Nitrile Disposable Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl (D5000)
Applicant
Shandong Xingyu Gloves Co., Ltd.
Product Code
LZA · General Hospital
Decision Date
Apr 1, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The Nitrile Disposable Examination Glove, Tested for Use with Chemotherapy Drugs and Fentanyl is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05.
Device Story
Nitrile disposable examination gloves (D5000); non-sterile; single-use; ambidextrous; blue; powder-free. Worn by healthcare examiners to prevent cross-contamination between patient and examiner. Tested for permeation resistance against various chemotherapy drugs and Fentanyl Citrate per ASTM D6978-05. Provides barrier protection in clinical settings. Performance verified via physical property testing (tensile strength, elongation), freedom from holes (ASTM D5151), and residual powder limits (ASTM D6124).
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993-10, -11, -23) and physical performance testing (ASTM D6319, D6124, D5151, D6978).
Technological Characteristics
Material: Nitrile. Form factor: Ambidextrous, blue, powder-free, six sizes (XS-XXL). Standards: ASTM D6319 (physical properties/dimensions), ASTM D6978 (chemotherapy permeation), ASTM D5151 (freedom from holes), ASTM D6124 (residual powder). Biocompatibility: ISO 10993-10, -11, -23. Non-sterile.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs and Fentanyl Citrate. Warning: Do not use with Carmustine or Thiotepa.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Non-Sterile, Single use, Powder-free examination glove, Blue, tested for use with Chemotherapy drugs and Fentanyl (K223903)
Reference Devices
Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs) (K212735)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 1, 2024
Shandong Xingyu Gloves Co., Ltd. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Co. Ltd. 14th Floor, 1500# Century Avenue Shanghai, Shanghai 200122 China
Re: K233990
Trade/Device Name: Nitrile Disposable Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl (D5000) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ, QDO Dated: March 11, 2024 Received: March 11, 2024
Dear Ivy Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Allan G
For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control
{2}------------------------------------------------
and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K233990
Device Name
Nitrile Disposable Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl (D5000)
Indications for Use (Describe)
The Nitrile Disposable Examination Glove, Tested for Use with Chemotherapy Drugs and Fentanyl is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05.
The tested chemotherapy drugs are:
Chemotherapy drug Concentration Minimum Breakthrough Detection Time
Bleomycin sulfate (15.0 mg/ml 15,000 ppm) >240 min
Carboplatin (10.0 mg/ml 10,000 ppm) >240 min
Carmustine (BCNU) (3.3mg/mL 3,300 ppm) 13.2
Cisplatin (1.0 mg/mL 1,000 ppm) >240 min
Cyclophosphamide (20.0 mg/mL 20,000 ppm) >240 min
Cytarabine (Cytosine) (100.0 mg/mL 100,000 ppm) >240 min
Dacarbazine (10.0 mg/mL 10.000 ppm) >240 min
Doxorubicin HCL (2.0 mg/mL 2,000 ppm) >240 min
Etoposide (Toposar) (20.0 mg/mL 20,000 ppm) >240 min
Fluorouracil (5 Flu) (50.0 mg/mL 50,000 ppm) >240 min
Idarubicin (1.0 mg/ml 1.000 ppm) >240 min
Ifosfamide (50.0 mg/mL 50,000 ppm) >240 min
Mechlorethamine HCL (1.0 mg/ml 1,000 ppm) >240 min
Methotrexate (25.0 mg/mL 25,000 ppm) >240 min
Mitomycin C (0.5 mg/mL 500 ppm) >240 min
Mitoxantrone HCL (2.0 mg/mL 2.000 ppm >240 min
Paclitaxel (6.0 mg/mL 6.000 ppm) >240 min
Thiotepa (10.0 mg/mL 10,000 ppm) 13.1
Vincristine Sulfate (1.0 mg/mL 1,000 ppm >240 min
The tested non-chemotherapy drugs are:
Fentanyl Citrate Injection (100 mcg/2mL) >240 min
MESNA (100.0 mg/mL 100,000 ppm) >240 min
Trisenox (1.0 mg/ml 1,000 ppm) >240 min
Note: Carmustine and Thiotepa have extremely low permeation times of 13.2 and 13.1 minutes respectively.
Warning: Do Not Use with Carmustine, Thiotepa
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(K) Summary
# K233990
(As required by 21 CFR 807.92)
Date prepared: 2024-04-01
## A. Applicant:
Name: Shandong Xingyu Gloves Co., Ltd. Address: NO. 2158 Yaoqian Road, Economic Development Zone, Gaomi City, Shandong Province, China Contact: Ms. Carrie Lee Title: QA Manager Tel: 0086-536-2580359 Email: carrie@xingyugloves.com
Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda@sungoglobal.com
## B. Device:
Trade Name: Nitrile Disposable Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl Common Name: Nitrile Patient Examination Gloves (Powder Free) Model: D5000 Size(s): XS, S, M, L, XL, XXL
Regulatory Information Classification Name: Polymer Patient Examination Glove Classification: Class I Product code: LZA (Primary), LZC, OPJ, QDO Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital
## C. Predicate device:
510K Number: K223903 Trade Name: Non-Sterile, Single use, Powder-free examination glove, Blue, tested for use with Chemotherapy drugs and Fentanyl Manufacturer: SEMPERIT INVESTMENTS ASIA PTE. LTD. Reference Device
{6}------------------------------------------------
#### 510K Number: K212735
Trade Name: Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs) Manufacturer: Xingyu Medical Tech Co., Ltd.
