K192039 · Johari Digital Healthcare Limited · NGX · Mar 29, 2020 · Physical Medicine
Device Facts
Record ID
K192039
Device Name
Torc Body
Applicant
Johari Digital Healthcare Limited
Product Code
NGX · Physical Medicine
Decision Date
Mar 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Torc Body is indicated to be used for: - Improvement of abdominal tone, for strengthening of the abdominal muscles, for development of firmer abdomen. - Strengthening, toning and firming of buttocks & thighs.
Device Story
Torc Body is a powered muscle stimulator used by clinicians in a clinical setting. The device delivers electrical stimulation via self-adhesive electrodes to contract muscles rhythmically. It features three modes: Classic Vector (warm-up/circulation), Power Sequence (strength/endurance), and Dynamic Rotation (toning/firming). The device is housed in a wheeled console with a 12-inch GUI for clinician operation. Clinicians adjust stimulation intensity to achieve muscle strengthening, firming, and toning. The device operates on an internal medical-grade power supply. By inducing rhythmic muscle contractions, the device assists in improving muscle tone and firmness, providing a non-invasive method for muscle conditioning.
Clinical Evidence
Bench testing only. Performance data demonstrated compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 14971, ISO 13485, IEC 62366, ISO 15233, and ISO 10993-1. Testing confirmed the modified device meets predetermined acceptance criteria for reliability and safety.
Technological Characteristics
Powered muscle stimulator; ABS plastic housing; 12-inch LCD GUI; internal medical-grade power supply (100-240VAC); 2 output channels; symmetrical biphasic waveform; transformer-based channel isolation; 2XMOPP isolation; prescription-use; wheeled console form factor. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, and ISO 14971.
Indications for Use
Indicated for adults for improvement of abdominal tone, strengthening of abdominal muscles, development of firmer abdomen, and strengthening, toning, and firming of buttocks and thighs.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 29, 2020
Johari Digital Healthcare Limited Pooja Johari Founder and Director Marketing G-582, 584 EPIP, Boranada Jodhpur, Rajsthan 342008 India
Re: K192039
Trade/Device Name: Torc Body Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: December 24, 2019 Received: December 30, 2019
Dear Pooja Johari:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192039
Device Name Torc Body
Indications for Use (Describe)
Torc Body is indicated to be used for:
- Improvement of abdominal tone, for strengthening of the abdominal muscles, for development of firmer abdomen.
- · Strengthening, toning and firming of buttocks & thighs.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (k) Summary As required by 21 CFR 807.92(c)
| Device Name | Torc Body |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Submitters name/<br>contact details | Nisha Johari<br>Johari Digital Healthcare Ltd<br>G-582, 584 EPIP, Boranada,<br>Jodhpur - India - 342008<br>Contact number: + 1-888-412-3160 |
| Summary Preparation Date | 20-Mar-2020 |
| Device Trade Name | Torc Body |
| Device / Proprietary Name | Torc Body |
| Common Name | Stimulator, Muscle, Powered, for Muscle Conditioning |
| Classification Name | Powered Muscle Stimulator,<br>Stimulator, Muscle, Powered, For Muscle Conditioning |
| Classification Regulation | 21 CFR 890.5850, Class II |
| Classification Product Code | NGX |
# Legally marketed Predicate Device
| Device Trade Name | Torc Body |
|--------------------------|-----------------------------------------------------------------------------------------|
| Classification Name | Powered Muscle Stimulator,<br>Stimulator, Muscle, Powered, For Muscle Conditioning |
| 510(K) No | K131291 |
| Address and Registration | Johari Digital Healthcare Ltd<br>G-582, 584 EPIP, Boranada,<br>Jodhpur - India - 342008 |
| Contact Person: | Nisha Johari |
| Contact Number: | +91-9829954228 |
| FDA Registration | 8040537 |
{4}------------------------------------------------
#### Device Description
Torc Body, the Powered Muscle Stimulator, is intended to improve tone and strengthen the muscles of the abdomen, buttocks, and thighs and to firm the abdomen muscles.
