The STERRAD® 100NX® Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to safely sterilize medical instruments and materials without leaving toxic residue. The STERRAD® 100NX® Sterilizer can sterilize instruments which have diffusionrestricted spaces, such as the hinged portion of forceps and scissors. Medical devices with the following materials and dimensions can be processed in the STERRAD® 100NX® Sterilizer Standard cycle: - Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a length of 500 mm or shorter* Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD® 100NX® Sterilizer Flex Scope cycle: - Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and length of 850 mm or shorter** Note: With the exception of the 1 x 850 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 lbs. The 1 x 850 mm flexible endoscopes were validated without any additional load. *A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle. ** A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load. The STERRAD® 100NX® EXPRESS Cycle is an additional optional cycle designed for surface sterilization of both metal and nonmetal medical devices at low temperatures. - It can sterilize instrument surfaces and instruments having diffusionrestricted spaces, such as the hinged portion of forceps and scissors - It can sterilize rigid and semi-rigid endoscopes without lumens Note: The validation studies for EXPRESS Cycle were performed using a validation load consisting of a single instrument tray weighing 10.7 Ibs placed on the bottom shelf. The STERRAD® 100NX® DUO Cycle is an additional optional cycle designed for sterilization of medical devices including most flexible endoscopes, with the following materials and dimensions: - Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1mm or larger and a length of 875mm or shorter - Accessory devices that are normally connected to a flexible endoscope during use - Flexible endoscopes without lumens Note: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 lbs.
Device Story
STERRAD 100NX Sterilizer is a self-contained, stand-alone system for low-temperature sterilization of medical instruments. It utilizes a hydrogen peroxide gas plasma process; aqueous hydrogen peroxide is injected, vaporized, introduced into a chamber under sub-ambient pressure, and transformed into gas-plasma via electrical energy. The system includes a sterilization chamber, covered frame, disposable sterilant cassette, and reusable trays. The new DUO cycle expands capabilities to include longer flexible endoscopes (up to 875mm) and associated accessories. Operated by clinical staff in healthcare settings, the system provides automated sterilization cycles. The process ensures instruments are safe for reuse by eliminating microbial contamination without toxic residue. The healthcare provider loads the instruments, selects the cycle, and the system manages the sterilization parameters, providing a validated, reproducible process for sensitive medical devices.
Clinical Evidence
Bench testing only. Validation studies utilized the 'overkill' approach with G. stearothermophilus spores. Testing included dose response, surface/mated surface sterilization, bacteriostasis, in-use testing, simulated use, material toxicity, device functionality/material compatibility, and process reproducibility. All studies passed.
Technological Characteristics
Hydrogen peroxide gas plasma sterilization system. Materials: stainless steel, polyethylene, Teflon. Sensing/Actuation: sub-ambient pressure chamber, electrical energy for plasma generation. Connectivity: standalone. Sterilization: hydrogen peroxide vapor/plasma. Software: embedded system for cycle control.
Indications for Use
Indicated for sterilization of metal and nonmetal medical devices, including flexible endoscopes (polyethylene/Teflon, ≥1mm ID, ≤875mm length), rigid/semi-rigid endoscopes, and accessory devices. Suitable for instruments with diffusion-restricted spaces (e.g., hinged forceps/scissors).
Regulatory Classification
Identification
An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.
{0}------------------------------------------------
K111377
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
SEP 1 3 2012
... ... ... ... ..
## 510(k) Summary
Applicant's Name, Address, Telephone, FAX, Contact Person Advanced Sterilization Products, Division of Ethicon, Inc. a Johnson & Johnson company 33 Technology Drive Irvine, CA 92618
## Contact Person
Nancy Chu Manager, Regulatory Affairs Tel: (949) 453-6435 Fax: (949) 789-3900
September 10, 2012
#### CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME 1
Classification Name: Common/Usual Name: Product Classification: Classification Number: Proprietary Name:
Sterilizer, Class II Hydrogen Peroxide Gas Plasma Sterilization System Sterilizer, Class II 21 CFR 880.6860 STERRAD® 100NX® Sterilizer with DUO Cycle
#### 2 PREDICATE DEVICES
STERRAD® 100NX® Sterilization System [K071385] STERRAD® 100S Sterilization System [K991999, K023290]
#### 3 INDICATIONS FOR USE
The STERRAD® 100NX® Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to safely sterilize medical instruments and materials without leaving toxic residue.
