Non-Sterile, Single use, Powder-free examination glove, Blue, tested for use with Chemotherapy drugs and Fentanyl SMALL SIZE (NGPF102); MEDIUM SIZE (NGPF103); LARGE SIZE (NGPF104); X-LARGE SIZE (NGPF105)
K223903 · Semperit Investments Asia Pte , Ltd. · LZA · Jan 27, 2023 · General Hospital
Device Facts
Record ID
K223903
Device Name
Non-Sterile, Single use, Powder-free examination glove, Blue, tested for use with Chemotherapy drugs and Fentanyl SMALL SIZE (NGPF102); MEDIUM SIZE (NGPF103); LARGE SIZE (NGPF104); X-LARGE SIZE (NGPF105)
Applicant
Semperit Investments Asia Pte , Ltd.
Product Code
LZA · General Hospital
Decision Date
Jan 27, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
This device is an ambidextrous patient examination glove that is a non-sterile, single use, disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Ambidextrous, non-sterile, single-use, powder-free nitrile examination glove; worn on examiner's hand/finger to provide barrier against infectious materials and contaminants. Tested for permeation resistance against chemotherapy drugs and Fentanyl per ASTM D6978. Available in S, M, L, XL sizes. Used in clinical settings by healthcare providers. Provides protection during patient examination and handling of hazardous drugs. Does not include software or algorithms.
Clinical Evidence
No clinical data. Bench testing only. Device demonstrated compliance with ASTM D6319 (dimensions, physical properties), ASTM D5151 (freedom from holes), ASTM D6124 (powder residue), ASTM D6978 (chemotherapy/Fentanyl permeation), and ISO 10993-10 (irritation/sensitization).
Technological Characteristics
Nitrile material; ambidextrous; powder-free; blue color. Complies with ASTM D6319, D5151, D6124, D6978, and ISO 10993-10. No energy source, connectivity, or software.
Indications for Use
Indicated for use as a patient examination glove to prevent contamination between patient and examiner. Tested for use with specific chemotherapy drugs and Fentanyl. Warning: Do not use with Carmustine or Thiotepa.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Non-Sterile, Single use, Powder-free examination glove, Blue, tested for use with Chemotherapy drugs (K171378)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 27, 2023
Semperit Investments Asia PTE. LTD. % Jay Mansour Regulatory Consultant / Principal Mansour Consulting LLC 845 Aronson Lake Court Roswell, Georgia 30075
Re: K223903
Trade/Device Name: Non-Sterile, Single use, Powder-free examination glove, Blue, tested for use with Chemotherapy drugs and Fentanyl SMALL SIZE (NGPF102); MEDIUM SIZE (NGPF103); LARGE SIZE (NGPF104); X-LARGE SIZE (NGPF105) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ, ODO Dated: December 27, 2022 Received: December 28, 2022
Dear Jay Mansour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Allan G
For Bifeng Qian M.D., Ph. D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223903
# Device Name Non-Sterile, Single use, Powder-free examination glove, Blue, tested for use with Chemotherapy drugs and Fentanyl SMALL SIZE (NGPF102); MEDIUM SIZE (NGPF103); LARGE SIZE (NGPF104); X-LARGE SIZE Indications for Use (Describe) This device is an ambidextrous patient examination glove that is a non-sterile, single use, disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between patient and examiner. The tested chemotherapy drugs are: Carmustine (BCNU) (3.3 mg/ml). Permeation time: Carmustine (BCNU) has extremely low permeation times of 14.7 minutes. Cisplatin (1.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Cyclophosphamide (Cytoxan) (20.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Cytarabine (100 mg/ml). Permeation time: no breakthrough up to 240 minutes Dacarbazine (DTIC) (10.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Doxorubicin Hydrochloride (2.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Etoposide (20.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Fluorouracil (50.0 mg/ml). Permeation time: no breakthrough up to 240 minutes lfosfamide (50.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Methotrexate (25.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Mitomycin C (0.5 mg/ml). Permeation time: no breakthrough up to 240 minutes Mitoxantrone (2.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Paclitaxel (Taxol) (6.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Thiotepa (10.0 mg/ml). Permeation time: Thiotepa has extremely low permeation times of 13.6 minutes Vincristine Sulfate (1.0 mg/ml). Permeation time: no breakthrough up to 240 minutes The tested Opioid is: Fentanyl Citrate Injection (100mcg/2mL). Permeation: no breakthrough up to 240 minutes Please note that the following drugs have extremely low permeation times:
Carmustine: 14.7 minutes Thiotepa: 13.6 minutes
