The envelope is intended to hold a pacemaker pulse generator or defibrillator securely in order to provide a stable environment when implanted in the body. The envelope contains the antimicrobial agents rifampin and minocycline, which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of the generator or defibrillator. The envelope is intended to be used in conjunction with pacemakers and implantable defibrillators.
Device Story
Sterile, absorbable mesh envelope designed to house cardiovascular implantable electronic devices (CIEDs), specifically pacemakers or ICDs. Constructed from a multifilament knitted mesh of glycolide, caprolactone, and trimethylene carbonate, coated with an absorbable tyrosine-based polyarylate polymer containing antimicrobial agents (rifampin and minocycline). Implanted by a physician during CIED surgery to secure the generator within the tissue pocket. The envelope provides a stable environment and releases antimicrobial agents to reduce infection risk. No electronic components or software; purely mechanical/chemical function.
Clinical Evidence
Bench testing only. Stability study data collected per ICH guidelines confirms the device meets all product requirements through the proposed extended shelf life.
Technological Characteristics
Multifilament knitted mesh composed of glycolide, caprolactone, and trimethylene carbonate. Coated with absorbable tyrosine-based polyarylate polymer containing rifampin and minocycline. Sterile, biocompatible. Supplied in two sizes: Medium (2.5 in x 2.7 in) and Large (2.9 in x 3.3 in). No electronic components, software, or energy source.
Indications for Use
Indicated for patients receiving a pacemaker or implantable cardioverter defibrillator (ICD) to hold the device securely and provide a stable environment in the tissue pocket.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 27, 2020
Medtronic, Inc. Nancy Cameron Senior Principal Regulatory Specialist 8200 Coral Sea Street NE Mounds View, Minnesota 55112
Re: K192389
Trade/Device Name: TYRX Absorbable Antibacterial Envelope (medium), TYRX Absorbable Antibacterial Envelope (large) Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: March 2, 2020 Received: March 3, 2020
Dear Nancy Cameron:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hetal Odobasic Assistant Director Implantable Electrophysiology Devices Team Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) TBD K192389
R172309
Device Name TYRX Absorbable Antibacterial Envelope
#### Indications for Use (Describe)
The envelope is intended to hold a pacemaker pulse generator or defibrillator securely in order to provide a stable environment when implanted in the body. The envelope contains the antimicrobial agents rifampin and minocycline, which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of the generator or defibrillator. The envelope is intended to be used in conjunction with pacemakers and implantable defibrillators.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
| Date Prepared: | 28 August 2019 |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner / Address: | Medtronic, Inc.<br>Cardiac Rhythm and Heart Failure (CRHF)<br>8200 Coral Sea St. NE<br>Mounds View, MN 55112 |
| Contact Person: | Primary Contact: |
| | Nancy Cameron<br>Senior Principal Regulatory Affairs Specialist<br>Medtronic Cardiac Rhythm Heart Failure<br>763.526.2355<br>nancy.e.cameron@medtronic.com |
| | Secondary Contact: |
| | Wendy Saunders<br>Regulatory Affairs Director<br>Medtronic Cardiac Rhythm Heart Failure<br>763.526.8134<br>wendy.a.saunders@medtronic.com |
| Submission Type: | Traditional 510(k): Shelf Life Extension |
| Device Trade Name: | TYRX™ Absorbable Antibacterial Envelope |
| Device Common Name: | Surgical Mesh |
| Regulation Number: | CFR 878.3300 |
| Product Code: | FTL |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
| Special Controls: | None |
| Predicate Device: | TYRX™ Absorbable Antibacterial Envelope, K180030 |
{4}------------------------------------------------
### Device Description
TYRXTM Absorbable Antibacterial Envelope (TYRX Envelope, or the envelope) is a sterile prosthesis comprised of two components; an absorbable substrate mesh, and an absorbable tyrosine based polyarylate polymer containing the antimicrobial agents, rifampin and minocycline, and is designed to hold a Cardiovascular Implantable Electronic Devices. CIED. (pacemaker or Implantable Cardioverter Defibrillator, ICD), securely to create a stable environment when the device is implanted in the body.
The TYRX Envelope is constructed of multifilament knitted mesh composed of glycolide, caprolactone, and trimethylene carbonate polymer, which is coated with an absorbable polyarylate polymer containing the drug substances rifampin and minocycline.
Like its predicate, the TYRX Envelope is supplied in two sizes, a 2.5 in. x 2.7 in. pacemaker size (Medium), and a 2.9 in. x 3.3 in. ICD size (Large). Details for the TYRX Envelopes are provided below.
| Description of Device/ Part Number | Label Claim |
|-------------------------------------------------------------------------|----------------------------------------|
| TYRX Absorbable Antibacterial Envelope (Medium)<br>Product ID: CMRM6122 | 5.1 mg Minocycline<br>8.0 mg Rifampin |
| TYRX Absorbable Antibacterial Envelope (Large)<br>Product ID: CMRM6133 | 7.6 mg Minocycline<br>11.9 mg Rifampin |
### Indications for Use
There are no changes to the Indications for Use as a result of this submission. The Indications for Use are provided below:
The envelope is intended to hold a pacemaker pulse generator or defibrillator securely in order to create a stable environment when implanted in the body. The envelope contains the antimicrobial agents, rifampin and minocycline, which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of a pacemaker or defibrillator. The envelope is intended to be used in conjunction with pacemakers or defibrillators.
### Technological Characteristics
TYRX Envelope is a biocompatible, sterile device intended to hold a pacemaker or defibrillator securely in the surgically created tissue pocket in order to create a stable environment for the pacemaker or defibrillator when implanted in the body. The TYRX Envelope is identical to its predicate device, cleared under K180030. The further extension of product shelf life presented in this submission does not impact the technical characteristics of the device as compared to the predicate device.
# Summary of Testing
Extension of the TYRX Envelope shelf life is supported by stability study data collected per ICH guidelines. Results of this study demonstrate the TYRX Envelope, both medium and large size, continue to meet all product requirements through the proposed shelf life. There are no changes
{5}------------------------------------------------
to the finished product TYRX Envelope analytical testing requirements as a result of the modifications described in this submission. The extended shelf life TYRX Envelope design, materials, mechanism of action, patient contact and intended use are the same as the predicate device.
# Substantial Equivalence
Substantial equivalence of the TYRX Envelope with the proposed shelf life is based on ICH stability studies conducted using the dual foil pouch package with desiccant. Other minor manufacturing changes have been incorporated per Quality System processes. There are no changes to the finished product TYRX Envelope analytical testing requirements, design, materials, mechanism of action, patient contact or intended use associated with the extended shelf life. The individual and cumulative impact of these changes does not alter the risk profile of the TYRX Envelopes. The modified device meets the same finished goods acceptance criteria, using the same analytical test methodologies, as the currently marketed device. Therefore, the TYRX Envelope device, as modified with extended shelf life, is substantially equivalent to the predicate device.
# Conclusion
The TYRX Absorbable Antibacterial Envelope, as modified, is substantially equivalent to the predicate device.
Overall, these modifications do not affect the intended use of the device or alter the fundamental scientific technology. There are no changes to the physical design, principles of operation, or mechanism of action of the current TYRX Envelope.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.