The STERRAD® 100NX® Sterilizer can sterilize instruments which have diffusionrestricted spaces, such as the hinged portion of forceps and scissors.
Medical devices with the following materials and dimensions can be processed in the STERRAD® 100NX® Sterilizer Standard cycle:
{1}------------------------------------------------
- a Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a length of 500 mm or shorter*
Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD® 100NX® Sterilizer Flex Scope cycle:
- > Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and length of 850 mm or shorter**
Note: With the exception of the 1 x 850 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 lbs. The 1 x 850 mm flexible endoscopes were validated without any additional load.
*A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle.
** A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load.
The STERRAD® 100NX® EXPRESS Cycle is an additional optional cycle designed for surface sterilization of both metal and nonmetal medical devices at low temperatures.
- > It can sterilize instrument surfaces and instruments having diffusionrestricted spaces, such as the hinged portion of forceps and scissors
- > It can sterilize rigid and semi-rigid endoscopes without lumens
Note: The validation studies for EXPRESS Cycle were performed using a validation load consisting of a single instrument tray weighing 10.7 Ibs placed on the bottom shelf.
The STERRAD® 100NX® DUO Cycle is an additional optional cycle designed for sterilization of medical devices including most flexible endoscopes, with the following materials and dimensions:
- > Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1mm or larger and a length of 875mm or shorter
- > Accessory devices that are normally connected to a flexible endoscope during use
- > Flexible endoscopes without lumens
Note: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 lbs.
{2}------------------------------------------------
ិងក្រុង "
#### DESCRIPTION OF DEVICE ব
The STERRAD® 100NX® Sterilizer is a self-contained stand-alone system of hardware and software designed to sterilize medical instruments and devices using a patented hydrogen peroxide gas plasma process. Hydrogen peroxide vapor is generated by injecting aqueous hydrogen peroxide into the vaporizer where the solution is heated and vaporized, introducing the vapor into the chamber under sub-ambient pressure and transforming the vapor into a gas-plasma using electrical energy. The STERRAD® 100NX® Sterilizer has three previously cleared cycles (STANDARD, FLEX, EXPRESS) and the new optional additional DUO Cycle.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The hardware for the STERRAD® 100NX® Sterilizer consists of a sterilization chamber and a variety of instruments and components which are housed in a covered frame. The sterilizer system also uses accessories such as a disposable sterilant cassette, reusable instrument trays, and printer paper.
| CHARACTERISTICS | SUBJECT DEVICE<br>STERRAD® 100NX®<br>DUO CYCLE | PREDICATE DEVICES | |
|-----------------------------------------------------------------------|--------------------------------------------------------------------------|---------------------------------------|--------------------------------------------------------------|
| | | STERRAD® 100S<br>CYCLE | STERRAD® 100NX® FLEX<br>CYCLE |
| Base System | STERRAD® 100NX®<br>Sterilizer | STERRAD® 100S<br>Sterilizer | STERRAD®<br>100NX® Sterilizer |
| Hydrogen peroxide, 59% | Yes | Yes | Yes, with<br>concentration<br>process |
| Hydrogen peroxide<br>delivery | STERRAD® 100NX®<br>Cassette and metered<br>through Delivery<br>Module | STERRAD® 100S<br>cassette | STERRAD®<br>100NX® Cassette |
| Disposal of hydrogen<br>peroxide after 10 days of<br>on-board storage | Hydrogen peroxide<br>clearance process,<br>initiated by user | Cassette has to be<br>ejected by user | Cassette has to be<br>ejected by user |
| Cycle Time, approximate | 60 minutes | 54 minutes | 42 minutes |
| Load Type | Two flexible<br>endoscopes with<br>accessories; 1x875mm<br>PE/PTFE lumen | No flexible<br>endoscopes | Two flexible<br>endoscopes only;<br>1x850mm<br>PE/PTFE lumen |
| Load Weight, lbs | 13.2 | Not defined | 21.4 |
# 5 TABLE 5-1: COMPARISION BETWEEN SUBJECT DEVICE AND . PREDICATES
5
{3}------------------------------------------------
#### SUMMARY OF NONCLINICAL TESTS 6
### Validation Testing
Testing was performed using the "overkill" approach utilizing G. stearothermophilus spores. Table 6-1 identifies the validation studies performed and the results obtained.
