Geistlich Bio-Oss® is intended for the following uses: - Augmentation or reconstructive treatment of the alveolar ridge; - Filling of infrabony periodontal defects; - Filling of defects after root resection, apicoectomy, and cystectomy; - Filling of extraction sockets to enhance preservation of the alveolar ridge; - Elevation of the maxillary sinus floor; - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).
Device Story
Geistlich Bio-Oss is a biocompatible bone mineral matrix derived from purified bovine cancellous bone. Provided as granules or blocks, it features interconnected macro- and micropores, high porosity, and hydrophilicity. Used by dental clinicians in surgical settings for bone augmentation, ridge preservation, and defect filling. Acts as an osteoconductive scaffold to support bone regeneration. Clinicians place the material into the surgical site, often alongside GTR/GBR membranes. The device provides a structural matrix for new bone growth, facilitating clinical outcomes in periodontal and implant dentistry.
Clinical Evidence
Bench testing only. Evidence includes sterilization validation (ISO 11137), biocompatibility (ISO 10993), stability testing (ICH Q1A), and viral inactivation studies (ISO 22442-3, ICH Q5A). Physical and chemical characterization of validation lots from new suppliers confirmed equivalence to finished product specifications.
Technological Characteristics
Bovine-derived bone mineral matrix. Highly porous, hydrophilic, interconnected macro/micropores. Configurations: granules (0.25-1.0mm, 1.0-2.0mm) or blocks (1x1x2cm). Sterilization: Gamma or X-ray irradiation. Standards: ISO 11137 (sterilization), ISO 10993 (biocompatibility), ISO 22442-3 (viral inactivation).
Indications for Use
Indicated for patients requiring alveolar ridge augmentation, reconstruction, or bone defect filling (infrabony, post-resection, post-extraction, or peri-implant) in dental/periodontal procedures, often in conjunction with GTR/GBR products.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 12, 2024
Geistlich Pharma AG % Roshana Ahmed Sr. Regulatory Specialist TELOS Partners LLC 2850 Frontier Drive Warsaw, Indiana 46582
Re: K240661
Trade/Device Name: Geistlich Bio-Oss® Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPM Dated: June 21, 2024 Received: June 21, 2024
Dear Roshana Ahmed:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia,
{2}------------------------------------------------
Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
Device Name
Geistlich Bio-Oss®
Indications for Use (Describe)
Geistlich Bio-Oss® is intended for the following uses:
- · Augmentation or reconstructive treatment of the alveolar ridge;
- · Filling of infrabony periodontal defects;
- · Filling of defects after root resection, apicoectomy, and cystectomy;
- · Filling of extraction sockets to enhance preservation of the alveolar ridge;
- · Elevation of the maxillary sinus floor;
- · Filling of periodontal defects in coniunction with products intended for Guided Tissue Regeneration
- (GTR) and Guided Bone Regeneration (GBR); and
· Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### I. Submitter
Geistlich Pharma AG Bahnhofstrasse 40 CH-6110 Wolhusen Switzerland Phone: +41 41 492 55 55
Contact Person: Marco Steiner, Deputy Director Regulatory Affairs Date Prepared: July 12, 2024
#### II. Device
| Device Proprietary Name: | Geistlich Bio-Oss® |
|--------------------------|---------------------------------------|
| Common or Usual Name: | Bone Grafting Material |
| Classification Name: | Bone grafting material, animal source |
| Regulation Number: | 872.3930 |
| Product Code: | NPM |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following device:
- Geistlich Bio-Oss®, K122894, Geistlich Pharma AG ●
The following device is cited as a reference device:
- Orthoss®, K190754, Geistlich Pharma AG .
#### Device Description IV.
Geistlich Bio-Oss® is a biocompatible bone mineral matrix and is manufactured from purified spongiosa (cancellous) bovine bone mineral. The product is provided in granules or block form.
