CRANIAL LOOP, CRANIAL LOOP L AND CRANIAL LOOP XL AND CRANIAL BONE FIXATION SYSTEM
Applicant
Neos Surgery S.L
Product Code
GXR · Neurology
Decision Date
Sep 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cranial LOOP Cranial Bone Fixation Systems: Cranial LOOP, Cranial LOOP L and Cranial LOOP XL, are long-term implantable devices indicated for post-craniotomy bone flap fixation. In cranial bone fixation procedures, the Cranial LOOP (FC050000) and Cranial LOOP L (FC050100) are for use within the calvarial gap while the Cranial LOOP XL (FC050200) is to be used for covering a standard 14 mm cranial burr hole only.
Device Story
Biocompatible, postoperative cranial bone fixation system; fixes bone flap to skull. Device consists of upper and lower platforms with ratcheting mechanism; no specific surgical instruments required for implantation. Cranial LOOP and LOOP L used in calvarial gaps; Cranial LOOP XL used for 14 mm burr holes. Surgeon implants device during craniotomy; device provides mechanical fixation of bone flap. Benefits include rapid, simple application; low profile; adaptation to varying cranium curvatures and thicknesses. Mechanical locking mechanism secures bone flap; prevents lateral movement. Sterile, single-use device.
Clinical Evidence
No clinical data required. Evidence provided via bench testing: functional verification of locking mechanism, cadaveric testing for ease of use and anatomical adaptation, and biomechanical testing (push-in, pull-out, and dynamic load testing for 0.5 million cycles). Accelerated aging testing (ASTM F 1980) confirmed shelf-life integrity.
Technological Characteristics
Biocompatible cranial bone fixation system. Consists of upper and lower platforms with ratcheting locking mechanism. Dimensions vary by model (LOOP, LOOP L, LOOP XL). Sterile, single-use. No software or energy source. Packaging validated per ISO 11607.
Indications for Use
Indicated for post-craniotomy bone flap fixation in patients requiring cranial bone fixation. Cranial LOOP and Cranial LOOP L are for use within the calvarial gap; Cranial LOOP XL is for covering a standard 14 mm cranial burr hole.
Regulatory Classification
Identification
A burr hole cover is a plastic or metal device used to cover or plug holes drilled into the skull during surgery and to reattach cranial bone removed during surgery.
Predicate Devices
Cranial LOOP Cranial Bone Fixation System (K101235)
Submission Summary (Full Text)
{0}------------------------------------------------
K132044
## 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for the Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System 510(k) premarket notification.
| Applicant: | Neos Surgery S.L. C/Ceramistes. 2 08290-Cerdanyola del Vallès Barcelona, Spain |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 3008974316 |
| US Contact: | Cherita James M Squared Associates, Inc. 901 King Street, Suite 101 Alexandria, VA 22314 Ph: 703-562-9800 ext 257 Fax: 703-562-9797 E-mail: CJames@msquaredassociates.com |
| Date submitted: | August 27, 2013 |
| Proprietary Name: | Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System |
| Common Name: | BurrHole cover |
| Classification Status: | II |
| Regulation: | 21 CFR 882.5250 Burrhole cover |
| Panel: | Neurology |
| Product Code: | GXR |
| Predicate Device: | Cranial LOOP Cranial Bone Fixation System K101235 |
| Device Description: | The Cranial LOOP. Cranial LOOP L and Cranial LOOP XL Cranial Bone |
SEP 3 0 2013Fixation System is a biocompatible, postoperative cranial bone fixation system that fixes the bone flap to the skull, without any specific surgical instrument for its handling or implantation. It is provided sterile, for single use. This submission includes a line extension for the Cranial LOOP L
Cranial LOOP L and Cranial LOOP XL
{1}------------------------------------------------
| N° | Test | Methodology |
|----|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| A | Functional verification | Test 1: Functionality of the implantable parts (P01 (lower platform with ties) + P02 (upper platform)): Determine the breaking strength of P01+P02 union (ratchets). A load was applied to the platforms, in an opposite direction, in order to assess the strength of the locking mechanism. Test 2: Functionality of the handle: Determine the breaking strength of P01 (lower platform with ties) + P03 (handle) union (ratchets). A load was applied to the lower platform and the handle, in an opposite direction, in order to assess the strength of the locking mechanism. |
| B | Cadaveric test | Three different craniotomies were tested: Frontal-Parietal-Sphenoid wing craniotomy (left side) Frontal lateral (right side) Frontal-parietal (right side). One burr hole on the temporal bone. The craniotomies were obtained using Linvatec cranitome with saws (router) from 1.4 to 2.2 mm cutting diameter and 14 mm diameter perforator.FC050200 (Cranial LOOP XL ) are applied in each of the craniotomies according the IFUs. The following aspects were evaluated, Sufficient space for device placement No danger for the surgeon during device application. Correct positioning of the device Rapid and simple application of the device with minimum instruments No danger for duramater. Finally the duramater and the brain will be examined after the removal of the cranial fixation, detecting any possible incidence. Low epicranial and subcranial platform profiles. Adaptation to different cranium curvatures. Adaptation to different cranium thicknesses. Qualitative evaluation of the lateral movement. Qualitative evaluation of cranial fixation strength. Evaluation of the ease with which the system is removed. |
| C | Biomechanical tests | Test 1: Strength to bone flap compression (Push-in). A push in load was applied to a model which simulates the cranium and a bone flap with three Cranial LOOPs representing clinical use. Test 2: Pull-out strength test. A pull out load was applied to a model which simulates the cranium and a bone flap with three Cranial LOOPs representing clinical use. Test 3: Strength to dynamic load over the bone flap. A push in load to produce a displacement of 0.5 mm on the bone flap, was applied to a model which simulates the cranium and a bone flap with three Cranial LOOPs representing clinical use for 1/2M cycles. |
| D | Accelerated ageing test | The following parameters were verified after an aging: 'Blister seal strength' (5 samples). 'Leak integrity' (5 samples). 'Sterility' (5 samples). 'Tensile strength' (5 samples) The accelerated aging was conducted according to the ASTM F 1980 standard. Q10 Factor = 2.0 Real Time aging. RT = 5 years. AAT temperature = 60 °C, AAT = 131,02 days to simulate 5 years of real aging time |
・
{2}------------------------------------------------
and Cranial LOOP XL which have new device dimensions and design features that reduce lateral movement after implantation..
