K233991 · Elutia, Inc. · FTM · Jun 14, 2024 · General, Plastic Surgery
Device Facts
Record ID
K233991
Device Name
CanGaroo RM Antibacterial Envelope
Applicant
Elutia, Inc.
Product Code
FTM · General, Plastic Surgery
Decision Date
Jun 14, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CanGaroo® RM Antibacterial Envelope is intended to securely hold a cardiac implantable electronic device or an implantable neurostimulator to create a stable environment when implanted in the body. The cardiac implantable electronic devices that may be used with the Caroo RM Antibacterial Envelope include pacemaker pulse generators, defibrillators, and other cardiac implantable electronic devices. The implantable neurostimulator devices that may be used with the CanGaroo RM Antibacterial Envelope include vagus nerve stimulators, spinal cord neuromodulators, deep brain stimulators, sacral nerve stimulators, and other neurostimulator devices. The CanGaroo RM Antibacterial Envelope contains the antibiotics rifampin and minocycline, which have been shown in preclinical testing to reduce bacterial colonization on the envelope. Overall clinical benefit has not yet been evaluated.
Device Story
Surgical mesh envelope; holds cardiac implantable electronic devices (IEDs) or neurostimulators; creates stable environment; reduces bacterial colonization via antibiotic elution. Composed of decellularized, lyophilized porcine small intestinal submucosa (SIS) extracellular matrix; contains resorbable PLGA discs impregnated with rifampin and minocycline. Implanted by physicians during surgical procedures. Antibiotics elute from discs to inhibit bacterial growth on envelope surface. Single-use; sterile; resorbable. Benefits include device stabilization and potential reduction in bacterial colonization.
Clinical Evidence
No clinical data. Evidence includes bench testing (physical characteristics, seam strength, drug elution), in vitro antibacterial activity (Modified AATCC 100 against 4 gram-positive and 3 gram-negative organisms), and in vivo animal studies (pharmacokinetics and rabbit bacterial challenge model showing clearance at day 7).
Technological Characteristics
Materials: Porcine small intestinal submucosa (SIS) extracellular matrix, resorbable PLGA discs. Antibiotics: Rifampin, minocycline. Form factor: Multilaminate sheets with perforated design. Sterilization: E-beam or gamma irradiation. Resorbable. No software or electronic components.
Indications for Use
Indicated for patients requiring implantation of cardiac electronic devices (pacemakers, defibrillators) or neurostimulators (vagus nerve, spinal cord, deep brain, sacral nerve stimulators) to securely hold the device and create a stable environment.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
Elutia Inc. Erica Elchin Vice President, Quality & Regulatory Affairs 1100 Old Ellis Road Suite 1200 Roswell, Georgia 30076
# Re: K233991
Trade/Device Name: CanGaroo RM Antibacterial Envelope Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM Dated: May 17, 2024 Received: May 17, 2024
# Dear Erica Elchin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Jessica L. Batista -S Date: 2024.06.14 16:31:50 Batista -S -04'00
for
Sara Royce Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Jessica L
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233991
Device Name CanGaroo RM Antibacterial Envelope
#### Indications for Use (Describe)
The CanGaroo® RM Antibacterial Envelope is intended to securely hold a cardiac implantable electronic device or an implantable neurostimulator to create a stable environment when implanted in the body. The cardiac implantable electronic devices that may be used with the Caroo RM Antibacterial Envelope include pacemaker pulse generators, defibrillators, and other cardiac implantable electronic devices. The implantable neurostimulator devices that may be used with the CanGaroo RM Antibacterial Envelope include vagus nerve stimulators, spinal cord neuromodulators, deep brain stimulators, sacral nerve stimulators, and other neurostimulator devices.
The CanGaroo RM Antibacterial Envelope contains the antibiotics rifampin and minocycline, which have been shown in preclinical testing to reduce bacterial colonization on the envelope. Overall clinical benefit has not yet been evaluated.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(K) Summary is provided per the requirements of Section 807.92(c)
#### Company Information
| Company Name: | Elutia Inc.<br>1100 Old Ellis Road, Suite 1200, Roswell, GA 30076 |
|----------------------|-------------------------------------------------------------------|
| Contact Name: | Erica Elchin |
| Contact Email: | eelchin@elutia.com |
| Contact Title: | VP, Quality & Regulatory Affairs |
| Phone: | 877-651-2628 |
| Date Prepared: | June 14, 2024 |
| Device Information | |
| Trade Name: | CanGaroo® RM Antibacterial Envelope |
| Common Name: | Surgical Mesh Envelope |
| Classification Name: | Surgical Mesh |
| Regulation Number: | 21 CFR 878.3300 |
| Product Code: | FTM |
| Device Class: | Class II |
# Predicate Device
The CanGaroo® RM Antibacterial Envelope is substantially equivalent to the following device:
- . TYRX™ Absorbable Antibacterial Envelope, K192389
CanGaroo® (K201313), LUOFUCON® PHMB Alginate Dressing (K201016), and XenMatrix™ AB Surgical Graft (K162193) are reference devices.
