SprintRay OnX Tough is a tooth shade ceramic-hybrid resin used for the fabrication of hybrid denture prosthetics, implant-supported denture prosthetics, monolithic full and partial removable dentures, and preformed denture teeth to be used in a denture.
Device Story
OnX Tough is a curable dental acrylate resin for 3D printing dental prosthetics. Input: 3D scanned images of patient teeth. Process: Clinician uses CAD/CAM system to design appliance; 3D printer polymerizes liquid resin into final shape; UV lamp curing; manual trimming/polishing. Output: Hybrid dentures, implant-supported dentures, monolithic full/partial dentures, preformed denture teeth. Used in dental offices by clinicians. Benefits: Custom-fit, durable, comfortable alternative to traditional prostheses. Device is non-sterile; requires cleaning/verification before patient use.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing (ISO 10993-1, ISO 7405) for genotoxicity, cytotoxicity, acute systemic toxicity, sensitization, and irritation passed. Bench testing per ISO 20795-1/2 confirmed compliance for flexural strength (71.03 MPa), flexural modulus (2271 MPa), water sorption (31 µg/mm³), solubility (4.0 µg/mm³), and fracture toughness. Additive manufacturing validation performed per FDA guidance.
Technological Characteristics
Ceramic-hybrid acrylate resin. 3D printing (additive manufacturing) polymerization. ISO 20795-1/2 compliant. Non-sterile. UV lamp curing. Dimensions/form factor determined by patient-specific CAD design. No integrated software/connectivity; requires external CAD/CAM system, 3D printer, and curing equipment.
Indications for Use
Indicated for fabrication of hybrid denture prosthetics, implant-supported denture prosthetics, monolithic full and partial removable dentures, and preformed denture teeth for use in a denture by dental professionals.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 14, 2023
SprintRay Inc. Sara Moghtadernejad Regulatory Affairs 2705 Media Center Drive, Suite 100A Los Angeles, California 90065
Re: K230445
Trade/Device Name: OnX Tough Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI, PZY Dated: May 18, 2023 Received: May 19, 2023
Dear Sara Moghtadernejad:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bobak Shirmohammadi -S
For Michael E. Adjodha, M. ChE., CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K230445
Device Name
OnX Tough
Indications for Use (Describe)
SprintRay OnX Tough is a tooth shade ceramic-hybrid resin used for the fabrication of hybrid denture prosthetics, implant-supported denture prosthetics, monolithic full and partial removable dentures, and preformed denture teeth to be used in a denture.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather, and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY OnX Tough K230445
Submitter: SprintRay Inc. 2705 Media Center Drive, Suite 100A Los Angeles. CA 90065
Phone: (800) 914-8004
Contact Person: Sara Moghtadernejad
Date Prepared: July 12, 2023
| Name of Device: | OnX Tough | | |
|---------------------------------|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------|--|
| Common or Usual Name | Denture, Prescription | | |
| Regulation Number and<br>Names: | 21 CFR 872.3760,<br>Denture Relining,<br>Repairing, or Rebasing<br>Resin | 21 CFR 872.3590,<br>Additively Manufactured,<br>Preformed, Resin<br>Denture Tooth | |
| Product Codes | EBI | PZY | |
| Device Class | II | II<br>510(k)-Exempt | |
| Primary Predicate | K221678, SprintRay High Impact Denture Base,<br>(EBI product code) | | |
| Secondary Predicate | K151142, IvoBase CAD for Zenotec,<br>(EBI product code) | | |
#### Device Description
OnX Tough resin consists of a curable dental acrylate resin that is manufactured in a dental office based on a 3D scanned image of a patient's teeth. The acrylate resin material is designed to be used in conjunction with a scanned 3D image, and 3D printer assembly, to locally manufacture out a dental appliance based on the clinician's judgment of patient need.
Fabrication of dental prosthetics with OnX Tough resin requires computer-aided design and CAD/CAM manufacturing system that includes the following components not part of the device: oral casting impression, digital denture file created in an optical impression system, 3D printer, and curing light equipment.
The material is an alternative to traditional dental prostheses material. OnX Tough resin is intended exclusively for professional dental work.
