang, Shixia village, Dalang town,Dongguan city, Guangdong, China
- � Phone: 0769-83117755
- � Fax: 0769-83117759
- Contact Person (including title): Barry Yuan (Quality Director) �
- E-mail:quality5@li-tek.com �
#### Application Correspondent:
- � Guangzhou LETA Testing Technology Co., Ltd.
- � Address: 6F, No.1 TianTairoad, Science City, LuoGang District, GuangZhou City, China
- � Contact Person: Mr. Jet Li
- � Tile: Requlation Manager
- Tel: +86-20-22325619 �
- Email: med-jl@foxmail.com �
#### 2. Subject Device Information
Type of 510(k) submission: Traditional
Common Name: Laser surgical instrument for use in general and plastic surgery and in dermatology.
Light Therapy system
Trade Name: LED Phototherapy Device
Classification Name: Light Based over the Counter Wrinkle Reduction, Over-the-counter
powered light based for acne
Review Panel: General&Plastic Surgery
Product Code: OLP, OHS
Regulation Number: 878.4810
Regulation Class: 2
{3}------------------------------------------------
#### 3. Predicate Device Information
| Sponsor | Home Skinovations Ltd. | Silver Bay, LLC | NutraLuxe MD, LLC |
|----------------------|------------------------|--------------------------|-------------------|
| Device Name | Acne Treatment device | Quasar Calypso | Nutra Light Red |
| 510(k) Number | K121435 | K111286 | K141308 |
| Product Code | OLP | OLP | OHS |
| Regulation<br>Number | 878.4810 | 878.4810 and<br>890.5500 | 878.4810 |
| Regulation Class | 2 | 2 | 2 |
#### 4. Device Description
The LED Phototherapy Device (Model: PL-120) directly applies photon light onto skin surface by making using of specific photon spectrum. Each device is equipped with two LED lamp holder, one is emitting blue light which wavelength at 415nm±3nm. The other lamp emits red light which wavelength at 630nm±3nm. The red light is intended for the treatment of periorbital wrinkles. The blue light is intended for the treatment of the mild to moderate inflammatory acne. The user can change the treatment lamp according to their own needs. It employed a 3.7 V Li-ion battery to provide power, and the internal battery is rechargeable. And the device has five options for setting the auto-off time, as below table listed:
| Mode Selection | Default Duration |
|----------------|-------------------------------------|
| 1 | Auto Off after running for 1 minute |
| 2 | Auto Off after running for 2 minute |
| 3 | Auto Off after running for 3 minute |
| 4 | Auto Off after running for 4 minute |
| L | Continuous working mode |
#### 5. Intended Use / Indications for Use
The red light is intended for the treatment of periorbital wrinkles, and the blue light is intended for the treatment of the mild to moderate inflammatory acne.
#### 6. Materials
There are no patient directly components in the subject device as the following list.
| Component of<br>Device Requiring<br>Biocompatibility | Material of<br>Component | Body Contact<br>Category<br>(ISO 10993-1) | Contact Duration<br>(ISO 10993-1) |
|------------------------------------------------------|--------------------------|-------------------------------------------|-----------------------------------|
| - | - | - | - |
{4}------------------------------------------------
The holding part is made of ABS plastic, which is widely used in various medical devices, and its biological compatibility has been proved to be acceptable in many marketed device. So biological compatility test was waived here for this material. The users should place the device 2-3cm away from face before use it. So the LED Phototherapy device is not used in skin contact.