# Device Description
The CanGaroo® RM Antibacterial Envelope consists of decellularized, non-crosslinked, Iyophilized extracellular matrix (ECM) and resorbable ring-shaped poly(lactide-co-qlycolide) (PLGA) discs containing the antibiotics rifampin and minocycline. The ECM material is derived from porcine small intestinal submucosa (SIS). The envelope is constructed with four multilaminate sheets, perforated to allow for exudate. The drug-eluting polymer discs are secured between the multilaminate sheets on each side of the envelope.
The CanGaroo RM Antibacterial Envelope is intended to securely hold an implantable electronic device (IED) to create a stable environment when implanted in the body, while reducing bacterial colonization with rifampin and minocycline eluted from resorbable polymer discs in the envelope. The CanGaroo RM Antibacterial Envelope is based on the existing CanGaroo Envelope (K201313), with the addition of drug-eluting disc components with the antibiotics rifampin and minocvoline.
The CanGaroo RM Antibacterial Envelope is provided sterile in two sizes and is intended for single use in a single patient only.
{4}------------------------------------------------
| CanGaroo RM<br>Antibacterial Envelope | Envelope Size<br>W x L | # of<br>discs | Rifampin | Minocycline |
|---------------------------------------|------------------------|---------------|----------|-------------|
| Medium | 6.9 cm x 6.5 cm | 2 | 10.5 mg | 9.3 mg |
| Large | 6.9 cm x 8.0 cm | 2 | 10.5 mg | 9.3 mg |
# Indications for Use
The CanGaroo RM Antibacterial Envelope is intended to securely hold a cardiac implantable electronic device or an implantable neurostimulator to create a stable environment when implanted in the body. The cardiac implantable electronic devices that may be used with the CanGaroo RM Antibacterial Envelope include pacemaker pulse generators, defibrillators, and other cardiac implantable electronic devices. The implantable neurostimulator devices that may be used with the CanGaroo RM Antibacterial Envelope include vagus nerve stimulators, spinal cord neuromodulators, deep brain stimulators, sacral nerve stimulators, and other neurostimulator devices.
The CanGaroo RM Antibacterial Envelope contains the antibiotics rifampin and minocycline, which have been shown in preclinical testing to reduce bacterial colonization on the envelope. Overall clinical benefit has not yet been evaluated.
#### Summarv of Technological Characteristics
The CanGaroo RM Antibacterial Envelope is designed to hold IEDs securely to create a stable environment when implanted in the body. The intended use of the CanGaroo RM Antibacterial Envelope is identical to the predicate device (K192389) and its technical characteristics are substantially equivalent.
The CanGaroo RM Antibacterial Envelope is based on the CanGaroo Envelope (K201313), with the addition of rifampin and minocycline druq-eluting discs. The envelope is made of the same material, porcine SIS, as the predicate CanGaroo Envelope. CanGaroo RM is a resorbable device, as is the predicate device.
The drug eluting discs contain the same antibacterial drugs, rifampin and minocycline, as the predicate device K192389. Antibiotic concentrations for CanGaroo RM span the range of the predicate device K192389 and the reference device K162193.
The device and the predicate are all terminally sterilized (e-beam or gamma irradiation).
The performance of CanGaroo RM and the predicate device all have been assessed through bench, in vitro, and in vivo testing. The results demonstrate that CanGaroo RM does not raise any new questions of safety and effectiveness.
# Performance Data
The following performance data is provided in support of the substantial equivalence determination.
#### Performance Standards:
No performance standards have been established for this device under Section 514 of the Federal Food, Druq, and Cosmetic Act.
{5}------------------------------------------------
#### Biocompatibility Testing:
Biocompatibility testing in accordance with the current ISO 10993 series and FDA guidance was conducted on the subject device and the results indicate that the device is biocompatible per these standards. The testing included cytotoxicity, sensitization, acute systemic toxicity, pyrogenicity, and implantation studies.
#### Bench Testing:
Bench testing was completed for the CanGaroo RM Antibacterial Envelope. The subject device passed all test requirements. The following tests were conducted for the subject device:
- . Physical Characteristics
- Appearance o
- Dimensional o
- . Functional Characteristics
- Seam Strength o
- . Antibacterial Effectiveness
- Antibacterial Activity, in vitro (Modified AATCC 100) o Utilizing AATCC methodology, CanGaroo RM demonstrated bacterial reduction against a panel of 4 gram positive and 3 gram-negative organisms.
- Drug Elution
- o Drug assay and drug elution/release testing.
#### Animal Testing:
- . Pharmacokinetics, in vivo
- . Bacterial Challenge model, in vivo
A bacterial challenge model in rabbits showed clearance at day 7 following implantation for a panel of 3 gram positive and 3 gram negative organisms.
# Conclusion
Performance data utilizing in vivo testing demonstrates substantial equivalence to the predicate device (K192389). The CanGaroo RM device is as safe and as effective, with equivalent performance, to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.