ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit)
Applicant
SurGenTec, LLC
Product Code
KNW · Gastroenterology, Urology
Decision Date
Aug 9, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is indicated for aspiration of bone marrow or autologous blood. The ALARA BMA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.
Device Story
Device consists of single-use stainless steel needles used in minimally invasive spinal surgery. Needles function as access tools for bone marrow/autologous blood aspiration and as probes for pedicle pilot hole preparation. During spinal approach, device provides nerve/root proximity feedback to surgeon to support guidewire placement. Surgeon operates device manually; proximity feedback assists in identifying spinal roots/nerves, reducing risk of nerve injury. Output is physical access to bone marrow/blood and sensory/monitoring feedback for navigation. Benefits include safe, guided access to target anatomy and efficient aspiration.
Clinical Evidence
Bench testing only. Testing included cannula/stylet handle torque and pull tests, needle strength (impaction/compression), Luer lock verification (ISO 80369), needle particle aspiration, needle cell viability, packaging validation, sterilization validation, and biocompatibility assessment.
Technological Characteristics
Materials: Stainless steel and ABS. Form factor: 11 Gauge (3mm) diameter, 12.8cm length. Principle: Mechanical aspiration needle with integrated neuromonitoring proximity feedback. Single-use, sterile. Connectivity: Not applicable (mechanical/electrical interface for neuromonitoring).
Indications for Use
Indicated for patients requiring bone marrow or autologous blood aspiration, and for patients undergoing spinal procedures requiring pedicle pilot hole preparation and nerve/root identification via proximity feedback.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 9, 2024
SurGenTec, LLC % Mr. Michael Coladonato Associate Director, Regulatory Affairs MCRA. LLC 803 7th Street NW Washington DC 20001
Re: K240664
Trade/Device Name: ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology-Biopsy Instrument Regulatory Class: Class II Product Code: KNW, PDQ Dated: May 30, 2024 Received: May 30, 2024
Dear Mr. Michael Coladonato:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long H. Chen -S = Digitally signed by Long H.Chen-S
Date: 2024.08.09 15:21:26 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240664
Device Name
ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit)
#### Indications for Use (Describe)
The ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is indicated for aspiration of bone marrow or autologous blood.
The ALARA BMA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span> Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | <div style="display:flex; align-items:center;"><span> Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitter: | SurGenTec, LLC<br>911 Clint Moore Rd<br>Boca Raton, FL 33487 |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Mr. Michael Coladonato<br>Associate Director, Regulatory Affairs<br>MCRA, LLC<br>803 7th Street NW<br>Washington DC 20001<br>202-552-5800<br>mcoladonato@mcra.com |
| Date Prepared: | August 9, 2024 |
| Trade Name: | ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA<br>BMAC Kit) |
| Common Name: | Aspiration Needle, BMA Needle |
| Classification: | 21 CFR 876.1075, 21 CFR 874.1820 |
| Product Code: | KNW, PDQ |
| Primary Predicate: | Ranfac Bone Marrow Aspiration Needle, Ranfac Corp. (K131157) |
| Additional Predicate: | ALARA Neuro Access Kit, SurGenTec LLC (K190163) |
### Device Description:
The ALARA BMA Neuro Access Kit utilizes a neuromonitoring platform in providing the surgeon with nerve / root location feedback during the approach to a pedicle to support a guidewire placement.
The ALARA BMAN and ALARA BMAC Kit includes various single-use needles designed for bone marrow aspiration and aspiration of autologous blood.
### Indications for Use:
The ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is indicated for aspiration of bone marrow or autologous blood.
The ALARA BMA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.
{4}------------------------------------------------
# Substantial Equivalence:
The ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is substantially equivalent to the predicate device.
| Information | Subject Device | Primary Predicate | Secondary predicate |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | ALARA BMA Neuro Access Kit<br>(ALARA BMAN and ALARA BMAC Kit)<br>SurGenTec, LLC | Ranfac Bone Marrow<br>Aspiration Needle<br>Ranfac Corporation | ALARA Neuro Access Kit<br>SurGenTec, LLC |
| 510(k) number | Subject<br>Device | K131157 | K190163 |
| Regulation | 21 CFR 876.1075<br>21 CFR 874.1820 | 21 CFR 876.1075 | 21 CFR 874.1820 |
| Classification | II | II | II |
| Product Code | PDQ, KNW | KNW | PDQ |
| Indications for<br>use | The ALARA BMA Neuro Access Kit is<br>indicated for pedicle pilot hole preparation,<br>locating, and identifying spinal roots / nerves<br>by providing proximity feedback.<br>The ALARA BMA Neuro Access Kit (ALARA<br>BMAN and ALARA BMAC Kit) is indicated<br>for aspiration of bone marrow or autologous<br>blood. | The Ranfac Bone Marrow<br>Aspiration Needle is intended<br>for use in aspirating bone<br>marrow. | The ALARA Neuro Access<br>Kit is indicated for pedicle<br>pilot hole preparation,<br>locating, and identifying<br>spinal roots / nerves by<br>providing proximity<br>feedback. |
| Surgical<br>Approach | Minimally Invasive | Minimally Invasive | Minimally Invasive |
| Material | Stainless Steel | Stainless Steel | Stainless Steel |
| Diameter | 11 Gauge - 3mm | 8 Gauge, 11 Gauge | 11 Gauge - 3mm |
| Lengths | 12.8cm | 11cm - 15cm | 12.8cm |
| Use of the Needle | Needle depth stop at bone | Needle depth stop at skin | Needle depth stop at bone |
| Packaged<br>Sterilized | Yes | Yes | Yes |
| Single Use Device | Yes | Yes | Yes |
The ALARA BMAN and BMAC Kit are similar in materials and design to the predicate device. Both are manufactured from stainless steel and ABS. Both devices include a cannulated version and include similar lengths, and diameters. Both devices have plastic molded handles for the cannula and stylet that mate with each other to prevent separation.
# Non-Clinical Data:
The following tests have been performed on the ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit):
- Cannula Handle Torque Test
- Cannula Handle Pull Test
- Stylet Handle Pull Test ●
- Needle Strength from Impaction and Compression Tests
- Luer Lock Verification per ISO 80369
- Needle Particle Aspiration ●
- Needle Cell Viability Test
- Packaging Validation
- Sterilization Validation
{5}------------------------------------------------
- . Biocompatibility Assessment
# Conclusion:
Based on the similarities in materials, design, principles of operation, and results of the non-clinical testing, the ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is substantially equivalent to the predicate device for the requested indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.