Retrospective analysis of clinical data from 59 hospitalized subjects
A retrospective analysis of hospitalized patient data was used to validate the performance of the respiration rate (RRp) measurement feature of the device.
Validation of respiration rate (RRp) measurement performance
Indications for Use
The Masimo MightySat Rx Fingertip Pulse Oximeter is intended for hospitals, hospital-type facilities, home environments, and transport. The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for the noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult and pediatric patients during both no motion and motion conditions, and for patients who are well or poorly perfused. The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for the noninvasive spot checking of respiration rate (RRp) for adult patients.
Device Story
Fingertip pulse oximeter; spot-check monitoring; hospital, home, and transport use. Inputs: photoplethysmogram (PPG) signals via two-wavelength sensor on finger. Processing: Masimo SET technology; analyzes cyclic variations in PPG amplitude, frequency, and intensity to derive SpO2, pulse rate, perfusion index, pleth variability index, and respiration rate (RRp). Output: OLED display of vital signs and waveforms. Operation: automatic power-on/off via finger insertion; optional Bluetooth for data transfer to mobile devices. Benefits: noninvasive, portable vital sign assessment; aids clinical decision-making by providing respiratory and oxygenation status without alarms.
Two-wavelength optical sensor; OLED display; plastic housing; powered by two AAA alkaline batteries. Measures SpO2, PR, Pi, PVi, and RRp. Connectivity: Bluetooth LE. Standards: ISO 80601-2-61 (SpO2/PR), IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use). Software: moderate level of concern.
Indications for Use
Indicated for noninvasive spot checking of SpO2 and pulse rate in adult and pediatric patients (well or poorly perfused, motion or no motion) and respiration rate (RRp) in adult patients.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Nellcor Bedside Respiratory Patient Monitoring System (K141518)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Masimo Corporation Sindura Penubarthi Regulatory Affairs Manager 52 Discovery Irvine, California 92618
Re: K181956
Trade/Device Name: Masimo MightySat Rx Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA, BZQ Dated: December 20, 2018 Received: December 26, 2018
# Dear Sindura Penubarthi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd D. Courtney -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K181956
Device Name
Masimo MightySat Rx Fingertip Pulse Oximeter
Indications for Use (Describe)
The Masimo MightySat Rx Fingertip Pulse Oximeter is intended for hospital-type facilities, home environments, and transport.
The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for the noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult and pediatric patients during both no motion and motion conditions, and for patients who are well or poorly perfused.
The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for the noninvasive spot checking of respiration rate (RRp) for adult patients.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in a bold, sans-serif font. To the right of the logo is the text "MASIMO CORPORATION" in a smaller, lighter font, followed by the address "52 Discovery, Irvine, CA 92618".
| Submitter and Address of<br>Manufacturing Facility: | Masimo Corporation<br>52 Discovery<br>Irvine, CA 92618<br>Phone: (949) 297-7000<br>FAX: (949) 297-7199 |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Sindura Penubarthi |
| Date: | December 20, 2018 |
| Contact: | Sindura Penubarthi |
| Trade Name: | Masimo MightySat Rx Fingertip Pulse Oximeter |
| Common Name: | Oximeter |
| Classification Regulation/ Product<br>Code: | 21 CFR 870.2700, Class II/DQA |
| Additional Product Code: | BZQ |
| Establishment Registration<br>Number: | 2031172 |
| Reason for Premarket Notification: | Additional or Expanded Indications |
| Primary Predicate Device: | K141518 - Nellcor Bedside Respiratory Patient Monitoring<br>System |
| Secondary Predicate Device: | K150314 – Masimo MightySat Rx Fingertip Pulse Oximeter |
| Performance Standards | No performance standards for the above device have been<br>promulgated pursuant to Section 514 of the Food and Drug<br>Administration Modernization Act of 1997 (FDAMA). |
# Device Description
The subject device, MightySat Rx, was previously cleared under K150314 as a fingertip pulse oximeter that includes Masimo SET technology for the measurement of functional oxygen saturation of arterial hemoglobin (SpOz), pulse rate (PR), calculation of Perfusion Index (Pi) and optional Pleth Variability Index (PVi) in adults and pediatrics. The current submission concerns the MightySat Rx's measurement of respiration rate through photoplethysmogram analysis (designated as RRp).
Like the secondary predicate (K150314), the device is a spot check pulse oximeter and does not include alarms. The device has the combined function of a pulse oximeter monitor and a reusable sensor. It includes an OLED color display, enclosed by plastic housing and powered by two alkaline AAA batteries.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".
The MightySat Rx also includes optional Bluetooth wireless technology for the wireless transfer of patient data to mobile devices, such as a smartphone.
### Intended Use/Indications for Use
The Masimo MightySat Rx Fingertip Pulse Oximeter is intended for hospitals, hospital-type facilities, home environments, and transport.
The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for the noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult and pediatric patients during both no motion and motion conditions, and for patients who are well or poorly perfused.
