AMADEO, M-UK1015 (incl. attachments and accessories)
Applicant
W&H Dentalwerk Buermoss GmbH
Product Code
ERL · Ear, Nose, Throat
Decision Date
May 26, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The drive unit for surgical transmission instruments is indicated for: drilling, milling, sawing, screwing (for positioning) of osteosynthesis screws, implants and plate systems in soft and hard tissue. Including: ENT surgery and Maxillofacial surgery
Device Story
AMADEO is an electrical surgical drive unit comprising a control unit, motor, attachments, and accessories. It converts electrical energy into mechanical rotary motion for drilling, cutting, sawing, and screwing in ENT and maxillofacial surgery. A displacement pump delivers physiological saline to the surgical site via a single-use sterile irrigation tubing set. The system features an integrated touch display for monitoring and adjusting operating parameters (motor speed, torque, coolant flow) and a foot control for activation and parameter toggling. Used in clinical settings by surgeons, the device assists in the precise placement of implants and osteosynthesis hardware. The system provides controlled mechanical power to surgical handpieces, facilitating bone and soft tissue manipulation, thereby supporting surgical efficiency and procedural accuracy.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included functional verification of application settings, touchscreen features, biocompatibility (ISO 10993 series), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and reprocessing validation.
Technological Characteristics
Electrical drive unit for surgical handpieces; converts electrical energy to mechanical rotary motion. Features integrated displacement pump for irrigation. Includes touch display and wired/wireless foot control. Intermittent duty cycle (S3). Complies with IEC 60601-1 and IEC 60601-1-2. Provided non-sterile.
Indications for Use
Indicated for drilling, milling, sawing, and screwing of osteosynthesis screws, implants, and plate systems in soft and hard tissue for ENT and maxillofacial surgery.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency responsible for regulating food and drug products in the United States.
May 26, 2022
W&H Dentalwerk Buermoss GmbH Weidler Gerhard Regulatory Affairs Ignaz-Glaser-Strasse 53 Buermoos, Salzburg 5111 AUSTRIA
Re: K213221
Trade/Device Name: AMADEO, M-UK1015 (incl. attachments and accessories) Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, Nose, And Throat Electric Or Pneumatic Surgical Drill Regulatory Class: Class II Product Code: ERL, DZI Dated: April 21, 2022 Received: April 27, 2022
Dear Weidler Gerhard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha. M.ChE Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213221
Device Name
AMADEO (M-UK1015, incl. attachments and accessories)
Indications for Use (Describe)
The drive unit for surgical transmission instruments is indicated for: drilling, milling, sawing, screwing (for positioning) of osteosynthesis screws, implants and plate systems in soft and hard tissue.
Including: ENT surgery and Maxillofacial surgery
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <svg height="12" width="12"> <polygon points="0,0 12,0 12,12 0,12" style="fill:none;stroke:black;stroke-width:1"></polygon> <polygon points="2,2 10,2 10,10 2,10" style="fill:black"></polygon> </svg> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <svg height="12" width="12"> <polygon points="0,0 12,0 12,12 0,12" style="fill:none;stroke:black;stroke-width:1"></polygon> </svg> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K213221
Image /page/3/Picture/2 description: The image shows a logo with the letters W&H inside a gray, rounded hexagon. To the right of the logo is a green arrow pointing to the right. Inside the green arrow are the words "People have Priority" in white text.
