K192978 · Andersen Sterilizers, Inc. · FLF · Nov 12, 2020 · General Hospital
Device Facts
Record ID
K192978
Device Name
EOGas 4 Ethylene Oxide Gas Sterilizer
Applicant
Andersen Sterilizers, Inc.
Product Code
FLF · General Hospital
Decision Date
Nov 12, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6860
Device Class
Class 2
Indications for Use
The EOGas 4 Ethylene Oxide Gas Sterilizer is designed to sterilize reusable medical devices that are sensitive to moisture, heat, chemical corrosion, or radiation.
Device Story
EOGas 4 is an ethylene oxide (EO) gas sterilizer for heat/moisture-sensitive reusable medical devices. System uses unit-dose 100% EO cartridges and gas-impervious flexible sterilization bags as chambers. Operator places devices in bag with Humidichip and appropriate Process Challenge Device (PCD); system executes automated cycle (3-hour or 6-hour exposure at 50°C). Output is sterilized medical equipment. Used in hospitals/clinics; operated by healthcare staff. Sterility verified by inactivation of biological indicators (BI) in PCDs (SteriTest or Endo-SteriTest). Post-sterilization, devices require mandatory aeration per manufacturer instructions to meet ANSI/AAMI/ISO 10993-7 residual limits. Benefits include effective sterilization of complex endoscopes without heat/moisture damage.
Clinical Evidence
Bench testing only. Validation included minimum sterilization parameter definition, half-dose repeatability studies, simulated-use testing with artificial soil, and in-use testing in hospital/clinic settings. Primary endpoint was 6-log inactivation of Bacillus atrophaeus biological indicators. Results confirmed 10^-6 sterility assurance level for specified endoscope loads. EO residual analysis confirmed compliance with ANSI/AAMI/ISO 10993-7 after specified aeration times.
Technological Characteristics
Ethylene oxide gas sterilization system; 100% EO unit dose cartridge; flexible gas-impervious sterilization bag chamber. Critical parameters: EO concentration, relative humidity (35-70%), temperature (50°C ± 3°C), and time. Connectivity: Standalone. Sterilization: Ethylene oxide gas. Compliance: ANSI/AAMI/ISO 10993-7 for residuals; electromagnetic compatibility and electrical safety standards.
Indications for Use
Indicated for sterilization of reusable medical devices sensitive to moisture, heat, chemical corrosion, or radiation. 3-hour cycle for surface sterilization of instruments (including hinges/mated surfaces) and endoscopes ≤ 1100 mm working length. 6-hour cycle for duodenoscopes and colonoscopes > 1100 mm working length.
Regulatory Classification
Identification
An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 12, 2020
Andersen Sterilizers, Inc. William Andersen President 3154 Caroline Drive Haw River, North Carolina 27258
Re: K192978
Trade/Device Name: EOGas 4 Ethylene Oxide Gas Sterilizer Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene Oxide Gas Sterilizer Regulatory Class: Class II Product Code: FLF Dated: October 8, 2020 Received: October 13, 2020
#### Dear William Andersen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT. Elizabeth Claverie, MS Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K192978
Device Name EOGas 4 Ethylene Oxide Gas Sterilizer
#### Indications for Use (Describe)
The EOGas 4 Ethylene Oxide Gas Sterilize is designed to sterilize reusable medical devices that are sensitive to moisture, heat, chemical corrosion, or radiation. The critical process parameters for the two available cycles are summarized below:
Table 1. EOGas 4 Ethylene Oxide Sterilizer cycle parameters
| EO Exposure Time | Total Cycle Time | EO Amount | Temperature | Relative Humidity |
|------------------|------------------|-----------------|----------------|-------------------|
| 3 hours | 3.5 hours | 17.6 g $\pm$ 5% | 50°C $\pm$ 3°C | 35-70% |
| 6 hours | 7 hours | | | |
The differences between the two options are the length of EO gas exposure and the length of mandatory ventilation after gas exposure; the gas exposure is chosen based on the devices to be sterilized. The appropriate purge probe and process challenge device (PCD) must be used: EOGas 4 SteriTest for a 3-hour gas exposure, EOGas 4 Endo-SteriTest for a 6-hour gas exposure.