The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for the noninvasive spot checking of respiration rate (RRp) for adult patients.
The Indications for Use statement for the MightySat Rx is similar to the primary predicate (K141518). Both the subject and predicate device have the same intended use for vital sign monitoring, by transmitting and detecting light through tissue in the finger.
### Device Specifications
The MightySat Rx specifications are as follows:
| FEATURE | SPECIFICATION |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display | |
| Display Type | OLED color display |
| Display Range | Oxygen Saturation (SpO2): 0-100%<br>Pulse Rate (PR): 25-240 beats per minute (BPM)<br>Perfusion Index (Pi): 0.02-20%<br>Pleth Variability Index (PVi): 0-100%<br>Respiration rate (RRp): : 4-70 respirations per minute<br>(RPM) |
| Display Waveform | Photoplethysmogram<br>Signal IQ |
| Display Resolution | SpO2: 1%<br>PR: 1 BPM<br>RRp: 1 RPM |
| Measurement Accuracy in<br>Accordance with ISO 80601-2-61 | |
| SpO2, No Motion | 70 – 100%, 2%, ARMS, adults/pediatrics |
| SpO2, Motion | 70 – 100%, 3% ARMS, adults/pediatrics |
| SpO2, Low Perfusion | 70 – 100%, 2%, ARMS, adults/pediatrics |
| Pulse Rate, No Motion | 25 – 240 BPM, 3 BPM ARMS, adults/pediatrics |
| Pulse Rate, Motion | 25 – 240 BPM, 5 BPM ARMS, adults/pediatrics |
| Pulse Rate, Low Perfusion | 25 – 240 BPM, 3 BPM ARMS, adults/pediatrics |
| RRp | 4-70 RPM, 3 RPM ARMS, 1 RPM Mean Error, adults |
| Power | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.
| FEATURE | SPECIFICATION |
|-----------------------------|-----------------------------------------------|
| Internal battery | Alkaline “AAA” batteries |
| Interface | |
| Wireless | Bluetooth LE |
| Mechanical | |
| Enclosure Material | Plastic |
| Dimensions/Weight | 2.9" x 1.6" x 1.2" (7.4 cm x 4.1 cm x 3.0 cm) |
| Weight | 0.16 lbs (73 g) |
| Environmental | |
| Operating Temperature | 5°C to +40°C, ambient humidity |
| Storage Temperature | -40°C to +70°C, ambient humidity |
| Operating/ Storage Humidity | 10% to 95%, non-condensing |
| Atmospheric Pressure | 540 mBar to 1060 mBar |
| Mode of Operation | |
| Mode of Operation | Spot Check |
### Technological Characteristics
### Principle of Operations
The MightySat Rx's principle of operation for Masimo SET technology is similar to the primary predicate (K141518). Specifically, the Masimo SET technology uses a two-wavelength sensor to measure the indicated parameters based on light absorption principles of oxygenated blood and deoxygenated blood which generates a photoplethysmogram. Respiration rate (RRp) measures the respiration rate by analyzing cyclic variations in the photoplethysmogram due to respiration.
### Mechanism of Action for Achieving the Intended Effect
The MightySat Rx's mechanism of action is similar to the primary predicate (K141518) and the same as the secondary predicate (K150314). MightySat Rx automatically turns on when the device is opened. The device is then positioned on the patient's finger. Once the device is applied to the finger, it collects and processes physiological signals, and then displays Masimo SET measurements on the device's display screen. The MightySat Rx automatically turns off after removing the device from the finger.
### Summary of Technological Characteristics of Subject Device Compared to the Primary Predicate
#### Similarities and Differences between Predicate and Subject Device, MightySat Rx
The subject device, MightySat Rx, and the Primary Predicate (K141518), have the following key similarities:
- · Both devices have the same intended use as vital sign monitors;
- · Both devices have the same technological characteristics with respect to pulse oximetry with SpO2 and PR measurements;
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red symbol on the left, followed by the word "Masimo" in black, bold letters. To the right of the word "Masimo" are the words "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.
- · Both devices can display photoplethysmogram and Signal IQ waveforms;
- · The subject device can calculate a respiration rate (RRp);
- · Both devices have the same measurement site;
The subject device. MightySat Rx. and the Primary Predicate (K141518), have the following kev difference:
- · The subject device is intended for home environments;
- · The subject device's intended use has been updated to indicate for transport, whereas the secondary predicate (K150314) used the term mobile environments;
- · Device includes optional wireless Bluetooth feature for transferring patient information from the device to a mobile device (e.g. smartphone), the same as the secondary predicate (K150314);
- · Both devices are internally powered, except subject device uses "AAA" alkaline batteries the same as the secondary predicate (K150314); and
- · Subject device provides Perfusion Index (Pi) and Pleth Variability Index (PVi) the same as the secondary predicate (K150314).
