K233735 · Globus Medical, Inc. · KWQ · Jan 25, 2024 · Orthopedic
Device Facts
Record ID
K233735
Device Name
ADIRA™ Lateral Plate System
Applicant
Globus Medical, Inc.
Product Code
KWQ · Orthopedic
Decision Date
Jan 25, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ADIRA™ 2-Hole and 4-Hole Plates, when used with screws only, are intended for use in the treatment of thoracolumbar (T1-L5) spinal instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (DDD, defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, lordosis, spinal stenosis, or failed previous spinal surgery. The ADIRA™ 1-Hole Plate is intended to stabilize allograft at one level (T1-L5), aiding in spinal fusion and to provide temporary stabilization and augment of a spinal fusion. It may be used alone or with other anterior, lateral, anterolateral, or posterior spinal systems. The device is not intended for load bearing applications. ADIRA™ Plates may be assembled to a lateral lumbar interbody fusion device (HEDRON® L, TransContinental®, RISE®-L, Modulus® XLIF, CoRoent®, or Cohere® XLIF Spacers) to create a plate-spacer assembly. When assembled to HEDRON® L, TransContinental®, or RISE®-L Spacers, the plate-spacer assembly is indicated for use at one or more levels of the lumbosacral spine (L1-L5), as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. When ADIRA™ Plates are assembled to Modulus® XLIF, CoRoent®, or Cohere® XLIF Spacers, the plate-spacer assembly takes on the indications of the interbody device. ADIRA™ Plate-Spacers are intended to be used with screws and/or anchors which accompany the implants. These devices are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). In addition, the 2-Hole or 4-Hole Plate-Spacers are intended for stand-alone use in patients with DDD at one or two levels only when <20° lordotic implants are used with two or four screws, respectively. ADIRA™ Plate-Spacers are to be filled with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone.
Device Story
ADIRA™ Lateral Plate System comprises 1-, 2-, and 4-hole thoracolumbar plates; used with bone screws for anterolateral/lateral vertebral body fixation. Plates attach to lateral lumbar interbody fusion spacers (HEDRON® L, TransContinental®, RISE®-L, Modulus® XLIF, CoRoent®, Cohere® XLIF) via alignment screw to form plate-spacer assemblies. Assemblies filled with autograft/allogenic bone graft. Used in surgical settings by surgeons to stabilize spine, aid fusion, or provide supplemental fixation. Output is mechanical stabilization of spinal segments; assists in fusion development; benefits patients by addressing instability, deformity, and degenerative conditions.
Clinical Evidence
Bench testing only. Mechanical testing performed per ASTM F1717 and FDA guidance for spinal systems (static/dynamic compression, bending, torsion, and expulsion). MR compatibility simulations performed.
Technological Characteristics
Thoracolumbar plate system; includes 1-, 2-, and 4-hole plates. Materials and design similar to predicate spinal intervertebral body fixation orthoses. Mechanical testing conducted per ASTM F1717. Standalone or assembly-based fixation using screws/anchors.
Indications for Use
Indicated for skeletally mature patients with thoracolumbar (T1-L5) spinal instability due to fracture, tumor, DDD, scoliosis, lordosis, spinal stenosis, or failed surgery. Plate-spacer assemblies indicated for lumbosacral (L1-L5) fusion adjunct in patients with DDD, disc herniation, spondylolisthesis, deformity, spinal stenosis, or pseudarthrosis, requiring 6 months of failed non-operative treatment. Stand-alone use for DDD at 1-2 levels permitted under specific lordotic/screw configurations.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
January 25, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Globus Medical Inc Jennifer Antonacci Group Manager, Regulatory Affairs Valley Forge Business Center 2560 General Armistead Ave Audubon. Pennsylvania 19403
## Re: K233735
Trade/Device Name: ADIRA™ Lateral Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ, OVD, MAX Dated: November 21, 2023 Received: November 22, 2023
### Dear Jennifer Antonacci:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows a digital signature. The signature is from Eileen Cadel and was signed on January 25, 2024, at 11:52:11 -05'00'. The signature includes the name Eileen Cadel, followed by -S.
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233735
Device Name ADIRA™ Lateral Plate System
#### Indications for Use (Describe)
ADIRATM Lateral Plate System
The ADIRA™ 2-Hole and 4-Hole Plates, when used with screws only, are intended for use in the treatment of thoracolumbar (T1-L5) spinal instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (DDD, defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, lordosis, spinal stenosis, or failed previous spinal surgery.
The ADIRA™ 1-Hole Plate is intended to stabilize allograft at one level (T1-L5), aiding in spinal fusion and to provide temporary stabilization and augment of a spinal fusion. It may be used alone or with other anterior, lateral, anterolateral, or posterior spinal systems. The device is not intended for load bearing applications.
#### ADIRA™ Plate-Spacer Assemblies
ADIRA™ Plates may be assembled to a lateral lumbar interbody fusion device (HEDRON® L, TransContinental®), RISE®-L, Modulus® XLIF, CoRoent®, or Cohere® XLIF Spacers) to create a plate-spacer assembly. When assembled to HEDRON® L, TransContinental®, or RISE®-L Spacers, the plate-spacer assembly is indicated for use at one or more levels of the lumbosacral spine (L1-L5), as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. When ADIRA™ Plates are assembled to Modulus® XLIF, CoRoent®, or Cohere® XLIF Spacers, the plate-spacer assembly takes on the indications of the interbody device.
