Enteral Feeding Sets are intended to dispense liquid nutrition (feeding solutions) at a user controlled rate. These enteral feeding sets interface with a patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The enteral feeding sets include a bag to contain the feeding solution and/or a EnPlus spike to connect to a pre-filled container. The device is used for infants, children, adolescents and adults.
Device Story
Enteral Feeding Sets consist of tubing, drip chambers, roller clamps, and either a formula bag or an EnPlus spike for connection to pre-filled containers. Used in clinical or home settings to deliver liquid nutrition to patients via feeding tubes. Operated by clinicians or patients/caregivers. Gravity-fed or pump-driven flow; user controls rate via roller clamp or pump settings. Provides a closed or semi-closed system for nutrition administration. Benefits include reliable, controlled delivery of enteral nutrition, reducing risk of misconnection via ISO 80369-3 compliant connectors.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including metrology, leakage, tensile strength, flow rate, and biocompatibility (cytotoxicity, sensitization, irritation) per ISO 10993 series.
Technological Characteristics
Materials: PVC, ABS, Silicone, LDPE, paper. Components: EnFit connectors (ISO 80369-3:2016), ENPlus spikes (ISO 18250-3:2018), tubing, roller clamps, drip chambers. Non-sterile, single-use, disposable. Connectivity: ISO 80369-3 compliant small-bore connectors. No software or electronic components.
Indications for Use
Indicated for infants, children, adolescents, and adults requiring enteral nutrition delivery via feeding tube using gravity or pump-assisted systems.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 17, 2024
Hangzhou Primecare Medical Co., Ltd. Li Xuedong Regulatory Director Room 408-409, Zancheng Center West, Shangcheng District Hang zhou, Zhe jiang 310008 China
Re: K240039
Trade/Device Name: Enteral Feeding Sets Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube And Accessories Regulatory Class: Class II Product Code: PIF, KNT, Dated: March 18, 2024 Received: March 18, 2024
#### Dear Li Xuedong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
3 Page
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sivakami Venkatachalam -S
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K240039
Device Name
Enteral Feeding Sets
#### Indications for Use (Describe)
Enteral Feeding Sets are intended to dispense liquid nutrition (feeding solutions) at a user controlled rate. These enteral feeding sets interface with a patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The enteral feeding sets include a bag to contain the feeding solution and/or a EnPlus spike to connect to a pre-filled container. The device is used for infants, children, adolescents and adults.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a cursive, blue font. Below "PrimeCare" is the word "MEDICAL" in a smaller, sans-serif font, also in blue. A curved line underlines the word "MEDICAL".
## 510(k) Summary - K240039
Device Trade Name: Enteral Feeding Sets
Common Name: Enteral Feeding Sets
Classification Name: Gastrointestinal tube and accessories
Product Code: KNT, PIF
Regulation: 876.5980
Regulatory Class: Class II
#### Submitter/Manufacturer:
Hangzhou Primecare Medical Co., Ltd. Room 408-409, Zancheng Center West, Shangcheng District, Hangzhou Zhejiang 310008 China Tel: 86-571-81958620 Regulatory Contact: Li Xuedong, Regulatory Director
## Predicate Device
Device Trade Name: CONOD Enteral Feeding Sets 510(k) Number: K181276 Classification: Class II, 876.5980, Gastrointestinal tube and accessories Product Code: KNT, PIF
## Device Description
The Enteral Feeding Sets have four models, Enteral Feeding Pump bag Set, Enteral Feeding Pump EnPlus spike Set, Enteral Feeding Gravity bag Set, Enteral Feeding Gravity EnPlus spike Set. The Enteral Feeding Pump bag Set and Enteral Feeding Gravity bag Set both have a structure of a formula bag with 100ml graduations and a protective closure cap. The difference is the former is droved by pump and the latter is droved by gravity. Enteral Feeding Pump EnPlus spike Set and Enteral Feeding Gravity EnPlus spike Set feature a piercing spike (with or without screw cap) used to connect the set to a prefilled container of enteral feeding solution. It is a single-use, non-sterile device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a cursive, blue font. Below that, the word "MEDICAL" is written in a smaller, sans-serif font, also in blue. A curved line underlines the word "PrimeCare".
# Intended Use / Indications for Use
Enteral Feeding Sets are intended to dispense liquid nutrition (feeding solutions) at a user controlled rate. These enteral feeding sets interface with a patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The enteral feeding sets include a bag to contain the feeding solution and/or a EnPlus spike to connect to a pre-filled container. The device is used for infants, children, adolescents and adults.