The modified Torc Body, as described in this submission, is an upgrade from its previous version; which includes User- Friendly GUI, Internal power supply, and a wheeled console allowing clinicians to navigate and operate the device effortlessly. We intend to market this device to clinicians, as a prescription device. For ease of use and operation for the clinicians, we have 3 modes (suggestive):
- . CLASSIC VECTOR is only necessary for non-active patients. It creates a twisting motion to warm up, build blood circulation and slowly build a tolerance to muscle contractions.
- POWER SEQUENCE contracts the muscles, holds it and then relaxes it to increase strength and muscle endurance.
- DYNAMIC ROTATION - Fast, sequential contractions of the muscles which leads to toning and firming.
The modified device generates electrical stimulation to contract the muscles rhythmically to achieve the intended use of strengthening, firming, and toning the muscles of the abdomen, thighs, and buttocks
While redesigning the Torc Body, complete care and considerable measures have been taken to retain its safety and effectiveness. Torc Body device complies with voluntary standards.
#### Intended Use
Torc Body is indicated to be used for:
- . Improvement of abdominal tone, for strengthening of the abdominal muscles, for development of firmer abdomen.
- Strengthening, toning and firming of buttocks & thighs.
The device safety and efficacy was demonstrated by performance data and comparison of technical characteristics between the modified device and the predicate device.
#### Substantial Equivalence:
The intended use and indication for use of the modified device are the same as the predicate device (Torc Body , K131291). The modification made to the device include:
- User Friendly GUI- LCD with wide angle view that significantly facilitates the use of the device.
- Power Supply- Internal power supply in place of external AC/DC adaptor and rechargeable battery. Wheeled Console- Allows clinicians to navigate and operate the device effortlessly
- Prescriptive device- we intend to market this device to clinicians, we propose making it a prescription device.
{5}------------------------------------------------
### Comparison of Technological Characteristics
A comparison below given below identifies all the changes between the modified and the predicate device:
| Device Features | Modified Device | Predicate Device | Comparison |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | K192039 | K131291 | N/A |
| Device Name,<br>Model | Torc Body | Torc Body | Design revision updated<br>to V2.0 |
| Manufactured For | 7E LLC | 7E LLC | Identical |
| Manufactured By | Johari Digital<br>Healthcare Ltd. | Johari Digital<br>Healthcare Ltd. | Identical |
| Classification Name | Powered muscle<br>stimulator | Powered muscle<br>stimulator | Identical |
| Product Code | NGX | NGX | Identical |
| Regulation Number | 21 CFR 890.5850 | 21 CFR 890.5850 | Identical |
| Panel | Physical Medicine | Physical Medicine | Identical |
| Class | Class II | Class II | Identical |
| Prescription/OTC | Prescription | OTC | Modified;<br>We intend to market this<br>device to clinicians and<br>propose making it a<br>prescription device. |
| Indication for use | Improvement of<br>abdominal tone,<br>for strengthening<br>of the abdominal<br>muscles, for<br>development of<br>firmer abdomen.<br>Strengthening,<br>toning and<br>firming of<br>buttocks &<br>thighs. | Improvement<br>of abdominal<br>tone, for<br>strengthening<br>of the<br>abdominal<br>muscles, for<br>development<br>of firmer<br>abdomen.<br>Strengthening,<br>toning and<br>firming of<br>buttocks &<br>thighs. | Identical |