Geistlich Bio-Oss® serves as a matrix consisting of interconnected macro- and micropores. The material is highly porous, hydrophilic and has a large inner surface area. Geistlich Bio-Oss® is provided sterile via gamma irradiation or x-ray irradiation in the following configurations:
{5}------------------------------------------------
- Geistlich Bio-Oss® spongiosa (cancellous) granules (0.125 g, particle size 0.25 1.0 mm) ●
- Geistlich Bio-Oss® spongiosa (cancellous) granules (0.25 g, particle size 0.25 1.0 mm) ●
- Geistlich Bio-Oss® spongiosa (cancellous) granules (0.5 g, particle size 0.25 1.0 mm) ●
- Geistlich Bio-Oss® spongiosa (cancellous) granules (1.0 g, particle size 0.25 1.0 mm) ●
- Geistlich Bio-Oss® spongiosa (cancellous) granules (2.0 g, particle size 0.25 - 1.0 mm)
- Geistlich Bio-Oss® spongiosa (cancellous) granules (5.0 g, particle size 0.25 - 1.0 mm)
- Geistlich Bio-Oss® spongiosa (cancellous) granules (0.5 g, particle size 1.0 - 2.0 mm)
- Geistlich Bio-Oss® spongiosa (cancellous) granules (1.0 g, particle size 1.0 -2.0 mm)
- Geistlich Bio-Oss® spongiosa (cancellous) granules (2.0 g, particle size 1.0 - 2.0 mm)
- Geistlich Bio-Oss® spongiosa (cancellous) block (approx. 1 x 1 x 2 cm) .
#### V. Indications for Use
Geistlich Bio-Oss® is intended for the following uses:
- Augmentation or reconstructive treatment of the alveolar ridge; ●
- . Filling of infrabony periodontal defects;
- Filling of defects after root resection, apicoectomy, and cystectomy; ●
- Filling of extraction sockets to enhance preservation of the alveolar ridge; ●
- Elevation of the maxillary sinus floor; .
- . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and
- Filling of peri-implant defects in conjunction with products intended for Guided Bone ● Regeneration (GBR).
#### Comparison of Technological Characteristics VI.
The Indications for Use Statement is identical to the predicate device.
The subject device is identical to the predicate device with respect to materials characteristics, manufacturing methods, packaging, and size. The subject device and the predicate device have identical final product specifications. A comparison table is provided below.
{6}------------------------------------------------
| | Subject Device | Geistlich Bio-Oss®<br>(K122894) | Analysis |
|----------------|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|-----------|
| Material | Mineral of bovine origin | Mineral of bovine origin | Same |
| Shape | Granules<br>Block | Granules<br>Block | Same |
| Particle Sizes | 0.25 - 1.0<br>1.0 - 2.0 | 0.25 - 1.0<br>1.0 - 2.0 | Same |
| Configurations | Granules:<br>0.125 g<br>0.25 g<br>0.5 g<br>1.0 g<br>2.0 g<br>5.0 g<br>Block: 1 x 1 x 2 cm | Granules:<br>0.25 g<br>0.5 g<br>2.0 g<br>5.0 g<br>Block: 1 x 1 x 2 cm | Different |
| Single-Use | Yes | Yes | Same |
| Sterilization | Gamma<br>X-ray | Gamma | Different |
The purpose of this submission is to obtain clearance for the use of x-ray irradiation as an alternative terminal sterilization method and the use of animal sourced materials from a new additional supplier in New Zealand and new abattoir in Australia. In addition, the product is being offered in two new volumes: 0.125 g (particle size 0.25 - 1.0 mm), 1.0 g (particle size 0.25 - 1.0 mm), and 1.0 g (particle size 1.0 - 2.0 mm). These changes do not raise different questions of safety and effectiveness and are addressed by the performance data cited below and provided within the submission.
Further to these changes, the product shelf life has been extended from 3 years to 4 years, storage temperatures for raw bones were updated, the product specification was clarified, the secondary packaging for the 20 mL vial was changed to a blister pack, and the gamma sterilization load was increased to improve production capacity. These changes were assessed, validated as appropriate, and documented under the quality system.
{7}------------------------------------------------
#### VII. Performance Data
Mechanical testing (K122984), biocompatibility (K190754), sterilization (K122894), shelf-life (K190754), and non-clinical performance testing (K122894) from the applicant's own predicate and reference device were leveraged in support of substantial equivalence.
Based on the results of assessment, the following evaluations were undertaken:
- . sterilization validation per ISO 11137-1, ISO 11137-2, and ISO 11137-3
- biocompatibility assessments per ISO 10993-1 and ISO 10993-5
- stability testing per ICH Q1A (R) stability testing guidelines ●
- characterization of structural and mechanical properties
- viral inactivation studies per ISO 22442-3 and ICH Q5A(R2) Draft Version ●
In addition, validation of raw materials from the new supplier was undertaken. The physical and chemical composition and appearance of the validation lots were characterized using the same tests and acceptance criteria as the final, finished product.
## VIII. Conclusion
The subject device is identical to the predicate device. The addition of x-ray irradiation as an alternative terminal sterilization method and introduction of an additional raw material supplier do not raise different questions of safety and effectiveness. Therefore, it is concluded that Geistlich Bio-Oss® is substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.