CUTIVA(TM) Topical Skin Adhesive (RM-1700); CUTIVA(TM) PLUS Skin Closure System (RM-1739)
K234114 · Okapi Medical, LLC Dba Resivant Medical · MPN · Aug 20, 2024 · General, Plastic Surgery
Device Facts
Record ID
K234114
Device Name
CUTIVA(TM) Topical Skin Adhesive (RM-1700); CUTIVA(TM) PLUS Skin Closure System (RM-1739)
Applicant
Okapi Medical, LLC Dba Resivant Medical
Product Code
MPN · General, Plastic Surgery
Decision Date
Aug 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
CUTIVA™ Topical Skin Adhesive (RM1700) is intended for topical application only, to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed trauma-induced lacerations. CUTIVA™ Topical Skin Adhesive (RM1700) should be used in conjunction with, but not in place of, deep dermal stitches. The CUTIVA™ PLUS Skin Closure System (RM1739) is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. CUTIVA™ PLUS Skin Closure System (RM1739) should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the application of the liquid adhesive.
Device Story
CUTIVA™ Topical Skin Adhesive (RM1700) and CUTIVA™ PLUS Skin Closure System (RM1739) are 2-octyl cyanoacrylate-based liquid adhesives. RM1700 is a standalone adhesive; RM1739 includes a self-adhering mesh for skin edge alignment. Devices are supplied in single-use pen-style applicators. Upon application, liquid adhesive polymerizes to form a thin, strong film providing a microbial barrier. Used by clinicians in surgical or clinical settings to close wounds. The adhesive acts as a physical barrier to microbial penetration while maintaining skin edge approximation to facilitate natural healing. Benefits include secure wound closure and reduced risk of microbial contamination when the film remains intact.
Clinical Evidence
No human clinical studies performed. Evidence includes bench testing (ASTM F2458-05, F2258-05, F2255-05, D3330, D882, D3654) and a 14-day porcine full-thickness linear wound healing model. Porcine study confirmed safety, performance, and local tissue response comparable to predicates. Microbial barrier properties demonstrated in vitro for 72 hours.
Technological Characteristics
2-octyl cyanoacrylate adhesive; benzalkonium chloride initiator. Pen-style single-use applicator. Sterilization: dry heat (adhesive) and ethylene oxide (final package) to SAL 10^-6. Biocompatibility per ISO 10993-1:2018 (prolonged contact). Bench standards: ASTM F2458-05, F2258-05, F2255-05, D3330, D882, D3654.
Indications for Use
Indicated for topical closure of easily approximated skin edges of surgical incisions (including minimally invasive surgery punctures) and simple, cleansed trauma-induced lacerations. Must be used with deep dermal stitches. No specific age or gender restrictions noted.
Regulatory Classification
Identification
A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
*Tissue adhesive for non-topical use* —(1)*Identification.* A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 20, 2024
Okapi Medical LLC dba Resivant Medical Tom Stephens Sr. Director, Regulatory Affairs and Quality Assurance 526 S. Main St. Suite 124M Akron, Ohio 44311
Re: K234114
Trade/Device Name: CUTIVA(TM) Topical Skin Adhesive (RM-1700); CUTIVA(TM) PLUS Skin Closure System (RM-1739) Regulation Number: 21 CFR 878.4010 Regulation Name: Tissue Adhesive Regulatory Class: Class II Product Code: MPN, OMD Dated: July 18, 2024 Received: July 18, 2024
Dear Tom Stephens:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Wilmarie Flores -S
Julie Morabito, Ph.D. Assistant Director
{2}------------------------------------------------
DHT4B: Division of Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K234114
Device Name
CUTIVA™ Topical Skin Adhesive (RM1700); CUTIVA™ PLUS Skin Closure System (RM1739)
### Indications for Use (Describe)
CUTIVA™ Topical Skin Adhesive (RM1700) is intended for topical application only, to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed trauma-induced lacerations. CUTIVA™ Topical Skin Adhesive (RM1700) should be used in conjunction with, but not in place of, deep dermal stitches.
