K241086 · Avidhrt, Inc. · DQA · May 9, 2025 · Cardiovascular
Device Facts
Record ID
K241086
Device Name
Avidhrt Sense SpO2
Applicant
Avidhrt, Inc.
Product Code
DQA · Cardiovascular
Decision Date
May 9, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
Avidhrt Sense is a handheld spot-checking measurement device, indicated to do the following: • Non-invasive spot checking of functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse waveform (when prescribed by a physician). Avidhrt Sense SpO2 is intended for use in adult patients in the home environment. The Avidhrt Sense SpO2 is not for continuous monitoring, use during motion or for patients with low perfusion. The Avidhrt Sense SpO2 is not intended for pediatric use.
Device Story
Handheld spot-checking pulse oximeter; uses reflectance pulse oximetry to measure functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse waveform. Device pairs with smartphone app for data display. Operated by adult lay users in home settings. Output provides immediate SpO2 and waveform visualization for user/clinician review; assists in health status assessment. Benefits include non-invasive, convenient monitoring of oxygen saturation levels.
Clinical Evidence
Controlled hypoxia/desaturation study per ISO 80601-2-61:2017. 12 healthy volunteers (ages 21-49, 67% female, diverse skin tones). 291 samples collected across 70-100% SaO2 range. ARMS results: 2.44% (70-80%), 1.49% (80-90%), 1.43% (90-100%), and 1.82% (70-100%).
Technological Characteristics
Handheld reflectance pulse oximeter. Connectivity via smartphone app. Complies with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ANSI C63.27, and ISO 10993-1. Performance validated per ISO 80601-2-61.
Indications for Use
Indicated for non-invasive spot checking of functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse waveform in adult patients in home environments. Contraindicated for pediatric use, continuous monitoring, use during motion, or patients with low perfusion.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 9, 2025
Avidhrt Inc.
% Prabhu Raghavan
Principal Consultant
MDQR, LLC
1790 Montemar Way
San Jose, California 95125
Re: K241086
Trade/Device Name: Avidhrt Sense SpO2
Regulation Number: 21 CFR 870.2700
Regulation Name: Oximeter
Regulatory Class: Class II
Product Code: DQA
Dated: May 5, 2025
Received: May 6, 2025
Dear Prabhu Raghavan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K241086 - Prabhu Raghavan
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K241086 - Prabhu Raghavan
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer W. Shih -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K241086 | |
| Device Name Avidhrt Sense SpO2 | |
| Indications for Use (Describe) Avidhrt Sense is a handheld spot-checking measurement device, indicated to do the following: • Non-invasive spot checking of functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse waveform (when prescribed by a physician). | |
| Avidhrt Sense SpO2 is intended for use in adult patients in the home environment. The Avidhrt Sense SpO2 is not for continuous monitoring, use during motion or for patients with low perfusion. The Avidhrt Sense SpO2 is not intended for pediatric use. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Vidhrt
# 510(k) Summary for K241086
Prepared in accordance with the requirements of 21 CFR 807.92
## Submitter Information [807.92(a)(1)]
Submitter/Applicant
Avidhrt Inc.
325 E. Grand River Avenue #225
East Lansing, MI 48823
Applicant Contact:
Prabode Weebadde
Email: prabode@avidhrt.com
Phone: (517) 214-9041
Primary Contact Person
Prabhu Raghavan
Regulatory Consultant for Avidhrt Inc.
Principal Consultant, MDQR LLC.
1790 Montemar Way
San Jose, CA 95125
Phone: 408-316-5707
Email: prabhu@mdqr.solutions
Date Prepared: May 09, 2025
## Device Information [807.92(a)(2)]
Trade Name: Avidhrt Sense SpO2
Common Name: Oximeter
Regulation: 21 CFR §870.2700
Device Class: II
Product Code: DQA
Subsequent Product Code: N/A
## Predicate Information [807.92(a)(3)]
Predicate(s): K181956, Masimo MightySat Rx Fingertip Pulse Oximeter by Masimo Corporation
Common Name: Oximeter
Regulation: 21 CFR §870.2700
Device Class: II
Product Code: DQA
Subsequent Product Code: BZQ
## Device Description [807.92(a)(4)]
Avidhrt Sense SpO2 is a handheld spot-checking measurement device designed for non-invasive spot checking of functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse
Page 1
{5}
510(k) Summary
Traditional 510(k) Premarket Notification
K241086 Avidhrt Sense SpO2
waveform (when prescribed by a physician). Avidhrt Sense SpO2 is used in conjunction Avidhrt Sense SpO2 smart phone app that displays the SpO2.
## Indications for use [807.92(a)(5)]
Avidhrt Sense SpO2 is a handheld spot-checking measurement device, indicated to do the following:
- Non-invasive spot checking of functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse waveform (when prescribed by a physician).
Avidhrt Sense SpO2 is intended for use in adult patients in the home environment. The Avidhrt Sense SpO2 is not for continuous monitoring, use during motion or for patients with low perfusion. The Avidhrt Sense SpO2 is not intended for pediatric use.
