K233926 · Mettler Electronics Corporation · IPF · Mar 21, 2024 · Physical Medicine
Device Facts
Record ID
K233926
Device Name
accufit
Applicant
Mettler Electronics Corporation
Product Code
IPF · Physical Medicine
Decision Date
Mar 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The accufit is indicated for use for Relaxation of muscle spasms, Prevention or retardation of disuse atrophy, Increase local blood circulation, Muscle re-education, Maintaining or increasing range of motion, Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, Improvement of abdominal tone, for strengthening of the abdominal muscles, for development of firmer abdomen, and Improvement of muscle tone and firmness, for strengthening muscles in arms, thighs and buttocks areas.
Device Story
Accufit is a powered muscle stimulator console; provides direct electrical stimulation to muscles via IntelliSTIM electrodes; produces muscle contractions for conditioning and therapeutic purposes. Device features Touch LCD/GUI; offers two waveforms (Biphasic, Interferential 4P) and four protocols (twist, hold, grip, tap) plus an automated 'IntelliCycle' program. Used in clinical settings by healthcare providers; console includes emergency stop and key switch. Output intensity is adjustable (0-100 mA). Device supplements existing therapies; does not replace traditional treatment. Benefits include improved muscle tone, firmness, and strength; clinical study demonstrated increased arm strength via dynamometer measurements.
Clinical Evidence
Prospective clinical study; 45 subjects; assessed effect of bioelectric muscle activation (BMA) on arm strength and appearance. Strength measured via dynamometer at baseline, final treatment, and 30/90-day follow-ups. Results showed improved muscle strength from initial to final treatment; no significant adverse events reported.
Technological Characteristics
Powered muscle stimulator; ABS plastic housing; 2 output channels; 100-240VAC power. Waveforms: Biphasic and Interferential (4P). Safety: Type BF classification, automatic over-current/no-load trip, patient interrupt switch. Standards: IEC 60601-1, 60601-1-2, 60601-2-10, ISO 14971, UL 60601, CSA C22.2 No 606.1. Biocompatibility: ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), ISO 10993-23 (irritation).
Indications for Use
Indicated for patients requiring muscle stimulation for relaxation of spasms, prevention of disuse atrophy, increased local blood circulation, muscle re-education, range of motion maintenance, post-surgical calf stimulation to prevent venous thrombosis, and strengthening/toning of abdominal, arm, thigh, and buttock muscles.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 21, 2024
Mettler Electronics Corporation Kevin O'Connell VP Global Regulatory Affairs Lutronic Corporation Lutronic Center 219, Sowon-Ro Deogyang-Gu, Goyang-Si 410220 South Korea
Re: K233926
Trade/Device Name: accufit Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, NGX Dated: March 7, 2024 Received: March 7, 2024
## Dear Kevin O'Connell:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Heather L. Dean -S
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233926
Device Name accufit
Indications for Use (Describe)
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- Increase local blood circulation
- Muscle re-education
- Maintaining or increasing range of motion
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Improvement of abdominal tone, for strengthening of the abdominal muscles, for development of firmer abdomen.