The sole technological difference between the subject device and the secondary predicate (K150314) is the addition of the RRp measurement. The technology change is a software change which utilizes the same hardware as the secondary predicate. The difference does not raise different questions of safety and effectiveness. This is supported by the primary predicate, the Nellcor™ Bedside Respiratory Patient Monitoring System (K141518). The primary predicate and subject device are both part of the same DQA and BZQ product codes. Both the subject device and predicate device utilize similar principles of operation which measures respiration rate from the photoplethysmogram obtained by transmitting and detecting of light though a finger. Both devices rely on the technological principle of respiration-induced variations in the photoplethysmogram. Specifically, both devices rely on the general principle that respiration modulates the frequency, amplitude, and intensity of the photoplethysmograph as described in Karlen, W., Raman, S., Ansermino, J.M., Dumont, G.A .: Multiparameter respiratory rate estimation from the photoplethysmogram. IEEE Transactions on Biomedical Engineering 60(7), 1946-1953 (2013). The subject and primary predicate measure respiration rate from these modulations of the photoplethysmogram; however, they may differ in the method in which the signal is processed. Both signal processing methods, have similar performance.
### Performance Data
The following performance data were provided in support of substantial equivalence determination.
#### Biocompatibility Testing
There was no change in the patient contacting material as part of this submission. The MightySat Rx includes patient contacting materials as the sensor is positioned on the patient's finger. It is substantially the same as the cleared secondary predicate (K150314) with respect to biocompatibility. Therefore, the testing performed and submitted with K150314 with respect to ISO 10993-1, ISO 10993-5 and ISO 10993-10 supports the substantial biocompatibility
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the company name is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.
equivalence evaluation.
### Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and Cleaning
There was no change to the hardware as part of this submission. The subject device is substantially the same as previously cleared under K150314. Therefore, the testing performed and submitted with K150314 with respect to the IEC 60601-1 standard for basic safety and essential performance, the IEC 60601-2-1 standard for EMC, environmental, mechanical, and cleaning applies to the subject device. Because the subject device includes the added environments of home and transport, additional testing was performed regarding EMC immunity for the home and transport environment in accordance with the FDA's guidance, Design Considerations for Devices Intended for Home Use dated November 24, 2014 and IEC 60601-1-2:2014.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005. The software for this device was considered as a "moderate" level of concern, because a failure or latent flaw in the software could directly result in minor to moderate injury to the patient similar to the predicate.
#### Wireless Testing
There is no difference between the wireless capabilities of the subject device and the predicate device. The subject device relies on the wireless testing demonstrating the connectivity (i.e. Bluetooth pairing) between MightySat Rx and mobile devices, such as smartphones, submitted in secondary predicate device, cleared MightySat Rx (K150314). That testing was done in accordance with FDA Guidance, Radio-Frequency Wireless Technology in Medical Devices, dated August 14, 2013.
#### Human Factors Usability Testing
The use-related changes from the predicate device to the subject device are: (1) the addition of the RRp measurement to the MightySat Rx graphical interface, and (2) the expanded intended use to cover home use. Because the addition of the RRp measurement does not impact the mechanism of action of the MightySat Rx and only makes a minor change to the layout to accommodate the additional parameter, the human factors and usability risk were found to be acceptable based upon the previous testing conducted as part of the predicate device, cleared MightySat Rx (K150314).
To address the use by home users, the human factors and usability testing was conducted using non-clinicians to support the acceptability of the human factors and usability risks associated with home users. To further minimize risks associated with use in the home environment, the subject device was subjected to additional testing for home use per IEC 60601-1-11:2010 and in accordance with FDA Guidance, Guidance for Industry and Food and Drug Administration Staff
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".
- Design Considerations for Devices Intended for Home Use, dated November 24, 2014.
## Non-clinical Testing
In K150314. Masimo included non-clinical testing of the predicate device in accordance with ISO 80601-2-61 for SpO2, pulse rate and perfusion index. Masimo performed additional, non-clinical testing in accordance with Masimo design control requirements and quality system with respect to RRp to ensure that the specifications of the subject device were met.
## Clinical Testing
In K150314, Masimo included clinical testing of the predicate device for the SpO2 measurement. Because there was no change to the SpO2 measurement function of the device as part of this submission, the clinical study for SpO2 was not repeated.
For this submission to add the RRp. Masimo performed two additional clinical tests of the MightySat Rx to validate the measurement performance of respiration rate (RRp) measurement. The first study was a prospective analysis to validate the accuracy of the RRp measurement on 28 healthy volunteers. The second study was a retrospective analysis of data from 59 hospitalized subjects to validate the performance of the RRp measurement on subjects with different clinical conditions. Masimo's clinical tests measured RRp against manual, clinician-scored capnograms (obtained from a Capnostream20, Oridion, Needham, MA, K060065). The results from the prospective clinical study and retrospective clinical study demonstrated substantial equivalence of the MightySat Rx to the primary predicate.
### Conclusion
The results of the testing demonstrate that all requirements and performance specifications were satisfied and the subject device is substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.