OnX Tough resin is designed to meet appropriate ISO standards for flexibility and sorption, to withstand prolonged use in the oral cavity. It is delivered non-sterile,
{4}------------------------------------------------
and instructions are provided on cleaning the material prior to providing it to a patient. Curing is performed with a UV lamp. The appliance is then cleaned, trimmed, and verified to fit in the dental office before the patient leaves.
#### Intended Use / Indications for Use
SprintRay OnX Tough is a tooth shade ceramic-hybrid resin used for the fabrication of hybrid denture prosthetics, implant-supported denture prosthetics, monolithic full and partial removable dentures, and preformed denture teeth to be used in a denture.
The intended use of fabrication of monolithic full and partial removable dentures is the same as the primary predicate device (K221678) (product code EBI). The fabrication of hybrid denture prosthetics, implant-supported denture prosthetics is the same as the secondary predicate device (K151142) (product code EBI). The fabrication of preformed denture teeth to be used in a denture is Class II, but 510(k)-exempt (product code PZY), therefore no predicate is necessary.
The intended use, technological characteristics, and critical specifications, of OnX Tough are similar to both predicate devices.
### Summary of Technological Characteristics
OnX Tough and the primary predicate device (K221678) are both 3D printed denture devices. The liquid resin is polymerized in the 3D printer. which creates the final denture device. The secondary predicate device (K151142) is a prepolymerized denture disk that is milled to the final shape of the impression. Here is a more detailed explanation of the two processes:
#### OnX Tough and Primary Predicate Device
- 1. An impression of the patient's mouth is taken.
- 2. The impression is scanned and sent to a 3D printer.
- 3. The 3D printer creates a mold of the denture.
- 4. The mold is filled with liquid resin.
- 5. The resin is polymerized, which creates the final denture device.
#### Secondary Predicate Device
- 1. An impression of the patient's mouth is taken.
- 2. The impression is scanned and sent to a computer.
- 3. A pre-polymerized denture disk is created on the computer.
- 4. The disk is milled to the final shape of the impression.
- 5. The milled disk is placed in the patient's mouth.
Both OnX Tough and the primary predicate device are created using 3D printing technology. This technology allows for the creation of custom-fit dentures that are more comfortable and durable than traditional dentures. The secondary predicate device is a less expensive option that is still effective in providing patients with a new set of teeth. The principles of operation are essentially similar; therefore, the OnX Tough is substantially equivalent to its predicate devices.
{5}------------------------------------------------
The following technological differences exist between the subject and predicate devices:
- Differences in acrylate resin material. ●
### Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility Testing
The biocompatibility evaluation for OnX Tough was conducted in accordance with the FDA Blue Book Memorandum #G95-1 and International Standard ISO 10993-1 and ISO 7405, as recognized by FDA. The battery of testing included the following tests:
- Genotoxicity ●
- Cytotoxicity ●
- . Acute Systematic Toxicity
- Sensitization ●
- . Irritation
OnX Tough is considered tissue contacting for a period longer than 30 days (a removable prosthesis).
#### Bench Testing
OnX Tough was tested for conformity with the industry consensus standard ISO 20795-1. The battery of testing included the following tests:
- Flexural Strength and Modulus ●
- Water Sorption and Solubility .
- . Stability
- Residual Methyl Methacrylate Monomers ●
- Homogeneity ●
- Surface Characteristics ●
- Shape Capability, Translucency, and Polishability
- Freedom from Porosity
- Color Stability ●
#### Additive Manufacturing
Testing, according to FDA's guidance Technical Considerations for Additive Manufactured Medical Devices, was performed and results were provided in the 510(k). These tests included evaluation of all relevant properties of the printed resin using the permitted machines. Further, tests based on considerations of the orientation during manufacturing were also performed.
In all instances, OnX Tough functioned as intended and the outcomes were as expected.
