STERRAD® 100NX Sterilizer with ALLClear™ Technology and ULTRA GI™ Cycle (10104); APTIMAX™ Instrument Tray for ULTRA GI™ Cycle (011077)
Applicant
Advanced Sterilization Products
Product Code
MLR · General Hospital
Decision Date
Jul 5, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The STERRAD 100NX Sterilizer with ALLClear Technology is designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture. The APTIMAX Tray for ULTRA GI Cycle is designed to encase surgical instruments for sterilization in STERRAD 100NX Sterilization System with ALLCLEAR Technology.
Device Story
STERRAD 100NX Sterilizer is a self-contained system using hydrogen peroxide gas plasma to sterilize heat/moisture-sensitive medical instruments. Process involves injecting aqueous hydrogen peroxide, vaporizing it under sub-ambient pressure, and converting to gas plasma via electrical energy. System includes sterilizer chamber, reusable instrument trays (APTIMAX), and sterilant cassettes. Operated by healthcare staff in clinical settings. New ULTRA GI cycle specifically processes flexible multi-channel duodenoscopes. ALLClear technology integrates load check, system check, and load conditioning. Output is a sterilized device ready for clinical use; benefits include safe, residue-free sterilization of complex, diffusion-restricted instruments like hinged forceps and multi-channel endoscopes.
Clinical Evidence
Bench testing only. No clinical data generated. Performance testing included sterilization verification (SAL 10^-6), growth inhibition, in-use testing, biocompatibility, simulated use, and device functionality/material compatibility post-processing. All tests passed.
Indicated for sterilization of metal and nonmetal medical devices, including flexible endoscopes, rigid/semi-rigid endoscopes, and duodenoscopes. Compatible with hydrogen peroxide-compatible flexible multi-channel duodenoscopes (up to 4 channels, ID ≥1mm x ≤1500mm or ≥2mm ID x ≤1630mm). Not for use with non-compatible materials.
Regulatory Classification
Identification
An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.
Predicate Devices
STERRAD 100NX Sterilizer with ALLClear Technology (K212174)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 5, 2024
Advanced Sterilization Products Ayse Erkan Sr. Regulatory Affairs Program Lead 33 Technology Dr. Irvine, California 92618
Re: K234082
Trade/Device Name: STERRAD® 100NX Sterilizer with ALLClear™ Technology (10104); APTIMAX™ Instrument Tray for ULTRA GITM Cycle (011077) Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene oxide gas sterilizer Regulatory Class: Class II Product Code: MLR, FRG Dated: December 22, 2023 Received: June 7, 2024
Dear Ayse Erkan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Christopher K. Dugard -S
Christopher K. Dugard, M.S. Assistant Director DHT4B: Division of Infection Control
{2}------------------------------------------------
and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K234082
#### Device Name
STERRAD™ 100NX Sterilizer with ALLClear™ Technology and ULTRA GIT™ Cycle (10104) APTIMAX™ Instrument Tray for ULTRA GI Cycle (011077)
#### Indications for Use (Describe)
The STERRAD 100NX Sterilizer with ALLClear Technology is designed for sterilization of both metal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterlization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to safely sterilize medical instruments and materials without leaving toxic residue.
The STERRAD 100NX Sterilize can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.
Medical devices with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer STANDARD cycle:
· Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a length of 500 mm or shorter. A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle.
Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer FLEX Scope cycle:
• Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and length of 1065 mm or shorter. A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load.
Note: With the exception of the 1 x 1065 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 lbs. The 1 x 1065 mm flexible endoscopes were validated without any additional load.
The STERRAD 100NX EXPRESS Cycle is an additional optional cycle designed for surface sterilization of both metal and nonmetal medical devices at low temperatures.
• It can sterilize instrument surfaces and instruments having diffusion-restricted spaces, such as the hinged portion of forceps and scissors
· It can sterilize rigid and semi-rigid endoscopes without lumens.
Note: The validation studies for EXPRESS Cycle were performed using a validation load consisting of a single instrument tray weighing 10.7 lbs placed on the bottom shelf.
The STERRAD 100NX DUO Cycle is an additional optional cycle designed for sterilization of medical devices including most flexible endoscopes, with the following materials and dimensions:
• Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and a length of 875 mm or shorter
· Accessory devices that are normally connected to a flexible endoscope during use
• Flexible endoscopes without lumens
Note: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 lbs.