The CUTIVA™ PLUS Skin Closure System (RM1739) is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. CUTIVA™ PLUS Skin Closure System (RM1739) should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the application of the liquid adhesive.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------------------------------------|
| | <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This 510(k) summary is prepared in accordance with the requirements of 21 CFR §807.92.
| Submitter<br>1. | |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by: | Okapi Medical, LLC dba Resivant Medical<br>526 S. Main St. Suite 124M<br>Akron, OH 44311<br>Phone: +1-234-738-4959<br>Web: resivant.com |
| Contact Person: | Tom Stephens<br>Sr. Director, Regulatory Affairs & Quality Assurance<br>Email: tstephens@resivant.com<br>Phone: +1-919-616-1723 |
| Date of Summary: | August 20, 2024 |
#### 2. Device Name and Classification
| | Trade/Proprietary Name: CUTIVA™ Topical Skin Adhesive (RM1700) |
|--------------------|----------------------------------------------------------------|
| Regulation Number: | 21 CFR 878.4010(a) |
| Regulation Name: | Tissue adhesive |
| Product Code: | MPN |
| Classification: | Class II |
| Panel: | General and Plastic Surgery |
| Trade/Proprietary Name: | CUTIVA ™ PLUS Skin Closure System (RM1739) |
|-------------------------|--------------------------------------------------------------------------------------------------|
| Regulation Number: | 21 CFR 878.4011 |
| Regulation Name: | Tissue adhesive with adjunct wound closure device intended for the topical approximation of skin |
| Product Code: | OMD |
| Classification: | Class II |
| Panel: | General and Plastic Surgery |
#### 3. Predicate Device
| Device Name: | DERMABOND™ NX Topical Skin Adhesive (currently marketed as<br>DERMABOND™ ADVANCED™ Topical Skin Adhesive) |
|---------------------|-----------------------------------------------------------------------------------------------------------|
| 510(k) File Number: | K100423 |
| Device Name: | DERMABOND™ PRINEO™ Skin Closure System |
| 510(k) File Number: | K133864 |
{5}------------------------------------------------
#### 4. Device Description
CUTIVA™ Topical Skin Adhesive (RM1700) and the CUTIVA™ PLUS Skin Closure System (RM1739) are skin closure devices that are comprised of a 2-octyl cyanoacrylate liquid adhesive formulation. The liquid adhesive is supplied sterile within a single use dispensing applicator, which is used to deliver the adhesive to the skin. The CUTIVA™ PLUS Skin Closure System (RM1739) also incorporates a self-adhering mesh component that is applied to the wound prior to the application of the liquid adhesive to align the skin edges. The liquid adhesive is then applied to the mesh with the adhesive applicator to complete the device application.
Once applied, the liquid adhesive polymerizes to form a thin film with strong bonding and tensile properties. CUTIVA™ Topical Skin Adhesive (RM1700) and the CUTIVA™ PLUS Skin Closure System (RM1739) provide a physical barrier to microbial penetration as long as the adhesive film remains intact. In vitro studies have been performed to demonstrate the microbial barrier properties of CUTIVA™ Topical Skin Adhesive (RM1700) and the CUTIVA™ PLUS Skin Closure System (RM1739) for 72 hours after device application. No clinical studies have been performed and no clinical benefit associated with the in vitro microbial barrier performance of the device has been demonstrated.
#### 5. Indications for Use
CUTIVA™ Topical Skin Adhesive (RM1700) is intended for topical application only, to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed trauma-induced lacerations. CUTIVA™ Topical Skin Adhesive (RM1700) should be used in conjunction with, but not in place of, deep dermal stitches.
The CUTIVA™ PLUS Skin Closure System (RM1739) is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. CUTIVA™ PLUS Skin Closure System (RM1739) should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound during the application of the liquid adhesive.
#### 6. Comparison of Technological Characteristics with Predicate
The technological characteristics of CUTIVA™ Topical Skin Adhesive (RM1700) and the CUTIVA™ PLUS Skin Closure System (RM1739) are closely comparable to their respective predicate devices, DERMABOND™ NX Topical Skin Adhesive and the DERMABOND™ PRINEO™ Skin Closure System. Minor differences in technical characteristics do not raise new questions of safety and effectiveness. These technological characteristics are compared in the table on the following page.