| Target population | It is to be used by<br>adults only. | It is to be used by<br>adults only. | Identical |
| Power source | 100-240AC,<br>50/60Hz, 75VA | AC/DC Adaptor &<br>24V DC Ni MH<br>Rechargeable<br>Battery | Modified;<br>The modified device uses<br>an internal medical grade |
| Method of Line<br>Current Isolation | (a) AC Power supply<br>is converted to DC<br>Power supply<br>through a medical<br>grade PSU, which<br>has isolation of<br>2XMOPP (IEC60601-<br>1)<br>(b) Isolation thru<br>transformer in<br>between device and<br>patient | (a) AC Power<br>supply is<br>converted to DC<br>Power supply<br>through Adaptor,<br>which uses a<br>transformer in<br>between the Low<br>and High voltage<br>sides. Hence there<br>is insulation of<br>mains from<br>Circuitry.<br>(b) The device<br>operates on Ni-Mh<br>rechargeable<br>battery | power supply, in place of<br>AC/DC adaptor and<br>rechargeable battery<br>used in the predicate<br>device. The internal<br>power supply complies<br>with all necessary<br>voluntary standards.<br>Modified;<br>The new version of Torc<br>Body has 2XMOPP level<br>isolation as required by<br>the IEC standards. |
| Patient Leakage<br>Current<br>Normal condition<br>Single Fault<br>condition | Normal condition =<br>less than<br>$100\mu A$<br>Single fault<br>condition = less than<br>$300\mu A$ | Normal condition<br>= less than $100\mu A$<br>Single fault<br>condition = less<br>than $300\mu A$ | Identical |
| Components | Main unit, 8 self<br>adhesive electrodes<br>, 4 electrode lead<br>wires, , 1 AC Power<br>Cord | Main unit, 8<br>electrodes, 4<br>electrode lead<br>wires, 6 elastic<br>belts (2 large, 2<br>medium, 2 small),<br>1 AC Adapter. | Modified;<br>The modified device has<br>an internal medical grade<br>power supply with an AC<br>power cord; the<br>predicate device has a<br>rechargeable battery<br>with an AC adapter. |
| Display | 12" LCD | 5.7" LCD | Modified; |
| | | | Modified for better visual<br>which gives a wide view<br>angle. |
| Number of Output<br>Modes | 03<br>(Classic Vector,<br>Power Sequence,<br>Dynamic Rotation) | 03<br>(ABS, BTS, CVL) | Modified;<br>Name of the modes<br>(Classic Vector, Power<br>Sequence, Dynamic<br>Rotation) has been<br>modified for a clear<br>understanding of the<br>clinician. |
| Number of Output<br>Channels | 2 (Two) | 2 (Two) | Identical |
| Synchronous or<br>Alternating? | Synchronous<br>(a) Output 1 to 2 are<br>completely isolated.<br>Only power supply<br>and ground are<br>common | Synchronous<br>(a) Output 1 to 2<br>are completely<br>isolated. Only<br>power supply and<br>ground are<br>common | Identical |
| Method of Channel<br>Isolation | Transformer | Transformer | Identical |
| Regulated Current<br>or<br>Regulated Voltage | Tran conductance | Tran conductance | Identical |
| Software/Firmware/<br>Microprocessors<br>Controls? | YES | YES | Identical |
| Automatic Overload<br>Trip? | YES | YES | Identical |
| Automatic No-Load<br>Trip | NO | NO | Identical |
| Automatic Shut off? | YES | YES | Identical |
| Patient Override<br>Control? | YES | YES | Identical |
| Indicator Display:<br>On/Off Status? | YES | YES | Identical |
| Low Battery? | N/A | YES | Modified;<br>The modified device does<br>not have a battery and,<br>therefore, does not have |
| | | | a battery indicator<br>display. |
| Voltage/Current<br>Level? | YES | YES | Identical |