The STERRAD 100NX Sterilizer ULTRA GI Cycle is designed for sterilization of the following:
• Hydrogen peroxide compatible flexible multi-channel duodenoscopes, with no more than 4 channels, with lumen
dimensions having an inside diameter (ID) of ≥1mm x ≤1500mm in length, or ≥2mm ID x ≤1630mm in length.
· One flexible duodenoscope per tray, and no more than two flexible duodenoscope per cycle
Notel : The STERRAD 100NX Sterilizer ULTRA GI Cycle was validated using a load weight of 15.4 lbs (2 x 7.7 lbs), one endoscope per shelf. Note 2: Only duodenoscopes that have been cleared as compatible with vaporized hydrogen peroxide are acceptable. Check STERRAD User Guide "STERRAD Sterilizer Cycle Selection table" for ULTRA GI Cycle compatible duodenoscopes.
The APTIMAX™ Tray for ULTRA GI™ Cycle is designed to encase surgical instruments for sterilization in
{4}------------------------------------------------
## STERRAD™ 100NX Sterilization System with ALLCLEAR™ Technology:
## • ULTRA GI™ Cycle
Type of Use (Select one or both, as applicable)
7 Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(k) Summary for K234082 Advanced Sterilization Products, Inc. STERRAD™ 100NX Sterilizer with ALLClear™ Technology and ULTRA GI™ Cycle APTIMAX™ Instrument Tray for ULTRA GI Cycle
This summary of 510(k) Safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### General Information
| Submitter Name: | Advanced Sterilization Products, Inc. |
|-----------------|---------------------------------------------------------------------------------------------------------------|
| Address: | 33 Technology Drive<br>Irvine, CA 92618 |
| Contact Person: | Ayse Erkan<br>Senior Regulatory Affairs Program Lead<br>Telephone: (714)580-2880<br>Email: ayse.erkan@asp.com |
Date Prepared: July 1, 2024
#### Devices Names
| Proprietary Name: | STERRAD™ 100NX<br>Sterilizer with ALLClear™<br>Technology and ULTRA<br>GI™ Cycle (10104) | APTIMAX™ Instrument<br>Tray for ULTRA GI Cycle<br>(011077) |
|----------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Device Common Name: | Hydrogen Peroxide Gas<br>Plasma Sterilization System | Sterilizable Instrument Tray<br>and Accessories |
| Classification Name: | Ethylene Oxide Gas Sterilizer | Wrap, sterilization |
| Product Code: | MLR | FRG |
| Device Class: | Class II | Class II |
| CFR Section: | 21 CFR 880.6860 | 21 CFR 880.6850 |
#### Predicate Devices
STERRAD 100NX Sterilizer with ALLClear Technology K212174 cleared October 9, 2021 APTIMAX Instrument Tray K013003 cleared August 28, 2002.
#### Reference Device
EOGas 4 Ethylene Oxide Gas Sterilizer K192978 cleared November 12, 2020.
#### Device Description
The STERRAD 100NX Sterilization System with ALLClear Technology is a self-contained stand-alone system of hardware and software designed to sterilize medical instruments and devices using a patented hydrogen peroxide gas plasma process. Hydrogen peroxide vapor is generated by injecting aqueous hydrogen peroxide into the vaporizer where the solution is heated and vaporized, introducing the vapor into the chamber under sub-ambient pressure and transforming the vapor into a gas plasma using electrical energy. The sterilization process is a multiphase process that utilizes a combination of
{6}------------------------------------------------
exposure to hydrogen peroxide vapor and plasma to achieve sterilization. The hardware for the STERRAD 100NX Sterilizer consists of a sterilizer chamber and a variety of instruments and components which are housed in a covered frame. The sterilizer also uses accessories such as reusable instrument trays, biological indicator products and sterilant cassettes. The STERRAD 100NX Sterilizer has four cleared sterilization cycles: STANDARD, FLEX, EXPRESS, and DUO Cycles.
The STERRAD 100NX Sterilizer described within this submission introduces the new ULTRA GI Cycle along with the new APTIMAX Instrument Tray for ULTRA GI Cycle.
The STERRAD 100NX Sterilizer ULTRA GI Cycle is an additional cycle designed for sterilization of the following:
- Hydrogen peroxide compatible flexible multi-channel duodenoscopes, with no more ● than 4 channels, with lumen dimensions having an inside diameter (ID) of ≥1mm x ≤1500mm in length, or ≥2mm ID x ≤1630mm in length.
