Dental Zirconia Blank are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals.
Device Story
Dental Zirconia Blanks are pre-sintered zirconia blocks/rods used to fabricate dental restorations; crowns, bridges, veneers, inlays, and onlays. Device consists of yttria-stabilized tetragonal zirconia (Y-TZP) powder; pre-shaded versions include inorganic pigments (Fe2O3, Er2O3, MnO) for tooth-color matching. Used in dental laboratories or clinics; operated by dental professionals. Blanks are milled via CAD/CAM or manual systems to desired patient-specific geometry; then sintered at temperatures >1480°C to achieve final material properties. Final restoration provides durable, biocompatible tooth replacement. Benefits include structural integrity and aesthetic color matching to natural dentition.
Clinical Evidence
No clinical data. Bench testing performed per ISO 6872:2015 (flexural strength, density, fracture toughness, etc.) and biocompatibility testing per ISO 10993 series (cytotoxicity, irritation, sensitization, systemic toxicity, genotoxicity, and implantation) demonstrated safety and performance equivalent to the predicate.
Technological Characteristics
Materials: Yttria-stabilized zirconia (ZrO2, Y2O3, HfO2, Al2O3 >=99.0%) per ISO 13356; inorganic pigments (Fe2O3, Er2O3, MnO). Form factor: Blocks and rods. Processing: CAD/CAM or manual milling followed by sintering >1480°C. Non-sterile. Performance conforms to ISO 6872.
Indications for Use
Indicated for dental restorations (crowns, bridges, veneers, inlays/onlays) in patients requiring prosthetic dental devices, processed by dental professionals or laboratories using CAD/CAM or manual milling.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
May 20, 2024
Hunan Ceramaster Material Technology Co., Ltd Bin Wang General Manager 103, Building B4, Dreamworks, No.1 Lantian North Road, Xingsha Industrial Base Changsha, Hunanng 410137 CHINA
Re: K240772
Trade/Device Name: Dental Zirconia Blank (HT-plus, ST-C, ST-ML, SHT-C, SHT-ML, UT-C, UT-ML, Pro- ML) Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: March 21, 2024 Received: March 21, 2024
Dear Bin Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K240772
#### Device Name
Dental Zirconia Blank (HT-plus, ST-C, ST-ML, SHT-C, SHT-ML, UT-C, UT-ML, Pro- ML)
Indications for Use (Describe)
Dental Zirconia Blank are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals.
| Type of Use (Select one or both, as applicable)<br>Over-The-Counter Use (21 CFR 801 Subpart C)<br> X Prescription Use (Part 21 CFR 801 Subpart D) | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995.<br>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | |
| Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff(@fda.hhs.gov | | | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number." | | | |
{4}------------------------------------------------
#### K240772
HUNAN CERAMASTER MATERIAL TECHNOLOGY CO.,LTD
#### Section 4 - 510(k) Summary
#### Date of Summary Preparation: May 09, 2024
#### 1. Submitter's Identifications
Submitter's Name: HUNAN CERAMASTER MATERIAL TECHNOLOGY CO.,LTD Address: 103, Building B4, Dreamworks, No.1 Lantian North Road, Xingsha Industrial Base, Changsha ,Hunan CHINA Zip Code: 410137 Contact Person: Bin Wang Contact Title: General Manager Contact E-mail Address: eckes2(@gg.com Tel: +86-731-88218650
#### 2. Correspondent's Identifications
Correspondent's Name: Guangzhou Junyi Information Technology Co., Ltd. Address: Room 304, Building A, No. 62 Nanyun 2nd Road, Science Town, Huangpu District, Guangzhou City, Guangdong, 510663, China ZIP Code: 510663 Contact Person: Shanfeng Jiang Contact Title: Regulation Control Manager Contact E-mail Address: jiang13620586569@126.com Tel: +86-20-82329549 Fax: +86-20-82329549
#### 3. Name of the Device
Device Classification Name: Powder, Porcelain Regulation Description: Porcelain powder for clinical use Common Name: Dental Zirconia Ceramics Trade Name: Dental Zirconia Blank (HT-plus, ST-C, ST-ML, SHT-C, SHT-ML, UT-C, UT-ML, Pro- ML ) Regulation Medical Specialty: Dental Review Panel: Dental Product Code: EIH Regulation Number: 21 CFR 872.6660 Device Classification: Class II
#### 4. The Predicate Devices
Predicate device: K141724 Upcera Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank Liaoning Upcera Company Limited
{5}------------------------------------------------
### 5. Device Description
Dental Zirconia Blank are derived from zirconia powder that has been processed into their final shapes. These blanks are then being further fabricated into all-ceramic restorations such as crowns, bridges, veneers, inlay/onlay. The zirconia powder is composed of ZrO2+ Y2O3+ HfO2+ Al2O3 with its composition conforms to ISO 13356, Implants for surgery - Ceramic materials based on yttriastabilized tetragonal zirconia (Y-TZP). The performance of the dental blanks conforms to ISO 6872. Dentistry - Ceramic materials.
