K211562 · ShenB Co., Ltd. · GEI · Nov 23, 2021 · General, Plastic Surgery
Device Facts
Record ID
K211562
Device Name
Virtue RF
Applicant
ShenB Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 23, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Virtue RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. Smart RF Handpiece: The Smart RF Handpiece with 1 MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis, and the percutaneous treatment of facial wrinkles. The VirtueRF System for facial wrinkles is intend for use with Skin types I-V. The Smart RF Handpiece with 2 MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Smart RF with 2MHz functionality is not intended to treat wrinkles. Exact RF Handpiece: The Exact RF Handpiece with 1MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Exact RF Handpiece is not intended to treat wrinkles. Deep RF Handpiece: The Deep RF Handpiece with 1MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 1MHz frequency is not intended to treat wrinkles. The Deep RF Handpiece with 2MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 2MHz frequency is not intended to treat wrinkles. The Deep RF Handpiece with 0.5MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 0.5MHz frequency is not intended to treat wrinkles.
Device Story
VirtueRF is a radiofrequency (RF) electrosurgical system for dermatologic and general surgical procedures. System comprises a main console, foot switch, and three types of handpieces (SmartRF, ExactRF, DeepRF) using disposable needle arrays (1, 12, or 36 needles). Device delivers RF energy at 0.5MHz, 1MHz, or 2MHz to human tissue. ExactRF operates in monopolar mode with a grounding plate; SmartRF and DeepRF operate in bipolar mode. DeepRF includes a detachable TEC cooling plate with internal water cooling. Operated by clinicians in a clinical setting. Energy delivery induces thermal effects for coagulation and hemostasis; percutaneous needle insertion with RF energy treats facial wrinkles. Output parameters (power, duration) are controlled by the clinician to achieve desired tissue effects. Benefits include precise tissue coagulation and wrinkle reduction.
Clinical Evidence
Bench testing only. Provided data includes IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-2 (high-frequency surgical equipment), and ISO 10993-1 (biocompatibility). Thermal effect tests were conducted on liver, kidney, muscle, and skin tissues to validate performance against the predicate.
Technological Characteristics
RF electrosurgical system; 0.5MHz, 1MHz, 2MHz frequencies. Bipolar and monopolar modalities. Disposable microneedle arrays (1, 12, 36 needles). Max power up to 220W. Features TEC cooling plate with water cooling for DeepRF. Complies with IEC 60601-1, 60601-1-2, 60601-2-2, and ISO 10993-1.
Indications for Use
Indicated for dermatologic and general surgical electrocoagulation and hemostasis. Smart RF (1MHz) also indicated for percutaneous treatment of facial wrinkles in Skin Types I-V. Not indicated for wrinkle treatment with other handpieces or frequencies.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
November 23, 2021
ShenB Co Ltd % Connie Hoy Consultant Hoy and Associates 3916 N. Potsdam Ave. #4676 Sioux Falls, South Dakota 57104
Re: K211562
Trade/Device Name: Virtue RF Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: October 19, 2021 Received: October 20, 2021
Dear Connie Hoy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. for Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211562
Device Name
Virtue RF
#### Indications for Use (Describe)
The Virtue RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
#### Smart RF Handpiece:
The Smart RF Handpiece with 1MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis, and the percutaneous treatment of facial wrinkles. The VirtueRF Electrosurgical System for facial wrinkles is intended for use with Skin Types I-V.
The Smart RF Handpiece with 2MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The with 2MHz functionality is not intended to treat wrinkles.
#### Exact RF Handpiece:
The Exact RF Handpiece with 1MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Exact RF Handpiece is not intended to treat wrinkles.
Deep RF Handpiece:
The Deep RF Handpiece with 1MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 1MHz functionality is not intended to treat wrinkles.
The Deep RF Handpiece with 2MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 2MHz functionality is not intended to treat wrinkles.
