Clinical outcomes registry study; Clinical literature analysis
The sponsor utilized registry data from 75 patients treated with the Modulus XLIF device and Attrax Putty, supplemented by a literature review comparing Attrax Putty with other interbody fusion devices (e.g., CoRoent PEEK), to support the safety and effectiveness of using Attrax Putty with the NuVasive Thoracolumbar Interbody Systems.
Registry study; Literature review; Clinical outcomes; Fusion rates; Adverse events
Patients treated with NuVasive Modulus XLIF Device and Attrax Putty; Sample Size: 75 patients
Not applicable for this study
Fusion rates, adverse event rates
Clinical literature analysis; Literature review/comparison
Patients treated with NuVasive Attrax Putty and interbody fusion devices
Autograft or allograft, and posterolateral fusion
Fusion rates, adverse event rates
Indications for Use
NuVasive CoRoent Thoracolumbar System The NuVasive CoRoent Thoracolumbar System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive CoRoent Thoracolumbar System (XL platform) implants are intended for use in interbody fusions in the thoracic spine, from T1 to T12, and at the thoracolumbar junction (T12-L1), and the CoRoent Thoracolumbar System (XL and L platforms) implants are intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive CoRoent Thoracolumbar System (XL and L platforms) can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. NuVasive CoRoent XL Interfixated System The NuVasive CoRoent XL Interfixated System implants are indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive CoRoent XL Interfixated System implants are intended for use in interbody fusions in the lumbar spine, from L2 to L5, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive CoRoent XL Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
Device Story
NuVasive Thoracolumbar Interbody Systems comprise various intervertebral body fusion implants (interfixated and non-interfixated) with hollow graft apertures; used with autogenous/allogeneic bone graft or Attrax Putty (synthetic, resorbable, osteoconductive bone void filler). Implants are placed via various surgical approaches to facilitate spinal fusion; often used with supplemental internal spinal fixation (e.g., pedicle screws). Attrax Putty is molded by clinicians to fill bony voids or graft apertures. Systems are used in clinical settings by surgeons to treat symptomatic disc degeneration, spondylolisthesis, and spinal deformities. Benefits include promotion of solid fusion and stabilization of the spinal segment. Attrax Putty resorbs and is replaced by bone during healing.
Clinical Evidence
Clinical evidence includes a registry study of 75 patients treated with the Modulus XLIF device and Attrax Putty, supplemented by a literature comparison of Attrax Putty used with CoRoent (PEEK) devices. Comparisons against autograft/allograft and posterolateral fusion supported performance based on fusion and adverse event rates. Bacterial endotoxin testing (BET) performed per ANSI/AAMI ST72:2011/(R)2016.
Technological Characteristics
Intervertebral body fusion devices available in various footprints, heights, and lordotic angles. Attrax Putty consists of ceramic granules and a polymeric binder. No software or electrical equipment. Sterilization provided for implants; accessory instruments provided non-sterile for end-user sterilization. BET testing per ANSI/AAMI ST72:2011/(R)2016.
Indications for Use
Indicated for skeletally mature patients requiring intervertebral body fusion of the spine (thoracic T1-T12, thoracolumbar T12-L1, or lumbar L1-S1) following at least six months of failed non-operative treatment. Conditions include symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, thoracic disc herniation, multilevel degenerative scoliosis, and sagittal deformity. Attrax Putty is indicated to fill bony voids or gaps in the skeletal system (posterolateral spine, intervertebral disc space, pelvis).
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
NuVasive Thoracolumbar Interbody Systems (K201820)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 23, 2021
NuVasive, Incorporated
Martin Yahiro, MD
Director, Medical Affairs
7475 Lusk Blvd.
San Diego, California 92121
Re: K203714
Trade/Device Name: NuVasive Thoracolumbar Interbody Systems: NuVasive CoRoent Thoracolumbar System, NuVasive CoRoent XL Interfixated System, Brigade System, Brigade Lateral System, BASE Interfixated Titanium System, Coalesce Thoracolumbar Interbody Fusion System, NuVasive Cohere Thoracolumbar Interbody System, NuVasive Modulus XLIF Interbody System, NuVasive Modulus TLIF Interbody System, NuVasive Modulus ALIF System
NuVasive Attrax Putty
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: MAX, PHM, OVD, MQV
Dated: December 8, 2021
Received: December 9, 2021
Dear Dr. Yahiro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K203714 – Dr. Martin Yahiro
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose -S
Laura C. Rose, Ph.D.