Dental Zirconia Blank are derived from the same Zirconia powder as the regular Dental Zirconia Blank with the addition of very small amount of inorganic pigments before the composite material is processed into their final net shapes. These blanks are then being further fabricated into all-ceramic restorations such as crowns, bridges, veneers, inlay/onlay. The purpose of the inorganic pigments is to generate the color on the prosthetic dental devices, after sintering at dental labs, that matches natural color of patient's teeth. The performance of the dental blanks conforms to ISO 6872, Dentistry - Ceramic materials.
#### 6. Indications for Use
Dental Zirconia Blank are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals.
### 7. Summary of Substantial Equivalence
| | Proposed Device | Predicate device | Comparison |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510k Number | K240772 | K141724 | ------ |
| Product Code | EIH | EIH | Same |
| Proprietary Name | Dental Zirconia<br>Blank (HT-plus,<br>ST-C, ST-ML,<br>SHT-C, SHT-ML,<br>UT-C, UT-ML,<br>Pro- ML) | Upcera Dental<br>Zirconia Blank &<br>Dental Zirconia<br>Pre-Shaded Blank | ------ |
| Model: | HT-plus, ST-C,<br>ST-ML, SHT-C,<br>SHT-ML, UT-C,<br>UT-ML, Pro- ML | ------ | ------ |
| Manufacturer | HUNAN<br>CERAMASTER<br>MATERIAL<br>TECHNOLOGY | Liaoning Upcera<br>Company Limited | ------ |
| | CO.,LTD | | |
| Indications for<br>Use | Dental Zirconia<br>Blank are used for<br>dental restorations<br>using different<br>CAD/CAM or<br>manual milling<br>machines. All<br>blanks are<br>processed through<br>dental laboratories<br>or by dental<br>professionals. | Upcera Dental<br>Zirconia Blank &<br>Dental Zirconia<br>Pre-Shaded Blank<br>are used for dental<br>restorations using<br>different<br>CAD/CAM or<br>manual milling<br>machines. All<br>blanks are<br>processed though<br>dental laboratories<br>or by dental<br>professionals. | Same |
| Basic design | Blocks, and rods | Blocks, disc, and<br>rod | Same |
| Materials | Regular:<br>Zirconia (ZrO2+<br>Y2O3+ HfO2+<br>Al2O3≥99.0%)<br><br>Pre-shaded:<br>Zirconia (ZrO2+<br>Y2O3+ HfO2+<br>Al2O3≥99.0%)<br><br>Inorganic pigments<br>(Fe2O3, Er2O3 and<br>MnO <2.0%) | Regular:<br>Zirconia (ZrO2+<br>Y2O3+ HfO2+<br>Al2O3≥99.0%)<br><br>Pre-Shaded:<br>Zirconia (ZrO2+<br>Y2O3+ HfO2+<br>Al2O3≥98.0%)<br><br>Inorganic pigments<br>(Fe2O3, Pr2O3, and<br>Er2O3 <2.0%) | Similar¹ |
| Processing | Sintering at<br>temperature: ><br>1480 °C | Sintering at<br>temperature: ><br>1500 °C | Similar² |
| Dimension | Various | Various | Same |
| Single Use | Yes | Yes | Same |
| Color | None, and<br>Pre-shaded (for<br>pre-shaded series) | None, and<br>Pre-shaded (for<br>pre-shaded series) | Same |
| Sterile | Non-sterile | Non-sterile | Same |
{6}------------------------------------------------
The proposed device is essentially identical to the predicate devices in terms of
{7}------------------------------------------------
indication for use, design between our device and the predicate devices. The minor differences are that as below:
Note 1: Zirconia (ZrO2+ Y2O3+ HfO2+ Al2O3) of proposed device which is not less than 99.0% is higher than that of the predicate device. The three inorganic pigments of the proposed device contain Fe203, Er2O3 and MnO, while those of the predicate device are Fe2O3, Pr2O3, and Er2O3. These are minor differences and the biocompatibility testing of the overall proposed device passed.
