Orthoss(R)

K190754 · Geistlich Pharma AG · MQV · Jun 23, 2019 · Orthopedic

Device Facts

Record IDK190754
Device NameOrthoss(R)
ApplicantGeistlich Pharma AG
Product CodeMQV · Orthopedic
Decision DateJun 23, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3045
Device ClassClass 2
AttributesTherapeutic

Indications for Use

ORTHOSS® is an implant intended to fill bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine and pelvis). ORTHOSS® can be used with autograft as a bone graft extender in a posterolateral spine fusion. These osseous defects may be surgically created or be the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. The device resorbs and is replaced with bone during the healing process.

Device Story

Orthoss is an inorganic bovine-derived bone matrix with interconnected macro- and microporous structure; supports new bone ingrowth; undergoes physiological remodeling by osteoclasts. Provided sterile in block or granular forms. Used by surgeons to fill non-structural bony voids or as a bone graft extender in posterolateral spine fusion. Implanted during orthopedic procedures; resorbs over time as replaced by host bone. Benefits include facilitating bone healing in defects. Performance validated via animal spinal fusion models comparing radiographic, micro-CT, and histological outcomes against autograft and predicate ceramic granules.

Clinical Evidence

Bench testing included sterilization (ISO 11137), pyrogenicity/endotoxin (USP <151>/<85>), shelf-life (ICH Q1A), packaging (ISO 11607), biocompatibility (ISO 10993), and material characterization (X-ray, FTIR, Porosimetry). Animal study: Boden rabbit spinal fusion model compared Orthoss, MASTERGRAFT, and autograft control. Evaluated at 6, 9, 12 weeks via radiography, micro-CT, and histology. Results showed similar bone formation, resorption, and remodeling patterns between Orthoss and predicate. No significant differences in manual palpation or spinal flexibility.

Technological Characteristics

Inorganic bovine bone matrix; interconnected macro/microporous structure. Supplied as blocks or granules. Sterilized via gamma irradiation. Packaging: double-blister packs or glass vials in blister packs. Characterization via X-ray, FTIR, and HG-Pressure Porosimetry. Complies with ISO 11137 (sterilization), ISO 11607 (packaging), ISO 10993 (biocompatibility), and ICH Q1A (shelf-life).

Indications for Use

Indicated for patients requiring filling of bony voids or gaps in the skeletal system (extremities, posterolateral spine, pelvis) resulting from surgery or trauma. Used as a bone graft extender with autograft in posterolateral spine fusion. Not for use in defects intrinsic to bony structure stability.