Warning: DO NOT USE WITH CARMUSTINE OR THIOTEPA
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K223903- 510(k) summary
# I Submitter
Device submitter: SEMPERIT INVESTMENTS ASIA PTE. LTD. 8 Jurong Town Hall Road, #29-03 To 06 The JTC Summit Singapore 609434
- Contact person: Jay Mansour Mansour Consulting LLC Regulatory consultant Phone (678) 908-8180 Email jay@mansourconsulting.com
Date of preparation: January 27, 2023
#### II Proposed device
| 510(k) Number: | K223903 |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade/Device Name: | Non-Sterile, Single use, Powder-free examination glove, Blue, tested for use with Chemotherapy drugs and Fentanyl<br>SMALL SIZE (NGPF102); MEDIUM SIZE (NGPF103); LARGE SIZE (NGPF104)<br>X-LARGE SIZE (NGPF105) |
| Regulation number: | 21 CFR 880.6250 |
| Regulation name: | Non-powdered Patient Examination Glove |
| Regulatory class: | Class I |
| Product code: | LZA (primary), LZC, OPJ, QDO |
| Review panel: | General Hospital |
# III Predicate devices
| 510(k) Number: | K171378 |
|--------------------|------------------------------------------------------------------------------------------------------|
| Trade/Device name: | Non-Sterile, Single use, Powder-free examination glove, Blue, tested for use with Chemotherapy drugs |
| Regulation number: | 21 CFR 880.6250 |
| Regulation name: | Non-powdered Patient Examination Glove |
| Regulatory class: | Class I |
{5}------------------------------------------------
#### Product code: LZA (primary), LZC SEMPERIT INVESTMENTS ASIA PTE. LTD. Manufacturer:
# IV Device description
The Non-Sterile, Single use, Powder-free Examination glove, Blue, tested for use with Chemotherapy Drugs and fentanyl is provided in blue. It meets all the requirements of ASTM D6319. It is a medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves were tested for use with chemotherapy drugs as well as Fentanyl, per ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drug. It can be available in 4 specifications: S, M, L and XL.
# V Indications for use
This device is an ambidextrous patient examination glove that is a non-sterile, single use, disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between patient and examiner.
The tested chemotherapy drugs are:
Carmustine (BCNU) (3.3 mg/ml). Permeation time: Carmustine (BCNU) has extremely low permeation times of 14.7 minutes.
Cisplatin (1.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Cyclophosphamide (Cytoxan) (20.0 mg/ml). Permeation time: no breakthrough up to 240 minutes
Cytarabine (100 mg/ml). Permeation time: no breakthrough up to 240 minutes Dacarbazine (DTIC) (10.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Doxorubicin Hydrochloride (2.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Etoposide (20.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Fluorouracil (50.0 mg/ml). Permeation time: no breakthrough up to 240 minutes lfosfamide (50.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Methotrexate (25.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Mitomycin C (0.5 mg/ml). Permeation time: no breakthrough up to 240 minutes Mitoxantrone (2.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Paclitaxel (Taxol) (6.0 mg/ml). Permeation time: no breakthrough up to 240 minutes Thiotepa (10.0 mg/ml). Permeation time: Thiotepa has extremely low permeation times of 13.6 minutes
Vincristine Sulfate (1.0 mg/ml). Permeation time: no breakthrough up to 240 minutes The tested Opioid is:
Fentanyl Citrate Injection (100mcg/2mL). Permeation: no breakthrough up to 240 minutes
Please note that the following drugs have extremely low permeation times:
Carmustine: 14.7 minutes
Thiotepa: 13.6 minutes
Warning: DO NOT USE WITH CARMUSTINE OR THIOTEPA
{6}------------------------------------------------
# VI Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was same/similar to the predicate devices. The test results demonstrated that the proposed device complies with the following standards:
- > ASTM D5151 Standard Test Method for Detection of Holes in Medical Gloves