### D. Indications for use of the device:
The Nitrile Disposable Examination Glove, Tested for Use with Chemotherapy Drugs and Fentanyl is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05. The tested chemotherapy drugs are:
| Chemotherapy drug | Concentration | Minimum Breakthrough Detection Time |
|----------------------------------------|---------------------------|-------------------------------------|
| Bleomycin sulfate | (15.0 mg/ml 15,000 ppm) | >240 min |
| Carboplatin | (10.0 mg/ml 10,000 ppm) | >240 min |
| Carmustine (BCNU) | (3.3mg/mL 3,300 ppm) | 13.2 |
| Cisplatin | (1.0 mg/mL 1,000 ppm) | >240 min |
| Cyclophosphamide | (20.0 mg/mL 20,000 ppm) | >240 min |
| Cytarabine (Cytosine) | (100.0 mg/mL 100,000 ppm) | >240 min |
| Dacarbazine | (10.0 mg/mL 10,000 ppm) | >240 min |
| Doxorubicin HCL | (2.0 mg/mL 2,000 ppm) | >240 min |
| Etoposide (Toposar) | (20.0 mg/mL 20,000 ppm) | >240 min |
| Fluorouracil (5 Flu) | (50.0 mg/mL 50,000 ppm) | >240 min |
| Idarubicin | (1.0 mg/ml 1,000 ppm) | >240 min |
| Ifosfamide | (50.0 mg/mL 50,000 ppm) | >240 min |
| Mechlorethamine HCL | (1.0 mg/ml 1,000 ppm) | >240 min |
| Methotrexate | (25.0 mg/mL 25,000 ppm) | >240 min |
| Mitomycin C | (0.5 mg/mL 500 ppm) | >240 min |
| Mitoxantrone HCL | (2.0 mg/mL 2,000 ppm | >240 min |
| Paclitaxel | (6.0 mg/mL 6,000 ppm) | >240 min |
| Thiotepa | (10.0 mg/mL 10,000 ppm) | 13.1 |
| Vincristine Sulfate | (1.0 mg/mL 1,000 ppm | >240 min |
| The tested non-chemotherapy drugs are: | | |
| Fentanyl Citrate Injection | (100 mcg/2mL) | >240 min |
| MESNA | (100.0 mg/mL 100,000 ppm) | >240 min |
Note: Carmustine and Thiotepa have extremely low permeation times of 13.2 and 13.1 minutes respectively. Warning: Do Not Use with Carmustine, Thiotepa
>240 min
(1.0 mg/ml 1,000 ppm)
#### E. Device Description:
Trisenox
The Nitrile Disposable Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl (D5000) are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the patient and the examiner. The gloves are blue color, powder free, nitrile ambidextrous gloves are offered in six sizes, extra small, small, medium, large, extra-large and extra extra-large, packed in a paper box.
{7}------------------------------------------------
The gloves are designed and manufactured in accordance with the ASTM D6319 standard and are tested for use with chemotherapy drugs and Fentanyl per ASTM D6978 standard.