## Technological Comparison
- > The subject device has the same intended use as the predicate device, dispense liquid nutrition (feeding solutions) at a user controlled rate. These enteral feeding sets interface with a patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution, and uses the same technology.
- > The subject device has the same technological characteristics and provide the same principle of operation as the predicate device.
- > The differences:
- The predicate device has only 1200ml volume bags, the subject device adds different volumes (1200ml,1000ml,500ml,100ml), Material and intended used are the same.
- The subject device adds one universal screw cap(with or without) to the spike set to extend the connectivity for different pre-filled formula containers. Material of the screw cap is exactly same as ENPlus spike.
- The subject device adds a new model that replaces the built-in enteral bag with ENPlus spike and screw cap so that it can connects to the pre-filled nutrition containers instead of having to prepare formula by the user. the ENPlus spike and screw cap introduced in this set is the same as those used in Enteral feeding pump EnPlus Spike sets.
- The subject device removes the EnFit transitional connector.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a stylized, cursive font in blue. Below "PrimeCare" is the word "MEDICAL" in smaller, block letters, also in blue. A curved line underlines the word "MEDICAL".
#### Shown below is a side by side comparison of the subject device with the predicate device. A
| Device<br>Characteristic | Subject device | Predicate device<br>(K181276) | Results |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Enteral Feeding Sets | CONOD Enteral Feeding Sets | same |
| Indications for<br>Use | Enteral Feeding Sets are intended to dispense liquid nutrition<br>(feeding solutions) at a user controlled rate. These enteral<br>feeding sets interface with a patient's feeding tube and may use<br>gravity or an enteral feeding pump to dispense feeding solution.<br>The enteral feeding sets include a bag to contain the feeding<br>solution and/or a EnPlus spike to connect to a pre-filled<br>container. The device is used for infants, children, adolescents<br>and adults. | CONOD Enteral Feeding Sets are intended to dispense<br>liquid nutrition (feeding solutions) at a user controlled rate.<br>These enteral feeding sets interface with a patient's<br>feeding tube and may use gravity or an enteral feeding<br>pump to dispense feeding solution. The enteral feeding<br>sets include a bag to contain the feeding solution and/or a<br>piercing spike to connect to a pre-filled container. The<br>device is used for infants, children, adolescents and adults. | same |
| Product code | PIF, KNT | PIF, KNT | same |
| Classification<br>regulation | 21 CFR 876.5980 | 21 CFR 876.5980 | same |
| Device<br>configurations | ● Enteral Feeding Pump bag Set (1200ml、1000ml、<br>500ml、100ml)<br>■ 1200ml/1000ml/500ml/100ml Enteral Bag with<br>closure cap<br>■ PVC tubing<br>■ Roller clamp<br>■ Drip chamber<br>■ Silicone tube<br>■ Silicone tube connector<br>■ Warning label<br>■ EnFit connector<br>■ Protection cap for the EnFit connector | ● CONOD Feeding Pump Set (1200ml)<br>■ 1200ml Enteral Bag with closure cap<br>■ PVC tubing<br>■ Roller clamp<br>■ Drip chamber<br>■ Silicone tube<br>■ Silicone tube connector<br>■ Warning label<br>■ EnFit connector<br>■ EnFit transitional connector<br>■ Protection cap for the transition connector | Similar<br>1.Adding different bag volume<br>.Material and intended used are the<br>same.<br>2. Remove the EnFit<br>transitional connector. |
| | ● Enteral Feeding Pump EnPlus spike Set<br>■ Universal screw cap (with or without)<br>■ Protection cap for the spike<br>■ ENPlus spike | ● CONOD Safety spike plus Pump Set<br>■ Protection cap for the spike<br>■ ENPlus spike<br>■ PVC tubing<br>■ Roller clamp | Similar<br>1.Adding one universal screw cap<br>to the spike set to extend the |