| Timer Range<br>(minutes) | 1 – 60 Minutes<br>In step of 1 minute | 1 – 60 Minutes<br>In step of 1<br>minute | Identical |
| Compliance with<br>Voluntary<br>Standards? | YES<br>IEC 60601-1, IEC<br>60601-1-<br>2,IEC60601-2-10,<br>and ISO14971 | YES<br>IEC 60601-1, IEC<br>60601-1-<br>2,IEC60601-2-10,<br>and ISO14971 | Identical |
| Compliance With 21<br>CFR 898 | YES, the electrode<br>cable can never be<br>plugged in the AC<br>socket, not even<br>accidentally | YES, the electrode<br>cable can never be<br>plugged in the AC<br>socket, not even<br>accidentally | Identical |
| Weight | 32.66 Kgs | 3.0 KG | Modified;<br>The modified device has<br>been upgraded to a<br>wheeled console. |
| Dimension<br>(L X B X H) | 14″(L) X 17.5″ (W) X<br>40″(H) | 8.2″(L) x 6.0″ (W) x<br>4.0" (H) | Modified;<br>The modified device has<br>been upgraded to a<br>wheeled console. |
| Housing Material<br>and construction | ABS Plastic Body | ABS Plastic Body | Identical |
| Operating<br>Temperature | Temperature: +15°C<br>to +35°C | Temperature:<br>+15°C to +35°C | Identical |
| | Relative Humidity:<br>30 % to 75 % (non-<br>condensing)<br>Barometric | Relative Humidity:<br>30 % to 75 % (non-<br>condensing)<br>Barometric | |
| | pressure:<br>700 hPa to 1060 hPa | pressure:<br>700 hPa to 1060 hPa | |
| Transport and<br>storage<br>environment | Temperature: +5°C<br>to +45°C<br><br>Relative Humidity:<br>10% to 85% (non-<br>condensing) | Temperature:<br>+5°C to +45°C<br><br>Relative Humidity:<br>10% to 85% (non-<br>condensing) | Identical |
## Basic Device Characteristics – Comparison with Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Comparison
# Output Specification – Comparion with Predicate Devcies
| Characteristics | MODIFIED DEVICE | PREDICATE DEVICE | Comparison |
|---------------------------|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| | Torc Body (K192039) | Torc Body (K131291) | |
| Output Waveform | Classic Vector | ABS-6 | N/A |
| Waveform | Symmetrical Biphasic | Symmetrical Biphasic | Identical |
| Shape | Step Sine Wave | Square Wave | Modified;<br>The modified devices uses sinewave |
| Maximum Output<br>Voltage | 51.6 Vpp @ 500Ω<br>(± 10%)<br>62.4 Vpp @ 2KΩ<br>(± 10%)<br>66 Vpp @ 10KΩ<br>(± 10%) | 51 Vpp @ 500Ω<br>(± 10%)<br>63 Vpp @ 2KΩ<br>(± 10%)<br>66 Vpp @ 10KΩ<br>(± 10%) | Modified;<br>The difference between the devices at 500 and 2KΩ values is negligible. |
| Maximum Output<br>Current | 103.2 mA pp @ 500 Ohm<br>31.2 mA pp @ 2K Ohm<br>6.6 mA pp @ 10K | 102 mA pp @ 500 Ohm<br>31.5 mA pp @ 2K Ohm<br>6.6 mA pp @ 10K | Modified;<br>The difference between the devices at 500 |
# Mode: Classic Vector
| | | | sinewave |
|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Maximum Output<br>Voltage | 51.6 Vpp @ 500Ω<br>(± 10%)<br>62.4 Vpp @ 2KΩ<br>(± 10%)<br>66 Vpp @ 10KΩ<br>(± 10%) | 51 Vpp @ 500Ω<br>(± 10%)<br>63 Vpp @ 2KΩ<br>(± 10%)<br>66 Vpp @ 10KΩ<br>(± 10%) | Modified;<br>The difference<br>between the two<br>devices at 500Ω<br>and 2KΩ values, is<br>negligible. |
| Maximum Output<br>Current | 103.2 mA pp @ 500<br>Ohm<br>31.2 mA pp @ 2K<br>Ohm<br>6.6 mA pp @ 10K<br>Ohm | 102 mA pp @ 500<br>Ohm<br>31.5 mA pp @ 2K<br>Ohm<br>6.6 mA pp @ 10K<br>Ohm | Modified;<br>The difference<br>between the two<br>devices at 500Ω<br>and 2KΩ values, is<br>negligible. |
| Pulse Width | 124.7 μS (± 10%)<br>@500Ω | 350 μS (± 10%) | Modified;<br>The pulse width is<br>less than the<br>predicate device,<br>however the<br>waveshape is<br>sinewave with<br>duty cycle of 50%,<br>which implies that<br>the strength of<br>the stimulation is<br>essentially the |