- > ASTM D6124 Standard Test Method for Residual Powder on Medical Gloves
- > ASTM D6319 Standard Specification for Nitrile Examination Gloves for Medical Application
> ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs
> ISO 10993-10: 2010 Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization
{7}------------------------------------------------
| Test<br>method | Purpose | Acceptance Criteria | Result | |
|-----------------|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|---------------------------------------------------------------|
| ASTM<br>D6319 | Dimensions | Extra-Small | Length: ≥220mm;<br>Width: 70±10 mm | |
| | | Small | Length: ≥220mm;<br>Width: 80±10mm | |
| | | Medium | Length: ≥230mm;<br>Width: 95±10mm | Pass |
| | | Large | Length: ≥230mm;<br>Width: 110±10mm | |
| | | Extra- Large | Length: ≥230mm;<br>Width: 120±10mm | |
| | | Thickness (mm):<br>Finger: ≥0.05<br>Palm: ≥0.05 | | Pass |
| | Physical properties | Before<br>Aging | Tensile<br>Strength ≥14MPa<br>Ultimate<br>Elongation ≥500% | Pass |
| | | After<br>Aging | Tensile<br>Strength ≥14MPa<br>Ultimate<br>Elongation ≥400% | Pass |
| | | | | |
| ASTM<br>D5151 | Freedom from holes | Shall comply with freedom from holes (AQL = 2.5)<br>when tested in accordance with ASTM D5151. | | Pass |
| ASTM<br>D6124 | Powder-free Residue | Have a powder residue limit of 2.0 mg in<br>accordance with ASTM D6124. | | Pass |
| ISO<br>10993-10 | To determine if the<br>finished device<br>material is an irritant. | Non-irritating | | Under the conditions of<br>the study not an irritant/<br>Pass |
| ISO<br>10993-10 | To determine if the<br>finished device<br>material is a sensitizer. | Non-sensitizing | | Under conditions of the<br>study, not a sensitizer. /<br>Pass |
| | Resistance to<br>permeation by<br>Fentanyl | No breakthrough when tested for permeation with<br>Fentanyl in accordance with ASTM D6978. | | Pass |
| ASTM<br>D6978 | Resistance to<br>Permeation by<br>Chemotherapy Drugs | Covered in predicate device 510(k) K171378 for<br>Carmustine (BCNU) (3.3 mg/ml)<br>Cisplatin (1.0 mg/ml)<br>Cyclophosphamide (Cytoxan) (20.0 mg/ml)<br>Cytarabine (100 mg/ml)<br>Dacarbazine (DTIC) (10.0 mg/ml)<br>Doxorubicin Hydrochloride (2.0 mg/ml)<br>Etoposide (20.0 mg/ml)<br>Fluorouracil (50.0 mg/ml)<br>Ifosfamide (50.0 mg/ml)<br>Methotrexate (25.0 mg/ml)<br>Mitomycin C (0.5 mg/ml)<br>Mitoxantrone (2.0 mg/ml)<br>Paclitaxel (Taxol) (6.0 mg/ml) | | Pass |
Table 1 Summary of Non-Clinical Performance Testing
{8}------------------------------------------------
| | Vincristine Sulfate (1.0 mg/ml)<br>Thiotepa (10.0 mg/ml) | |
|--|----------------------------------------------------------|--|
| | | |
{9}------------------------------------------------
# VII Clinical Test Conclusion
No clinical study is included in this submission.
# VIII Summary of Technological characteristics
| | Subject device | Predicate device | Comments |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | SEMPERIT INVESTMENTS ASIA PTE. LTD. | SEMPERIT INVESTMENTS ASIA PTE. LTD. | N/A |
| 510(k) number | TBD | K171378 | N/A |
| Device trade or<br>proprietary name | Non-Sterile, Single use, Powder-free<br>examination glove, Blue, tested for use<br>with Chemotherapy drugs and fentanyl | Non-Sterile, Powder-free Examination<br>glove, Blue, tested for use with<br>Chemotherapy Drugs | Different: except the addition of Fentanyl for<br>subject device, it is identical |
| Device Classification<br>Name/ Regulation number | Patient Examination Glove<br>21 CFR Part 880.6250 | Patient Examination Glove<br>21 CFR Part 880.6250 | Identical |
| Product Code | LZA (primary), LZC, OPJ, QDO | LZA (primary), LZC | Different: added two product codes |