| PVC tubing Roller clamp Drip chamber Silicone tube Silicone tube connector Warning label EnFit connector Protection cap for the EnFit connector | Drip chamber Silicone tube Silicone tube connector Warning label EnFit connector EnFit transitional connector Protection cap for the transition connector | connectivity for different pre-<br>filled formula containers.<br><br>2.Material of the screw cap is<br>exactly same as ENPlus spike.<br><br>3.Remove the EnFit transitional<br>connector. | |
| Enteral Feeding Gravity bag Set (1200ml、1000ml、<br>500ml、100ml) 1200ml/1000ml/500ml/100ml Enteral Bag with<br>closure cap PVC tubing Roller clamp Drip chamber Warning label EnFit connector Protection cap for the EnFit connector | CONOD Gravity Feeding Set (1200ml) 1200ml Enteral Bag with closure cap PVC tubing Roller clamp Drip chamber Warning label EnFit connector EnFit transitional connector Protection cap for the transition connector | Similar<br><br>1.Adding different bag volume<br><br>Material and intended used are the<br>same.<br><br>2.Remove the EnFit transitional<br>connector. | |
| Enteral Feeding Gravity EnPlus spike Set Universal screw cap (with or without) Protection cap for the spike ENPlus spike PVC tubing Roller clamp Drip chamber Warning label EnFit connector Protection cap for the EnFit connector | CONOD Gravity Feeding Set (1200ml) 1200ml Enteral Bag with closure cap PVC tubing Roller clamp Drip chamber Warning label EnFit connector EnFit transitional connector Protection cap for the transition connector | Different<br><br>1.Replacing the built-in enteral<br>bag with ENPlus spike and screw<br>cap so that it can connects to the<br>pre-filled nutrition containers<br>instead of having to prepare<br>formula by the user. the ENPlus<br>spike and screw cap introduced in<br>this set is the same as those used in<br>Enteral feeding pump EnPlus<br>Spike sets | |
| | | | 2.Remove the EnFit transitional<br>connector. |
| Design feature | Non DEHP material sets featuring ISO 80369-3:2016 connector | Non DEHP material sets featuring ISO 80369-3:2016 connector | same |
| Materials | PVC、ABS、Silicone、LDPE、paper | PVC、ABS、Silicone、LDPE、paper | same |
| Prescription use or not | Prescription use | Prescription use | same |
| Disposable or not | Disposable for single use | Disposable for single use | same |
| Intended use time | No longer than 24 hours | No longer than 24 hours | same |
| Sterile or not | Non-sterile | Non-sterile | same |
| Shelf-life | 5 years | 5 years | same |
| Available tube lengths | Enteral Feeding Pump bag Set: 60cm+157cm<br>Enteral Feeding Pump EnPlus spike Set: 60cm+157cm<br>Enteral Feeding Gravity bag Set: 170cm<br>Enteral Feeding Gravity EnPlus spike Set: 170cm | Pump set: 60cm+157cm<br>Spike set: 60cm+157cm<br>Gravity set: 170cm | Similar |
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image contains the logo for PrimeCare Medical. The logo is in blue and consists of the word "PrimeCare" in a script font, with the word "MEDICAL" in smaller, block letters underneath. A curved line underlines the word "MEDICAL".
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a stylized blue font. Below "PrimeCare" is the word "MEDICAL" in a smaller, sans-serif font, also in blue. A curved blue line underlines the word "MEDICAL".
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a cursive, blue font. Below "PrimeCare" is the word "MEDICAL" in a smaller, sans-serif font, also in blue. A curved line underlines the word "MEDICAL".
# Summary of Non-Clinical Testing
The non-clinical bench performance testing of the Enteral Feeding Sets is tested according to ISO 80369-3:2016(5-123), ISO20695:2020(9-138), ISO18250-3:2018(5-139), IEC62366-1:2020(5-129), ISO 10993 series of biological standards, Packaging test related standards and the guidance: Safety Considerations to Mitigate the Risks of Misconnections with Small-bore Connectors Intended for Enteral Applications.
