K240248 · Classys, Inc. · GEI · Apr 29, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240248
Device Name
Volnewmer™
Applicant
Classys, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 29, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Volnewmer™ is intended for use in dermatologic procedures for electrocoagulation and hemostasis of soft tissue.
Device Story
Volnewmer™ is a noninvasive monopolar radiofrequency (RF) therapy system used in dermatologic procedures. The device operates at 6.78 MHz with a maximum output power of 115 W. It delivers RF energy to soft tissue via a disposable/reusable tip for selective coagulation and hemostasis, while simultaneously providing conductive water-based cooling to the epidermis. The system includes a console, handpieces, treatment tips, return pad, and footswitch. The operator interacts with the device via an LCD touchscreen to control energy delivery. By tuning RF output across varying tissue impedances, the device ensures consistent energy delivery. The system is intended for use by healthcare professionals in clinical settings to achieve tissue coagulation, potentially benefiting patients through noninvasive dermatologic treatment.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including electrical safety and EMC testing per IEC 60601-1, 60601-1-2, and 60601-2-2; biocompatibility testing per ISO 10993-1; software verification and validation; and in-vivo thermal effect and safety testing.
Indicated for dermatologic procedures requiring electrocoagulation and hemostasis of soft tissue.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
April 29, 2024
Classys Inc. James Hoon Lim Regulatory Affairs Team Manager, Classys Inc. 208 Teheran-ro. Gangnam-gu Seoul. 06220 Korea, South
#### Re: K240248
Trade/Device Name: Volnewmer™ Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: January 30, 2024 Received: January 30, 2024
#### Dear James Hoon Lim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
# Sincerely,
Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.04.29 09:23:20 -04'00 Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Submission Number (if known)
Device Name
Volnewmer TM
Indications for Use (Describe)
Volnewmer™ is intended for use in dermatologic procedures for electrocoagulation and hemostasis of soft tissue.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
The assigned 510(k) Number: K240248
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
# 1. Submitter's Information
- Name: CLASSYS Inc.
- . Address: 208 Teheran-ro, Gangnam-gu, Seoul, Republic of Korea
- . Postal Code: 06220
- General Telephone: +82-2-1544-3481 .
### Contact Person:
- Contact Person: Mr. James Hoon Lim (RA Team Manager, Classys)
- Tel: +82-10-3351-1414
- Email: h.lim@classys.com
### 2. Date of the summary prepared: April 25, 2024
### 3. Subject Device Information
- Type of 510(k): Traditional ●
- Trade Name: Volnewmer™ ●
- Classification Name: Electrosurgical Cutting and Coagulation Device and Accessories ●
- Common Name: Radio Frequency Therapy System ●
- Review Panel: General & Plastic Surgery ●
- Product Code: GEI ●
- Regulation: 21 CFR 878.4400 ●
- . Regulatory Class: Class II
# 4. Predicate Device Information
- 510(k) Number: K170758
- Trade/Device Name: Thermage FLX System .
- Regulation Number: 21 CFR 878.4400
- Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories ●
- Regulatory Class: Class II ●
- Product Code: GEI, ISA ●
- Manufacturer: Solta Medical, Inc.
{4}------------------------------------------------
#### 5. Reference Device Information
- . 510(k) Number: K211562
- Trade/Device Name: Virtue RF ●
- Regulation Number: 21 CFR 878.4400 .
- Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories ●
- Regulatory Class: Class II ●
- . Product Code: GEI
- Manufacturer: ShenB Co., Ltd. .
# 6. Device Description
Volnewmer™ is a noninvasive monopolar radiofrequency (RF) therapy system. The radiofrequency output of the device is 6.78 MHz, and the maximum output power is 115 W.
- . Volnewmer™ delivers radiofrequency energy for selective coagulation of tissue while conductively cooling the epidermis.
- Volnewmer™ delivers energy from the disposable/reusable tip to the patient.
- . Volnewmer™ employs radiofrequency tuning to provide radiofrequency energy across a range of impedances for delivery to the patient through the disposable/reusable tip.