{6}------------------------------------------------
| Technical Characteristic | CUTIVA™<br>(RM1700) | DERMABOND™<br>NX | CUTIVA™ PLUS<br>(RM1739) | DERMABOND™<br>PRINEOTM |
|---------------------------------------------------------------------|------------------------|------------------------|--------------------------|------------------------|
| 2-Octyl cyanoacrylate-based<br>adhesive | Yes | Yes | Yes | Yes |
| Adhesive in primary packaging<br>sterilized by dry heat to SAL 10-6 | Yes | Yes | Yes | Yes |
| Final packaged device sterilized<br>by ethylene oxide to SAL 10-6 | Yes | Yes | Yes | Yes |
| Pen-style single use applicator | Yes | Yes | Yes | Yes |
| Sterile peel-open final package | Yes | Yes | Yes | Yes |
| Flexible tip on adhesive applicator | Yes | Yes | Yes | Yes |
| Polymerization initiating agent is<br>benzalkonium chloride | Yes | Yes | Yes | Yes |
| Polymerizes within minutes | Yes | Yes | Yes | Yes |
| Forms a polymer film with strong<br>bonding and tensile properties | Yes | Yes | Yes | Yes |
| Maintains skin edge<br>approximation for natural healing | Yes | Yes | Yes | Yes |
| Provides a microbial barrier when<br>the adhesive film is intact | Yes | Yes | Yes | Yes |
| Applicator actuation mechanism | Depress<br>lever | Squeeze<br>applicator | Depress<br>lever | Twist<br>collar |
| Color additive | None | D&C Violet<br>#2 | None | D&C Violet<br>#2 |
| Location of polymerization | Adhesive<br>applicator | Adhesive<br>applicator | Adhesive<br>applicator | Mesh<br>Component |
Comparison Table of Technical Characteristics
#### 7. Performance Data
The testing plan for the CUTIVA™ Topical Skin Adhesive (RM1700) device was developed in consideration of the FDA guidance document "Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Tissue Adhesive for the Topical Approximation of Skin", May 30, 2008. The testing plan for the CUTIVA™ PLUS Skin Closure System (RM1739) device was developed in consideration of the FDA guidance document "Tissue Adhesive with Adjunct Wound Closure Device Intended for the Topical Approximation of Skin - Class II Special Controls Guidance Document for Industry and FDA Staff", issued November 10, 2010. Both devices were also developed according to the requirements of design control and risk analysis. Testing to establish substantial equivalence of the CUTIVA™ Topical Skin Adhesive (RM1700) and the CUTIVA™ PLUS Skin Closure System (RM1739) to their respective predicate devices has been completed according to Resivant Medical procedures in compliance with the requirements of 21 CFR 820.30 – Design Control. Completed testing is listed on the following page.
{7}------------------------------------------------
### Performance Testing:
Bench studies have been performed to evaluate the device properties:
- . Viscosity
- . Setting time
- . Purity
- Moisture
- Hydrolytic degradation
- 트 Microbial barrier
Performance studies have been performed to demonstrate substantial equivalence to the predicate devices. In all cases, the CUTIVA™ Topical Skin Adhesive (RM1700) and the CUTIVA™ PLUS Skin Closure System (RM1739) devices demonstrated substantial equivalence to their respective predicates.
- . Wound closure strength (ASTM F2458-05)
- . Tensile strength (ASTM F2258-05)
- Lap shear strength (ASTM F2255-05)
- 트 Heat of polymerization
- Applicator Functionality
In addition, the mesh component of the CUTIVA™ PLUS Skin Closure System (RM1739) has been tested against the performance of the mesh component of the DERMABOND™ PRINEO™ Skin Closure System predicate device. In all cases, the CUTIVA™ PLUS Skin Closure System (RM1739) mesh demonstrated substantial equivalence to the predicate mesh:
- Mesh Peel Adhesion Strength (ASTM D3330) .
