K203311 · Choice Spine, LLC · OVE · May 10, 2021 · Orthopedic
Device Facts
Record ID
K203311
Device Name
ChoiceSpine Blackhawk Ti Cervical Spacer System
Applicant
Choice Spine, LLC
Product Code
OVE · Orthopedic
Decision Date
May 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Blackhawk™ Ti Cervical Spacer System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease at one disc level from C2-T1. Degenerative Disc Disease (DDD) is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The Blackhawk™ Ti Cervical Spacer System is to be used with autogenous bone and/or allogenic bone graft composed of cancellous and /or corticocancellous bone graft, and supplemental fixation (i.e., an anterior cervical plate), and is implanted via an open, anterior approach.
Device Story
Interbody fusion device; implanted via open, anterior approach to treat degenerative disc disease. Device consists of titanium alloy (Ti-6Al-4V ELI) cage, nitinol locking clips, titanium anchors, and titanium locking cam. Chamber filled with autogenous or allogenic bone graft to promote fusion. Used with supplemental anterior cervical plate. Surgeon implants device to provide stability during fusion process. Benefits patient by restoring disc height and facilitating spinal fusion. Mechanical performance validated via static subsidence, expulsion, and dynamic axial compression/torsion testing.
Clinical Evidence
Bench testing only. Mechanical testing performed per ASTM F2077-18 (dynamic axial compression, compression-shear, torsion) and ASTM F2267-04 (static subsidence). Sterilization validated via gamma radiation to achieve SAL 10^-6.
Technological Characteristics
Implant materials: Ti-6Al-4V ELI (ASTM F3001/F136) and Nitinol (ASTM F2063). Instrumentation materials: 17-4 Stainless Steel (ASTM F899/A693), 465 Stainless Steel (ASTM F899), and polyphenylsulfone. Device is an interbody cage with internal locking cam and anchors. Provided sterile. No software or electronics.
Indications for Use
Indicated for anterior cervical interbody fusion in skeletally mature patients with degenerative disc disease (DDD) at one level (C2-T1) who failed six weeks of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
ChoiceSpine Blackhawk™ Cervical Spacer System (K191621, K171489)
ChoiceSpine Tomcat™ Cervical Spacer System (K170953)
Exactech Acapella One Cervical Spacer System (K132582)
ChoiceSpine Tiger Shark C™ Cervical Spacer System (K201643)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 10, 2021
ChoiceSpine, LLC Ms. Kim Finch Director of Regulatory Affairs 400 Erin Drive Knoxville, Tennessee 37919
Re: K203311
Trade/Device Name: ChoiceSpine Blackhawk™ Ti Cervical Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE Dated: April 20, 2021 Received: April 22, 2021
Dear Ms. Finch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
#### 510(k) Number (if known)
#### K203311
#### Device Name
Blackhawk™ Ti Cervical Spacer System
#### Indications for Use (Describe)
The Blackhawk™ Ti Cervical Spacer System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease at one disc level from C2-T1. Degenerative Disc Disease (DDD) is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The Blackhawk™ Ti Cervical Spacer System is to be used with autogenous bone and/or allogenic bone graft composed of cancellous and /or corticocancellous bone graft, and supplemental fixation (i.e., an anterior cervical plate), and is implanted via an open, anterior approach.
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of infonmation unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
| Date: | March 19, 2021 |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | ChoiceSpine, LLC<br>400 Erin Drive<br>Knoxville, TN 37919 |
| Phone: | 865-246-3333 |
| Fax: | 865-246-3334 |
| Contact Person: | Kim Finch, Director of Regulatory Affairs |
| Proposed<br>Proprietary Trade<br>Name: | ChoiceSpine Blackhawk™ Ti Cervical Spacer System |
| Product Class: | Class II |
| Classification<br>Name: | ChoiceSpine Blackhawk™ Ti Cervical Spacer System<br>888.3080 - Intervertebral Fusion Device with Integrated Fixation, Cervical<br>• |
| Device Product<br>Code: | ChoiceSpine Blackhawk™ Ti Cervical Spacer System<br>OVE<br>• |
| Purpose of Submission: | The purpose of this submission is to gain clearance for the new ChoiceSpine<br>Blackhawk™ Ti Cervical Spacer System, which consists of implant spacers made<br>of Ti-6Al-4V ELI per ASTM F3001 (Class C). The subject implant devices &<br>instrumentation will be used with additional previously cleared instrumentation.<br>The subject implant introduces material, design, and manufacturing method<br>changes to the previously cleared ChoiceSpine Blackhawk Cervical Spacer<br>System. |