#### 7. Physical characteristics
| Main Unit Weight | 150g |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| dimensions (L*W*H) | 187*65*51mm |
| Power Source | DC 5V, 1A |
| Maximum power | 1.5W |
| light source | LED |
| Maximum emission wavelength | The red light: 630nm±3nm |
| | The blue light: 415nm±3nm |
| Power density | The red light: 80mW/cm2±10% |
| | The blue light: 65mW/cm2±10% |
| Irradiation area | 30cm2±5% |
| The distance between the lamp to<br>treatment skin surface | 2-3cm |
| Li battery | 3.7V, 1050mAh |
| Maximum charging time | 3 hours |
| work<br>environment | Temperature 5°C-40°C<br>relative humidity 10%-80%<br>atmospheric pressure 700hPa~1060hPa |
| Transportati<br>on and<br>storage<br>conditions | Temperature -10°C-40°C<br>relative humidity 5%-95%<br>atmospheric pressure 600hPa~1060hPa |
### 8. Test Summary
The LED Phototherapy Device has been evaluated the safety and performance by lab bench testing as following:
- � Electrical safety and essential performance test according to IEC 60601-1; IEC 60601-2-57 IEC60601-1-11 standards
- Electromagnetic compatibility test according to IEC 60601-1-2 standard �
- Photobiological safety test according to IEC62471 �
{5}------------------------------------------------
- � Software verification and validation test according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Device"
#### 9. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, and intended use of the LED Phototherapy Device is substantially equivalent to the predicate device quoted above. The differences between the subject device and predicate device do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | | | Remark |
|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device Name<br>and Model | LED Phototherapy<br>Device. Model:PL-120 | Silkn Blue | Quasar Calypso | Nutra Light<br>Red | -- |
| 510(k) Number | Applying | K121435 | K111286 | K141308 | -- |
| Manufacturer | Li-Tek Electronics<br>Technologies | Home<br>Skinovations Ltd | PhotoActif, LLC | NutraLuxe MD,<br>LLC | -- |
| Product Code | OLP | OLP | OLP | OHS | -- |
| Regulation<br>Number | 878.4810 | 878.4810 | 878.4810 and<br>890.5500 | 878.4810 | -- |
| Intended Use | The red light is<br>intended for the<br>treatment of periorbital<br>wrinkles, and the blue<br>light is intended for the<br>treatment of the mild to<br>moderate inflammatory<br>acne. | The Silkn Blue is<br>indicated as an<br>over the counter<br>phototherapy<br>device for the<br>treatment of mild<br>to moderate acne. | The Quasar Calypso<br>C5O is intended to<br>emit energy in the red<br>and blue region of the<br>spectrum, specifically<br>indicated to treat mild<br>to moderate acne<br>vulgaris. | The Nutra Light<br>Red is a non-<br>invasive LED light<br>device is<br>intended/indicated<br>for over- the -<br>counter use for the<br>treatment of<br>periorbital<br>wrinkles, and<br>rhytides | SE |
| Basic Unit Characteristics | | | | | |
| Power<br>Source(s) | 3.7V 1050mAh Li<br>battery | battery | 12 volt wall mount<br>power | internal NI-MH<br>rechargeable<br>battery | SE<br>Note 1 |
| Software/Firm<br>ware/Micropro<br>cessor<br>Control? | Yes | Yes | Yes | No | SE<br>Note1 |
{6}------------------------------------------------
| Elements of<br>Comparison | Subject Device | Predicate Device | | | Rema<br>rk |
|------------------------------------------|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|--------------|
| Power<br>(mW/cm²) | Red light 80±10%<br>Bule light 65±10% | 50 | 65 | 80 | SE<br>Note 1 |
| treatment<br>duration | 3 minutes per target<br>area; 2 treatments per<br>week for 6 weeks | --; | 3 minutes daily,<br>minimum 5 days per<br>week | 3min | SE |
| Compliance<br>with 21 CFR<br>898 | Yes | Yes | Yes | Yes | SE |
| Housing<br>Materials and<br>Construction | ABS plastic | Stainless steels<br>17-4H,<br>Rigid ABS | polycarbonate | medical grade<br>biocompatibility<br>plastics via<br>injection molding | SE |
| wavelength | blue light:<br>415nm±3nm<br>red light: 630nm±3nm | 415±15nm | Blue light 405-<br>420nm,red light<br>628±10nm | The output<br>wavelength of Red<br>is 650 +/- 5 at 80<br>mW/cm2 | SE<br>Note 2 |
| Standards | | | | | |
| Biocompatibilit<br>y | ABS plastic in hand<br>hold part can be<br>considered safety | All user directly<br>contacting<br>materials are<br>compliance with<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | All user directly<br>contacting materials<br>are compliance with<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | All user directly<br>contacting<br>materials are<br>compliance with<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | SE |
| Electrical<br>Safety | Comply with IEC<br>60601-1 | Comply with IEC<br>60601-1 | Comply with IEC / EN<br>6060 1-1 | Comply with IEC<br>60601-1 | SE |
| EMC | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | SE |
# Comparison in Detail(s):
# Note 1:
"Power Source(s)""Power (mW/cm)"is belonging to basic characteristics. Although they are a little different from the predicate device, it will not affect the main function and the intended use of the device. They all also comply with IEC 60601-1, IEC 60601-1-2, IEC62471, for safety requirement, and comply with IEC60601-2-57 for essential requirements. So the differences will not raise any safety or effectiveness issue.
{7}------------------------------------------------
Note 2: "wavelength" of subject device is a little different from the predicate device, they all comply with IEC62471, IEC60601-2-57requirement, so the differences of function specification will not raise any safety or effectiveness issue.
# Finial Conclusion:
The subject device LED Phototherapy Device is substantial equivalence to all predicate devices.
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K162098
Device Name LED Phototherapy Device
Indications for Use (Describe)
The red light is intended for the treatment of periorbital wrinkles, and the blue light is intended for the treatment of the mild to moderate inflammatory acne.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.