- I Mesh Tensile Strength (ASTM D882)
- 트 Mesh Shear Adhesion Strength (ASTM D3654)
## Animal Studies:
A porcine study was conducted to compare the safety and effectiveness of CUTIVA™ Topical Skin Adhesive (RM1700) and the CUTIVA™ PLUS Skin Closure System (RM1739) to their respective predicate devices in a surgically induced, full thickness, linear wound healing model over a period of 14 days. This study evaluated overall animal health, test article performance, and local tissue response to the test article. All success criteria were met for Endpoint 1 (overall animal health), Endpoint 2 (test article performance), and Endpoint 3 (local tissue response to the test article). Overall, CUTIVA™ Topical Skin Adhesive (RM1700) and the CUTIVA™ PLUS Skin Closure System (RM1739) performed comparably to DERMABOND™ NX Topical Skin Adhesive and the DERMABOND™ PRINEO™ Skin Closure System, respectively.
## Biocompatibility:
The biocompatibility of the CUTIVA™ Topical Skin Adhesive (RM1700) and the CUTIVA™ PLUS Skin Closure System (RM1739) devices has been evaluated according to the studies recommended in ISO 10993-1:2018, "Biological evaluation of medical devices - part 1: Evaluation and testing within a risk management process" for a device intended for prolonged contact (24 hours – 30 days) with breached skin. The results of these studies, listed below,
{8}------------------------------------------------
demonstrate that the CUTIVA™ Topical Skin Adhesive (RM1700) and the CUTIVA™ PLUS Skin Closure System (RM1739) devices are safe for their intended use.
- Cytotoxicity (ISO 10993-5:2009) .
- . Sensitization/Kligman Maximization (ISO 10993-10:2021)
- Irritation/Intracutaneous Reactivity (ISO 10993-10:2021)
- Material Mediated Pyrogenicity (ISO 10993-11:2017)
- . Acute Systemic Injection (ISO 10993-11:2017)
- 트 Subcutaneous Implantation (ISO 10993-6:2016)
- Subacute Systemic Toxicity (ISO 10993-11:2017)
### Sterilization and Shelf-Life:
The CUTIVA™ Topical Skin Adhesive (RM1700) and the CUTIVA™ PLUS Skin Closure System (RM1739) devices are sterilized in two steps during device production. The liquid adhesive formulation is sterilized in its primary packaging by dry heat in compliance with the requirements of ISO 20857:2010. After device assembly and packaging, the finished device is sterilized by exposure to ethylene oxide gas in compliance with the requirements of ISO 11135:2014. Both sterilization processes are validated to provide a Sterility Assurance Level of 106. The shelf life of the device has been confirmed through real-time aging studies.
#### 8. Conclusions
The CUTIVA™ Topical Skin Adhesive (RM1700) device and its predicate device, DERMABOND™ NX Topical Skin Adhesive (K100423), have identical indications for use, function according to the same fundamental scientific technology, and are closely comparable with regard to device design and performance. Minor differences between CUTIVA™ Topical Skin Adhesive (RM1700) device and its predicate with regard to the mechanism for device actuation and the incorporation of a color additive in the predicate formulation do not raise different questions of safety or effectiveness. In vitro and in vivo testing results provided in the 510(k) submission demonstrate that the CUTIVA™ Topical Skin Adhesive (RM1700) device meets its specifications and is substantially equivalent to the DERMABOND™ NX Topical Skin Adhesive predicate.
The CUTIVA™ PLUS Skin Closure System (RM1739) device and its predicate device, DERMABOND™ PRINEO™ Skin Closure System (K133864), have identical indications for use, function according to the same fundamental scientific technology, and are closely comparable with regard to device design and performance. Minor differences between CUTIVA™ Topical Skin Adhesive (RM1700) device and its predicate with regard to the mechanism for device actuation, the incorporation of a color additive in the predicate formulation, and the location of the polymerization initiator do not raise different questions of safety or effectiveness. In vitro and in vivo testing results provided in the 510(k) submission demonstrate that the CUTIVA™ PLUS Skin Closure System (RM1739) device meets its specifications and is substantially equivalent to the DERMABOND™ PRINEO™ Skin Closure System predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.