K233762 · Ideate Medical, Inc. · MLR · Jun 27, 2024 · General Hospital
Device Facts
Record ID
K233762
Device Name
SteroScope® Sterilization Technology System
Applicant
Ideate Medical, Inc.
Product Code
MLR · General Hospital
Decision Date
Jun 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6860
Device Class
Class 2
Indications for Use
The SteroScope® Sterilization Technology System is indicated for the terminal sterilization of cleaned reusable Flexible Endoscopes with up to 8 internal lumens with lumen dimensions of: ID of 1.0 mm or larger and a length of 3580 mm or shorter and ID of 1.2 mm or larger and a length of 4095 mm or shorter
Device Story
SteroScope® Sterilization Technology System is a stand-alone vaporized hydrogen peroxide (VHP) gas plasma sterilizer for flexible endoscopes. Device uses proprietary gas diffusion technology creating pressure differentials in internal endoscope lumens to actively pull 59% H2O2 vapor through channels, facilitating sterilization of inner surfaces while minimizing peroxide dose exposure. System operates at 57°C. Used in healthcare facilities; operated by trained personnel. Output is a terminally sterilized endoscope. Benefits include effective sterilization of long multi-lumen endoscopes with reduced oxidation damage compared to traditional methods.
Clinical Evidence
Bench testing only. Testing included sporicidal efficacy (AOAC 966.04), microbial inactivation kinetics, simulated use testing (ASTM E1837-96) with 6-log reduction of Geobacillus stearothermophilus, material compatibility, and biocompatibility (ISO 10993-5). System functional, environmental, and H2O2 emission testing confirmed performance against internal specifications and safety standards.
Technological Characteristics
VHP gas plasma sterilizer; 59% H2O2 sterilant; 91.0L chamber volume; 57°C operating temperature. Proprietary gas diffusion mechanism creates pressure differential for lumen penetration. Complies with IEC 60601-1-2 (EMC), IEC 61010-1 (Electrical Safety), and ISO 10993-5 (Biocompatibility).
Indications for Use
Indicated for terminal sterilization of cleaned reusable flexible endoscopes (up to 8 internal lumens; ID ≥1.0mm/length ≤3580mm or ID ≥1.2mm/length ≤4095mm).
Regulatory Classification
Identification
An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
June 27, 2024
IDEATE Medical, Inc. % Kevin Corrigan Regulaory Consultant Corrigan Regulatory Consulting P.O. Box 6284 Anaheim, California 92816
Re: K233762
Trade/Device Name: SteroScope® Sterilization Technology System Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene Oxide Gas Sterilizer Regulatory Class: Class II Product Code: MLR Dated: November 22, 2023 Received: November 24, 2023
Dear Kevin Corrigan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Christopher K. Dugard -2
Christopher K. Dugard, M.S. Assistant Director DHT4B: Division of Infection Control
{2}------------------------------------------------
and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K233762
#### Device Name
SteroScope® Sterilization Technology System
#### Indications for Use (Describe)
The SteroScope® Sterilization Technology System is indicated for the terminal sterilization of cleaned reusable
- Flexible Endoscopes with up to 8 internal lumens with lumen dimensions of:
- ID of 1.0 mm or larger and a length of 3580 mm or shorter and ●
- . ID of 1.2 mm or larger and a length of 4095 mm or shorter
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------|--|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | |
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY K233762
26 June 2024
I. SUBMITTER
IDEATE Medical, Inc. 2821 N. Ballas Road C55 St. Louis, MO 63131
Phone: 760-402-8322
Contact Person: William Wong, CEO bill.wong@ideatemedical.com
II. DEVICE
| Name of Device: | SteroScope® Sterilization Technology System |
|----------------------|-------------------------------------------------|
| Common Name: | Vaporized Hydrogen Peroxide Sterilizer |
| Classification Name: | Ethylene oxide gas sterilizer (21 CFR 888.6860) |
| Regulatory: | Class II |
| Product Code: | MLR, Sterilizer, Chemical |
#### III. PREDICATE DEVICE
## STERRAD 100NX DUO Cycle (K111377)
Reference Device: STERRAD 100NX EXPRESS Cycle (K092622)
#### IV. DEVICE DESCRIPTION
The SteroScope® Sterilization Technology System is a vaporized hydrogen peroxide based (VHP) sterilizer for the terminal sterilization of flexible endoscopes with the following characteristics:
Flexible Endoscopes with up to 8 internal lumens with lumen dimensions of:
- · ID of 1.0 mm or larger and a length of 3580 mm or shorter and
- · ID of 1.2 mm or larger and a length of 4095 mm or shorter.