Note 2: Processing (Sintering at temperature) of the proposed device is slightly lower 5°C than that of the predicate device. This is minor difference and the performance testing of the proposed device passed.
#### 8. Summary of Non-Clinical Testing
Bench testing was performed per ISO 6872:2015 and internal procedures to ensure that the Dental Zirconia Blank met its specifications. All tests were verified to meet acceptance criteria. Test results on radioactivity, flexural strength, chemical solubility, linear thermal expansion coefficient, freedom from extraneous materials, uniformity, shrinkage factor, pre-sintered density, sintered density, and fracture toughness of the subject device are very similar to predicate device. Biocompatibility testing was performed to verify the equivalent safety of the materials that are used.
According to ISO 10993-1:2018 and ISO7405:2018, we evaluated and conducted the compatibility test for the proposed device. The following table shows the biocompatibility testing results.
| Item | Proposed device | Result |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Cytotoxicity<br>(ISO 10993-5:2009) | Under the conditions of this study, the test article<br>was non cytotoxic for 2 h and 24 h in the filter<br>diffusion method.<br>Under the conditions of this study, the test article<br>was accepted in the agar diffusion method.<br>Under the conditions of this study, the test article<br>has no potential toxicity to L-929 cells. | Pass |
| Oral Mucosa Irritation<br>(ISO 10993-23:2021) | The test article has no potential oral mucosa<br>irritation in the Syrian hamsters. | Pass |
| Skin Sensitization<br>Test<br>(ISO 10993-10:2021) | The test article showed no evidence of causing<br>delayed dermal contact sensitization in the guinea<br>pig. | Pass |
| Subacute Toxicity<br>Test<br>(ISO 10993-11:2017) | The test result showed that the test article did not<br>induce subacute systemic toxicity in rats under<br>this condition . | Pass |
| Subacute Systemic<br>Toxicity Test<br>(ISO 10993-11:2017) | The test result showed that the test article did not<br>induce subacute systemic toxicity in rats under<br>this condition. | Pass |
| Acute Systemic | The test article showed no evidence of causing | Pass |
| | Table 2 Biocompatibility testing | |
|--|----------------------------------|--|
| | | |
{8}------------------------------------------------
## HUNAN CERAMASTER MATERIAL TECHNOLOGY CO.,LTD
| Toxicity<br>(ISO 10993-11:2017) | acute system toxicity in the ICR mice. | |
|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| In Vitro Mammalian<br>Cell Gene Mutation<br>Test<br>(ISO 10993-3:2014) | Under the conditions of this study, the test article<br>is considered non-mutagenic. | Pass |
| In vitro Mammalian<br>Chromosome<br>Aberration Test<br>(ISO 10993-3:2014) | Under the conditions of this study, the test article<br>did not induce structural chromosome aberrations<br>in cultured CHL cells. | Pass |
| Bacterial<br>Reverse<br>Mutation Test<br>(ISO 10993-3:2014) | Under the conditions of this study, the number of<br>reverting colonies in the test article group is not<br>equal to or greater than 2 times that of the<br>spontaneous control, so the test article have no<br>potential mutagenesis. | Pass |
| Muscle Implant<br>(ISO 10993-6:2016) | The test result showed that the test article did not<br>induce local effects after implantation of<br>biomaterials in rabbits under this condition. | Pass |
Note: Testing were Performed on Dental Zirconia Blank ST-C with Pre-Shaded powder C4 (C4 is deep color, see the VITA 16 color shade guide picture) to cover the regular and pre-shaded zirconia blanks.
#### 9. Clinical Test Conclusion
No clinical study is included in this submission.
#### 10. Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that subject device Dental Zirconia Blank performs as well as or better than the legally marketed predicate device K141724 Upcera Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank. Dental Zirconia Blank is substantial equivalent to the legally marketed predicate device K141724 Upcera Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.