- Improvement of muscle tone and firmness, for strengthening muscles in arms, thighs and buttocks areas.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
## K233926
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| 807.92(a)(1) - Submitter Information | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Mettler Electronics Corporation |
| Address | 1333 South Claudina Street<br>Anaheim, CA USA 92805 |
| Phone number | 978-888-1426 |
| Fax number | N/A |
| Name of contact person | An Le<br>QS/RA Manager, Mettler Electronics Corporation<br>1333 South Claudina Street<br>Anaheim, CA 92805 |
| Date prepared | March 19, 2024 |
| 807.92(a)(2) - Name of device | |
| Trade or proprietary name | accufit |
| Common or usual name | Powered Muscle Stimulator |
| Classification name | Stimulator, Muscle, Powered, For Muscle Conditioning |
| Classification panel | Physical Medicine |
| Regulation | 21 CFR 890.5850 |
| Product Code(s) | IPF, NGX |
| 807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed | |
| Mettler Electronics Corporation K223802 (10/06/2023) Predicate<br>Johari Digital Healthcare Limited K192039 (03/29/2020) Reference<br>Sport- Elec K092476 (05/07/2010) Reference | |
| 807.92(a)(4) - Device description | |
| The principle of operation of the accufit device is to provide direct<br>electrical stimulation of muscles which produces muscle contractions in<br>and around the area of the treatment electrodes.<br><br>The accufit generates electrical stimulation to contract the muscles to<br>achieve the desired intended use using unique IntelliPhase waveforms<br>and IntelliSTIM electrodes. It offers two unique IntelliPhase<br>waveforms, Biphasic and Interferential (4P), that are designed to<br>optimize muscle re-education and activation for patients. The four<br>IntelliPhase waveform protocols are twist, hold, grip, and tap. Each | |
| protocol may be used for 15, 30, or 45 minutes. Not only are there four<br>IntelliPhase protocols, but there is also a fully automated protocol<br>(IntelliCycle) that integrates all the above listed waveforms in a<br>comprehensive, interlaced series. This engages all the targeted muscles<br>in a continuous program to maximize muscle effects with the touch of a<br>button. | |
| accufit is both the proprietary-trade name and generic name. accufit is a<br>supplement to existing treatments and does not replace any traditional<br>treatment or therapy. | |
| The system console is the heart of the accufit and contains the Touch<br>LCD/GUI, electrodes, and power supply module. The main console<br>also includes a key switch used to turn the power on and off, and an<br>emergency stop push button that quickly de-energizes the system in<br>emergency situations. There are 4 casters in the console base that can<br>be used when moving the system. | |
| 807.92(a)(5) Intended use of the device | |
| Indications for use | The accufit is indicated for use for Relaxation of muscle spasms,<br>Prevention or retardation of disuse atrophy, Increase local blood<br>circulation, Muscle re-education, Maintaining or increasing range of<br>motion, Immediate post-surgical stimulation of calf muscles to prevent<br>venous thrombosis, Improvement of abdominal tone, for strengthening<br>of the abdominal muscles, for development of firmer abdomen, and<br>Improvement of muscle tone and firmness, for strengthening muscles in<br>arms, thighs and buttocks areas. |
{5}------------------------------------------------
{6}------------------------------------------------
| 807.92(a)(6) Summary of the technological characteristics of the device compared to the predicate | | | | | |
|---------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| Model Name | accufit | accufit | Torc Body | Body Control<br>System "4M" | Notes |
| 510(k) # | Subject device | K223802 | K192039 | K092476 | |
| Manufacturer | Mettler | Mettler | Johari Digital<br>Healthcare Limited | Sport-Elec S.A | |