The EOGas 4 Ethylene Oxide Gas Sterilizer 3-hour gas exposure is used for surface sterilization of medical devices, including instruments with diffusion-restricted spaces (hinges or mated surfaces), as well as for the sterilization of endoscopes with working length shorter than 1100 mm as specified in the labeling. The EOGas 4 SteriTest PCD is used with the 3-hour gas exposure.
The EOGas 4 Ethylene Oxide Gas Sterilizer 6-hour gas exposure is used for sterilization of duodenoscopes and colonoscopes with working length longer than 1100 mm as specified in the labeling. The EOGas 4 Endo-SteriTest PCD is used with the 6-hour gas exposure.
{3}------------------------------------------------
| Device Type | Maximum Load | Device Examples | Required Aeration |
|--------------------------------------------------------------|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| 3-Hour EO Exposure, EOGas 4 SteriTest PCD (Blue Purge Probe) | | | No additional aeration required; Follow pouch or wrap manufacturer's instructions (Example: Tyve pouches require ≥ 6 hours at 50°C) |
| Metal | 24 lbs (11 kg) | Surgical instruments,<br>delicate sharps,<br>including those with<br>hinges and mated<br>surfaces | |
| Plastic | 7.0 lbs (3.2 kg) | Reusable power cords,<br>trocars, and similar<br>devices | 24 hours at 50°C;<br>Follow manufacturer's<br>instructions |
| Fabric | 6.1 lbs (2.8 kg) | Reusable cloth gowns,<br>towels, and similar devices | |
| ≤ 1100 mm<br>Working<br>Lumen<br>Length | One (1)<br>≥ 2.0 mm ID biopsy channel<br>≤ 1100 mm working length | Gastrovideoscopes,<br>gastrointestinal videoscopes,<br>and similar devices | 8 hours at 50°C;<br>Follow manufacturer's<br>instructions |
| Endoscopes | Four (4)<br>≥ 1.2 mm ID biopsy channel<br>≤ 700 mm working length | Bronchoscopes,<br>bronchovideoscopes, cystoscopes,<br>ureteroscopes, choledocoscopes,<br>and similar devices | |
Table 2. Load and material types validated in the EOGas 4 Ethylene Oxide Gas Sterilizer
6-Hour EO Exposure, EOGas 4 Endo-SteriTest PCD (Gold Purge Probe)
| >1100 mm | Two (2) Duodenoscopes* | Olympus TJF-Q180V, | 6 hours at 50°C for |
|------------|------------------------------------------------------------|----------------------|--------------------------|
| Working | ≥ 2.0 mm ID biopsy channel | Olympus TJF-Q160VF , | Olympus and Pentax |
| Lumen | ≤ 1250 mm working length | Olympus TJF-Q190V, | endoscopes in Sterisheet |
| Length | ≥ 1.2 mm ID, ≤ 3530 mm<br>maximum length of any<br>channel | Olympus PJF-160, | |
| Endoscopes | Two (2) Colonoscopes* | Fujifilm ED-530XT, | 8 hours at 50°C for |
| | ≥ 3.7 mm ID biopsy channel | Pentax ED34-i10T2, | Fujifilm endoscopes |
| | ≤ 1700 mm working length | Pentax ED-3490TK | in Sterisheet |
| | ≥ 1.2 mm ID, ≤ 3530 mm<br>maximum length of any channel | Olympus CF-Q180AL, | Follow manufacturer's |
| | | Fujifilm EC-600HL, | instructions |
| | | Pentax EC-3490Li | |
* One (1) duodenoscope may also be paired with one (1) colonoscope
Reusable medical devices must be aerated following the instructions of the device manufacturer and the packaging material manufacturer. Devices are released for use after sterilization based on successful inactivation of a biological indicator (BI) in the Andersen EOGas 4 SteriTest (3-hour gas exposure) or EOGas 4 Endo-SteriTest (6-hour gas exposure) process challenge devices.
{4}------------------------------------------------
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary K192978
#### Applicant's Name and Address
Andersen Sterilizers, Inc. 3154 Caroline Drive Haw River, NC 27258
## Contact Person
William K. Andersen, BE, MD, FAAOS President Phone: 336-376-8622, Fax: 336-376-5428
### Date of Preparation
October 8, 2020
### Device
Proprietary Name EOGas 4 Ethylene Oxide Gas Sterilizer Common Name Ethylene oxide gas sterilizer Classification Class II (21 CFR 880.6860) Medical Specialty General Hospital Product Code FLF
The refill kits for the EOGas 4 Ethylene Oxide Gas Sterilizer, including the accessories (sterilization bags, EOGas 4 cartridges, Dosimeters, and Humidichips), are registered with the US Environmental Protection Agency (EPA #69340-7).