| Device Description: | The Blackhawk™ Ti Cervical Spacer System is an anterior cervical interbody<br>device consisting of a titanium alloy (Ti-4Al-6V ELI) implant cage per ASTM<br>F3001, nitinol internal locking components per ASTM F2063, two internal<br>titanium alloy (Ti-6Al- 4V ELI) anchors per ASTM F136, and a titanium alloy (Ti-<br>6Al-4V ELI) locking cam per ASTM F136. They are intended for use as interbody<br>fusion devices and offered in a variety of heights, footprints, and lordotic angles<br>to accommodate varying anatomical conditions. The device features a chamber<br>intended to be filled with autogenous bone and/or allogenic bone graft<br>material. The Blackhawk™ Ti Cervical Spacer System is used with two internal<br>anchors that lock on deployment and provide additional fixation. |
| Indications for Use: | The Blackhawk™ Ti Cervical Spacer System is indicated for anterior cervical<br>interbody fusion procedures in skeletally mature patients with degenerative disc<br>disease at one disc level from C2-T1. Degenerative Disc Disease (DDD) is defined<br>as discogenic pain with degeneration of the disc confirmed by history and<br>radiographic studies. These patients should have had six weeks of non-operative<br>treatment. The Blackhawk™ Ti Cervical Spacer System is to be used with<br>autogenous bone and/or allogenic bone graft composed of cancellous and /or<br>corticocancellous bone graft, and supplemental fixation (i.e., an anterior cervical<br>plate), and is implanted via an open, anterior approach. |
| Materials: | The Blackhawk™ Ti Cervical Spacer System implants are composed of a titanium<br>alloy (Ti-6Al-4V ELI per ASTM F3001, Class C) implant spacer, nitinol (per ASTM<br>F2063) internal locking clips, titanium alloy (Ti-6Al-4V ELI per ASTM F136)<br>anchors, and a titanium alloy (Ti-6Al-4V ELI per ASTM F136) locking cam. The<br>implant body, cam, and anchors are composed of Titanium alloy Ti-6Al-4V per<br>either ASTM F136 or ASTM 3001, which has a long history of safe and effective<br>use in orthopedic implants. The locking clips are composed of nitinol per ASTM<br>F2063, which has a long history of safe and effective use in orthopedic<br>implants. The Blackhawk™ Ti Cervical Spacer System implants will be provided<br>sterile. Instruments will be provided non-sterile but will be steam sterilized<br>before use. The instrumentation is made from 17-4 Stainless Steel (ASTM<br>F899/A693), 465 Stainless Steel (ASTM F899), titanium alloy (Ti-6Al-4V ELI<br>Grade 5 per ASTM F136), and polyphenylsulfone (Radel R5000 Series). |
| Predicate Devices: | Primary Predicate:<br>ChoiceSpine Blackhawk™ Cervical Spacer System (K191621 and K171489)<br>Additional Predicates:<br>ChoiceSpine Tomcat™ Cervical Spacer System (K170953)<br>Exactech Acapella One Cervical Spacer System (K132582)<br>ChoiceSpine Tiger Shark C™ Cervical Spacer System (K201643) |
| Non-clinical Testing: | Static subsidence per ASTM F2267-04<br>Expulsion (standard N/A)<br>Dynamic axial compression per ASTM F2077-18<br>Dynamic compression-shear per ASTM F2077-18<br>Dynamic torsion per ASTM F2077-18 |
{4}------------------------------------------------
### Substantial Equivalence Conclusion:
The implants proposed in this submission are similar to the predicate devices in principle of operation, indications for use, stabilization method, anatomic location and approach, product code and classification, and biocompatibility.
The subject device has similar geometry and identical footprints as the primary predicate (Blackhawk Cervical Spacer System). The subject device's spacer component is made out of titanium alloy whereas the primary predicate's spacer component is made out of PEEK. The anchor components of the subject and primary predicate devices are made out of the same titanium alloy material and are deployed in an equivalent fashion. The locking mechanism of the subject device utilizes a new cam component whereas the primary predicate utilized pins and clips. While the designs are different, the mechanical strength and stabilization performance are equivalent as demonstrated through ASTM F2077 and ASTM F2267 testing.
The subject device's spacer component and predicate Tiger Shark Cervical Spacer (K201643) are both manufactured at the same supplier using identical additive and traditional manufacturing processes.
{5}------------------------------------------------
The subject device and predicates are all provided sterile and are processed at the same supplier. ChoiceSpine's sterilization process has been validated through gamma validation and distribution testing and the results demonstrate that the predetermined acceptance criteria were met. The minimum radiation dose of 25kGy was sufficient to meet a sterilization assurance level (SAL) of 10° and the package system remained intact while also maintaining the hermetic barrier.
The worst-case device among the proposed system was identified and mechanical testing was conducted to prove the new design is substantially equivalent to the predicates. Therefore, after including all other similarities and performing mechanical testing, the proposed device is shown to be equivalent to the predicate devices in safety, effectiveness, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.