| | | | same in both |
| | | | devices. |
| Frequency | 4010 Hz (± 10%)<br>@500Ω<br>Resultant: 1 – 100 Hz | 99 Hz (± 10%) | Modified;<br>The pulse rate is<br>higher in the<br>modified device,<br>but due to the<br>interferential<br>effect the<br>resultant<br>frequency is only<br>1 – 100 Hz<br>(sweeping), which<br>is virtually<br>identical to the<br>predicate devcie<br>with frequency<br>(99 Hz). |
| For Interferential<br>modes only:<br>-Beat frequency | N/A | N/A | Identical |
| For multiphasic<br>Waveform<br>- Symmetrical<br>Phases? | Yes | Yes | Identical |
| - Phase duration | 124.7 μS | 350 μS | Modified;<br>The phase<br>duration is less<br>than the predicate<br>device, however<br>the waveshape is<br>sinewave with<br>duty cycle of 50%,<br>which implies that<br>the strength of<br>the stimulation is<br>essentially the<br>same in both<br>devices. |
| Net charge | 0 μC @500Ω (Being<br>Biphasic in nature<br>the net charge would<br>be Zero) | 0 μC @500Ω (Being<br>Biphasic in nature<br>the net charge would<br>be Zero) | Identical |
| Maximum<br>Phase Charge | 6.435 μC | 35.7 μC | Modified; |
| | | | The phase charge<br>is less than the<br>predicate device,<br>however the<br>waveshape is<br>sinewave with<br>duty cycle of 50%,<br>which implies that<br>the strength of<br>the stimulation is<br>essentially the<br>same in both<br>devices. |
| Maximum<br>Current Density | 1.15 mA/cm²<br>*measured with 50 x 90mm<br>rectangular electrodes | 1.14 mA/cm²<br>*measured with 50 x 90mm<br>rectangular electrodes | Modified;<br><br>The difference<br>between the<br>values of both<br>devices is<br>negligible. |
| Maximum<br>Power Density | 0.0295 Watt/cm²<br>*measured with 50 x 90mm<br>rectangular electrodes | 0.0289 Watt/cm²<br>*measured with 50 x 90mm<br>rectangular electrodes | Modified;<br><br>The difference<br>between the<br>values of both<br>devices is<br>negligible. |
| Burst Mode<br>-Pulses Per Burst<br>-Burst Per second<br>-Burst Duration<br>-Duty Cycle | N/A | N/A | Identical |
| ON Time | N/A | 5.5 seconds | Modified;<br><br>Modified device<br>does not have ON<br>time. |
| OFF Time | N/A | 5.5 seconds | Modified;<br><br>Modified device<br><br>does not have OFF<br>time. |
| Additional Features | Sweep Frequency 1-<br>100 Hz | N/A | The modified<br>device has feature<br>of sweep<br>frequency |
| Characteristics | MODIFIED DEVICE<br>Torc Body (K192039) | PREDICATE DEVICE<br>Torc Body (K131291) | Comparison |
| Output Waveform | Power Sequence | ABS-6 | N/A |
| Waveform | Symmetrical Biphasic | Symmetrical Biphasic | Identical |
| Shape | Square wave | Square Wave | Identical |
| Maximum Output<br>Voltage | 50.8 Vpp @ 500Ω<br>(± 10%)<br>64 Vpp @ 2KΩ<br>(± 10%)<br>65.6 Vpp @ 10KΩ<br>(± 10%) | 51 Vpp @ 500Ω<br>(± 10%)<br>63 Vpp @ 2KΩ<br>(± 10%)<br>66 Vpp @ 10KΩ<br>(± 10%) | Modified;<br>The difference<br>between the<br>values of both<br>devices at<br>different is<br>negligible. |
| Maximum Output<br>Current | 101.6 mA pp @ 500<br>Ohm<br>32 mA pp @ 2K Ohm<br>6.56 mA pp @ 10K<br>Ohm | 102 mA pp @ 500<br>Ohm<br>31.5 mA pp @ 2K<br>Ohm<br>6.6 mA pp @ 10K<br>Ohm | Modified;<br>The difference<br>between the<br>values of both<br>devices is<br>negligible. |
| Pulse Width | 350 μS (± 10%) @<br>500Ω | 350 μS (± 10%) | Identical |
| Frequency | 99.91 Hz (± 10%) @<br>500Ω | 99 Hz (± 10%) | Modified;<br>The difference<br>between the<br>values of both<br>devices is<br>negligible. |
| Beat frequency<br>For multiphasic<br>Waveform<br>Symmetrical Phases? | N/A | N/A | Identical |