The Deep RF Handpiece with 0.5MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 0.5MHz functionality is not intended to treat wrinkles.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
This 510(K) Summary of safety and effectiveness for the VirtueRF is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.
| Applicant: | ShenB Co Ltd. |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Shenb Bldg 148<br>Seongsui-ro<br>Seongdong-Gu, Seoul , KR 04796 |
| Contact Person: | Bora Kim |
| Contact Information: | kimbora@shenb.com<br>+82-70-4814-2978 |
| Preparation Date: | November 18, 2021 |
| Device Trade Name: | VirtueRF |
| Common Name: | Electrosurgical, cutting & coagulation & accessories |
| Regulation Name: | Electrosurgical cutting and coagulation device and accessories |
| Regulation Number: | 21 CFR 878.4400 (Product Code: GEI) |
| Legally Marketed Predicate Device: | Virtue RF System<br>K202415 |
| Legally Marketed Reference Device: | Primaeva Medical Miratone System<br>K082391<br>Agnes<br>K192728 |
| Regulatory Class: | Class II Prescription Use |
| Description of the VirtueRF : | The VirtueRF device is a radio frequency output device using<br>0.5MHz, 1MHz & 2Mhz to deliver radio frequency energy to the<br>human body with 3 applicators that attach to the connecting<br>cable. The energy is applied to the human body using sterile<br>disposable needle arrays consisting of 36 needles, 12 needles, o<br>1 needle.<br>This product consists of main body, SmartRF Bipolar handpiece,<br>DeepRF Bipolar handpiece, ExactRF Monopolar handpiece,<br>connector cable, grounding plate for ExactRF, cooling plate for<br>the DeepRF, foot switch and power cord. |
{5}------------------------------------------------
Intended use / Indication for Use of Virtue RF :
The Virtue RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Smart RF Handpiece:
The Smart RF Handpiece with 1 MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis, and the percutaneous treatment of facial wrinkles. The VirtueRF System for facial wrinkles is intend for use with Skin types I-V.
The Smart RF Handpiece with 2 MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Smart RF with 2MHz functionality is not intended to treat wrinkles.
Exact RF Handpiece:
The Exact RF Handpiece with 1MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Exact RF Handpiece is not intended to treat wrinkles.
Deep RF Handpiece:
The Deep RF Handpiece with 1MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 1MHz frequency is not intended to treat wrinkles.
The Deep RF Handpiece with 2MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 2MHz frequency is not intended to treat wrinkles.
The Deep RF Handpiece with 0.5MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 0.5MHz frequency is not intended to treat wrinkles.
{6}------------------------------------------------
Performance Testing: The following performance data was provided in support of the substantial equivalence determination:
> IEC 60601-1 Test for Medical Electrical equipment was performed for General Requirements for basic safety and essential performance
> IEC 60601-1-2 Test for Medical Equipment for General Requirements for basic safety and essential performance: electromagnetic compatibility
IEC 60601-2-2 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
IEC 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
Thermal Effect Tests on 4 types of tissues (liver, kidney, muscle and skin) per FDA guidance Premarket Notification (510(K)) Submissions for Electrosurgical Devices for General Surgery. Test shows that the VirtueRF is substantially equivalent to the predicate device.
Comparison to the predicate and reference devices:
| Comparison for Smart RF handpiece - 1MHz functionality | | | |
|--------------------------------------------------------|---------------------------------------|------------------------------------------------------|------------|
| | Proposed Device<br>ShenB VirtueRF | Predicate device<br>for 1MHz<br>Virtue RF<br>K202415 | Comparison |
| Source of<br>energy/Modality | Bipolar Fractional RF | Bipolar Fractional RF | Same |
| Radiofrequency | 1MHz | 1MHz | Same |
| Output energy type | High Frequency | High Frequency | Same |
| Electrode type | Bipolar microneedle | Bipolar microneedle | Same |
| Max Power | 35.9 W | 35.9 W | Same |
| RF Duration | 100ms-800ms, with<br>100ms increments | 100ms-800ms, with<br>100ms increments | Same |
{7}------------------------------------------------
| Comparison for Smart RF Handpiece - 2MHZ Functionality | | | |
|--------------------------------------------------------|------------------------------------------------------------|------------------------------------------|------------|