| Indications for use | This device is an ambidextrous patient<br>examination glove that is a non-sterile,<br>single use, disposable device intended<br>for medical purposes, worn on the<br>examiner's hand or finger to prevent<br>contamination between patient and<br>examiner.<br>The tested chemotherapy drugs are:<br>Carmustine (BCNU) (3.3 mg/ml).<br>Permeation time: Carmustine (BCNU)<br>has extremely low permeation times of<br>14.7 minutes.<br>Cisplatin (1.0 mg/ml). Permeation time:<br>no breakthrough up to 240 minutes<br>Cyclophosphamide (Cytoxan) (20.0 mg/<br>ml). Permeation time: no breakthrough<br>up to 240 minutes Cytarabine (100 mg/<br>ml). Permeation time: no breakthrough<br>up to 240 minutes Dacarbazine (DTIC)<br>(10.0 mg/ml). Permeation time: no<br>breakthrough up to 240 minutes<br>Doxorubicin Hydrochloride (2.0 mg/ml).<br>Permeation time: no breakthrough up to<br>240 minutes<br>Etoposide (20.0 mg/ml). Permeation<br>time: no breakthrough up to 240<br>minutes Fluorouracil (50.0 mg/ml).<br>Permeation time: no breakthrough up to<br> | This device is an ambidextrous patient<br>examination glove that is a non-sterile,<br>single use, disposable device intended<br>for medical purposes, worn on the<br>examiner's hand or finger to prevent<br>contamination between patient and<br>examiner. The tested chemotherapy<br>drugs are:<br>Carmustine (BCNU) (3.3 mg/ml).<br>Permeation time: Carmustine (BCNU) has<br>extremely low permeation times of 14.7<br>minutes.<br>Cisplatin (1.0 mg/ml). Permeation time:<br>no breakthrough up to 240 minutes<br>Cyclophosphamide (Cytoxan) (20.0<br>mg/ml). Permeation time: no<br>breakthrough up to 240 minutes<br>Cytarabine (100 mg/ml). Permeation<br>time: no breakthrough up to 240 minutes<br>Dacarbazine (DTIC) (10.0 mg/ml).<br>Permeation time: no breakthrough up to<br>240 minutes<br>Doxorubicin Hydrochloride (2.0 mg/ml).<br>Permeation time: no breakthrough up to<br>240 minutes<br>Etoposide (20.0 mg/ml). Permeation<br>time: no breakthrough up to 240 minutes<br>Fluorouracil (50.0 mg/ml). Permeation<br>time: no breakthrough up to 240 minutes<br>Ifosfamide (50.0 mg/ml). Permeation<br>time: no breakthrough up to 240 minutes<br>Methotrexate (25.0 mg/ml). Permeation<br>time: no breakthrough up to 240 minutes<br>Mitomycin C (0.5 mg/ml). Permeation<br>time: no breakthrough up to 240 minutes<br>Mitoxantrone (2.0 mg/ml). Permeation<br>time: no breakthrough up to 240 minutes<br>Paclitaxel (Taxol) (6.0 mg/ml).<br>Permeation time: no breakthrough up to<br>240 minutes<br>Thiotepa (10.0 mg/ml). Permeation time:<br>Thiotepa has extremely low permeation<br>times of 13.6 minutes<br>Vincristine Sulfate (1.0 mg/ml).<br>Permeation time: no Breakthrough up to<br>240 minutes.<br>DO NOT USE WITH CARMUSTINE OR<br>THIOTEPA | Different<br>Subject device includes Fentanyl, while<br>predicate device does not. Other than for<br>Fentanyl, the indications for use are identical.<br>Note: Adding the Fentanyl to the indication<br>does not affect the intended use. The device's<br>intended use is it would be worn on the hand<br>for medical purposes to provide a barrier<br>against potentially infectious materials and<br>other contaminants. Therefore, there are no<br>differences to the intended therapeutic,<br>diagnostic, prosthetic, or surgical use of the<br>device, and therefore the change in the<br>indication for use (by adding Fentanyl) does<br>not affect the safety and effectiveness of the<br>device when used as labeled. |
| Materials | Nitrile | Nitrile | Identical |
| | | | |
| Standards used | 1. ISO 10993-10<br>2. ASTM D6319-19<br>3. ASTM D6124-06 (reapproved 2017)<br>4. ASTM D5151-19<br>5. ASTM D6978-05 (reapproved 2019)<br>6. ISO 2859-1<br>7. ASTM D412-16e1<br>8. ASTM D3767-03 (2020)<br>9. ASTM D573-04 (2019) | 1. ISO 10993-10<br>2. ASTM D6319-10 (reapproved 21015)<br>3. ASTM D6124-06 (2011)<br>4. ASTM D5151-06<br>5. ASTM D6978-05 (re-approved 2013)<br>6. ISO 2859-1<br>7. ASTM D412-16<br>8. ASTM D3767-03 (2014)<br>9. ASTM D573-04 (2015) | Identical, with justification provided below, following the same numbering sequence:<br>1. See biocompatibility below<br>2. The changes did not affect product design, production and compliance<br>(1st change: section 1.5. 2nd change: adding calculation options instead of actual cutting. 3rd change: limitation on accelerated aging option based on natural aging. 4th change: size XXL is introduced in the revised standard BUT it is NOT added in this submission). Recognition number is updated from 6-244 to 6-446 |
| | | | 3. No changes in the standard.<br>Recognition number 6-178 remains the same. |
| | | | 4. The changes did not affect product design, production and compliance<br>(1st change: section 1.7. 2nd change: temperature of the water leak test). Recognition number is updated from 6-175 to 6-424 |
| | | | 5. No change in the standard. Recognition number 6-147 remains the same |
| | | | 6. No change in the standard. Recognition number was updated from 5-62 to 5-88 |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.