Regulatory Classification

Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. Geistlich Pharma AG % Roshana Ahmed, M.A., RAC Associate Director, Regulatory Affairs TELOS Partners, LLC 571 Christina Lake Drive Lakeland, Florida 33813 Re: K190754 Trade/Device Name: Orthoss® Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: March 22, 2019 Received: March 25, 2019 Dear Ms. Ahmed: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Stereotaxic, Trauma and Restorative Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K190754 Device Name Orthoss(R) ### Indications for Use (Describe) ORTHOSS® is an implant intended to fill bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine and pelvis). ORTHOSS® can be used with autograft as a bone graft extender in a posterolateral spine fusion. These osseous defects may be surgically created or be the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. The device resorbs and is replaced with bone during the healing process. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--------------------------------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span> <span style="font-size:16px;">☑</span> </span> </div> | | Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> <span style="font-size:16px;">☐</span> </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary #### I. Submitter Geistlich Pharma AG Bahnhofstrasse 40 CH-6110 Wolhusen Switzerland Phone: +41 41 492 67 64 Contact Person: Marco Steiner, Deputy Director Regulatory Affairs Date Prepared: May 20, 2019 #### II. Device | Device Proprietary Name: | Orthoss® | |--------------------------|------------------------------------------| | Common or Usual Name: | Resorbable Bone Void Filler | | Classification Name: | Resorbable calcium salt bone void filler | | Regulation Number: | 21 CFR 888.3045 | | Product Code: | MQV | | Device Classification | II | #### III. Predicate Device Substantial equivalence is claimed to the following device: - ORTHOSS® Resorbable Bone Void Filler, K090401, Geistlich Pharma AG . The following reference devices are cited within the submission: - Geistlich Bio-Oss, K122894, Geistlich Pharma AG ● - MASTERGRAFT® Resorbable Ceramic Granules, K082918, Medtronic Sofamer Danek - Geistlich Bio-Oss Pen, K120601, Geistlich Pharma AG ● #### Device Description IV. Orthoss® is an inorganic bone matrix, manufactured from bovine bone, with an interconnected macro- and microporous structure that supports the formation and ingrowth of new bone. Over time, Orthoss® is partially remodeled by osteoclasts (physiological remodeling). The single-use product is provided sterile (via gamma irradiation) in block (1 x 1 x 2 cm and 2 x 2 x 1.3 cm) or granular (1 - 2 mm and 2 - 4 mm) form in double-blister packs or glass vials in a blister pack, respectively. {4}------------------------------------------------ #### V. Indications for Use Orthoss is an implant intended to fill bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine and pelvis). Orthoss® can be used with autograft as a bone graft extender in a posterolateral spine fusion. These osseous defects may be surgically created or be the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. The device resorbs and is replaced with bone during the healing process. #### VI. Comparison of Technological Characteristics The subject device is similar to the predicate device with respect to materials characteristics, manufacturing and sterilization methods, packaging, and size. Both the subject and predicate device have identical final product specifications. Since prior clearance of the device, the following minor changes were made: - . change from an aluminum vial cap to a polyethylene vial cap; - slight changes in manufacturing and processes; ● - . addition of alternative raw material supplier; - addition of a new filling volume of 1 - 2 mm, 1 g (~2.5 cc); and - . packaging change for Orthoss granules from glass vial in a peel pouch to glass vial in blister pack These changes do not raise different questions of safety and effectiveness and are addressed by the performance testing identified below. #### VII. Performance Data The following performance data were provided in support of the substantial equivalence determination: - sterilization validation per ISO 11137-1:2006, ISO 11137-2:2013, and ISO 11137-● 3:2017; - pyrogenicity per USP <151> and endotoxin per USP <85>; ● - shelf-life studies per ICH Q1A (R2); ● - packaging validation per ISO 11607-1:2009/ Amd 2014 and ISO 11607-2:2006/ Amd ● 2014; - biocompatibility per ISO 10993-1:2018; ● - X-ray, Fourier Transform Infrared, and HG-Pressure Porosimetry analysis; and ● - validation of extended indication in a suitable study model. ● {5}------------------------------------------------ ## Summary of Animal Study The performance of Orthoss® and MASTERGRAFT® Resorbable Ceramic Granules (K082918) as autograft extenders were evaluated in a validated Boden rabbit spinal fusion model by determining the effect of the test article relative to iliac crest bone graft control (autograft) and MASTERGRAFT® Resorbable Ceramic Granules. Rabbits were randomized to receive Orthoss®, MASTERGRAFT® Resorbable Ceramic Granules, or autograft control. Prior to surgery, lateral radiographs were taken for evidence of tibial growth plate closure and blood was drawn for routine biochemistry and hematology tests. Evaluation of the local biological effects after implantation was conducted at 6, 9, and 12 weeks after implantation and examined at the transverse process and in the middle of the fusion mass. There were no adverse events or device related failures noted during harvesting. Results of the study demonstrate that the subject and predicate device performed in a similar manner with respect to radiographic appearance, micro-computed tomography, and histology, and support a determination of substantial equivalence. Evidence of new bone formation as well as resorption and remodeling of the graft materials were observed via radiographs and microcomputed tomography. No differences with respect to the distribution of findings for manual palpation between the subject device and predicate device test groups were observed and the multidirectional flexibility of the lumbar spine in animals treated with the subject device and predicate device were similar. Histological analysis showed normal patterns of bone healing in the test group and similar to the predicate group as well as the positive control. New bone formation and bone remodeling occurred with time reflecting a normal in vivo bone formation pathway for the test material as well as the predicate. There was presence of some inflammatory cells as the subject and predicate devices degraded over time. The performance of positive control group (autograft) was consistent with the published data on the model as well as with the experience of the study site. ## VIII. Conclusion The information provided above supports that Orthoss® is as safe and effective as the predicate device. Non-clinical studies support the expansion to the indications for use statement and the safe and effective performance of the product. Therefore, it is concluded that Orthoss® is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...