The Surgical Drive System (Model: ES70, ES90, E8) is indicated for: drilling, milling, sawing, screwing (for positioning) of osteosynthesis screws, implants and plate systems in soft and hard tissue. Including: ENT surgery and Maxillofacial surgery
Device Story
Surgical Drive System (ES70, ES90, E8) is an electrical drive unit comprising a control console, motor, foot control, and cables. It converts electrical energy into mechanical rotary motion to power surgical handpieces. Used in ENT and maxillofacial surgery, the device is operated by clinicians to perform incision, cutting, drilling, and sawing of bone and soft tissue. The control unit features an integrated touch display for monitoring and adjusting operating parameters (e.g., speed, pump state, direction). The foot control allows hands-free activation and parameter adjustment during procedures. By providing precise mechanical power to surgical attachments, the system facilitates the placement of osteosynthesis screws and implants, aiding in surgical reconstruction and tissue management.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1, IEC 8060-2-60), electromagnetic compatibility (IEC 60601-1-2), and software verification/validation (IEC 62304).
Technological Characteristics
Electrical drive unit with control console, motor, and foot control. Motor speed up to 40,000 min-1; ISO 3964 Type 3 connector. Power input 110-220V AC, 50/60Hz. Intermittent duty cycle (S3: 3 min load/10 min rest). Software-controlled parameters via touch display. Non-sterile delivery.
Indications for Use
Indicated for drilling, milling, sawing, and screwing (positioning) of osteosynthesis screws, implants, and plate systems in soft and hard tissue for patients undergoing ENT or Maxillofacial surgery.
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
July 18, 2024
Guangdong Jinme Medical Technology Co., Ltd. % Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road Huangpu District, Guangzhou Guangdong. CHINA
Re: K240340
Trade/Device Name: Surgical Drive System (Model: ES70, ES90, E8) Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument And Accessories Regulatory Class: Class II Product Code: DZI, ERL Dated: February 5, 2024 Received: February 5, 2024
Dear Cassie Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K240340
Device Name
Surgical Drive System (Model: ES70, ES90, E8)
Indications for Use (Describe)
The Surgical Drive System (Model: ES70, ES90, E8) is indicated for: drilling, milling, sawing, screwing (for positioning) of osteosynthesis screws, implants and plate systems in soft and hard tissue.
Including: ENT surgery and Maxillofacial surgery
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary of K240340
#### 1. Date of the summary prepared: July 17, 2024
#### 2. Submitter's Information
Company Name: GUANGDONG JINME MEDICAL TECHNOLOGY CO., LTD. Address: Room 301, 401, 501, Block No.3, Wanyang Science Park, 84 Taoyuan East Rd. Shishan Town, Nanhai District, Foshan, Guangdong, CHINA Contact name: Ying Yang Title: Manager Tel: +86-13929914992 Fax: / E-mail: 1109607962@qq.com
#### Application Correspondent
Contact Person: Ms. Cassie Lee Company: Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Title: Manager Tel: +86 20 8266 2446 Email: regulatory@glomed-info.com
#### 3. Subject Device Information
Company Name: GUANGDONG JINME MEDICAL TECHNOLOGY CO., LTD. Trade/Device Name: Surgical Drive System (Model: ES70, ES90, E8) Common Name: Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece Classification Name: Ear, nose, and throat electric or pneumatic surgical drill. Product Code: ERL, DZI Regulation Number: 21 CFR 874.4250, 21 CFR 874.4360 Review Panel: Ear Nose & Throat Regulatory Class: II
#### 4. Predicate Device Information
510(k) Number: K213221 Company Name: W&H Dentalwerk Bürmoos GmbH Trade/Device Name: AMADEO, M-UK1015 (incl. attachments and accessories) Common Name: Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece Classification Name: Ear, Nose, and Throat Electric or Pneumatic Surgical Drill Product Code: ERL, DZI Regulation Number: 21 CFR 874.4250 Regulatory Class: II
{5}------------------------------------------------
#### 5. Device Description
The Surgical Drive System is an electrical drive unit, including a motor, power console, foot control, connection cables, and other accessories. It is intended for use in surgical procedures involving incision/cutting, removal, drilling, and sawing of soft and hard tissue and bone.
The basic function of the Surgical Drive System is the conversion of electrical energy into a mechanical rotary motion. The control unit is used to control the connected motor and the integrated pump. The integrated touch display is used to monitor the actual settings and to change, within predetermined limits, the operating parameters. The foot control is used for activation of the motor and changing parameters e.g. program, pump state, and motor direction. The motor's function is to provide power for handpieces with the gear ratio (1:1, 1:4.2, 1:5, 3.2:1, 3.4:1, 40:1), the Max. speed of the motor is 40000 min-1, and it is designed with an ISO 3964 Type 3 connector.