| DATE OF APPLICATION | May 25, 2022 |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | W&H Dentalwerk Bürmoos GmbH<br>Ignaz-Glaser-Strasse 53<br>5111 Bürmoos<br>Austria<br>0043 - 6274/6236-0<br>0043 - 6274/6236-55 |
| CONTACT PERSON | Mag. Dr. Gerhard Weidler<br>Regulatory Affairs Manager<br>0043 - 6274/6236-9339<br>gerhard.weidler@wh.com |
#### 1 Device Name
| Trade Name: | AMADEO (M-UK1015 incl. attachments and<br>accessories) |
|-----------------------------|---------------------------------------------------------------|
| Common Name: | AMADEO |
| Device Classification Name: | Ear, nose, and throat electric or<br>pneumatic surgical drill |
# 2 Classification / Product Code
The AMADEO can be classified according to following device names and product codes:
| Primary<br>Product<br>Code | Device | Regulation<br>Description | Regulation<br>Medical<br>Specialty | Review<br>Panel | Regulation<br>Number | Device<br>Classification |
|----------------------------|---------------------------------------------------------------------------|----------------------------------------------------------------------|------------------------------------|----------------------|----------------------|--------------------------|
| ERL | Drill, Surgical, Ent<br>(Electric or<br>Pneumatic) Including<br>Handpiece | Ear, nose, and<br>throat electric or<br>pneumatic surgical<br>drill. | Ear Nose &<br>Throat | Ear Nose<br>& Throat | 874.4250 | 2 |
{4}------------------------------------------------
| Secondary<br>Product<br>Code | Device | Regulation<br>Description | Regulation<br>Medical<br>Specialty | Review<br>Panel | Regulation<br>Number | Device<br>Classification |
|------------------------------|----------------------|------------------------------------------------|------------------------------------|-----------------|----------------------|--------------------------|
| DZI | Drill, Bone, Powered | Bone cutting<br>instrument and<br>accessories. | Dental | Dental | 872.4120 | 2 |
#### Predicate Device / Reference Device 3
| Device | Predicate<br>Device | Reference<br>Device | 510(k)<br>Number | 510(k) Holder |
|--------|---------------------|---------------------|------------------|----------------|
| AMADEO | Primado2 | -------- | K132264 | NSK |
| | -------- | Implantmed | K161957 | W&H Dentalwerk |
#### Device Description 4
The AMADEO is an electrical drive unit, including motor, attachments and accessories, for surgical transmission attachments which is indicated for: drilling, cutting, sawing, screwing (for placement) of osteosynthesis screws, implants and plates, in soft and hard tissue.
The foreseen areas of application will be:
- > ENT (ear, nose, throat) surgery
- > Maxillofacial surgery
The basic function is the conversion of electrical energy into a mechanical rotary motion. The control unit is used to control the connected motor and the integrated pump. Depending on the treatment, a physiological saline solution is pumped to the treatment site by a displacement pump. A single-use sterile disposable irrigation tubing set is supplied within the scope of delivery and offered as an accessory.
The integrated touch display is used to monitor the actual settings and to change, within predetermined limits, the operating parameters.
The foot control is used for activation/deactivation of the motor and for changing parameters e.g. program, pump state and motor direction.
#### 5 Indications for Use
The drive unit for surgical transmission instruments is indicated for: drilling, cutting, sawing, screwing (for positioning) of osteosynthesis screws, implants and plate systems in soft and hard tissue.
Including: ENT surgery and Maxillofacial surgery
{5}------------------------------------------------
# 6 Technological Characteristics
The technological characteristics of the AMADEO are equivalent to the technological characteristics of the predicate device.
| | W&H Dentalwerk Bürmoos<br>GmbH - AMADEO<br>(New Device) | NSK - Primado2<br>(Predicate Device) | Result |
|------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Device Name | AMADEO | Primado2 | ----- |
| Indications for<br>Use | The drive unit for surgical<br>transmission instruments is<br>indicated for: drilling, milling,<br>cutting, sawing; screwing (for<br>positioning) of<br>osteosynthesis screws,<br>implants and plate systems<br>in soft and hard tissue.<br>Including: ENT surgery and<br>Maxillofacial surgery | The Primado2 is an electrically<br>powered total surgical system<br>which is intended for cutting,<br>drilling, sawing and otherwise<br>manipulating soft tissue, hard<br>tissue, bone, bone cement,<br>prosthesis, implant and other<br>bone related tissue in a variety<br>of surgical procedures,<br>including but not limited to<br>Cranial (craniofacial and<br>maxillofacial), ENT,<br>Endoscopic/Arthroscopic,<br>Neuro, Orthopedic, Spinal, and<br>General surgical procedures. | Equivalent 1) |
| Regulation<br>Number | 874.4850 | 874.4850 | Identical1) |