| Indications for Use | • Relaxation of muscle<br>spasms<br>• Prevention or<br>retardation of disuse<br>atrophy<br>• Increase local blood<br>circulation<br>• Muscle re-education<br>• Maintaining or<br>increasing range of<br>motion<br>• Immediate post-<br>surgical stimulation of<br>calf muscles to prevent<br>venous thrombosis<br>• Improvement of<br>abdominal tone, for<br>strengthening of the<br>abdominal muscles, for<br>development of firmer<br>abdomen.<br>Improvement of muscle<br>tone and firmness, for<br>strengthening muscles in<br>arms, thighs and<br>buttocks areas. | • Relaxation of muscle<br>spasms<br>• Prevention or<br>retardation of disuse<br>atrophy<br>• Increase local blood<br>circulation<br>• Muscle re-education<br>• Maintaining or<br>increasing range of<br>motion<br>• Immediate post-<br>surgical stimulation of<br>calf muscles to<br>prevent venous<br>thrombosis | • Improvement of<br>abdominal tone, for<br>strengthening of the<br>abdominal muscles,<br>for development of<br>firmer abdomen.<br>• Strengthening, toning<br>and firming of<br>buttocks & thighs. | Improvement of<br>muscle tone and<br>firmness, for<br>strengthening<br>muscles in arms,<br>abdomen, thighs and<br>buttocks areas. | Identical to predicate<br>plus reference<br>devices |
| Classification | 890-5850 - IPF, NGX | 890-5850 - IPF | 890-5850 - NGX | 890-5850 - NGX | Identical to<br>predicate plus<br>reference devices |
| Power Source | 100-240VAC ,<br>50/60Hz, 1.0A (Fuse:<br>250V / 6.3A) | 100-240VAC ,<br>50/60Hz, 1.0A<br>(Fuse: 250V / 6.3A) | 100-240AC,<br>50/60Hz, 75VA | Battery powered | Identical |
| Therapeutic<br>Modality | Electrical muscle<br>stimulator | Electrical muscle<br>stimulator | Electrical muscle<br>stimulator | Electrical muscle<br>stimulator | Identical |
| Treatment Output<br>Mode | Electrode | Electrode | Electrode | Electrode | Identical |
| Method of Line<br>Current<br>Isolation | Double Insulated Wire<br>Non- Conductive<br>Enclosure | Double Insulated<br>Wire Non-<br>Conductive<br>Enclosure | (a) AC Power supply<br>is converted to DC<br>Power supply<br>through a medical<br>grade PSU.<br>(b) Isolation thru<br>transformer in<br>between device and<br>patient | Not available | Same as predicate |
| Measured Patient<br>Leakage: | | | | | |
| Normal Condition<br>(μ Α) | Less than 100μA | Less than 100μA | Less than 100μA | Not available | Identical |
| Single Fault<br>Condition (μΑ) | Less than 100μA | Less than 100μA | Less than 300μA | Not available | Similar |
| Number of Output<br>Modes | 4 | 4 | 3 | Not available | 1 additional mode |
| Number of Output<br>Channels | 2 | 2 | 2 | 2 | Identical |
| Regulated Current<br>or Voltage | Current = 100mA<br>maximum | Current = 100mA<br>maximum | 103.2 mA pp @ 500<br>Ohm 31.2 mA pp @<br>2K Ohm 6.6 mA pp<br>@ 10K Ohm | Not available | Identical when<br>compared at 500<br>Ohm |
| Pulse Intensities | Adjustable, 0-100 mA<br>peak into 500ohm load<br>each Electrode channel | Adjustable, 0-100<br>mA peak into<br>500ohm load each<br>Electrode channel | Not available | Not available | Same as predicate |
| Software/Firmware/<br>Microprocessor<br>Control | yes | yes | yes | Not available | Identical |
| Automatic Over<br>Current Trip? | yes | yes | yes | Not available | Identical |
| Automatic No Load<br>Trip? | yes | yes | no | Not available | Same as predicate |
| Automatic Shut<br>Off? | yes | yes | yes | Not available | Identical |
| User Override<br>control? | Patient Interrupt (Stop)<br>Switch | Patient Interrupt<br>(Stop) Switch | yes | Not available | Identical |
| Indicator Display: | yes | yes | yes | Not available | Identical |
| On/ Off Status? | yes | yes | yes | Not available | Identical |
| Voltage/ Current<br>Level? | yes | yes | yes | Not available | Identical |
| Timer range | 15, 30, 45 mins | 15, 30, 45 mins | 1 – 60 Minutes In<br>step of 1 minute | Not available | Similar |
| Compliance:<br>Voluntary<br>Standards | IEC 60601-1; 60601-<br>1-2; 60601-2-10, ISO<br>14971, UL 60601,<br>CSA C22.2 No 606.1 | IEC 60601-1;<br>60601-1-2; 60601-2-<br>10, ISO 14971, UL<br>60601, CSA C22.2<br>No 606.1 | YES IEC 60601-1,<br>IEC 60601-1-<br>2,IEC60601-2-10,<br>and ISO14971 | Not available | Similar |
| Compliance: 21<br>CFR 898 | yes | yes | yes | Not available | Identical |
| Weight | 27 Kg | 27 Kg | 32.66 Kgs | Not available | Similar |
| Dimensions (mm)<br>W x L x H | 452 x 582 x 1101 | 452 x 582 x 1101 | 444 x 356 x 1016 | 84x126x30 | Similar |
| Housing, Materials<br>and Construction | ABS plastic | ABS plastic | ABS Plastic Body | Not available | Identical |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| Medical Equipment<br>Classification…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.