Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System
K240137 · Neos Surgery S.L · GXR · Feb 16, 2024 · Neurology
Device Facts
Record ID
K240137
Device Name
Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System
Applicant
Neos Surgery S.L
Product Code
GXR · Neurology
Decision Date
Feb 16, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cranial LOOP Cranial Bone Fixation Systems: Cranial LOOP (L) and Cranial Loop (XL), are longterm implantable devices indicated for post-craniotomy bone flap fixation. In cranial bone fixation procedures, the Cranial LOOP (FC050000) and Cranial LOOP (L) (FC050100) are for use within the osteotomy line (calvarial gap) while the Cranial LOOP (XL) (FC050200) is to be used for covering a standard 14 mm cranial burr hole only.
Device Story
Biocompatible, single-use, sterile cranial bone fixation system; fixes bone flap to skull post-craniotomy. Device consists of two platforms linked by two adjustable cable ties with a locking mechanism. Surgeon applies device by tightening upper platform toward lower platform via manual pressure and pulling on handle; mechanism allows forward movement but prevents backward movement, clamping bone flap to cranial bone. Used in neurosurgical procedures; no specific surgical instrument required for implantation. Benefits include secure bone flap fixation without complex hardware. Performance verified via biomechanical testing (displacement force and tensile load to failure) on synthetic and cadaveric craniums.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by bench-top biomechanical and functional testing, including displacement force and tensile load to failure, comparing the subject device to the predicate using synthetic and cadaveric cranium models.
Technological Characteristics
Biocompatible implantable fixation system. Consists of two platforms linked by two adjustable cable ties with a one-way locking mechanism. Dimensions accommodate cranial thicknesses from 1.5 mm to 24 mm (LOOP/LOOP L) or 4 mm to 24 mm (LOOP XL). Mechanical clamping principle; no energy source. Provided sterile for single use.
Indications for Use
Indicated for post-craniotomy bone flap fixation in patients requiring long-term cranial bone fixation. Cranial LOOP and Cranial LOOP L are for use within the osteotomy line (calvarial gap); Cranial LOOP XL is for covering a standard 14 mm cranial burr hole.
Regulatory Classification
Identification
A burr hole cover is a plastic or metal device used to cover or plug holes drilled into the skull during surgery and to reattach cranial bone removed during surgery.
Predicate Devices
Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System (K132044)
Submission Summary (Full Text)
{0}------------------------------------------------
February 16, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol of the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". The logo is simple and professional, and it is easily recognizable.
Neos Surgery S.L. Eduard Garcia QA & RA Manager Ceramistes 2 Cerdanyola, Spain 08290
Re: K240137
Trade/Device Name: Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System Regulation Number: 21 CFR 882.5250 Regulation Name: Burr hole cover Regulatory Class: Class II Product Code: GXR Dated: January 17, 2024 Received: January 18, 2024
Dear Eduard Garcia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Adam D. Pierce -S" in a large font on the left side of the image. On the right side of the image, it says "Digitally signed by Adam D. Pierce -S" and "Date: 2024.02.16 12:20:35 -05'00'". The text on the right side of the image is smaller than the text on the left side of the image.
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K240137
Device Name
Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System
Indications for Use (Describe)
The Cranial LOOP Cranial Bone Fixation Systems: Cranial LOOP (L) and Cranial Loop (XL), are longterm implantable devices indicated for post-craniotomy bone flap fixation.
In cranial bone fixation procedures, the Cranial LOOP (FC050000) and Cranial LOOP (L) (FC050100) are for use within the osteotomy line (calvarial gap) while the Cranial LOOP (XL) (FC050200) is to be used for covering a standard 14 mm cranial burr hole only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
< | Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for the Cranial LOOP, Cranial LOOP Land Cranial LOOP XL Cranial Bone Fixation System Special 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
- Sponsor: Neos Surgery S.L. Parc Tecnològic del Vallès C/Ceramistes, 2 08290 Cerdanyola del Vallès Barcelona, Spain
- Contact: Cherita James M Squared Associates, Inc. 127 West 30th Street, 9th Floor New York, New York 10001 Ph: 347-954-0624 Fax: 703-562-9797 Email: Cjames@MSquaredAssociates.com
Date: February 16, 2024
Proprietary Name: Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System Common Name: Cover, Burr Hole
Regulatory Class: II
Regulation: 882.5250 Burr hole cover
Product Codes: GXR
Predicate Device(s): Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone
Fixation System - K132044
Device Description: The Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System is a biocompatible, postoperative cranial bone fixation system that fixes the bone flap to the skull, without any specific surgical instrument for its handling or implantation. It is provided sterile, for single use. Cranial LOOP and Cranial LOOP L are applied in the craniotomy gap. They can fix cranial thicknesses ranging from 1.5 mm to 24 mm and
{5}------------------------------------------------
gaps ranging between 1.7 mm and those made using a craniotome standard cranial router. Cranial LOOP XL is applied in a burr hole made using a standard drill 14 mm. They can fix cranial thicknesses ranging from 4mm to 24 mm.
Indications for Use: The Cranial Bone Fixation Systems: Cranial LOOP, Cranial LOOP (L) and Cranial Loop (XL), are long-term implantable devices indicated for postcraniotomy bone flap fixation.
In cranial bone fixation procedures, the Cranial LOOP (FC050000) and Cranial LOOP (L) (FC050100) are for use within the osteotomy line (calvarial gap) while the Cranial LOOP (XL) (FC050200) is to be used for covering a standard 14 mm cranial burr hole only.
#### Technological characteristics, comparison to predicate device
The device under evaluation has the same technical characteristics as the predicate device. Both designs consist of two platforms linked by two adjustable cable ties. These cable ties are joined to the lower platform and have a locking system that allows the movement of the upper platform towards the lower platform but impedes backward movements. The upper platform is tightened to the cranial bone and bone flap by the surgeon with the aid of the applier by a simple pressure on the applier and pulling on the handle. Device remains adjusted to the bone thickness, and it fixes the bone flap to the cranial bone, like a clamp.
#### Performance testing
Biomechanical and functional testing (displacement force when fixated to a synthetic and cadaveric cranium and tensile load to device failure) were performed to evaluate the impact of the changes to Cranial LOOP with respect to its previously cleared version (identified as the predicate device). The acceptance criteria were met in both tests. The performance testing showed that the Cranial LOOP had equivalent or better performance compared to its predicate, thus demonstrating that the device is as safe and effective as it does not raise different questions of safety and effectiveness than the predicate.
#### Discussion of clinical testing
No clinical testing was performed to support this submission.
{6}------------------------------------------------
### Substantial Equivalence/ Conclusion
Based on the comparison of the indications for use, product technical characteristics, principles of operation and performance, the modified Cranial LOOP has demonstrated to be substantial equivalence to the previously cleared version of Cranial LOOP (identified as the predicate). The differences do not raise new issues of safety or effectiveness, based on the risk assessment performed and on the results of the verification / validation activities that have been conducted. The results of the performance testing demonstrate that the subject device meets its specifications.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.