Assistant Director
DHT6C: Division of Restorative, Repair and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{2}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
510(k) Number (if known)
K203714
Device Name
NuVasive Thoracolumbar Interbody Systems and NuVasive Attrax Putty
Indications for Use (Describe)
NuVasive CoRoent Thoracolumbar System
The NuVasive CoRoent Thoracolumbar System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive CoRoent Thoracolumbar System (XL platform) implants are intended for use in interbody fusions in the thoracic spine, from T1 to T12, and at the thoracolumbar junction (T12-L1), and the CoRoent Thoracolumbar System (XL and L platforms) implants are intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive CoRoent Thoracolumbar System (XL and L platforms) can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
NuVasive CoRoent XL Interfixated System
The NuVasive CoRoent XL Interfixated System implants are indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive CoRoent XL Interfixated System implants are intended for use in interbody fusions in the lumbar spine, from L2 to L5, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive CoRoent XL Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PISC Publishing Services (301) 443-6740
{3}
FORM FDA 3881 (7/17)
Page 2 of 7
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. |
| --- | --- |
510(k) Number (if known)
K203714
Device Name
NuVasive Thoracolumbar Interbody Systems and NuVasive Attrax Putty
Indications for Use (Describe)
NuVasive Brigade System
The Brigade System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Standalone System (lordotic angles of 8° and 12°) is a standalone system. The Brigade Hyperlordotic System (lordotic angles of 15° to 30°) must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.
The Brigade System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade System platform implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the Brigade System platform must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
FORM FDA 3881 (7/17)
Page 3 of 7
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K203714 | |
| Device Name NuVasive Thoracolumbar Interbody Systems and NuVasive Attrax Putty | |
| Indications for Use (Describe) NuVasive Brigade Lateral System The Brigade Lateral System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Lateral System must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous bone graft, allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The Brigade Lateral System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade Lateral System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
510(k) Number (if known)
K203714
Device Name
NuVasive Thoracolumbar Interbody Systems and NuVasive Attrax Putty
Indications for Use (Describe)
NuVasive BASE Interfixated Titanium System
The BASE Interfixated Titanium System is indicated for spinal fusion procedures in skeletally mature patients. The BASE Interfixated Titanium System 10° - 20° lordotic cages may be used as a standalone system. The BASE Interfixated Titanium System 25° - 30° lordotic cages must be used with supplemental internal spinal fixation systems (i.e., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.
The BASE Interfixated Titanium System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The BASE Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the BASE Interfixated Titanium System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 4 of 7
PSC Publishing Services (301) 443-6740
{6}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
510(k) Number (if known)
K203714
Device Name
NuVasive Thoracolumbar Interbody Systems and NuVasive Attrax Putty
Indications for Use (Describe)
Coalesce Thoracolumbar Interbody Fusion System
The Coalesce Thoracolumbar Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The Coalesce Thoracolumbar Interbody Fusion System is intended for use in interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Coalesce Thoracolumbar Interbody Fusion System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
NuVasive Cohere Thoracolumbar Interbody System
The NuVasive Cohere Thoracolumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. When used with or without the Cohere XLIF internal fixation, the system is indicated for use with supplemental spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive Cohere Thoracolumbar Interbody System is intended for use in interbody fusions in the thoracic spine, from T1 to T12, at the thoracolumbar junction (T12-L1), and in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Cohere Thoracolumbar Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 5 of 7
PSC Publishing Services (301) 443-6740
{7}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
510(k) Number (if known)
K203714
Device Name
