Rodin Titan 3D Resin is a tooth shade ceramic-hybrid resin used for the fabrication of hybrid denture prosthetics, implantsupported denture prosthetics, monolithic full and partial removable denture teeth to be used in a denture. It is indicated as a permanent restorative for both anterior restorations, including occlusal surfaces. It is used for fabricating permanent restorations such as inlays, veneers and full crown restorations.
Device Story
RODIN Titan 3D Resin is a light-curable dental acrylate resin; used by clinicians in dental offices to fabricate custom dental prosthetics. Input: 3D scanned image of patient teeth (via optical impression system). Process: CAD/CAM software designs appliance; 3D printer creates mold/structure; resin polymerized via UV lamp (380 nm). Output: Permanent dental restorations (crowns, veneers, dentures). Clinician cleans, trims, and verifies fit chairside. Benefits: Custom-fit, durable, and comfortable dental prosthetics compared to traditional methods.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 20795-1 and ISO 4049:2009. Key metrics: Flexural strength (136 MPa), Elastic Modulus (6.235 GPa), Fracture toughness (4.15 MPa m1/2), Water solubility (0.1 µg/mm³), and Water sorption (26 µg/mm³). Biocompatibility testing (ISO 7405) passed for cytotoxicity, sensitization, irritation, and acute systemic toxicity.
Technological Characteristics
Ceramic-hybrid methacrylate polymer resin (UDMA, HEMA, BHT, TPO, YbF3, Fumed Silica). 3D liquid light-cured printing. Curing wavelength 380 nm. Viscosity 1500 cP. Density 1.59 g/cm³. Complies with ISO 20795-1:2013 and ISO 4049:2009. Non-sterile. Standalone material for use with third-party CAD/CAM systems.
Indications for Use
Indicated for patients requiring permanent dental restorations, including anterior and posterior restorations, occlusal surfaces, inlays, onlays, veneers, full crowns, hybrid dentures, implant-supported dentures, and monolithic full/partial removable dentures.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 20, 2024
Belport Company, Inc., Gingi-Pak Daniel Gallegos Regulatory Affairs Specialist 670 Endeavor Circle Brea, California 92821
Re: K240688
Trade/Device Name: RODIN Titan 3D Resin Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI, EBF Dated: March 12, 2024 Received: March 13, 2024
Dear Daniel Gallegos:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240688
Device Name RODIN Titan 3D Resin
#### Indications for Use (Describe)
Rodin Titan 3D Resin is a tooth shade ceramic-hybrid resin used for the fabrication of hybrid denture prosthetics, implantsupported denture prosthetics, monolithic full and partial removable denture teeth to be used in a denture. It is indicated as a permanent restorative for both anterior restorations, including occlusal surfaces. It is used for fabricating permanent restorations such as inlays, veneers and full crown restorations.
Type of Use (Select one or both, as applicable):
| <span style="font-family: sans-serif;"> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Submission RODIN Titan 3D Resin 510(k) Summary K240688
### Submitter:
Belport Company, Inc., Gingi-Pak 670 Endeavor Circle Brea, CA 92821
Phone: (714) 990-5488
Contact Person: Daniel Gallegos
Date Prepared: April 25, 2024
| Name of Device: | RODIN Titan 3D Resin |
|-----------------------------|--------------------------------------------|
| Common or Usual Name: | Tooth shade resin material |
| Regulation Number and Name: | 21 CFR 872.3760, 21 CFR 872.3690 |
| Product Code | EBI, EBF |
| Device Class | Class II |
| Primary Predicate | K230445, SprintRay OnX Tough |
| Secondary Predicate | K210412, Pac-Dent Ceramic Nanohybrid Resin |
## Device Description:
RODIN Titan 3D Resin consists of a curable dental acrylate resin that is manufactured in a dental office based on a 3D scanned image of a patient's teeth. The acrylate resin material is designed to be used in conjunction with a scanned 3D image, and 3D printer assembly, to locally manufacture out a dental appliance based on the clinician's judgment of patient need. Fabrication of dental prosthetics with RODIN Titan 3D Resin requires computer-aided design and CAD/CAM manufacturing system that includes the following components not part of the device: oral casting impression, digital denture file created in an optical impression system, 3D printer, and curing light equipment. The material is an alternative to traditional dental prostheses material Rodin Titan 3D Resin is intended exclusively for professional dental work. RODIN Titan 3D Resin is designed to meet appropriate ISO standards for flexibility and sorption, to withstand prolonged use in the oral cavity. It is delivered non-sterile, and instructions are provided on cleaning the material prior to providing it to a patient. Curing is performed with a UV
{5}------------------------------------------------
lamp. The appliance is then cleaned, trimmed, and verified to fit in the dental office before the patient leaves.