### F. Summary of Technological Characteristics
Table 1 General Comparison of Proposed and Predicate Devices
| Device | Proposed Device | Predicate Device | Result |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K # | K233990 | K223903 | - |
| Product Name | Nitrile Disposable Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl (D5000) | Non-Sterile, Single use, Powder-free examination glove, Blue, tested for use with Chemotherapy drugs and Fentanyl | - |
| Product Code | LZA (primary), LZC, OPJ, QDO | LZA (primary), LZC, OPJ, QDO | Same |
| Classification | Class I | Class I | Same |
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Same |
| Indications for use | The Nitrile Disposable Examination Glove, Tested for Use with Chemotherapy Drugs and Fentanyl is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.<br>Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978.<br>The tested chemotherapy drugs are:<br>Chemotherapy drug Concentration<br>Minimum Breakthrough Detection Time<br>Bleomycin sulfate (15.0 mg/ml 15,000 ppm) >240 min<br>Carboplatin (10.0 mg/ml 10,000 ppm) > 240 min<br>Carmustine (BCNU) (3.3mg/mL 3,300 ppm) 13.2<br>Cisplatin (1.0 mg/mL 1,000 ppm) >240 min<br>Cyclophosphamide (20.0 mg/mL 20,000 ppm) >240 min<br>Cytarabine (Cytosine) (100.0 mg/mL 100,000 ppm) >240 min<br>Dacarbazine (10.0 mg/mL 10,000 ppm) >240 min<br>Doxorubicin HCl (2.0 mg/mL 2,000 | This device is an ambidextrous patient examination glove that is a non-sterile, single use, disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between patient and examiner.<br>The tested chemotherapy drugs are: Carmustine (BCNU) (3.3 mg/ml). Permeation time: Carmustine (BCNU) has extremely low permeation times of 14.7 minutes.<br>Cisplatin (1.0 mg/ml). Permeation time: no breakthrough up to 240 minutes<br>Cyclophosphamide (Cytoxan) (20.0 mg/ml). Permeation time: no breakthrough up to 240 minutes<br>Cytarabine (100 mg/ml). Permeation time: no breakthrough up to 240 minutes Dacarbazine (DTIC) (10.0 mg/ml). Permeation time: no breakthrough up to 240 minutes<br>Doxorubicin Hydrochloride (2.0 mg/ml). Permeation time: no breakthrough un | Similar<br>The proposed device and the predicate device have same indications, only minor differences on different drugs and the penetration time of carmustine and thiotepa between the proposed device and the predicate device, which will not raise any new safety or performance |
| | | | |
| | > 240 min | to 240 minutes | concerns. |
| | Etoposide (Toposar) (20.0 mg/mL<br>20,000 ppm) > 240 min | Etoposide (20.0 mg/ml). Permeation<br>time: no breakthrough up to 240<br>minutes | |
| | Fluorouracil (5 Flu) (50.0 mg/mL<br>50,000 ppm) >240 min | Fluorouracil (50.0 mg/ml).<br>Permeation time: no breakthrough up<br>to 240 minutes | |
| | Idarubicin (1.0 mg/ml 1,000 ppm) ><br>240 min | | |
| | Ifosfamide (50.0 mg/mL 50,000 ppm)<br>> 240 min | Ifosfamide (50.0 mg/ml). Permeation<br>time: no breakthrough up to 240<br>minutes | |
| | Mechlorethamine HCL (1.0 mg/ml 1,000<br>ppm) > 240 min | Methotrexate (25.0<br>mg/ml). Permeation time: no<br>breakthrough up to 240 minutes | |
| | MESNA (100.0 mg/mL 100,000 ppm) ><br>240 min | | |
| | Methotrexate (25.0 mg/mL 25,000<br>ppm) >240 min | Mitomycin C (0.5 mg/ml). Permeation<br>time: no breakthrough up to 240<br>minutes | |
| | Mitomycin C (0.5 mg/mL 500 ppm) ><br>240 min | | |
| | Mitoxantrone HCL (2.0 mg/mL 2,000 ppm<br>> 240 min | Mitoxantrone (2.0 mg/ml).<br>Permeation time: no breakthrough up<br>to 240 minutes | |
| | Paclitaxel (6.0 mg/mL 6,000 ppm) ><br>240 min | Paclitaxel (Taxol) (6.0<br>mg/ml). Permeation time: no<br>breakthrough up to 240 minutes | |
| | Thiotepa (10.0 mg/mL 10,000 ppm)<br>13.1 | Thiotepa (10.0 mg/ml). Permeation<br>time: Thiotepa has extremely low<br>permeation times of 13.6 minutes | |
| | Trisenox (1.0 mg/ml 1,000 ppm) > 240<br>min | Vincristine Sulfate (1.0 mg/ml).<br>Permeation time: no Breakthrough up<br>to 240 minutes. | |
| | Vincristine Sulfate (1.0 mg/mL 1,000 ppm<br>>240 min | | |
| | Fentanyl Citrate Injection (100 mcg/2mL) ><br>240 min | The tested Opioid is:<br>Fentanyl Citrate Injection<br>(100mcg/2mL). Permeation: no<br>breakthrough up to 240 minutes | |
| | Note: Carmustine and Thiotepa have<br>extremely low permeation times of 13.2<br>and 13.1 minutes respectively. | Please note that the following drugs<br>have extremely low permeation times:<br>Carmustine: 14.7 minutes<br>Thiotepa: 13.6 minutes<br>Warning: DO NOT USE WITH<br>CARMUSTINE OR THIOTEPA | |
| Powder free | Yes | Yes | Same |
| Design<br>feature | Ambidextrous | Ambidextrous | Same |
| Material | Nitrile | Nitrile | Same |
| Color | Blue | Blue | Same |
| OTC use | Yes | Yes | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Use | Singe use | Single use | Same |
| Labeling claim | Tested for Use with Chemotherapy Drugs<br>and Fentanyl | Tested for Use with Chemotherapy Drugs<br>and Fentanyl | Same |
{8}------------------------------------------------
{9}------------------------------------------------
# Table 2 Device dimension comparison
| Technological characteristics | Proposed<br>Device<br>(K233990) | Predicate<br>Device<br>(K223903) | Device | Result |
|-------------------------------|---------------------------------|----------------------------------|--------|-----------------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.