ADIRA™ Plate-Spacers are intended to be used with screws and/or anchors which accompany the implants. These devices are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). In addition, the 2-Hole or 4-Hole Plate-Spacers are intended for stand-alone use in patients with DDD at one or two levels only when <20° lordotic implants are used with two or four screws, respectively.
ADIRA™ Plate-Spacers are to be filled with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary: ADIRA™ Lateral Plate System
| Company: | Globus Medical Inc. |
|-----------------------------|---------------------|
| 2560 General Armistead Ave. | |
| Audubon, PA 19403 | |
| 610-930-1800 | |
- Primary Contact: Jennifer Antonacci, Ph.D. Group Manager, Regulatory Affairs
- Date Prepared: January 24, 2024
- ADIRA™ Lateral Plate System Device Name:
- Common Name: Thoracolumbar Plate System
- Classification: Per 21 CFR as follows: §888.3060 Spinal Intervertebral Body Fixation Orthosis §888.3080 Intervertebral Body Fusion Device Product Code(s): KWQ, OVD Regulatory Class II, Panel Code: 87
Primary Predicate: PLYMOUTH™ Thoracolumbar Plate (K120092)
## Additional
| Predicates: | PLYMOUTH™ Thoracolumbar Plate (K181846)<br>InterContinental™ Plate-Spacer (K103382, K203278)<br>4WEB Lateral Spine Truss System Plating (K203065)<br>ELSA™ Spacers (K161379, K203278) |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Reference Devices: | NuVasive Thoracolumbar Interbody Systems: CoRoent<br>Modulus XLIF, Cohere, and others (K203714) |
### Purpose:
The purpose of this submission is to request clearance for the ADIRA"" Lateral Plate System.
### Device Description:
The ADIRA™ Lateral Plate System consists of 4-, 2- and 1-hole thoracolumbar plates of various sizes to accommodate varying patient anatomy and surgical needs. The ADIRA™ plates are used with bone screws to attach to the anterolateral or lateral portion of the vertebral body of the thoracolumbar spine. ADIRA™ plates may also be attached to certain lateral lumbar interbody fusion devices (HEDRON L™, TransContinental™, RISE-L™, Modulus® XLIF, Cohere® XLIF, or CoRoent®) with an alignment screw. The plate-spacer assembly is used with bone screws and/or lateral anchors.
{5}------------------------------------------------
## Indications for Use:
### ADIRA'''' Lateral Plate System
The ADIRA™ 2-Hole and 4-Hole Plates, when used with screws only, are intended for use in the treatment of thoracolumbar (T1-L5) spinal instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (DDD, defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, lordosis, spinal stenosis, or failed previous spinal surgery.
The ADIRA™ 1-Hole Plate is intended to stabilize allograft or autograft at one level (T1-L5), aiding in spinal fusion and to provide temporary stabilization and augment development of a spinal fusion. It may be used alone or with other anterior, lateral, anterolateral, or posterior spinal systems. The device is not intended for load bearing applications.
### ADIRA™ Plate-Spacer Assemblies
ADIRA™ Plates may be assembled to a lateral lumbar interbody fusion device (HEDRON® L, TransContinental®, RISE®-L, Modulus® XLIF, CoRoent®, or Cohere® XLIF Spacers) to create a plate-spacer assembly. When assembled to HEDRON® L, TransContinental®, or RISE®-L Spacers, the plate-spacer assembly is indicated for use at one or more levels of the lumbosacral spine (L1-L5), as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. When ADIRA™ Plates are assembled to Modulus® XLIF, CoRoent®, or Cohere® XLIF Spacers, the plate-spacer assembly takes on the indications of the interbody device.
ADIRA™ Plate-Spacers are intended to be used with screws and/or anchors which accompany the implants. These devices are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). In addition, the 2-Hole or 4-Hole Plate-Spacers are intended for stand-alone use in patients with DDD at one or two levels only when <20° lordotic implants are used with two or four screws. respectively.
ADIRA™ Plate-Spacers are to be filled with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone.
## Performance Data:
Mechanical testing (static and dynamic compression bending and static torsion) was conducted in accordance with ASTM F1717, the "Guidance for Industry and FDA Staff, Guidance for Spinal System 510(k)s", May 3, 2004, and the "Guidance for Industry and FDA Staff: Spinal Plating Systems - Performance Criteria for
{6}------------------------------------------------
Safety and Performance Based Pathway," December 11, 2020. Expulsion testing was also conducted. Performance data demonstrate substantial equivalence to the predicate devices. Simulations demonstrated that the subject devices are not a worst case compared to the predicate devices in terms of MR compatibility.
### Technological Characteristics:
The ADIRA™ implants have similar technological characteristics as the predicate devices including overall design, intended use, material composition, function, and range of sizes.
### Basis of Substantial Equivalence:
The ADIRA™ Lateral Plate System has been found to be substantially equivalent to the predicate devices with respect to technological characteristics, performance, design, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.