## Table 1: Metrology (include EnFit connector, ENPlus spike & screw cap)
| No. | Items | Standards | Acceptance criteria |
|-----|------------------------------|------------------|-----------------------|
| 1 | Dimensions (EnFit connector) | ISO 80369-3:2016 | Meet the requirements |
| 2 | Dimensions (ENPlus spike) | ISO 18250-3:2018 | Meet the requirements |
| 3 | Dimensions (screw cap) | ISO 18250-3:2018 | Meet the requirements |
## Table 2: Appearance inspection
| No. | Items | Standards | Acceptance criteria |
|-----|---------------------------------------------------------------------------------------------|----------------|---------------------|
| 1 | Deformed parts, Cracking and<br>burr, Incomplete seal or<br>breakage of bag,<br>Kinked tube | Study in-house | None |
| 2 | Bag Printing | Study in-house | Present |
## Table 3: EnFit connector performance testing
| No. | Items | Standards | Acceptance criteria |
|-----|-------------------------------------------------|------------------|----------------------------------------------------|
| 1 | Fluid leakage<br>(Leakage by pressure<br>decay) | ISO 80369-3:2016 | shall not leak by more than<br>0.005 Pa.m³/s |
| 2 | Stress cracking | ISO 80369-3:2016 | No signs of leakage |
| 3 | Resistance to<br>separation from axial<br>load | ISO 80369-3:2016 | shall not separate from the<br>reference connector |
| 4 | Resistance to<br>separation from<br>unscrewing | ISO 80369-3:2016 | shall not separate from the<br>reference connector |
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image contains the logo for PrimeCare Medical. The word "PrimeCare" is written in a cursive, blue font. Below "PrimeCare" is the word "MEDICAL" in a smaller, sans-serif font, also in blue. A curved line underlines the word "MEDICAL".
| 5 | Resistance to<br>overriding | ISO 80369-3:2016 | shall not override the threads<br>or lugs of the reference<br>connector |
|---|--------------------------------|------------------|-----------------------------------------------------------------------------------------------------------------------|
| 6 | Disconnection by<br>unscrewing | ISO 80369-3:2016 | shall separate from the<br>reference connector with an<br>applied unscrewing<br>torque of no greater than 0.35<br>N·m |
# Table 4: ENPlus spike & screw cap performance testing
| No. | Items | Standards | Acceptance criteria |
|-----|---------------------------------------------|------------------|-------------------------------------------------------------------------|
| 1 | Positive pressure liquid<br>leakage | ISO 18250-3:2018 | shall show no signs of<br>leakage |
| 2 | Subatmospheric-pressure<br>air leakage | ISO 18250-3:2018 | shall not leak by more than<br>0.005 Pa.m3/s |
| 3 | Stress cracking | ISO 18250-3:2018 | No signs of leakage |
| 4 | Resistance to separation<br>from axial load | ISO 18250-3:2018 | shall not separate from the<br>reference connector. |
| 5 | Resistance to separation<br>from unscrewing | ISO 18250-3:2018 | shall not separate from the<br>reference connector |
| 6 | Resistance to overriding | ISO 18250-3:2018 | shall not override the<br>threads or lugs of the<br>reference connector |
# Table 5: Enteral Feeding Sets performance testing
| No. | Items | Standards | Acceptance criteria |
|-----|------------------|----------------------------------|--------------------------------------------------------------------------------------------------|
| 1 | Tensile strength | ISO 20695-2020 | shall withstand a tensile<br>force of 15 N before<br>breaking, becoming<br>detached, or cracking |
| 2 | Leakage | ISO 20695-2020 | shall not show signs of<br>leakage sufficient to<br>form a falling drop of<br>water |
| 3 | Flow rate | ISO 20695-2020<br>Study in-house | Meet the requirements |
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a cursive, blue font. Below "PrimeCare" is the word "MEDICAL" in a smaller, sans-serif, blue font. There is a curved line underneath the word "MEDICAL".
## Hangzhou Primecare Medical Co., Ltd. Table 6: Packaging performance testing
| No. | Items | Standards | Acceptance criteria |
|-----|--------------------|------------------------------------|------------------------------------------------------|
| 1 | Visual inspection | ASTMF1886/F1886M-16 | No defect of the integrity<br>of seals must be found |
| 2 | Seal strength test | ASTM F88/F88M-21<br>Study in-house | Should not be less than<br>12 N/15mm |
## Table 7: biological test
| No. | Items | Standards | Acceptance criteria |
|-----|------------------------------|-------------------|---------------------|
| 1 | Cytotoxicity | ISO 10993-5:2009 | Non-Cytotoxic |
| 2 | Sensitization | ISO 10993-10:2021 | Non-Sensitizer |
| 3 | Intracutaneous<br>Irritation | ISO 10993-23:2021 | Non-Irritant |
## Table 8: Usability Study
| No. | Items | Standards | Acceptance criteria |
|-----|-----------------|------------------|-----------------------|
| 1 | Usability Study | IEC 62366-1:2020 | Meet the requirements |
## Summary of Clinical Testing
Clinical testing was not required for device evaluation.
## Conclusion
The conclusions drawn from the nonclinical testing for the subject device, Enteral Feeding Sets, demonstrate that the device is as safe, as effective, and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.