ក្នុង -
| STUDY | RESULTS |
|-----------------------------------------------------------------------|---------|
| Dose Response with Biological Model (1 x 875 mm<br>Flexible Endoscope | Passed |
| Surface Sterilization | Passed |
| Mated Surface Sterilization | Passed |
| 1 x 875 mm Flexible Endoscope Validation | Passed |
| Bacteriostasis Testing in the DUO Cycle | Passed |
| In Use Testing – 1 x 875 mm Flexible Endoscopes and<br>Accessories | Passed |
| Simulated Use Testing | Passed |
| Toxicity Testing of Materials | Passed |
| Device Functionality and Material Compatibility | Passed |
| Process Reproducibility | Passed |
# TABLE 6-1: VALIDATION STUDIES
#### 7 OVERALL PERFORMANCE CONCLUSIONS
The nonclinical studies demonstrate that the STERRAD® 100NX® DUO Cycle is safe and effective for sterilization of medical devices within the indications for use for the sterilizer and establish equivalence of the STERRAD® 100NX® DUO Cycle to the predicate devices, the STERRAD® 100NX® Sterilizer FLEX Cycle and the STERRAD® 100S Sterilizer.
P. 4 of 4
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Nancy Chu Manager, Regulatory Affairs Advanced Sterilization Products, Division of Ethicon, Incorporated 33 Technology Drive Irvine, CA 92618
SEP
13 2012
Re: K111377
Trade/Device Name: STERRAD 100NX™ Sterilizer with DUO Cycle Regulation Number: 21 CFR §880.6860 Regulation Name: Ethylene Gas Sterilizer Regulatory Class: II Product Code: MLR Dated: September 10, 2012 Received: September 12, 2012
Dear Ms. Chu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Ms. Chu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology. General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{6}------------------------------------------------
# X 1113
្រី
## Indications for Use
## 510(k) Number (if known): K111377
STERRAD® 100NX® Sterilizer with DUO Cycle Device Name:
## Indications for Use:
The STERRAD® 100NX® Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to safely sterilize medical instruments and materials without leaving toxic residue.
The STERRAD® 100NX® Sterilizer can sterilize instruments which have diffusionrestricted spaces, such as the hinged portion of forceps and scissors.
Medical devices with the following materials and dimensions can be processed in the STERRAD® 100NX® Sterilizer Standard cycle:
- > Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a length of 500 mm or shorter*
Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD® 100NX® Sterilizer Flex Scope cycle:
- > Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and length of 850 mm or shorter**
Note: With the exception of the 1 x 850 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 lbs. The 1 x 850 mm flexible endoscopes were validated without any additional load.
*A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle.
** A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load.
Page 1 of 2
{7}------------------------------------------------
រដឹក
The STERRAD® 100NX® EXPRESS Cycle is an additional optional cycle designed for surface sterilization of both metal and nonmetal medical devices at low temperatures.
- > It can sterilize instrument surfaces and instruments having diffusionrestricted spaces, such as the hinged portion of forceps and scissors
- A It can sterilize rigid and semi-rigid endoscopes without lumens
The validation studies for EXPRESS Cycle were performed using a validation load Note: consisting of a single instrument tray weighing 10.7 lbs placed on the bottom shelf,
The STERRAD® 100NX® DUO Cycle is an additional optional cycle designed for sterilization of medical devices including most flexible endoscopes, with the following materials and dimensions:
- > Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1mm or larger and a length of 875mm or shorter
- > Accessory devices that are normally connected to a flexible endoscope during use
- > Flexible endoscopes without lumens
Note: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 Ibs.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
な
Over-The-Counter Use V (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Mn
Division Sign-Off) Division Sign-Om)
Division of Anesthesiology, General Hospital ntection Control, Dental Devices
510(k) Number: K 111377
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.