{6}------------------------------------------------
### EQUIVALENCE TO MARKETED DEVICES
| | Testing<br>Standards | High Impact<br>Denture Base<br>(Primary<br>Predicate) | IvoBase CAD<br>for Zenotec<br>(Secondary<br>Predicate) | OnX Tough<br>(Subject Device) |
|-------------------------------------------|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use &<br>Indications for<br>Use | N/A | The SprintRay<br>High Impact<br>Denture Base<br>resin is a light<br>curable<br>polymerizable<br>resin intended<br>to be used for<br>the fabrication<br>and repair, of<br>full and partial<br>removable<br>dentures and<br>baseplates. The<br>material is an<br>alternative to<br>traditional<br>denture base<br>material. | IvoBase CAD<br>for Zenotec,<br>IvoBase CAD<br>Bond and<br>Modelling<br>Liquid is a<br>system used:<br>For the<br>fabrication of<br>removable<br>dentures e.g.:<br>partial and<br>complete<br>denture<br>prosthetics,<br>hybrid denture<br>prosthetics,<br>combined<br>denture<br>prosthetics,<br>mouthguards,<br>implant-<br>supported<br>denture<br>prosthetics | SprintRay OnX<br>Tough is a tooth<br>shade ceramic-<br>hybrid resin used<br>for the fabrication<br>of hybrid denture<br>prosthetics,<br>implant-supported<br>denture prosthetics,<br>monolithic full and<br>partial removable<br>dentures, and<br>preformed denture<br>teeth to be used in<br>a denture. |
| User Population | N/A | Clinicians in<br>dental offices | Clinicians in<br>dental offices | Clinicians in dental<br>offices |
| Biocompatibility | ISO- 10993-1,<br>-3, -5, -10<br>and -11 | Passed | Passed | Passed |
| Composition | N/A | Methacrylate<br>Monomer/oligo<br>mer s that<br>polymerized to<br>Methymethacyl<br>ate Based<br>polymer | Traditional<br>PMMA<br>Polymer/mono<br>mer denture<br>base material<br>with<br>pigmentation<br>for shade. | Methacrylate<br>Monomer/oligomers<br>that polymerized to<br>Methymethacylate<br>Based polymer. The<br>subject device has<br>similar polymer<br>family to the<br>predicates except<br>for small changes in<br>pigments. |
| Principles of<br>operation | N/A | Additive<br>manufacturing<br>of the<br>polymerized<br>denture to the<br>final<br>shape of the<br>impression. | Polymerized<br>denture disk<br>milled to the<br>final shape of<br>the<br>impression. | Additive<br>manufacturing of<br>the polymerized<br>denture to the final<br>shape of the<br>impression. |
| Performance<br>Testing | | Testing<br>performed<br>according to<br>ISO standards | Testing<br>performed<br>according to<br>ISO standards | Testing performed<br>according to ISO<br>standards |
| Flexural Strength<br>(≥65 MPa) | ISO 20795-1 | $74.69 ± 1.55$ MPa | ≥65 MPa | $71.03 ± 1.45$ MPa |
| Flexural Modulus<br>(≥2000 MPa) | ISO 20795-1 | $2253 ± 78$ MPa | ≥2000 MPa | $2271 ± 118$ MPa |
| Sorption (≤32 µg/mm³) | ISO 20795-2 | $28 ± 1$ µg/mm³ | ≤32 µg/mm³ | $31 ± 1$ µg/mm³ |
| Solubility (≤5 µg/mm³) | ISO 20795-2 | $3.5 ± 0.5$ µg/mm³ | ≤5 µg/mm³ | $4.0 ± 0.5$ µg/mm³ |
| Monomer Methyl<br>Methacrylate<br>(≤2.2%) | ISO 20795-1 | Pass | Pass | Pass |
| Kmax ≥1.9<br>MPa*m1/2 | ISO 20795-1 | $3.328 ± 0.147$ | 2.5 | $3.176 ± 0.209$ |
| Total fracture<br>work,<br>≥900 J/m² | ISO 20795-1 | $1232 ± 119$ | 1172 | $945 ± 88$ |
{7}------------------------------------------------
## Conclusions
The OnX Tough resin is as safe and effective as its predicate devices. The OnX Tough has the same intended use and indication, and similar technological characteristics as its predicate devices. The minor technological differences between the OnX Tough and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the OnX Tough is as safe and effective as the predicate devices. Thus, the OnX Tough is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.