RANFAC BONE MARROW ASPIRATION NEEDLE

K131157 · Ranfac, Corp. · KNW · Oct 22, 2013 · Gastroenterology, Urology

Device Facts

Record IDK131157
Device NameRANFAC BONE MARROW ASPIRATION NEEDLE
ApplicantRanfac, Corp.
Product CodeKNW · Gastroenterology, Urology
Decision DateOct 22, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1075
Device ClassClass 2

Indications for Use

The Ranfac Bone Marrow Aspiration Needle is intended for use in aspirating bone marrow.

Device Story

Single-use bone marrow aspiration needle; consists of stainless steel cannula and stylet with molded plastic handles. User inserts needle into bone marrow cavity using clockwise-counterclockwise rotation; removes stylet; attaches luer syringe to cannula; applies negative pressure to aspirate marrow. Used in clinical settings by healthcare professionals. Benefits include access to bone marrow for diagnostic or therapeutic purposes.

Clinical Evidence

No clinical data.

Technological Characteristics

Materials: Stainless steel (ISO 9626). Design: Cannula and stylet with molded plastic handles; luer lock connections (ISO 594/1, ISO 594-2). Sterilization: Ethylene oxide (ISO 11135-1, ISO 10993-7). Packaging: ISO 11607-1.

Indications for Use

Indicated for the aspiration of bone marrow in patients requiring such procedures.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo with the text "Ranfac" on the top right. The logo is divided into two parts. The top part is a horizontal line above the text, and the bottom part is a semi-circle. The left half of the semi-circle is filled with a dark color, while the right half is white. Precision instruments for specialized applications #### SECTION 5 #### 510(k) Summary The contents of this 510(k) summary on the following pages have been provided in conformance with 21 CFR § 807.92 Content and format of a 510(k) summary. #### K131157 #### 510(k) Summary OCT 2 2 2013 Owner's Name and Address: Ranfac Corporation 30 Doherty Avenue Avon, MA 02322-0635 FDA Registration Number 1211566 Official Contact Person: Christopher P. Whelan Senior Vice President - Quality Telephone: 508-588-4400 extension: 106 Facsimile: 508-584-8588 e-mail: cwhelan@rantac.com Ranfac Bone Marrow Aspiration Needle Date Summary Prepared: Device Trade Name: Common Name: October 21, 2013 Aspiration Needle, BMA Needle Classification Name: Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II . Product Code: KNW > RANFAC CORP. | PO BOX 635 | 30 DOHERTY AVE | AVON, MA 02322 | USA Phone: 508-588-4400 | 800-272-6322 | Fax: 508-584-8588 | Email info@ranfac.com www.ranfac.com Since, 1888! {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a logo with the text "Ranfac" on the top right. There is a horizontal line above the text. Below the text is a semi-circle with the left quarter filled in black. The logo appears to be for a company or organization named Ranfac. #### Precision instruments for specialized applications ## 510(k) Summary #### Predicate Device: | 510(k) Number | Predicate Description | Manufactured By | |---------------|---------------------------------------------|-----------------| | K050795 | Imbibe Bone Marrow Aspiration Needle | Orthovita, Inc. | | K100665 | MarrowMax Bone Marrow Aspiration Needle Kit | NeedleTech | | K051506 | Sterylab Best Bone Harvesting System | Sterylab | | Device Description: | The Ranfac Bone Marrow Aspiration Needle is a single-use<br>aspiration needle comprised of a stainless steel cannula<br>with a molded plastic handle, and a stainless steel stylet<br>with a molded plastic handle which mates with the cannula<br>handle when the stylet is inserted through the cannula.<br>While holding the mated handles, the user rotates the<br>needle using a clockwise-counter clockwise motion while<br>applying pressure on the bone. The needle bores into the<br>marrow cavity. Once there the stylet is removed and<br>aspiration is accomplished by attaching a luer syringe<br>(not included) to the needle and applying negative pressure. | |--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use: | The Ranfac Bone Marrow Aspiration Needle is intended<br>for use in aspirating bone marrow. | | Technological Characteristics: | The Ranfac Bone Marrow Aspiration Needle is similar<br>in materials and design to the predicate device. Both<br>devices are comprised of stainless steel tubing for the<br>cannula and stainless steel wire for the stylet. Both devices<br>have plastic molded handles for the cannula and stylet that<br>mate with each other to prevent separation. Both devices<br>have side holes or no side holes to maximize aspiration and<br>both have luer lock connections enabling the use of<br>standard syringes. | RANFAC