### Predicate Device
| Device Name | EOGas 4 Ethylene Oxide Gas Sterilizer |
|---------------|---------------------------------------|
| 510(k) number | K150646 |
| Manufacturer | Andersen Sterilizers, Inc. |
The principles of operation for the EOGas 4 Ethylene Oxide Sterilizer, intended use, and technology are unchanged. The predicate EOGas 4 Ethylene Oxide Gas Sterilizer has the option of a 3-hour gas exposure or a 5-hour gas exposure, although only the 3-hour gas exposure is indicated for sterilization in hospitals and other human healthcare settings. To add the sterilization of duodenoscopes and colonoscopes to the indications for use, the 5-hour gas exposure was lengthened to 6 hours and a new process challenge device was developed appropriate for the longer cycle.
{6}------------------------------------------------
## Device Description
The EOGas 4 Ethylene Oxide Gas Sterilizer, model AN4000.60 (115V) or AN4000.61 (230V), is intended to sterilize moisture, temperature, chemical corrosion, or radiation-sensitive reusable medical devices in healthcare facilities. The sterilant is a unit dose of 100% ethylene oxide contained in a cartridge, and the sterilization chamber is a gas-impervious flexible sterilization bag.
### Indications for Use
The EOGas 4 Ethylene Oxide Gas Sterilizer is designed to sterilize reusable medical devices that are sensitive to moisture, heat, chemical corrosion, or radiation. The critical process parameters for the two available cycles are summarized below:
| EO Exposure<br>Time | Total Cycle<br>Time | EO Amount | Temperature | Relative<br>Humidity |
|---------------------|---------------------|-------------|-------------|----------------------|
| 3 hours | 3.5 hours | 17.6 g ± 5% | 50°C ± 3°C | 35-70% |
| 6 hours | 7 hours | | | |
Table 1. EOGas 4 Ethylene Oxide Sterilizer cycle parameters
The differences between the two options are the length of EO gas exposure and the length of mandatory ventilation after gas exposure; the gas exposure is chosen based on the devices to be sterilized. The appropriate purge probe and process challenge device (PCD) must be used: EOGas 4 SteriTest for a 3-hour gas exposure, EOGas 4 Endo-SteriTest for a 6-hour gas exposure.
The EOGas 4 Ethylene Oxide Gas Sterilizer 3-hour gas exposure is used for surface sterilization of medical devices, including instruments with diffusion-restricted spaces (hinges or mated surfaces), as well as for the sterilization of endoscopes with working length shorter than 1100 mm as specified in the labeling. The EOGas 4 SteriTest PCD is used with the 3-hour gas exposure.
The EOGas 4 Ethylene Oxide Gas Sterilizer 6-hour gas exposure is used for sterilization of duodenoscopes and colonoscopes with working length longer than 1100 mm as specified in the labeling. The EOGas 4 Endo-SteriTest PCD is used with the 6-hour gas exposure.
