K240699 · Spineart SA · OVE · May 10, 2024 · Orthopedic
Device Facts
Record ID
K240699
Device Name
SCARLET® AC-Ti
Applicant
Spineart SA
Product Code
OVE · Orthopedic
Decision Date
May 10, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
SCARLET® AC-Ti cages are indicated for use in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage. SCARLET® AC-Ti cages are intended to be used at one or two contiguous levels from C2-T1 to facilitate intervertebral body fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. When used with two bone screws, the SCARLET® AC-Ti Secured Anterior Cervical Cage is intended to be used as a standalone system and requires no additional supplementary fixation systems. When used with two anchors or without two bone screws, the SCARLET® AC-Ti cages are to be used with supplemental fixation which has been cleared by the FDA for use in the cervical spine. The SCARLET® AC-Ti hyperlordotic (12° lordosis) cages are to be used with bone screws and/or anchors and additional supplemental fixation system that has been cleared by the FDA for use in the cervical spine.
Device Story
SCARLET® AC-Ti is a cervical intervertebral body fusion device with integrated fixation. It provides mechanical support and stabilization to the cervical spine to facilitate fusion. The device is a box-shaped titanium alloy spacer with a central cavity for bone graft, implanted via an anterior approach. It features integrated fixation via bone screws and/or anchors guided through tunnels in the spacer. Used by surgeons in clinical settings, the device maintains disc space height. The surgeon selects appropriate size, height, footprint, and lordosis based on patient anatomy. The device is MR conditional. Benefits include stabilization of the cervical spine and promotion of intervertebral fusion.
Clinical Evidence
No clinical data. Evidence consists of bench testing (static/dynamic axial compression, shear-compression, torsion per ASTM F2077-22; subsidence per ASTM F2267-22; expulsion characterization) and human cadaveric trials. MRI safety evaluated per FDA guidance using ASTM F2052, F2213, F2119, F2182, and F2503.
Technological Characteristics
Medical grade titanium alloy (ASTM F136). Additive manufacturing (SLM) per ASTM F3001. Monolithic box-shaped spacer with central cavity and integrated fixation tunnels. Gamma sterilized. MR conditional.
Indications for Use
Indicated for skeletally mature patients with cervical degenerative disc disease (discogenic neck pain) at one or two contiguous levels (C2-T1) failing 6 weeks of non-operative treatment. Requires supplemental fixation unless used as a standalone system with two bone screws.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
ChoiceSpine Blackhawk Ti Cervical Spacer System (K203311, K223869)
Atlas Spine Expandable Cervical Standalone Interbody System (K192570)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
May 10, 2024
SPINEART SA Estelle Lefeuvre Regulatory & Market Access Manager Chemin du Pré-Fleuri 3 Plan-les-Ouates, 1228 Switzerland
Re: K240699
Trade/Device Name: SCARLET® AC-Ti Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE, ODP Dated: March 12, 2024 Received: March 14, 2024
Dear Estelle Lefeuvre:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K240699
Device Name SCARLET® AC-Ti
#### Indications for Use (Describe)
SCARLET® AC-Ti cages are indicated for use in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Patients should have at least six (6) weeks of non-operative treatment with an intervertebral cage. SCARLET® AC-Ti cages are intended to be used at one or two contiguous levels from C2-T1 to facilitate intervertebral body fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. When used with two bone screws, the SCARLET® AC-Ti Secured Anterior Cervical Cage is intended to be used as a standalone system and requires no additional supplementary fixation systems.
When used with two anchors or without two bone screws, the SCARLET® AC-Ti cages are to be used with supplemental fixation which has been cleared by the FDA for use in the cervical spine.