# Intended Use / Indications for Use
RODIN Titan 3D Resin is a tooth shade ceramic hybrid resin used for the fabrication of hybrid denture prosthetics, implant-supported denture prosthetics, monolithic full and partial removable dentures, and preformed denture teeth to be used in a denture. It is indicated as a permanent restorative for both anterior and posterior restorations, including occlusal surfaces. It is used for fabricating permanent restorations such as inlays, onlays, veneers and full crown restorations,
The intended use of fabrication of monolithic full and partial removable dentures is the same as the primary predicate device (K230445) (product code EBI) and the indication as a permanent restorative is the same as the secondary predicate device (K210412) (product code EBF). The intended use, technological characteristics, and critical specifications, of RODIN Titan 3D Resin are similar to the predicate devices.
# Summary of Technological Characteristics
RODIN Titan 3D Resin and the primary (K230445) and secondary (K210412) predicate devices are all 3D printed denture devices. The liquid resin is polymerized in the 3D printer, which creates the final denture device. Here is a more detailed explanation of the process:
# RODIN Titan 3D Resin, Primary Predicate Device, and Secondary Predicate Device
- 1. An impression of the patient's mouth is taken.
- 2. The impression is scanned and sent to a 3D printer.
- 3. The 3D printer creates a mold of the denture.
- 4. The mold is filled with liquid resin.
- 5. The resin is polymerized, which creates the final denture device.
RODIN Titan 3D Resin, the primary predicate device, and the secondary predicate device are created using 3D printing technology. This technology allows for the creation of custom-fit dentures that are more comfortable and durable than traditional dentures. The principles of operation are essentially similar; therefore, the RODIN Titan 3D Resin is substantially equivalent to its predicate devices.
# Performance Requirements:
RODIN Titan 3D Resin conforms to the requirements of ISO 4049:2009 (Dentistry -Polymer-based restorative materials), ISO 20795-1: 2013 (Dentistry - Base Polymers) and EN ISO10993-1/ISO 7405 (Biological evaluation of medical devices).
{6}------------------------------------------------
| Property | Specification | Result | Testing<br>Protocols |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|----------------------|
| Tooth Shades | N/A | BL, B1, A1, A2 | N/A |
| Compressive<br>Strength (MPa) | N/A | 368 MPa | ASTM D695-15 |
| Flexural Strength<br>(MPa) | ≥65 MPa | 136 MPa | ISO 20795-1:<br>2013 |
| Elastic Modulus<br>(GPa) | ≥2.0GPa | 6.235 GPa | ISO 20795-1:<br>2013 |
| Wavelength (nm) for<br>Curing | N/A | 380 nm | N/A |
| **Depth of Cure<br>(mm) | N/A | N/A | N/A |
| Filler particle size<br>distribution (μ) | N/A | 3 μ | N/A |
| **Curing Time (sec) | N/A | N/A | N/A |
| ***Release Profile (<br>μ g/mm³) | N/A | N/A | N/A |
| ****Working Time<br>(sec) | N/A | N/A | N/A |
| ****Setting Time<br>(min) | N/A | N/A | N/A |
| Stress Intensity<br>Factor | Kmax ≥ 1.9 MPa m1/2 | 4.15 MPa m1/2 | ISO 20795-1:<br>2013 |
| Density | N/A | 1.59 g/cm³ | N/A |
| Viscosity | N/A | 1500 cP | N/A |
| Total Fracture Work | ≥900 J/m² | 2442 (J/m²) | ISO 20795-1:<br>2013 |
| Hardness | N/A | 95 Shore D | ASTM D2240 |
| Water Solubility | ≤1.6 µg/mm³ | 0.1 µg/mm³ | ISO 20795-1:<br>2013 |
| Water Sorption | ≤32 µg/mm³ | 26 µg/mm³ | ISO 20795-1:<br>2013 |
| Opacity | N/A | 70% @ 1mm<br>thickness | N/A |
| Radiopacity | > 100 mm Al | 200 mm Al | ISO 4049: 2009 |
| Layer Thickness | N/A | 100μm | N/A |
| Biocompatibility | Cytotoxicity | Comply | ISO 7405 |
| Property | Specification | Result | Testing<br>Protocols |
| Biocompatibility | Sensitization | Comply | ISO 7405 |
| Biocompatibility | Irritation | Comply | ISO 7405 |
| Biocompatibility | Acute Systemic<br>Toxicity | Comply | ISO 7405 |
| Biocompatibility | Genotoxicity | Unknown | ISO 7405 |
| Biocompatibility | Subacute/Subchronic<br>Systemic Toxicity | Unknown | ISO 7405 |
| | *Post-cured in Otoflash for 5000 flashes | | |
| | **Depth of Cure and Curing Time are dependent on the<br>additive CAD/CAM printer, set by the manufacturer. | | |
| | ***The resin does not contain any releasable agent. | | |
| | ****Working time and setting time are not applicable because<br>the resin is instantly cured and solidified by a light source, the<br>photo initiation reaction occurs instantly without additional<br>working time. | | |
Table 1. Physical Properties of Subject Device*:
{7}------------------------------------------------
# Equivalence to Marketed Devices
The following table compares technological and other characteristics of the subject, primary predicate, and secondary predicate device.