| Phase duration | Yes, Symmetrical<br>Biphasic<br>350 μS (± 10%) | Yes, Symmetrical<br>Biphasic<br>350 μS (± 10%) | Identical |
| Net charge | 0 μC @500Ω (Being<br>Biphasic in nature<br>the net charge would<br>be Zero) | 0 μC @500Ω (Being<br>Biphasic in nature<br>the net charge would<br>be Zero) | Identical |
| Maximum Phase<br>Charge | 35.56 μC @ 500 Ω<br>Load | 35.7 μC @ 500 Ω<br>Load | Modified; |
| Maximum Current<br>density | 1.13 mA/cm²<br>*measured with 50 x 90mm<br>rectangular electrodes | 1.14 mA/cm²<br>*measured with 50 x 90mm<br>rectangular electrodes | The difference<br>between the<br>values of both<br>devices is<br>negligible.<br>Modified; |
| Maximum Power<br>Density | 0.0287 Watt/cm²<br>*measured with 50 x 90mm<br>rectangular electrodes | 0.0289 Watt/cm²<br>*measured with 50 x 90mm<br>rectangular electrodes | The difference<br>between the<br>values of both<br>devices is<br>negligible.<br>Modified; |
| Burst Mode<br>-Pulses Per Burst<br>-Burst Per second<br>-Burst Duration<br>-Duty Cycle | N/A | N/A | Identical |
| ON Time | 6 seconds | 5.5 seconds | Modified;<br>The difference<br>between the<br>values of both<br>devices is virtually<br>identical. |
| OFF Time | 4 seconds | 5.5 seconds | Modified;<br>The difference<br>between the<br>values of both<br>devices is virtually<br>identical. |
| Additional Features | | | Identical. |
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
## Mode: Power Sequence
{13}------------------------------------------------
# Mode: Dynamic Rotation
| Characteristics | MODIFIED DEVICE<br>Torc Body (K192039) | PREDICATE DEVICE<br>Torc Body (K131291) | Comparison |
|-----------------|----------------------------------------|-----------------------------------------|------------|
| Output Waveform | Dynamic Rotation | ABS-6 | N/A |
| Waveform | Symmetrical Biphasic | Symmetrical<br>Biphasic | Identical |
{14}------------------------------------------------
| Shape | Modulated Sine Wave | Square Wave | Modified; |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Maximum Output<br>Voltage | 51.2 Vpp @ 500Ω<br>(± 10%)<br>63.2 Vpp @ 2KΩ<br>(± 10%)<br>65.6 Vpp @ 10KΩ<br>(± 10%) | 51 Vpp @ 500Ω<br>(± 10%)<br>63 Vpp @ 2KΩ<br>(± 10%)<br>66 Vpp @ 10KΩ<br>(± 10%) | Modified;<br>The difference<br>between the two<br>devices values at<br>different load<br>values is<br>negligible. |
| Maximum Output<br>Current | 102.4 mA pp @ 500<br>Ohm<br>31.6 mA pp @ 2K<br>Ohm<br>6.56 mA pp @ 10K<br>Ohm | 102 mA pp @ 500<br>Ohm<br>31.5 mA pp @ 2K<br>Ohm<br>6.6 mA pp @ 10K<br>Ohm | Modified;<br>The difference<br>between the two<br>devices values at<br>different load<br>values is<br>negligible |
| Pulse Width | 124.9 μS (± 10%) @<br>500Ω | 350 μS (± 10%) | Modified;<br>The pulse width is<br>less than the<br>predicate device,<br>however the<br>waveshape is<br>sinewave with<br>duty cycle of 50%,<br>which implies that<br>the strength of<br>the stimulation is<br>essentially the<br>same in both<br>devices. |
| Frequency | 4002 Hz (± 10%) @<br>500Ω<br>Resultant: 1 – 100 Hz | 99 Hz (± 10%) | Modified;<br>The pulse rate is<br>higher in the<br>modified device,<br>but due to the<br>interferential<br>effect the<br>resultant<br>frequency is only |
| | | | 1 – 100 Hz<br>(sweeping), which<br>is virtually<br>identical to the<br>predicate devcie<br>with frequency<br>(99 Hz). |
| Beat frequency | NA | NA | Identical |
| For multiphasic<br>Waveform<br>Symmetrical<br>Phases? | Yes, Symmetrical<br>Biphasic…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.