Cranial LOOP and Cranial LOOP L are applied in the craniotomy gap. It can fix cranial thicknesses ranging from 1.5 mm to 24 mm and gaps ranging between 1.7 mm and those made using a craniotome standard cranial router. Cranial LOOP XL is applied in a burr hole made using a standard drill 14 mm. They can fix cranial thicknesses ranging from 4 mm to 24 mm.
Intended Use: The Cranial LOOP Cranial Bone Fixation Systems: Cranial LOOP, Cranial LOOP L and Cranial LOOP XL, are long-term implantable devices indicated for post-craniotomy bone flap fixation.
In cranial bone fixation procedures, the Cranial LOOP (FC050000) and Cranial LOOP L (FC050100) are for use within the calvarial gap while the Cranial LOOP XL (FC050200) is to be used for covering a standard 14 mm cranial burr hole only.
#### Summary of Technological Characteristics:
The claim of substantial equivalence of the Cranial Loop to the predicate devices is based on the comparison of the intended use, product technical characteristics, performance characteristics and product handling.
Summary of Nonclinical Testing: Mechanical and various performance testing confirms the Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System performs as intended and is substantially equivalent to the predicate device. A Declaration of Conformity with design controls is provided.
{3}------------------------------------------------
| Nº | Test | Methodology |
|----|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| E | Profile and range of<br>bone thickness<br>covered. | Dimensional test are performed in order to validate:<br>Epicranial and subcranial profile. The epicranial profile protruding from the plane of the bone was<br>measured to ensure it does not exceed the established limits. And<br>the subcranial profile protruding from the plane of the bone was<br>measured to ensure it does not exceed the established limits Range of bone thickness The upper platform (P02) is lowered as much as possible towards the<br>lower platform (P01). Measure the distance between both platforms in<br>order to check the minimum bone thickness required.<br>The distance between platforms on ready-to-use device is measured in<br>order to check the maximum bone thickness |
| F | Packaging integrity | Packaging validation according ISO 11607. |
Summary of Clinical Testing: Clinical data was not required in support of this submission.
### Substantial Equivalence Discussion:
The line extension for the Cranial LOOP L and Cranial LOOP XL has the following similarities to the previously cleared Cranial LOOP Cranial Bone Fixation System- K101235:
- . has the same indicated use.
- uses the same operating principle, .
- incorporates the same basic device design, .
- incorporates the same materials .
- incorporates comparable manufacturing and packaging processes .
- incorporates the same sterilization process. .
Based on these similarities, and mechanical and performance tests outcomes, the Cranial LOOP L and Cranial LOOP XL are substantially equivalent to the predicate Cranial LOOP Cranial Bone Fixation System.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### September 30, 2013
Neos Surgery S.L. c/o Cherita James, Regulatory Consultant M Squared Associates, Inc. 901 King Street, Suite 101 Alexandria, VA 22314
Re: K132044
Trade/Device Name: Cranial LOOP, Cranial LOOP L, Cranial LOOP XL Cranial Fixation System Regulation Number: 21 CFR 882.5250 Regulation Name: Burr hole cover Regulatory Class: Class II Product Code: GXR Dated: June 21, 2013 Received: July 2, 2013
Dear Ms. James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Ms. James
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
### 510(k) Number: K132044
Cranial LOOP, Cranial LOOP L, and Cranial LOOP XL Cranial Bone Device Name: Fixation System
Indications For Use:
The Cranial LOOP Cranial Bone Fixation Systems: Cranial LOOP, Cranial LOOP L and Cranial LOOP XL, are long-term implantable devices indicated for post-craniotomy bone flap fixation.
In cranial bone fixation procedures, the Cranial LOOP (FC050000) and Cranial LOOP L (FC050100) are for use within the calvarial gap while the Cranial LOOP XL (FC050200) is to be used for covering a standard 14 mm cranial burr hole only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Joyce M. Whang -S
Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)
510(k) Number K132044
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.