The SCARLET® AC-Ti hyperlordotic (12° lordosis) cages are to be used with bone screws and/or anchors and additional supplemental fixation system that has been cleared by the FDA for use in the cervical spine.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span> <span style="color: rgb(0, 0, 0);"> <span> <span style="font-variant-ligatures: normal;"> <span style="font-variant-caps: normal;"> <span style="letter-spacing: normal;"> <span style="orphans: 2;"> <span style="text-align:left;"> <span style="text-indent: 0px;"> <span style="text-transform: none;"> <span style="white-space: normal;"> <span style="widows: 2;"> <span style="word-spacing: 0px;"> <span style="display: inline !important;"> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span> <span style="color: rgb(0, 0, 0);"> <span> <span style="font-variant-ligatures: normal;"> <span style="font-variant-caps: normal;"> <span style="letter-spacing: normal;"> <span style="orphans: 2;"> <span style="text-align:left;"> <span style="text-indent: 0px;"> <span style="text-transform: none;"> <span style="white-space: normal;"> <span style="widows: 2;"> <span style="word-spacing: 0px;"> <span style="display: inline !important;"> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Traditional 510k SCARLET® AC-Ti
Image /page/3/Picture/2 description: The image shows a black silhouette of a butterfly. The butterfly's wings are spread open, and its body is visible in the center. The butterfly is facing slightly to the right. The image is simple and elegant.
## 510(k) SUMMARY
| 510k | Traditional | |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| Basis for submission | New devices | |
| Submitted by | SPINEART SA | |
| | | 3 Chemin du Pré Fleuri |
| | | 1228 Plan-les-Ouates |
| | | SWITZERLAND |
| Contacts | Max LOK, EVP RAQA | |
| | | Phone : +41 22 570 1200 Fax : +41 22 594 8306 |
| | | Mail : mlok@spineart.com |
| | | Regulatory contact : Estelle LEFEUVRE (elefeuvre@spineart.com) |
| Date Prepared | May 2nd, 2024 | |
| Common Name | Intervertebral body fusion device | |
| Trade Name | SCARLET® AC-Ti | |
| Classification Name | Intervertebral Fusion Device With Integrated Fixation, Cervical | |
| Class | II | |
| Product Code | OVE, ODP | |
| CFR section | 888.3080 | |
| Device panel | Orthopedic | |
| Legally marketed predicate devices | Primary predicate: SCARLET® AC-T Secured Anterior Cervical Cage manufactured by Spineart (K141314, K143214, K172065, K190322) | |
| | Additional predicates:<br>- TRYPTIK® Ti manufactured by Spineart (K200312, K230583)<br>- ChoiceSpine Blackhawk Ti Cervical Spacer System manufactured by ChoiceSpine LLC (K203311, K223869)<br>- Atlas Spine Expandable Cervical Standalone Interbody System manufactured by Atlas Spine, Inc. (K192570) | |
{4}------------------------------------------------
| Indications for use | SCARLET® AC-Ti cages are indicated for use in skeletally mature<br>patients with degenerative disc disease (defined as neck pain of<br>discogenic origin with degeneration of the disc confirmed by history and<br>radiographic studies) of the cervical spine. Patients should have at least<br>six (6) weeks of non-operative treatment prior to treatment with an<br>intervertebral cage. SCARLET® AC-Ti cages are intended to be used at<br>one or two contiguous levels from C2-T1 to facilitate intervertebral body<br>fusion with autogenous bone graft and/or allogenic bone graft comprised<br>of cancellous and/or corticocancellous bone graft. When used with two<br>bone screws, the SCARLET® AC-Ti Secured Anterior Cervical Cage is<br>intended to be used as a standalone system and requires no additional<br>supplementary fixation systems.<br>When used with two anchors or without two bone screws, the<br>SCARLET® AC-Ti cages are to be used with supplemental fixation which<br>has been cleared by the FDA for use in the cervical spine.