| Secondary<br>Regulation<br>Number | 872.4120 | 872.4120 | Identical 1) |
| Class | II | II | Identical |
| Product Code | ERL, DZI | ERL, DZI, GEY, HWE, HBC,<br>DZJ, HBE, GFF, EQJ | Equivalent1) |
| Regulation<br>Generic Name<br>(Primary Code,<br>Neurology<br>excluded) | Drill, Surgical, Ent (Electric or<br>Pneumatic) Including<br>Handpiece | Drill, Surgical, Ent (Electric or<br>Pneumatic) Including<br>Handpiece | Identical1) |
| Sterility | Provided non-sterile | Provided non-sterile | Equivalent |
| Use | Rx only | Rx only | Equivalent |
| Basic functions | The basic function is the<br>conversion of electrical<br>energy into a mechanical<br>rotary motion. In addition,<br>depending on the treatment,<br>a physiological saline<br>solution is pumped to the<br>treatment site by a<br>displacement pump.<br>Via control unit with<br>electronics, the user can<br>change the most crucial<br>operating parameters within<br>predetermined limits:<br>> Motor speed<br>> Torque setting<br>> Coolant flow rate via<br>pump speed control 0 to<br>100% (100%<br>corresponds to max.<br>amount of water)<br>> Transmission ratio with<br>various handpieces and<br>contra angle handpieces) | The Primado2 consists of the<br>control unit, the motor, the foot<br>control and various handpieces<br>for use with specific motors.<br>The control unit drives the<br>motors during procedures and<br>is used to control the functions<br>related to that motor such as<br>speed and rotational direction.<br>The control unit also<br>incorporates the irrigation pump<br>and controls the irrigation<br>functions.<br>The software allows for the<br>control of the devices features<br>such as:<br>> speed control<br>> rotational direction<br>> irrigation control<br>etc. | Equivalent |
| Mains voltage<br>Frequency | 120 V<br>50-60 Hz | 120 V<br>50/60 Hz | Identical |
| Operating Mode | intermittent duty S3 (load<br>time max. 3min / rest time<br>min. 5min) | intermittent duty (load/rest<br>depending on motors used) | Equivalent |
| Foot control | > wireless<br>> wired | > wired | Equivalent |
#### Device Characteristics Table 6.1
{6}------------------------------------------------
{7}------------------------------------------------
1) The indications of use of the AMADEO do not include neurological indications. Considering FDA's Guideline "Bundling Multiple Devices or Multiple Indications in a Single Submission, issued June 22, 2007" bundling of different review panels within different review division/group, OHT1 for ENT, Dental and OHT5 for Neurology, is not appropriate. The removal of neurological indications does not influence the device's substantial equivalence. Burs or other corresponding accessories are not part of the submission.
#### 6.2 Summary of Technological Characteristics
The proposed device is similar in terms of design, operating principles and intended use and has similar technological characteristics as the predicate device. The new device shares technological characteristics with the predicate device but also has some differences. The differences in the technological characteristics can be evaluated as minor and reflect market strategy and/or perceived user preferences. The differences do not impact substantial equivalence of the device.
The materials used on this device are also used in the legally marketed predicate device.
#### 7 Performance Data
Non-clinical testing has been performed showing that the device performs as intended and are substantially equivalent to the predicate device Primado2 (K132264).
#### 7.1 Biocompatibility
An evaluation of biocompatibility according to ISO 10993-1, ISO 10993-10, ISO 10993-10, ISO 10993-
11, ISO 10993-12, ISO 10993-17, ISO 10993-18 and ISO 10993-23 was performed.
#### 7.2 Electromagnetic Compatibility and Electrical Safety
Electrical safety and EMC testing were conducted. The AMADEO is in compliance with IEC 60601- 1 as well as IEC 60601-1-2.
#### 7.3 Reprocessing Validation
Reprocessing validation was provided per the FDA Guidance Document for "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling". Cleaning and intermediate level disinfection validation was provided for the control unit and foot controller. Cleaning and sterilization was provided for the motor cable and handpieces.
#### 7.4 Bench Testing
Functional testing of the AMADEO to test the application, settings, features, and touchscreen per the device specifications requirements.
### 8 Software
Software verification according to IEC 62304 and the FDA Guidance Document for Software Contained in Medical Device was conducted and the necessary software documentation according to the defined moderate level of concern was provided.
#### 9 Substantial Equivalence Summary / Conclusion
Based on available 510(k) information provided herein, our AMADEO is considered to be substantially equivalent to the predicate device Primado2 in terms of indication for use, materials and technology, design and performance specifications.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.