{7}------------------------------------------------
| Device Type | Maximum Load | Device Examples | Required Aeration |
|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 3-Hour EO Exposure, EOGas 4 SteriTest PCD (Blue Purge Probe) | | | |
| Metal | 24 lbs<br>(11 kg) | Surgical instruments,<br>delicate sharps,<br>including those with<br>hinges and mated<br>surfaces | No additional aeration<br>required;<br>Follow pouch or wrap<br>manufacturer's<br>instructions (Example:<br>Tyvek pouches require<br>$\ge$ 6 hours at 50°C) |
| Plastic | 7.0 lbs<br>(3.2 kg) | Reusable power cords,<br>trocars, and similar<br>devices | 24 hours at 50°C; |
| Fabric | 6.1 lbs<br>(2.8 kg) | Reusable cloth gowns,<br>towels, and similar<br>devices | Follow manufacturer's<br>instructions |
| ≤ 1100 mm<br>Working<br>Lumen<br>Length<br>Endoscopes | One (1)<br>$\ge$ 2.0 mm ID biopsy channel<br>≤ 1100 mm working length | Gastrovideoscopes,<br>gastrointestinal<br>videoscopes, and<br>similar devices | 8 hours at 50°C if in<br>Sterisheet; |
| | Four (4)<br>$\ge$ 1.2 mm ID biopsy channel<br>≤ 700 mm working length | Bronchoscopes,<br>bronchovideoscopes,<br>cystoscopes,<br>ureteroscopes,<br>choledocoscopes, and<br>similar devices | Follow manufacturer's<br>instructions |
| 6-Hour EO Exposure, EOGas 4 Endo-SteriTest PCD (Gold Purge Probe) | | | |
| >1100 mm<br>Working<br>Lumen<br>Length<br>Endoscopes | Two (2) Duodenoscopes*<br>$\ge$ 2.0 mm ID biopsy channel<br>≤ 1250 mm working length<br>$\ge$ 1.2 mm ID, ≤ 3530 mm<br>maximum length of any<br>channel | Olympus TJF-Q180V,<br>Olympus TJF-Q160VF,<br>Olympus TJF-Q190V,<br>Olympus PJF-160,<br>Fujifilm ED-530XT,<br>Pentax ED34-i10T2<br>Pentax ED-3490TK | 6 hours at 50°C for<br>Olympus and Pentax<br>endoscopes in<br>Sterisheet; |
| | Two (2) Colonoscopes*<br>$\ge$ 3.7 mm ID biopsy channel<br>≤ 1700 mm working length<br>$\ge$ 1.2 mm ID, ≤ 3530 mm<br>maximum length of any<br>channel | Olympus CF-Q180AL,<br>Fujifilm EC-600HL,<br>Pentax EC-3490Li | 8 hours at 50°C for<br>Fujifilm endoscopes in<br>Sterisheet;<br>Follow manufacturer's<br>instructions |
| | * One (1) duodenoscope may also be paired with one (1)<br>colonoscope | | |
Table 2. Load and material types validated in the EOGas 4 Ethylene Oxide Gas Sterilizer
Reusable medical devices must be aerated following the instructions of the device manufacturer and the packaging material manufacturer. Devices are released for use after sterilization based
{8}------------------------------------------------
on successful inactivation of a biological indicator (BI) in the Andersen EOGas 4 SteriTest (3hour gas exposure) or EOGas 4 Endo-SteriTest (6-hour gas exposure) process challenge devices.
## Technological Characteristics Comparison Table
The EOGas 4 Ethylene Oxide Gas Sterilizer is compared to the predicate device - the EOGas 4 Ethylene Oxide Gas Sterilizer (K150646), because the sterilizers are intended for the same use, designed in the way and use the same technology. The predicate EOGas 4 Ethylene Oxide Gas Sterilizer has the option of a 3-hour gas exposure or a 5-hour gas exposure. To sterilize a load consisting of two duodenoscopes or colonoscopes at a time, the 5-hour gas exposure was lengthened to 6 hours, and a process challenge device was created, the EOGas 4 Endo-SteriTest, that is appropriate for the longer gas exposure. A comparison between the sterilizers is listed in Table 3.
| | Subject<br>EOGas 4 Sterilizer K192978 | Predicate EOGas 4<br>Sterilizer K150646 | Comparison |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Intended<br>Use | Indoor EO sterilizer in a healthcare setting with a SAL of 10-6 | | Same |
| Design | Same design for exterior and cabinet | | |
| | Same unit dose EO; Same EO impervious sterilization bag | | Same |
| | Same flexible sterilization chamber | | |
| Technology | Use EO as sterilant; EPA registered | | Same |
| | Critical parameters: EO concentration, RH, temperature, and time | | |
| Safety | Verify compliance for electromagnetic compatibility and electrical<br>safety | | Same |
| Indications<br>for Use | 3-hour gas exposure at 50°C;<br>6-hour gas exposure at 50°C | 3-hour gas exposure at 50°C;<br>(5-hour gas exposure at 50°C;<br>not used for healthcare<br>sterilization) | Added a 6-hour<br>gas exposure for<br>sterilization of<br>endoscopes with<br>> 1100 mm<br>working length |