| | Proposed Device<br>ShenB VirtueRF<br>Smart RF<br>Handpiece | Predicate Device<br>Virtue RF<br>K202415 | Comparison |
| Source of<br>Energy/Modality | Bipolar Fractional RF | Bipolar Fractional RF | Same |
| Radiofrequency | 2MHz | 2MHz | Same |
| Output energy type | High Frequency | High Frequency | Same |
| Electrode type | Bipolar microneedle | Bipolar microneedle | Same |
| Max Power | 25 W | 25W | Same |
| RF Duration | 100ms-800ms, with<br>100ms increments | 100ms-800ms, with<br>100ms increments | Same |
| Comparison for ExactRF Handpiece | | | | |
|----------------------------------|----------------------------------------------------|---------------------------------------------|-----------------------------------------|---------------------------------------------------------|
| | Proposed Device<br>Virtue RF<br>Exact RF Handpiece | Predicate Device<br>Virtue RF<br>K202415 | Reference Device<br>Agnes<br>K160469 | Comparison |
| Source of<br>Energy/Modality | Monopolar | Bipolar | Monopolar | Similar; same as<br>reference device |
| Radiofrequency | 1MHz | 1Mhz & 2MHz | 1MHz | same |
| Electrode type | Monopolar single<br>needle RF electrode | Bipolar<br>microneedle | Monopolar single<br>needle RF electrode | Similar; same as<br>reference device |
| Number of Needles | 1 | 36 | 1 | Different; same as<br>reference device |
| Grounding<br>Mechanism | Electrode plate | N/A | Disposable neutral<br>electrode pad | Similar to<br>reference device |
| Max Power | 46W | 35.9W | 46W at 200Ω | Same as<br>reference; similar<br>to predicate<br>device |
| RF Duration | 100ms-800ms, with<br>100ms increments | 100ms-800ms,<br>with<br>100ms<br>increments | 50ms - 2,000ms | Same; similar to<br>reference device |
| Comparison for DeepRF Handpiece – 1MHz & 2MHz Functionality | | | |
|-------------------------------------------------------------|--------------------------------------------|-------------------------------------------------|-----------|
| | Proposed Device<br>Deep RF Handpiece- 1MHz | Predicate Device Virtue<br>RF - 1MHz<br>K202415 | |
| Source of<br>Energy/Modality | Bipolar | Bipolar | Same |
| Radiofrequency | 1MHz | 1MHz | Same |
| Electrode type | Bipolar microneedles with RF<br>Electrode | Bipolar microneedles<br>with RF Electrode | Same |
| Max Power | 155 W at 500Ω | 35.9W | Different |
| RF Duration | 100ms-800ms, with<br>100ms increments | 100ms-800ms, with<br>100ms increments | Same |
{8}------------------------------------------------
| | Proposed Device - Deep RF -<br>2MHz | Predicate Device<br>Virtue RF – 2MHz<br>K202415 | |
|------------------------------|-------------------------------------------|-------------------------------------------------|---------|
| Source of<br>Energy/Modality | Bipolar | Bipolar | Same |
| Radiofrequency | 2MHz | 2MHz | Same |
| Electrode type | Bipolar microneedles with RF<br>Electrode | Bipolar microneedles<br>with RF Electrode | Same |
| Number of<br>Needles | 12 or 36 | 36 | Similar |
| Max Power | 60W | 35.9W | Similar |
| RF Duration | 100ms-800ms, with<br>100ms increments | 100ms-800ms, with<br>100ms increments | Same |
| Comparison for DeepRF Handpiece - 0.5MHz Functionality | | | | |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|--------------------------------------------------------------------------------|------------|
| | Proposed Device<br>- Deep RF<br>0.5MHz | Predicate Device<br>Virtue RF | Reference Device<br>Primaeva<br>K082391 | Comparison |
| Source of<br>Energy/Modality | Bipolar | N/A. Approved<br>Virtue does not<br>have 0.5MHz<br>functionality | Bipolar | Same |
| Radiofrequency | 0.5MHz | N/A | .46MHz | Similar |
| Electrode type | Bipolar<br>Handpiece +<br>Micro needle<br>electrodes | N/A | Bipolar Handpiece +<br>Micro needle<br>electrodes | Same |
| Number of Needles | 12 or 36 | N/A | 10 (3 different tips all<br>with 10 needles) | Similar |
| Max Power | 220W | N/A | 25W | Different |
| RF Duration | 100-800ms and<br>can be adjusted<br>100ms | N/A | Adjustable 10-50 ms<br>on for every 100 ms | Same |
| Cooling Modality | Detachable TEC<br>Cooling plate.<br>Water cooling<br>system inside<br>handpiece<br>controls and<br>maintains<br>temperature of<br>cooling plate. | N/A | Cooler controller and<br>reusable cooler<br>controller<br>handpiece/applicator | Similar |
{9}------------------------------------------------
Conclusion: The proposed VirtueRF is the same system as the FDA cleared VirtueRF (K202415) and adds 2 additional handpieces, each with their own disposable cartridges. The indications for use and the technological characteristics are the same as the predicate device and the reference devices. Substantial equivalence is supported by the performance testing and there are no new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.