- . One flexible duodenoscope per tray, and no more than two flexible duodenoscope per cycle
Note 1: The STERRAD 100NX Sterilizer ULTRA GI Cycle was validated using a load weight of 15.4 lbs (2 x 7.7 lbs), one endoscope per shelf.
Note 2: Only duodenoscopes that have been cleared as compatible with vaporized hydrogen peroxide are acceptable. Check STERRAD Sterilizer Cycle Selection table for ULTRA GI Cycle compatible duodenoscopes.
The APTIMAX Instrument Tray for ULTRA GI Cycle is designed to encase surgical instruments for sterilization in STERRAD 100NX Sterilization System with ALLClear Technology:
- ULTRA GI Cycle .
This cycle is designed to run on STERRAD 100NX System with ALLClear Technology with the DUO Module enabled. ALLClear Technology features of Load Check, System Check, and Load Conditioning are integrated into the ULTRA GI cycle.
## Intended Use/Indications for Use
The intended use of the subject STERRAD 100NX Sterilizer with ALClear Technology has not changed with the proposed expanded indications for use in the ULTRA GI Cycle. The APTIMAX Tray for ULTRA GI Cycle is designed to encase surgical instruments for sterilization in STERRAD 100NX Sterilization System with ALLCLEAR Technology. Refer to Table 1, Comparison between the subject device and predicate devices and Table 2, APTIMAX ULTRA GI Instrument Tray Predicate Device Comparison for of Intended Use & Indications for Use Comparison.
## Technological Characteristics
The technological characteristics associated with the sterilization process of the proposed STERRAD 100NX Sterilizer with ALLClear Technology with expanded indications for the ULTRA GI Cycle in combination with the APTIMAX Tray for ULTRA GI Cycle is designed to run on STERRAD 100NX System with ALLClear Technology with the DUO Module enabled. ALLClear Technology features of Load Check, System Check, and Load Conditioning are integrated into the ULTRA GI cycle.
The subject devices continue to have the same technological characteristics, sterilization, performance, and physical traits as the predicate devices, STERRAD 100NX Sterilizer with ALL Clear Technology and the APTIMAX Instrument Tray.
{7}------------------------------------------------
| Feature | STERRAD 100NX Sterilizer with<br>ALLClear Technology with ULTRA<br>GI Cycle Subject Device | STERRAD 100NX Sterilizer with<br>ALLClear Technology<br>K212174 Predicate Device | Comparison |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Intended Use | Designed for sterilization of both metal<br>and nonmetal medical devices at low<br>temperatures. Because the cycle operates<br>within a dry environment and at low<br>temperatures, it is especially suitable for<br>instruments sensitive to heat and<br>moisture. | Designed for sterilization of both metal<br>and nonmetal medical devices at low<br>temperatures. Because the cycle operates<br>within a dry environment and at low<br>temperatures, it is especially suitable for<br>instruments sensitive to heat and<br>moisture. | Identical |
| Indications<br>for Use | The STERRAD 100NX Sterilizer with<br>ALLClear Technology is designed for<br>sterilization of both metal and nonmetal<br>medical devices at low temperatures. The<br>STERRAD sterilization process is a<br>multiphase sterilization process that<br>utilizes a combination of exposure to<br>hydrogen peroxide vapor and plasma to<br>safely sterilize medical instruments and<br>materials without leaving toxic residue.<br>The STERRAD 100NX Sterilizer can<br>sterilize instruments which have diffusion-<br>restricted spaces, such as the hinged<br>portion of forceps and scissors.<br>Medical devices with the following<br>materials and dimensions can be<br>processed in the STERRAD 100NX<br>Sterilizer STANDARD cycle:<br>Single channel stainless steel<br>lumens with an inside diameter of<br>0.7 mm or larger and a length of 500<br>mm or shorter. A maximum of ten<br>single channel stainless steel lumens,<br>five per tray per sterilization cycle.Medical devices, including most flexible<br>endoscopes, with the following materials<br>and dimensions can be processed in the<br>STERRAD 100NX Sterilizer FLEX Scope<br>cycle:<br>Single channel polyethylene and<br>Teflon (polytetrafluoroethylene) | The STERRAD 100NX Sterilizer with<br>ALLClear Technology is designed for<br>sterilization of both metal and nonmetal<br>medical devices at low temperatures. The<br>STERRAD sterilization process is a<br>multiphase sterilization process that<br>utilizes a combination of exposure to<br>hydrogen peroxide vapor and plasma to<br>safely sterilize medical instruments and<br>materials without leaving toxic residue.