| | Table 2. Comparison of Technological Characteristics with Predicates | |
|--|----------------------------------------------------------------------|--|
| | | |
| Parameter | Subject Device<br>RODIN Titan 3D<br>Resin (K240688) | Primary Predicate<br>Device<br>SprintRay OnX<br>Tough (K230445) | Secondary<br>Predicate<br>Device<br>Pac-Dent<br>Ceramic<br>Nanohybrid<br>Resin<br>(K210412) | Substantial<br>Equivalence<br>with<br>Predicate<br>Devices |
|--------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Manufacturer | Belport<br>Company, Inc.,<br>Gingi-Pak | SprintRay Inc. | Pac-Dent, Inc. | |
| Product Code | EBI, EBF | EBI, PZY | EBF | Substantially<br>equiv. |
| Regulation<br>Number | 872.3760,<br>872.3690 | 872.3760,<br>872.3590 | 872.3690 | Substantially<br>equiv. |
| Parameter | Subject Device<br>RODIN Titan 3D<br>Resin (K240688) | Primary Predicate<br>Device<br>SprintRay OnX<br>Tough (K230445) | Secondary<br>Predicate<br>Device<br>Pac-Dent<br>Ceramic<br>Nanohybrid<br>Resin<br>(K210412) | Substantial<br>Equivalence<br>with<br>Predicate<br>Devices |
| Regulatory<br>Class | Class II | Class II | Class II | Substantially<br>equiv. |
| Indication for<br>use | RODIN Titan 3D<br>Resin is a tooth<br>shade ceramic<br>hybrid resin used<br>for the<br>fabrication of<br>hybrid denture<br>prosthetics,<br>implant-<br>supported<br>denture<br>prosthetics,<br>monolithic full<br>and partial<br>removable<br>dentures, and<br>preformed<br>denture teeth to<br>be used in a<br>denture. It is<br>indicated as a<br>permanent<br>restorative for<br>both anterior and<br>posterior<br>restorations,<br>including<br>occlusal<br>surfaces. It is<br>used for<br>fabricating<br>permanent<br>restorations such<br>as inlays, onlays,<br>veneers and full<br>crown<br>restorations. | SprintRay OnX<br>Tough is a tooth<br>shade ceramic<br>hybrid resin used<br>for the fabrication of<br>hybrid denture<br>prosthetics,<br>implant-supported<br>denture prosthetics,<br>monolithic full and<br>partial removable<br>dentures, and<br>preformed denture<br>teeth to be used in<br>a denture. | Pac-Dent<br>Ceramic<br>Nanohybrid<br>Resin is<br>indicated as a<br>permanent<br>restorative for<br>both anterior<br>and posterior<br>restorations,<br>including<br>occlusal<br>surfaces. It is<br>used for<br>fabricating<br>permanent<br>restorations<br>such as inlays,<br>onlays, veneers<br>and full crown<br>restorations. | Substantially<br>equiv. |
| Parameter | Subject Device<br>RODIN Titan 3D<br>Resin (K240688) | Primary Predicate<br>Device<br>SprintRay OnX<br>Tough (K230445) | Secondary<br>Predicate<br>Device<br>Pac-Dent<br>Ceramic<br>Nanohybrid | Substantial<br>Equivalence<br>with<br>Predicate<br>Devices |
| | | | Resin<br>(K210412) | |
| User<br>Population | Clinicians in<br>dental offices | Clinicians in dental<br>offices | Clinicians in<br>dental offices | Substantially<br>equiv. |
| Technology | 3D liquid (light-<br>cured) print resin<br>for dental<br>CAD/CAM | 3D liquid (light-<br>cured) print resin<br>for dental<br>CAD/CAM | 3D liquid (light-<br>cured) print<br>resin for dental<br>CAD/CAM | Substantially<br>equiv. |