The SteroScope® Sterilization Technology System is a Gas Plasma VHP sterilization system designed specifically for flexible endoscope terminal sterilization, not for general surgical instrument loads. It is a self-contained stand-alone system of hardware and software system that uses technology to rapidly diffuse VHP into all the channels of the flexible endoscope to facilitate the sterilization of inner device surfaces, while being occupationally and environmentally safe to use. To enable the terminal sterilization of long multi-lumen endoscopes, while minimizing oxidation damage, the SteroScope® Sterilization Technology System incorporates a proprietary gas diffusion technology which creates a pressure differential
{5}------------------------------------------------
in each internal endoscope lumen channel to actively pull hydrogen peroxide vapor from the chamber of the VHP sterilizer through the distal end of each endoscope channel, while the endoscope is in a sterile container. Utilizing a separate mechanism for the rapid diffusion of VHP from the chamber into the individual SteroScope® Sterilization Technology System allows the entire endoscope to be sterilized with less overall peroxide dose exposure.
## V. INDICATIONS FOR USE
The SteroScope® Sterilization Technology System is indicated for the terminal sterilization of cleaned reusable Flexible Endoscopes with up to 8 internal lumens with lumen dimensions of:
- ID of 1.0 mm or larger and a length of 3580 mm or shorter and .
- ID of 1.2 mm or larger and a length of 4095 mm or shorter .
## VI. TECHNOLOGICAL CHARACTERISTICS COMPARISON
Table 1 (below) provides a summary of the device's technological characteristics comparing the subject device and the predicate device and reference device(s).
| Characteristic | Subject Device | Predicate Device | Reference Device | Comments |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| | STEROSCOPE<br>SYSTEM (K233762) | STERRAD100NX DUO<br>(K111377) | STERRAD 100NX<br>EXPRESS (K092622) | |
| Intend Use | SteroScope Sterilizer is<br>intended for the terminal<br>sterilization of properly<br>prepared (cleaned,<br>rinsed, and dried) flexible<br>endoscopes in Healthcare<br>Facilities. The<br>preprogrammed<br>sterilization cycle<br>operates at low pressure<br>and temperature, suitable<br>for processing<br>endoscopes without<br>leaving toxic residues. | The STERRAD Sterilizers<br>are designed for<br>sterilization of both metal<br>and nonmetal medical<br>devices at low<br>temperatures. Because the<br>cycle operates within a dry<br>environment and at low<br>temperatures, it is<br>especially suitable for<br>instruments sensitive to<br>heat and moisture. | The STERRAD Sterilizers<br>are designed for sterilization<br>of both metal and nonmetal<br>medical devices at low<br>temperatures. Because the<br>cycle operates within a dry<br>environment and at low<br>temperatures, it is especially<br>suitable for instruments<br>sensitive to heat and<br>moisture. | SIMILAR<br>Subject to<br>predicate &<br>reference |
| Indications for<br>Use | The SteroScope<br>Sterilization System is<br>indicated for the terminal<br>sterilization of cleaned<br>reusable flexible<br>endoscopes with up to 8<br>internal lumens with<br>lumen dimensions of: | The STERRAD®<br>100NXTM DUO<br>Cycle is an additional<br>optional cycle designed for<br>sterilization of medical<br>devices including most<br>flexible endoscopes, with<br>up to 1 internal lumen with<br>lumen dimension of:<br>1.0 mm or larger and a<br>length of 875mm or shorter | The STERRAD® 100NXTM<br>EXPRESS Cycle is an<br>additional optional cycle<br>designed for surface<br>sterilization of both metal<br>and nonmetal medical<br>devices at low temperatures.<br>It can sterilize instrument<br>surfaces and it can sterilize<br>rigid and semi-rigid<br>endoscopes without lumens | SIMILAR<br>Subject to<br>predicate &<br>reference |
## Table 2: TECHNOLOGY Comparison
{6}------------------------------------------------
| Characteristic | Subject Device<br>STEROSCOPE<br>SYSTEM (K233762)<br>ID of 1.0 mm or larger<br>and a length of 3580 mm<br>or shorter and<br>ID of 1.2 mm or larger<br>and a length of 4095 mm<br>or shorter | Predicate Device<br>STERRAD100NX<br>(K111377) | Reference Device<br>STERRAD 100NX<br>EXPRESS (K092622) | Comments |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------|---------------------------------------------------|