#### 6. Intended Use / Indications for Use
The Surgical Drive System (Model: ES70, ES90, E8) is indicated for: drilling, milling, sawing, screwing (for positioning) of osteosynthesis screws, implants and plate systems in soft and hard tissue. Including: ENT surgery and Maxillofacial surgery
#### 7. Comparison to predicate devices
Compare with the predicate devices, the subject device is very similar in technical design, intended use/indications for use, functions, etc. And the differences between the subject device and predicate devices do not affect substantial equivalence. Please refer to the detailed Substantial Equivalence Comparison of the subject device and predicate devices in the following table.
| Elements of Comparison | Subject Device | Predicate Device 1<br>(Primary Predicate device) | Remark |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Company | GUANGDONG JINME<br>MEDICAL TECHNOLOGY<br>CO., LTD. | W&H Dentalwerk Bürmoos<br>GmbH | -- |
| Device Name | Surgical Drive System<br>(Model: ES70, ES90, E8) | AMADEO, M-UK1015 (incl.<br>attachments and<br>accessories) | -- |
| 510(k) Number | K240340 | K213221 | -- |
| Regulation Number | 21 CFR 874.4360, 21 CFR<br>874.4250 | 21 CFR 874.4360, 21 CFR<br>874.4250 | Same |
| Product Code | ERL, DZI | ERL, DZI | Same |
| Class | II | II | Same |
| Intended Use / Indications<br>for Use | The Surgical Drive System<br>(Model: ES70, ES90, E8)<br>is indicated for: drilling,<br>milling, cutting, sawing,<br>screwing (for positioning) | The drive unit for surgical<br>transmission instruments is<br>indicated for: drilling, milling,<br>cutting, sawing, screwing<br>(for positioning) of | Same |
{6}------------------------------------------------
| | of osteosynthesis screws,<br>implants and plate<br>systems in soft and hard<br>tissue.<br>Including: ENT surgery<br>and Maxillofacial surgery | osteosynthesis screws,<br>implants and plate systems<br>in soft and hard tissue.<br>Including: ENT surgery and<br>Maxillofacial surgery | |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Sterility | Provided non-sterile | Provided non-sterile | Same |
| Use | Rx only | Rx only | Same |
| Mains voltage | 110~220Va.c., 50Hz/60Hz | 120 V, 50-60 Hz | Similar<br>Note 1 |
| Frequency | | | |
| Operating Mode | Intermittent duty S3 (load<br>time max. 3min / rest time<br>min. 10min) | Intermittent duty S3 (load<br>time max. 3min / rest time<br>min. 5min) | Similar<br>Note 2 |
| Foot control | Wired | Wireless<br>Wired | Similar<br>Note 3 |
| Components | Control Unit<br>Motor (with the cable)<br>Foot control<br>Power cord<br>Pipeline head<br>Power fuse<br>Instruction for use | Control Unit<br>Motor<br>Handpiece (in different<br>versions)<br>Foot control (in different<br>versions)<br>Power cord<br>Instruction for use | Similar<br>Note 4 |
| Motor coupling systems<br>classification (ISO 3964) | Type 3 | No publicly available | Similar<br>Note 5 |
#### Comparison in Detail(s):
Note 1: Although the " Mains voltage Frequency" of the subject device is slightly different from the predicate devices, both the subject device and the predicate devices conducted the safety test according to the IEC 60601 series standards, and it meets the power supply requirements of the US, and the test results are in compliance with safety standards' requirements. So, the difference between the subject and predicate devices will not affect substantial equivalence.
Note 2: Although the "Operating Mode" of the subject device is slightly different from the predicate devices, both the subject device and predicate device K213221 are designed in the operating mode as Intermittent duty S3 and the max. Ioad time is the same and the min rest time is similar. The Rest time will not affect the treatment. The Rest time only affect the temperature raise and both the subject device and predicate devices conducted the IEC safety test, and the test results are in compliance with safety standards' requirements. So, the difference between the subject and predicate devices will not affect substantial equivalence.
{7}------------------------------------------------
Note 3: Although the "Foot control" designed with wire of the subject device is slightly different from the predicate devices, the predicate device K213221 has both wireless Foot control. So, the difference between the subject and predicate devices will not affect substantial equivalence.
Note 4: Although the "Components" of the subject device are slightly different from the predicate devices because they will provide different accessories or attachments. But all of them will provide the main components such as Control Unit, Motor, Foot control, Power cord, and Instruction for use. So, the difference between the subject and predicate devices will not affect substantial equivalence.
Note 5: Although the "Motor coupling systems classification (ISO 3964)" of the predicate device is not published our subject device designed the coupling systems classification (ISO 3964) as Type 3. Exactly, both the subject device and predicate device should be designed with proper specifications per ISO 3964 or an equivalent method. Furthermore, we also conducted the IEC series standards with a connected handpiece, the test results show that the subject device's performance can meet the standards' requirements without any accidentally disconnected. So, we can believe this difference between the subject and predicate devices will not affect substantial equivalence.
#### 8. Test Summary
#### 8.1 Summary of Non-clinical tests
Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility:
- Electrical safety test according to IEC 60601-1 and IEC 8060-2-60 standard; ।
- Electromagnetic compatibility test according to IEC 60601-1-2 and IEC/TR 60061-4-2 standard; -
- -Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling;
- -Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile;
- -Validation Data in Premarket Notification Submissions (510(k)s) for Reprocessed Single-Use Medical Devices;
- -Software verification and validation test according to the requirements of IEC 62304 and the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
- -Validation of the effectiveness of reprocessing and validation of the maximum number of reprocessing according to ISO 17665-1:
#### 8.2 Summary of Clinical Testing
Clinical testing is not required for this submission. The non-clinical performance testing described above is sufficient to support substantial equivalence of the subject device to the predicate.
#### 9. Final conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device is substantially equivalent to the legally marketed predicate device K213221.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.