| Process<br>Challenge<br>Device | EOGas 4 Endo-SteriTest (K192980)<br>EZTest-Gas BI (K930683);<br>Bionova BT110 RRBI (K191021);<br>BI Receptacle:<br>Stainless steel cylinder<br>0.020" ID x 4.50" L lumen | EOGas 4 SteriTest (K151585)<br>EZTest-Gas BI (K930683);<br>BI Receptacle:<br>Stainless steel cylinder<br>0.020" ID x 0.25" L orifice | Similar:<br>appropriate for<br>the intended<br>length of gas<br>exposure |
| Performance | Sterilize reusable medical devices as labeled to a SAL of 10-6 | | Similar |
Table 3. Comparison between the EOGas 4 Ethylene Oxide Gas Sterilizer and the predicate EOGas 4 Ethylene Oxide Gas Sterilizer
{9}------------------------------------------------
# Summary of Non-Clinical Testing:
| Time | Purpose | Acceptance Criteria | Results |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Minimum<br>Sterilization<br>Parameters | To define and<br>validate the<br>endoscope loads that<br>can be sterilized using<br>the 6-hour gas<br>exposure at 50°C in<br>an EOGas 4 sterilizer | The EO gas exposure time at<br>50°C in an EOGas 4 sterilizer<br>must result in sterilization of<br>the endoscope loads. | Using a 6-hour EO gas exposure<br>in the EOGas 4 sterilizer,<br><br>6-Log biological indicators were<br>consistently inactivated for two<br>duodenoscopes, two<br>colonoscopes, or one colonoscope<br>and one duodenoscope in each<br>load.<br><br>The minimum parameters for<br>sterilization of two<br>duodenoscopes or colonoscopes<br>are 518 mg/L EO, 46.4°C, and<br>35% RH. |
| Half Dose<br>Validation | To demonstrate the<br>repeatability of the<br>EOGas 4 sterilization<br>process, and that a<br>SAL of 10-6 is<br>achieved for<br>duodenoscope and<br>colonoscope<br>sterilization | For consecutive half dose<br>cycles and full dose cycles:<br><br>EO concentration is half when<br>half the amount of EO is used.<br><br>All 6-Log <i>Bacillus atrophaeus</i><br>biological indicators,<br>inoculated at the worst-case<br>locations of the tested<br>endoscopes, are inactivated. | Consecutive half dose cycles and<br>full dose cycles were performed.<br><br>The EO concentration was half<br>when half the amount of EO was<br>used. The cumulative lethality of<br>half dose cycles was half the<br>lethality of full dose cycle.<br><br>The cycles consistently<br>inactivated all 6-Log <i>Bacillus atrophaeus</i> biological indicators<br>inoculated at the midpoint of the<br>tested channels and at the elevator<br>mechanism of duodenoscopes, as<br>well as the water jet channel of<br>the colonoscopes. |
| Simulated-<br>Use Testing | To demonstrate the<br>successful<br>sterilization of<br>duodenoscopes and<br>colonoscopes within<br>the claims under a<br>worst-case scenario | For consecutive full dose<br>cycles:<br><br>Biological indicators with 6-<br>Log <i>Bacillus atrophaeus</i> ,<br>prepared in an artificial soil and<br>inoculated at the worst-case<br>locations of the tested<br>endoscopes, are inactivated. | Biological indicators with 6-Log<br><i>Bacillus atrophaeus</i> were<br>prepared in an artificial soil and<br>inoculated at the center of the<br>tested channels and at the elevator<br>mechanism of duodenoscopes as<br>well as the water jet channel of<br>the colonoscopes.<br><br>Inactivated biological indicators<br>were obtained in all cycles for all<br>duodenoscope and colonoscope<br>loads tested. |
| | | | |
| In-Use<br>Testing | To demonstrate that<br>the 6-hour cycle at<br>50°C in an EOGas 4<br>sterilizer successfully<br>sterilizes<br>duodenoscopes and<br>colonoscopes used in<br>routine endoscopic<br>procedures in a<br>hospital or clinic<br>setting | Duodenoscopes and<br>colonoscopes, used on the<br>patients, cleaned but not<br>disinfected, are sterilized using<br>the 6-hour cycle at 50°C in an<br>EOGas 4 sterilizer. | Duodenoscopes and<br>colonoscopes, used on patients,<br>were cleaned per hospital protocol<br>but not disinfected, processed<br>using the 6-hour EO exposure at<br>50°C in the EOGas 4 sterilizer;<br>sterility was tested by a flush<br>method per USP. All test cultures<br>from the processed<br>duodenoscopes and colonoscopes<br>were sterile. |