| Sterilization<br>Technology | VHP Gas Plasma | VHP Gas Plasma | VHP Gas Plasma | SAME |
| Sterilant | 59% H2O2 | 59% H2O2 | 59% H2O2 | SAME |
| H2O2/cycle | 5.8 mL | 3.1 mL | 10.8 mL | SIMILAR<br>Subject to<br>predicate &<br>reference |
| Chamber<br>Temperature | 57° C | 57° C | 57° C | SAME |
| Chamber<br>Volume | 91.0 Liters | 93.4 Liters | 93.4 Liters | SIMILAR<br>Subject to<br>predicate &<br>reference |
| Load Type | (1) flexible endoscope<br>(8) internal lumens<br>(1.0mm ID x 3585mm) | (2) flexible endoscopes<br>(1) internal lumen<br>(1.0mm ID x 850mm) | (2) Semi-rigid/rigid<br>endoscopes<br>(no lumens up to 500mm) | SIMILAR<br>Subject to<br>predicate |
| Load Weight | 17.8 lbs. | 13.2 lbs. | 10.7 lbs. | SIMILAR<br>Subject to<br>predicate |
| H2O2 Delivery | Extracted from H2O2<br>consumable and metered | Extracted from H2O2<br>consumable and metered | Extracted from H2O2<br>consumable with no metering | SAME |
| Cycle Time | 43 mins. | 51 mins. | 26 mins | SIMILAR<br>Subject to<br>predicate |
## VII. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
Table 3 (below) provides a summary of the non-clinical performance testing conducted on the subject device, SteroScope Sterilization System, to establish safety and efficacy. Table 4 below provides a list of standards used in the testing of the SteroScope Sterilization System
| Table 3: | | | |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|---------|
| Testing Method | Purpose | Acceptance Criteria | Results |
| System Environmental<br>Testing | System environmental testing under<br>extreme conditions operating<br>conditions | Internal<br>Specification | Pass |
| SteroScope H2O2<br>Emissions Test | H2O2 emissions compliance to<br>required OSHA safety levels | OSHA H2O2<br>SEL/PEL | Pass |
| Testing Method | Purpose | Acceptance Criteria | Results |
| System H2O2<br>Concentration Mapping | Peroxide concentration mapping during operation | Internal<br>Specification | Pass |
| System Temperature<br>Mapping | Temperature mapping during operation | Internal<br>Specification | Pass |
| System Functional Test | Testing of system functional requirements to ensure conformance to system specifications | Internal<br>Specification | Pass |
| SteroScope System<br>Shipping | System shipping test for vibration, shock & drop to ensure system and packaging integrity | ISTA 2B | Pass |
| SteroScope Consumable<br>Shipping | Consumable shipping test for vibration, shock & drop to ensure consumable and packaging integrity | ISTA 2A | Pass |
| EMC & Electrical Safety | The sterilizer was tested in accordance with IEC 60601-1-2 and IEC 61010-1 standards | IEC 60601-1-2 and<br>IEC 61010-1 | Pass |
| Qualification of Pressure<br>Limits | Validation of sterilizer operational pressure limits | Internal<br>Specification | Pass |
| Human Factors | Validation of system ergonomics and human factors with clinical users | Internal<br>Specification | Pass |
| Software Validation | Software verification and validation of software requirements | Internal<br>Specification | Pass |
| Subsystem Design<br>Verifications | Testing of each subsystem to ensure functional performance met design requirements | Internal<br>Specification | Pass |
| H2O2 Consumable Shelf-<br>life | Long term stability study of H2O2 consumable to establish shelf life | Internal<br>Specification | Pass |
| H2O2 Consumable<br>Storage | Stability testing of H2O2 consumable in extreme conditions | Internal<br>Specification | Pass |
| Endoscope Temperature<br>Mapping | Mapping of endoscope temperatures during cycle to verify the internal and external temperatures of a flexible endoscope | Internal<br>Specification | Pass |
| Final Process<br>Qualification | Operational process qualification with multiple systems | Internal<br>Specification | Pass |
| Mated Surface Connector-<br>Port | Half cycle connector efficacy test | No growth of<br>Geobacillus<br>stearothermophilus | Pass |
| BI & CI Functionality | BI & CI functionality under worst case conditions | No Growth in BI &<br>color change in CI | Pass |