<br>The STERRAD 100NX Sterilizer can<br>sterilize instruments which have diffusion-<br>restricted spaces, such as the hinged<br>portion of forceps and scissors.<br>Medical devices with the following<br>materials and dimensions can be<br>processed in the STERRAD 100NX<br>Sterilizer STANDARD cycle:<br>Single channel stainless steel<br>lumens with an inside diameter of<br>0.7 mm or larger and a length of 500<br>mm or shorter. A maximum of ten<br>single channel stainless steel lumens,<br>five per tray per sterilization cycle.Medical devices, including most flexible<br>endoscopes, with the following materials<br>and dimensions can be processed in the<br>STERRAD 100NX Sterilizer FLEX Scope<br>cycle:<br>Single channel polyethylene and<br>Teflon (polytetrafluoroethylene) | Similar,<br>includes<br>additional<br>indications for<br>Use for ULTRA<br>GI cycle. |
| | flexible endoscopes with an inside<br>diameter of 1 mm or larger and<br>length of 1065 mm or shorter. A<br>maximum of two flexible<br>endoscopes, one per tray per<br>sterilization cycle. No additional load.<br>Note: With the exception of the 1 x 1065<br>mm flexible endoscopes, the validation<br>studies were performed using a validation<br>load consisting of two instrument trays<br>each weighing 10.7 lbs. The 1 x 1065 mm | flexible endoscopes with an inside<br>diameter of 1 mm or larger and<br>length of 1065 mm or shorter. A<br>maximum of two flexible<br>endoscopes, one per tray per<br>sterilization cycle. No additional load.<br>Note: With the exception of the 1 x 1065<br>mm flexible endoscopes, the validation<br>studies were performed using a validation<br>load consisting of two instrument trays<br>each weighing 10.7 lbs. The 1 x 1065 mm | |
| Feature | STERRAD 100NX Sterilizer with<br>ALLClear Technology with ULTRA<br>GI Cycle Subject Device | STERRAD 100NX Sterilizer with<br>ALLClear Technology<br>K212174 Predicate Device | Comparison |
| | flexible endoscopes were validated<br>without any additional load. | flexible endoscopes were validated<br>without any additional load. | |
| | The STERRAD 100NX EXPRESS Cycle is<br>an additional optional cycle designed<br>for surface sterilization of both metal<br>and nonmetal medical devices at low<br>temperatures. | The STERRAD 100NX EXPRESS Cycle is<br>an additional optional cycle designed<br>for surface sterilization of both metal<br>and nonmetal medical devices at low<br>temperatures. | |
| | It can sterilize instrument surfaces<br>and instruments having diffusion-<br>restricted spaces, such as the hinged<br>portion of forceps and scissors It can sterilize rigid and semi-rigid<br>endoscopes without lumens. Note: The validation studies for<br>EXPRESS Cycle were performed using<br>a validation load consisting of a single<br>instrument tray weighing 10.7 lbs<br>placed on the bottom shelf. | It can sterilize instrument surfaces<br>and instruments having diffusion-<br>restricted spaces, such as the hinged<br>portion of forceps and scissors It can sterilize rigid and semi-rigid<br>endoscopes without lumens. Note: The validation studies for<br>EXPRESS Cycle were performed using<br>a validation load consisting of a single<br>instrument tray weighing 10.7 lbs<br>placed on the bottom shelf. | |
| | The STERRAD 100NX DUO Cycle is an<br>additional optional cycle designed for<br>sterilization of medical devices<br>including most flexible endoscopes,<br>with the following materials and<br>dimensions: | The STERRAD 100NX DUO Cycle is an<br>additional optional cycle designed for<br>sterilization of medical devices<br>including most flexible endoscopes,<br>with the following materials and<br>dimensions: | |