<br>The SCARLET® AC-Ti hyperlordotic (12° lordosis) cages are to be used<br>with bone screws and/or anchors and additional supplemental fixation<br>system that has been cleared by the FDA for use in the cervical spine. |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description of the device | Spineart SCARLET® AC-Ti system is a Cervical Intervertebral Body<br>Fusion device with integrated fixation intended to provide mechanical<br>support and stabilization to the cervical spine and maintain adequate<br>disc space until fusion occurs. The system comprises a range of<br>intervertebral spacers implanted via an anterior approach, and having<br>various sizes, heights, footprints, and lordosis to adapt individual<br>pathology and different patients' anatomical conditions. The interbody<br>device is a box-shaped spacer with a central cavity that can receive bone<br>graft, which is intended to promote intervertebral fusion. It has a<br>monolithic design and is crossed by two tunnels that guide the bone<br>screws and/or the anchors insertion into the vertebral endplates.<br>Moreover, SCARLET® AC-Ti comprises integrated fixation by the mean<br>of bone screws and/or anchors that come in various diameters and<br>lengths.<br>The SCARLET® AC-Ti spacers are all made from medical grade titanium<br>alloy conforming to ASTM F136 and are produced by additive<br>manufacturing (SLM) according to ASTM F3001. Subsequently the<br>spacer is polished and thread tapping is machined. The screws and<br>anchors are made from Ti-6AI-4V ELI conforming to ASTM F136.<br>The SCARLET® AC-Ti spinal implants are delivered sterile (gamma<br>sterilization) and supplied with dedicated surgical instruments (reusable<br>– provided non-sterile). Bacterial endotoxin testing on final, finished<br>devices as specified in USP standard is used for pyrogenicity testing to<br>achieve the Endotoxin limit of 20 EU / device. |
| Technological<br>characteristics<br>compared to the<br>predicate devices | The SCARLET® AC-Ti spinal implants are manufactured using the same<br>manufacturing technology, i.e. additive manufacturing (SLM) as<br>predicate device TRYPTIK® Ti. The characterization of the chemical,<br>physical and mechanical properties of the material was performed in<br>accordance with ASTM F3001.<br>The SCARLET® AC-Ti implants are available in various sizes, heights,<br>footprints and lordosis so as to adapt individual pathology and different<br>patient's anatomical conditions. The SCARLET® AC-Ti implants are<br>implanted via an anterior approach. These features are similar to those<br>of the predicate devices (K141314, K143214, K172065, K190322,<br>K200312, K230583, K223869, K203311, K192570) |
| Mechanical testing: | Static and dynamic Axial Compression, Static and dynamic Shear-<br>compression and Static and dynamic Torsion according to ASTM F2077-<br>22 and Subsidence according to ASTM F2267-22 were conducted.<br>Results demonstrate comparable mechanical properties to the identified<br>predicate devices. Mass loss was measured on post-test run out<br>SCARLET® AC-Ti worst-case specimens dynamically tested.<br>Expulsion characterization was conducted. |
| Discussion of Testing | MRI Safety Evaluation:<br>In accordance with the FDA guidance “Testing and Labeling Medical<br>Devices for Safety in the Magnetic Resonance (MR) Environment”<br>(Issued October 2023), testing has been completed on the worst-case<br>implants. The following testing has been completed and provided a<br>determination that the subject Spineart SCARLET® AC-Ti in this 510(k)<br>submission has been determined to be MR conditional:<br>- ASTM F2052-2021: Standard Test Method for Measurement of<br>Magnetically Induced Displacement Force on Medical Devices in the<br>Magnetic Resonance Environment<br>- ASTM F2213-17: Standard Test Method for Measurement of<br>Magnetically Induced Torque on Medical Devices in the Magnetic<br>Resonance Environment<br>- ASTM F2119-07: Standard Test Method for Evaluation of MR Image<br>Artifacts from Passive Implants<br>- ASTM F2182-19e2: Standard Test Method for Measurement of Radio<br>Frequency Induced Heating On or Near Passive Implants During<br>Magnetic Resonance<br>- ASTM F2503-23e1:"Standard Practice for Marking Medical Devices<br>and Other Items for Safety in the Magnetic Resonance Environment" |
| | Human cadaveric testing:<br>Cadaver lab trials were conducted. |
| Conclusion | Based on the design features, technological characteristics, feature<br>comparisons, indications for use, and non-clinical performance testing,<br>the SCARLET® AC-Ti implants have demonstrated substantial<br>equivalence to the identified predicate devices |
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.