| EO<br>Residuals | To demonstrate that<br>endoscopes and their<br>accessories are safe<br>to use if the guidance<br>and instructions are<br>followed | After additional aeration<br>following the cycle, EO<br>residuals on duodenoscopes and<br>colonoscopes are evaluated,<br>and the residuals on the<br>endoscopes and accessories<br>meet the requirements of<br>ANSI/AAMI/ISO 10993-7. | Olympus TJF-Q180V, TJF-<br>Q160VF, TJF-Q190V, PJF-160,<br>CF-Q180AL, Pentax ED34-i10T2,<br>ED-3490TK duodenoscopes and<br>EC-3490Li colonoscopes wrapped<br>in Sterisheet must aerate for an<br>additional 6 hours after the cycle.<br>Fujifilm ED-530XT and EC-<br>600HL wrapped in Sterisheet<br>must aerate for an addition 8<br>hours after the cycle.<br>After 6 hours of additional<br>aeration, EO residuals on the<br>packaging materials met the<br>requirements of ANSI/AAMI/ISO<br>10993-7. |
## Table 4. Summary of Non-Clinical Testing
{10}------------------------------------------------
The EOGas 4 Ethylene Oxide Gas Sterilizer 6-hour gas exposure has been validated using applicable tests in the FDA 1993 "Guidance on Premarket Notification [510(k)] Submissions for Sterilizers Intended for Use in Health Care Facilities".
Duodenoscopes used in these studies include:
- Olympus TJF-Q180V ●
- Olympus TJF-Q160VF ●
- Olympus TJF-Q190V ●
- Olympus PJF-160 ●
- Fujifilm ED-530XT
- Pentax ED34-i10T2
- Pentax ED-3490TK ●
Colonoscopes used include:
- Olympus CF-Q180AL ●
- Fujifilm EC-600HL
- Pentax EC-3490Li .
The maximum loads of duodenoscopes or colonoscopes that may be routinely sterilized in the
{11}------------------------------------------------
EOGas 4 Ethylene Oxide Gas Sterilizer were defined and validated. Using a 6-hour gas exposure at 50°C, the EOGas 4 sterilization system reproducibly and effectively sterilizes two duodenoscopes, two colonoscopes, or one duodenoscope paired with one colonoscope, achieving a minimum sterility assurance level of 10-6.
The EOGas 4 Endo-SteriTest process challenge device, consisting of a BI receptacle with a selfcontained biological indicator, was developed for the 6-hour gas exposure at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. It represents a rigorous challenge to the EOGas 4 sterilization process as it impedes gas diffusion, heating, and humidification. Its resistance characteristics are greater than the same BI placed in the worst-case locations in the two duodenoscope or two colonoscope loads.
The validation testing demonstrated that exposure to EO gas under the defined load and physical parameters achieved a minimum sterility assurance level of 106 for two duodenoscopes, two colonoscopes, or one duodenoscope paired with one colonoscope, in each load. The effectiveness of the sterilization process for the loads was confirmed by successful sterilization in simulateduse testing. In addition, in-use testing confirmed the ability of the cycle to sterilize duodenoscopes or colonoscopes used clinically in a hospital setting.
Process residue analysis showed that the EO residuals remaining on the duodenoscopes or colonoscopes tested in the study, after an additional aeration of 6 hours (for Olympus duodenoscopes, Olympus colonoscopes, and Pentax duodenoscopes) or 8 hours (for Fujifilm duodenoscopes and colonoscopes), met the requirements of ANSI AAMI ISO 10993-7. demonstrating that the EOGas 4 Ethylene Oxide Gas Sterilizer and its accessories are safe to use if the guidance and instructions are followed.
Physical performance tests demonstrated that the EOGas 4 Ethylene Oxide Gas Sterilizer. EOGas cartridges, sterilization bags, and Humidichips met their performance specifications. The EOGas 4 sterilization system achieved and maintained the cycle specifications for EO concentration, temperature, time, and relative humidity. Both the sterilizer and the accessories consistently operated in accordance with predetermined criteria. The 6-hour gas exposure at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer was repeatable and reliable under the indicated test load conditions.
### Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that the EOGas 4 Ethylene Oxide Sterilizer is as safe, as effective, and performs as well as or better than the legally marketed predicate EOGas 4 Ethylene Oxide Sterilizer (K150646).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.