| AOAC Sporicidal Test | Process and sterilant efficacy for multiple organisms | AOAC Official<br>Method 966.04 | Pass |
| Mated Surface<br>Sterilization | Half cycle mated material efficacy using material coupons | No growth of<br>Geobacillus<br>stearothermophilus | Pass |
| Testing Method | Purpose | Acceptance Criteria | Results |
| Material Surface<br>Sterilization | Half cycle surface efficacy using<br>material coupons | No growth of<br>Geobacillus<br>stearothermophilus | Pass |
| Uniformity Microbial<br>Inactivation | Dosing study to assess uniformity of<br>microbial inactivation | Uniformity in<br>inactivation kinetics<br>based on SLR | Pass |
| Endoscope Dose<br>Response Curves | Dosing study with various flexible<br>endoscopes | Internal<br>Specification | Pass |
| Material Bacteriostasis<br>Validation | Confirmation that there are no<br>bacteriostatic effects that inhibit<br>growth for material coupons | No bacteriostatic<br>effect after<br>sterilization with<br>growth of defined<br>micro-organisms | Pass |
| Bacteriostasis &<br>Fungistasis Validation | Confirmation that there are no<br>bacteriostatic effects that inhibit<br>growth for flexible endoscopes | No bacteriostatic<br>effect after<br>sterilization with<br>growth of defined<br>micro-organisms | Pass |
| Simulated Use Testing | Worst-case testing with unwashed<br>flexible endoscopes inoculated with<br>test culture challenge | 6-log reduction in<br>Geobacillus<br>stearothermophilus | Pass |
| In-Use Testing | Worst-case testing with clinically used<br>flexible endoscopes. | No growth upon<br>recovery after<br>sterilization | Pass |
| Functionality & Material<br>Compatibility | Worst-case material compatibility<br>testing of multiple flexible endoscopes<br>with back to back cycles | Internal<br>Specification | Pass |
| Biocompatibility Coupons | Worst-case residual and cytotoxicity<br>(MTT) testing with material coupons | ISO 10993-5 | Pass |
| Biocompatibility Flexible<br>Endoscopes | Worst-case residual and cytotoxicity<br>(MTT) testing with flexible<br>endoscopes | ISO 10993-5 | Pass |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Table 4 Standards Used:
| Organization | Designation | Title |
|------------------|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC | 60601-1-2 Edition 4.0 2014-02 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| IEC | 61010-1 Edition 3.1 2017-01<br>CONSOLIDATED VERSION | Safety requirements for electrical equipment for measurement control and laboratory use - Part 1: General requirements [Including: Corrigendum 1 (2019)] |
| AAMI | TIR 12:2020 | Design, Testing and Labeling of Reusable Medical Devices |
| ANSI AAMI<br>ISO | 11737-1:2018 | Microbial Methods Determination of the population of microorganisms |
| ASTM | E1837-96 (2014) | Simulated Use Test |
| AAMI ANSI | ST77:2013(R)2018 | Containment Devices |
| ANSI AAMI<br>ISO | 17664:2017 | Info To Be Provided for the processing of medical devices |
| AAMI | TIR39:2009(R)2017 | Selecting a Microbial Challenge and Inoculation sites |
| ANSI AAMI<br>ISO | 11138-1:2017 | BIs part 1 |
| ISO | 10993-5 Third edition 2009-06-01 | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity |
| ISO | 10993-12 Fourth edition 2012-07-01 | Biological evaluation of medical devices - Part 12: Sample preparation and reference materials |
| USP-NF | 1035 | BIs for Sterilization |
| AOAC | 966.04 | Sporicidal Activity of Disinfectants |
| AAMI ANSI | ST14:2008 | EtO Sterilization in Health Care Facilities |
| IEC | 61000 | Electromagnetic Compatibility |
| ISTA | 2A and 2B | International Safe Transit Association |
| OSHA | CAS 7722-84-1 | H2O2 Handling Guidance |
| AOAC | Official Method 966.04 | Sporicidal Activity of Disinfectants. Official Methods of Analysis of the AOAC, 2013 Edition |
# Conclusion
The conclusion drawn from the non-clinical testing demonstrates that the subject device, SteroScope® Sterilization Technology System submitted under K233762, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K131407, the STERRAD 100NX DUO Cycle, and the legally marketed reference device cleared under K092622, the STERRAD 100NX EXPRESS Cycle.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.