| | Single channel polyethylene and<br>Teflon (polytetrafluoroethylene)<br>flexible endoscopes with an inside<br>diameter of 1 mm or larger and a<br>length of 875 mm or shorter Accessory devices that are<br>normally connected to a flexible<br>endoscope during use Flexible endoscopes without<br>lumens Note: The validation studies for DUO<br>Cycle were performed using a<br>validation load consisting of two<br>flexible endoscopes with their<br>accessory devices weighing a total of<br>13.2 lbs. | Single channel polyethylene and<br>Teflon (polytetrafluoroethylene)<br>flexible endoscopes with an inside<br>diameter of 1 mm or larger and a<br>length of 875 mm or shorter Accessory devices that are<br>normally connected to a flexible<br>endoscope during use Flexible endoscopes without<br>lumens Note: The validation studies for DUO<br>Cycle were performed using a<br>validation load consisting of two<br>flexible endoscopes with their<br>accessory devices weighing a total of<br>13.2 lbs. | |
| | The STERRAD 100NX Sterilizer ULTRA<br>GI Cycle is designed for sterilization of<br>the following: | | |
| | Hydrogen peroxide compatible<br>flexible multi-channel<br>duodenoscopes, with no more<br>than 4 channels, with lumen<br>dimensions having an inside<br>diameter (ID) of ≥1mm x<br>≤1500mm in length, or<br>≥2mm ID x ≤1630mm in<br>length. | | |
| Feature | STERRAD 100NX Sterilizer with<br>ALLClear Technology with ULTRA<br>GI Cycle Subject Device | STERRAD 100NX Sterilizer with<br>ALLClear Technology<br>K212174 Predicate Device | Comparison |
| | One flexible duodenoscope per tray, and no more than two flexible duodenoscope per cycle | | |
| | Note 1: The STERRAD 100NX Sterilizer<br>ULTRA GI Cycle was validated using a<br>load weight of 15.4 lbs (2 x 7.7 lbs),<br>one endoscope per shelf.<br>Note<br>2: Only duodenoscopes that have been<br>cleared as compatible with vaporized<br>hydrogen peroxide are<br>acceptable. Check STERRAD Sterilizer<br>Cycle Selection table for ULTRA GI<br>Cycle compatible duodenoscopes | | |
| Sterilization<br>Cycles | ULTRA GI<br>STANDARD<br>FLEX<br>EXPRESS<br>DUO | STANDARD<br>FLEX<br>EXPRESS<br>DUO | Identical |
| Dimensions | One Door:<br>30.5" W x 70.9" H x 41.5" D<br>(775 mm x 1800 mm x 1055<br>mm)<br>Two Door:<br>30.5" W x 70.9" H x 43.1" D<br>(775 mm x 1800 mm x 1095<br>mm) | One Door:<br>30.5" W x 70.9" H x 41.5" D<br>(775 mm x 1800 mm x 1055<br>mm)<br>Two Door:<br>30.5" W x 70.9" H x 43.1" D<br>(775 mm x 1800 mm x 1095<br>mm) | Identical |
| H2O2<br>Concentration<br>by Weight | 59-94%<br>Depending on Cycle | 59-94%<br>Depending on Cycle | Identical |
| Number Of<br>Half Cycles | 2 | 2 | Identical |
| Chamber<br>Volume | 152L | 152L | Identical |
| Key Critical<br>Process<br>Parameters | Pressure, temperature & H2O2<br>concentration, H2O2 Dose &<br>Exposure Time | Pressure, temperature & H2O2<br>concentration, H2O2 Dose &<br>Exposure Time | Identical |
| Use of Pre-<br>Conditioning<br>Step | Yes | Yes | Identical |
| Use of<br>Secondary<br>Step to<br>Remove<br>Residual H2O2<br>(technology) | Yes (plasma) | Yes(plasma) | Identical |
# Table 1. Comparison between the subject device and predicate devices
{8}------------------------------------------------
{9}------------------------------------------------
### Table 2. APTIMAX ULTRA GI Instrument Tray Predicate Device Comparison
| Feature | APTIMAX ULTRA GI<br>Instrument Tray<br>Subject Device | APTIMAX Instrument Tray<br>K013003 Predicate Device | Comparison |
|------------------------|--------------------------------------------------------|-----------------------------------------------------|------------|
| Indications for<br>Use | Designed to encase<br>instruments for<br>sterilization | Designed to encase instruments<br>for sterilization | Identical |
{10}------------------------------------------------
| Feature | APTIMAX ULTRA GI<br>Instrument Tray<br>Subject Device | APTIMAX Instrument Tray<br>K013003 Predicate Device | Comparison |
|------------------------------------------|---------------------------------------------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Reusable/Single<br>Use | Reusable | Reusable | Identical |
| Classification<br>Name…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.