| Material | Methacrylate<br>polymer resin<br>(dimethacrylate) | Methacrylate<br>Monomer/oligomers<br>that polymerized to<br>Methymethacrylate<br>Based polymer | Methacrylate<br>polymer resin<br>(dimethacrylate) | Substantially<br>equiv. |
| Material<br>Shades | BL, B1, A1, A2 | Unknown | Bleach, A1, A2,<br>A3, B1 | Substantially<br>equiv. |
| Biocompatible | Yes | Yes | Yes | Substantially<br>equiv. |
| OTC or Rx | Rx | Rx | Rx | Substantially<br>equiv. |
| Sterile | Non-sterile | Non-sterile | Non-sterile | Substantially<br>equiv. |
| <b>Chemical Composition</b> | | | | |
| Chemical<br>composition | UDMA<br>HEMA<br>BHT<br>TPO<br>TF 3.0<br>YbF3<br>Fumed Silica | Methacrylate<br>polymer resin with<br>photo initiator,<br>inhibitor and<br>pigments | UDMA<br>HEMA<br>BHT<br>TPO<br>TF 3.0<br>YbF3<br>Fumed Silica | Substantially<br>equiv. |
| Non-Clinical Performance Test Data | | | | |
| Performance<br>Testing | ISO 20795-1<br>ISO 4049:2009 | ISO 20795-1 | ISO 4049:2009 | Substantially<br>equiv. |
| Compressive<br>Strength<br>(MPa) | 368 MPa | Unknown | 378 MPa | Substantially<br>equiv. |
| Flexural<br>Strength<br>(MPa) | 136 MPa | 126 MPa | 112 MPa | Substantially<br>equiv. |
| Elastic<br>Modulus<br>(GPa) | 6.235 GPa | 4.281 GPa | 10.380 GPa | Substantially<br>equiv. |
| Parameter | Subject Device<br>RODIN Titan 3D<br>Resin (K240688) | Primary Predicate<br>Device<br>SprintRay OnX<br>Tough (K230445) | Secondary<br>Predicate<br>Device<br>Pac-Dent<br>Ceramic<br>Nanohybrid<br>Resin<br>(K210412) | Substantial<br>Equivalence<br>with<br>Predicate<br>Devices |
| Wavelength<br>(nm) for<br>Curing | 380 nm | 380 nm | 380 nm | Substantially<br>equiv. |
| Depth of Cure<br>(mm) | N/A | N/A | N/A | |
| Filler particle<br>size<br>distribution (<br>µ) | 3 µ | N/A | 2 µ | Substantially<br>equiv. |
| Curing Time<br>(sec) | N/A | N/A | N/A | |
| Light intensity<br>for curing<br>(mW/cm²)* | 8.0 +/-0.5<br>mW/cm2 | Unknown | 8.0 +/-0.5<br>mW/cm2 | Substantially<br>equiv. |
| Stress<br>Intensity<br>Factor /<br>Fracture<br>Toughness<br>(MPa • m1/2) | 4.15 MPa • m1/2 | 3.38 MPa • m1/2 | 3.4 MPa • m1/2 | Substantially<br>equiv. |
| Water<br>Sorption<br>(µg/mm³) | 26 µg/mm³ | 30 µg/mm³ | 15.8 µg/mm³ | Substantially<br>equiv. |
| Water<br>Solubility<br>(µg/mm³) | 0.1 µg/mm³ | 3.5 µg/mm³ | 0.4 µg/mm³ | Substantially<br>equiv. |
| Density | 1.59 g/cm³ | Unknown | 1.78 g/cm³ | Substantially<br>equiv. |
| Viscosity | 1500 cP | Unknown | 2000 cP | Substantially<br>equiv. |
| Total Fracture<br>Work | 2442 J/m² | 1033 J/m² | 964 J/m² | Substantially<br>equiv. |
| Hardness | 95 Shore D | Unknown | 99 Shore D | Substantially<br>equiv. |
| Opacity | 70% @ 1mm<br>thickness | Unknown | 68% @ 1mm<br>thickness | Substantially<br>equiv. |
| Radiopacity<br>(mm of Al) | 200 mm | Unknown | 200 mm | Substantially<br>equiv. |
| Parameter | Subject Device<br>RODIN Titan 3D<br>Resin (K240688) | Primary Predicate<br>Device<br>SprintRay OnX<br>Tough (K230445) | Secondary<br>Predicate<br>Device<br>Pac-Dent<br>Ceramic<br>Nanohybrid<br>Resin<br>(K210412) | Substantial<br>Equivalence<br>with<br>Predicate<br>Devices |
| Layer<br>Thickness | 100 µm | 100 μm | 100 μm | Substantially<br>equiv. |
| Monomer<br>Methyl<br>Methacrylate<br>(≤2.2%) | Pass…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.