## Substantial Equivalence
Avidhrt utilizes the Masimo MightySat Rx Fingertip Pulse Oximeter (K181956) as a predicate for pulse oximetry functions.
## Comparison of intended use and indications for use
The Avidhrt Sense SpO2 is indicated for noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) using pulse oximetry. The Masimo MightySat Rx Fingertip Pulse Oximeter is “indicated for the noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult and pediatric patients during both no motion and motion conditions, and for patients who are well or poorly perfused” and is also “indicated for the noninvasive spot checking of respiration rate (RRp) for adult patients”. As such, the subject device has a subset of intended use of the predicate device, i.e., “noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) in adults”.
With regards to the intended use patient population, the subject device is intended for use by adults. This is a subset of the intended use population of the K181956 predicate, which includes both adults and pediatric patients. Both devices are intended for use by adult lay users independently or guided by health care professionals (HCPs) in home and non-acute clinical environments. The predicate includes additional use in acute environments, which is not present in the subject device. The predicate device’s SpO2 function is intended for use during both motion and no motion, and for patients who are well and poorly perfusion. The subject device is indicated for a subset of the predicate’s functions, i.e., during conditions of no motion and for patients who are well perfused.
Page 2
{6}
510(k) Summary
Traditional 510(k) Premarket Notification
K241086 Avidhrt Sense SpO2
The predicate is also indicated for noninvasive spot checking of respiration rate of adult patients, which is not included in the subject device.
As such, the subject device has the same intended use and substantially equivalent indications for a subset of the indications for use of the predicate device.
## Comparison of technological characteristics [807.92(a)(6)]
Technological characteristics of the Avidhrt Sense SpO2 pulse oximetry functions closely align with those of the MightySat device. They both use pulse oximetry, with the primary difference being the minor differences in the method (reflectance vs. transmittance) used to perform this measurement. The difference of method of the pulse oximetry between the Avidhrt Sense SpO2 and MightySat devices do not raise different questions of safety or effectiveness. The technological characteristic difference is addressed by performance testing using the consensus standard ISO 80601-2-61, and FDA Guidance: Pulse Oximeters-Premarket Notification Submissions: Guidance for Industry and Food and Drug Administration Staff.
In conclusion, Avidhrt Sense SpO2 has the same intended use as the predicate device, and any differences in technological characteristics do not raise different questions of safety or effectiveness.
## Performance Data [807.92(b)]
All necessary testing was conducted on Avidhrt Sense SpO2 to support a determination of substantial equivalence to the predicate device, which included:
- Electrical safety per IEC 60601-1 and IEC 60601-1-11,
- Electromagnetic compatibility per IEC 60601-1-2, wireless coexistence testing per ANSI C63.27,
- Biocompatibility testing per ISO 10993-1,
- Pulse oximetry testing per IEC 80601-2-61, which included clinical performance testing to evaluate the accuracy of the SpO2 measurement (hypoxia study per IEC 80601-2-61),
- Evaluating the performance of the subject device to reject conditions of low perfusion,
- Evaluating the ability of the users to safely operate the device in a human factors study.
As noted above, a clinical study was performed to support the device:
- A controlled hypoxia or desaturation study, was conducted to provide human validation data of pulse oximeters according to ISO 80601-2-61:2017 and FDA guidance on pulse oximeters. The study included fourteen (12) healthy volunteer subjects, ages 21 - 49, with 8 subjects identified as female (67%) and 4 subjects as male (33%). The subjects included a range of ethnicities. The 12 subjects had a range of skin tones, including at least 25% (3/12) with dark skin tones, thus exceeding the requirements in the standard and FDA guidance of at least 2 or 15%. Hypoxia was then induced to different and stable levels of oxyhemoglobin saturation (between 70-100%) by having subjects breathe mixtures of nitrogen, room air, and carbon dioxide. Each plateau level of oxyhemoglobin saturation was maintained for at least 30 seconds or until reference pulse
Page 3
{7}
510(k) Summary
Traditional 510(k) Premarket Notification
K241086 Avidhrt Sense SpO2
oximeters readings were stable. Two arterial blood samples were then obtained, approximately 30 seconds apart. Each stable plateau therefore was maintained for at least 60 seconds with SpO2 fluctuating by less than 2-3%. The plateaus were nominally at 100%, room air saturation, 93%, 90%, 87%, 85%, 82%, 80%, 77%, 75% and 70%. A total of 291 samples were obtained at the saturation plateaus across this span.
Avidhrt Sense SpO2 oximeter data were taken as 5-second averages at each plateau to arterial blood analysis for comparison. According to equation 1 in ISO 80601-2-61:2017, $A_{\text{RMS}}$ for the Avidhrt Sense SpO2 was calculated for each range of the following SaO2 ranges of 70-80%, 80-90%, 90-100% and 70-100%. The results were 2.44%, 1.49%, 1.43%, and 1.82%, respectively, which satisfied the prespecified performance criteria.
## Conclusions [807.92(b)(3)]
The results of these testing therefore demonstrate that the device performs as intended and confirm that the technological differences between the subject device and the predicate device do not raise different questions of safety and effectiveness, and that the device is as safe and as effective for its intended use as the predicate device.
Page 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.