Volnewmer™ consists of the following components:
- Console ●
- . Handpiece x2
- Treatment Tips ●
- Accessories:
- Return Pad -
- Return Pad Cable -
- Coupling Gel -
- Footswitch -
### 7. Indications for Use
Volnewmer™ is intended for use in dermatologic procedures for electrocoagulation and hemostasis of soft tissue.
### 8. Comparison to Predicate Device
The differences between the subject device and predicate device do not raise new issues of safety or effectiveness.
{5}------------------------------------------------
| | | K240248 | Page 3 of 4 |
|-----------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison<br>Item | Proposed Device,<br>VolnewmerTM | Predicate Device,<br>Thermage FLX System, | Reference Device,<br>Virtue RF |
| 510(k)<br>Number | K240248 | K170758 | K211562 |
| Trade/Device<br>Name | VolnewmerTM | Thermage FLX System | Virtue RF |
| Regulation<br>Number | 21 CFR 878.4400 | 21 CFR 878.4400 | 21 CFR 878.4400 |
| Regulation<br>Name | Electrosurgical Cutting and<br>Coagulation Device and<br>Accessories | Electrosurgical Cutting and<br>Coagulation Device and<br>Accessories | Electrosurgical Cutting and<br>Coagulation Device and<br>Accessories |
| Regulatory<br>Class | Class II | Class II | Class II |
| Product Code | GEI | GEI, ISA | GEI |
| Manufacture<br>r | CLASSYS INC., Inc. | Solta Medical, Inc. | ShenB Co., Ltd. |
| Indication<br>For Use /<br>Intended Use | Dermatologic procedures for<br>electrocoagulation and<br>hemostasis of soft tissue. | Dermatologic and general<br>surgical procedures for<br>electrocoagulation and<br>hemostasis of soft tissue. | The Virtue RF System is<br>intended for use in<br>dermatologic and general<br>surgical procedures for<br>electrocoagulation and<br>hemostasis. |
| Output<br>Frequency | 6.78MHz | 6.78MHz | 0.5 MHz ~ 2 MHz |
| Max Output<br>Power | 115 W | 400 W | 220 W |
| User<br>Interface | LCD Touchscreen<br>Technology for user<br>interaction and controls. | LCD Touchscreen<br>Technology for user<br>interaction and controls. | LCD Touchscreen<br>Technology for user<br>interaction and controls. |
| Tip<br>Specification | • 0.25 cm2<br>• 3 cm2<br>• 4 cm2<br>• 16 cm2 | • 0.25 cm2<br>• 3 cm2<br>• 4 cm2<br>• 16 cm2 | • 2.78 cm2 / 36pin<br>• 3.12 cm2 / 12pin<br>• 6 cm2 / 36pin |
| Cooling<br>Mechanism | Water cooling system | R-134A Cryogen Gas<br>Cooling | Water cooling system |
{6}------------------------------------------------
#### 9. Summary of Non-Clinical Testing
Verification/validation activities from non-clinical testing as described below demonstrate that the differences do not raise any new issues of safety or effectiveness of the subject device compared to the predicate device.
EMC and Electrical safety of the subject device was tested in compliance with IEC 60601-1 Edition 3.2; IEC 60601-1-2 Edition 4.0 and IEC 60601-2-2 Edition 6.0
Thermal Effect & Safety In-Vivo Test were conducted for evaluating the safety and performance profile of the subject device. The test results support the substantial equivalence.
Bench testing was performed to ensure that the subject device performs as intended and meets design specifications.
Biocompatibility testing was performed in compliance with ISO 10993-1 and FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (issued on September 8, 2023) to demonstrate biocompatibility of the patient-contacting components of the device.
Software Verification and Validation testing were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions" (issued on June 14, 2023). The software documentation level for the subject device is considered as 'Enhanced Documentation'.
### 11. Conclusion
The Volnewmer™ shares the same indications for use, fundamental scientific technology, design and functional features. Non-clinical test results demonstrate that the Volnewmer™ safety and performance are substantially equivalent to the predicate device for the requested indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.