CORP. | PO BOX 635 | 30 DOHERTY AVE | AVON, MA 02322 | USA Phone: 508-588-4400 | 800-272-6322 | Fax: 508-584-8588 | Email info@ranfac.com www.ranfac.com Sincel 1888 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a logo with the text "Ranfac" on the top right. The logo is divided into two parts. The left part is a black semi-circle, and the right part is a white semi-circle. The logo is enclosed in a rectangular border. Precision instruments for specialized applications #### 510(k) Summary Non-clinical Data: Standards '' The Stainless Steel, which is the only patient contacting material, and contacts the patient in a limited use is in conformance with ISO 9626 First edition 1991-09-01, Amendment 1 2001-06-01 Stainless steel needle tubing for the manufacture of medical devices. FDA Standards Recognition Number 6-163. The luer lock connection that is molded as part of the plastic handle is in conformance with ISO 594/1 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements. FDA Standards Recognition Number 6-11. The luer lock connection that is molded as part of the plastic handle is in conformance with ISO 594-2 Second edition 1998-09-01, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings. FDA Standards Recognition Number 6-129. The following standards apply to the sterilization of the finished device. ISO 11135-1, Sterilization of health care products -Ethylene Oxide - Partl: Requirements for development. validation and routine control of a sterilization process for medical devices. FDA Recognition Number 14-331. ISO 10993-7, Biological evaluation of medical devices -Part 7: Ethylene oxide sterilization residuals. FDA Recognition Number 14-335. AAMI/ANSI/ISO 14161:2009,Sterilization of health care products - Biological indicators - Guidance for the selection, use and interpretation of results. FDA Recognition Number 14-285 RANFAC CORP. | PO BOX 635 | 30 DOHERTY AVE | AVON. MA 02322 | USA Phone: 508-588-4400 | 800-272-6322 | Fax: 508-584-8588 | Email info@ranfac.com www.ranfac.com since | xxx {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo with the text "Ranfac" on the right side. On the left side of the text is a semi-circular shape that is half black and half white. Above the text and the semi-circle is a horizontal line. Page 4 of 4 #### Precision instruments for specialized applications ISO 11607-1:2006, Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems. FDA Recognition Number 14-355 Clinical Data: Conclusion: Not applicable Based on the similarities in materials, design, principles of function, biocompatibility and sterilization between the Ranfac Bone Marrow Aspiration Needle, subject of this premarket notification and the predicate device, the Ranfac Bone Marrow Aspiration Needle has been shown to be substantially equivalent to the predicate device under the Federal Food, Drug and Cosmetic Act. RANFAC CORP. | PO BOX 635 | 30 DOHERTY AVE | AVON. MA 02322 | USA Phone: 508-588-4400 | 800-272-6322 | Fax: 508-584-8588 | Email info@ranfac.com www.ranfac.com {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized symbol that resembles a bird or a person with outstretched arms, composed of three curved lines. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002 October 22, 2013 Ranfac Corporation Mr. Christopher P. Whelan Senior Vice President - Quality 30 Doherty Avenue, P.O. Box 635 Avon, Massachusetts 02322 Re: K131157 Trade/Device Name: Ranfac Bone Marrow Aspiration Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: August 23, 2013 Received: August 29, 2013 Dear Mr. Whelan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {5}------------------------------------------------ Page 2 - Mr. Christopher P. Whelan forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ,: · ﺘ . ### SECTION 4 #### Indications for Use 510(k) Number (if known): K131157 Device Name: Ranfac Bone Marrow Aspiration Needle Indications for Use: The Ranfac Bone Marrow Aspiration Needle is intended for use in aspirating bone marrow. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Long H. Chen - A | for MXM | Page 1 of 2 | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|-------------| | <div> <img alt="Digital Signature" src="data:image/png;base64,iVBORw0KGgoAAAANSUhEUgAAAAUAAAAFCAYAAACNbyblAAAAHElEQVQI12P4//8/AxMDAwMDAwMDwwMDw4wYAAK+CQgG+rYlAAAAAElFTkSuQmCC"/> Digitally signed by Long H. Chen - A<br/> DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Long H. Chen - A, 0.9.2342.19200300.100.1.1=1300359056<br/> Date: 2013.10.22 06:51:26 -04'00' </div> | | | | (Division Sign-off) | | | | Division of Surgical Devices | | | | 510(k) Number: K131157 | | |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...