NuVasive Thoracolumbar Interbody Systems and NuVasive Attrax Putty
Indications for Use (Describe)
NuVasive Modulus XLIF Interbody System
The NuVasive Modulus XLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. When used with or without Modulus XLIF internal fixation, the system is intended for use with supplemental spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive Modulus XLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus XLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
NuVasive Modulus TLIF Interbody System
The NuVasive Modulus TLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive Modulus TLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus TLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 6 of 7
PSC Publishing Services (301) 443-6740
{8}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
510(k) Number (if known)
K203714
Device Name
NuVasive Thoracolumbar Interbody Systems and NuVasive Attrax Putty
Indications for Use (Describe)
NuVasive Modulus ALIF System
The NuVasive Modulus ALIF System is indicated for spinal fusion procedures in skeletally mature patients. The Modulus ALIF System 10°-20° lordotic cages may be used as a standalone system. The Modulus ALIF System 25°-30° lordotic cages must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive Modulus ALIF System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Modulus ALIF System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the Modulus ALIF System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
NuVasive Attrax Putty
Attrax Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., posterolateral spine, intervertebral disc space, and pelvis). These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bone structure. NuVasive Attrax Putty resorbs and is replaced with bone during the healing process. NuVasive Attrax Putty may be used in combination with autogenous bone in the posterolateral spine and pelvis. When used in intervertebral body fusion procedures, NuVasive Attrax Putty must be used on its own with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 7 of 7
PSC Publishing Services (301) 443-6740
{9}
510(k) Number: K203714
Dated: December 21, 2021
K203714
Page 1 of 7
# 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
## A. Submitted by:
Ms. Jessica LeBlanc
Manager, Regulatory Affairs
NuVasive, Incorporated
7475 Lusk Blvd.
San Diego, California 92121
(951) 816-0973
Martin Yahiro, M.D.
Director, Medical Affairs
NuVasive, Incorporated
7475 Lusk Blvd.
San Diego, California 92121
(858) 320-4550
Date Prepared: December 21, 2021
## B. Device Name
Trade or Proprietary Name: NuVasive Thoracolumbar Interbody Systems, which includes the following systems:
- CoRoent Thoracolumbar System, including CoRoent Thoracolumbar System (Non-Interfixated) and CoRoent XL (Interfixated System)
- Brigade System, including Brigade Hyperlordotic, Brigade Standalone, and Brigade Lateral
- BASE Interfixated Titanium System
- Coalesce Thoracolumbar Interbody Fusion System
- Cohere Thoracolumbar Interbody System
- Modulus Interbody System, including Modulus XLIF Interbody System, Modulus TLIF Interbody System, and Modulus ALIF System
Common or Usual Name: Intervertebral Body Fusion Device
Classification Name: Intervertebral Body Fusion Device
Device Class: Class II
Classification: 21 CFR 888.3080
Product Code: MAX, PHM, OVD
Trade or Proprietary Name: NuVasive Attrax Putty
Common Name: Bone void filler
Classification Name: Resorbable calcium salt bone void filler device
Device Class: Class II
Classification: 21 CFR 888.3045
Product Code: MQV
{10}
NUVASIVE
K203714
Page 2 of 7
## C. Predicate Devices
The subject *NuVasive Thoracolumbar Interbody Systems* is substantially equivalent to the primary predicate device *NuVasive Thoracolumbar Interbody Systems* (K201820) and the additional predicate devices, *Modulus XLIF Interbody System* (K201692), and *Cohere Thoracolumbar Interbody System* (K200953).
The subject *NuVasive Attrax Putty* is substantially equivalent to the predicate device, *NuVasive Attrax Putty* (K191974).
DePuy Synthes T-PAL (K100089, K151276) was used as a reference device.
## D. Device Description
The *NuVasive Thoracolumbar Interbody Systems* is inclusive of several intervertebral body fusion devices, interfixated and non-interfixated. The hollow core, or graft aperture, of the devices allows for packing of graft to aid in the promotion of a solid fusion. The devices are available in varied footprints, heights, and lordotic angles to suit the individual pathology and anatomical conditions of the patient, utilizing placement through various surgical approaches. *NuVasive Attrax Putty* is a synthetic, osteoconductive and resorbable bone void filler device consisting of ceramic granules premixed with a polymeric binder that provides cohesion between the granules. Pressure applied by user manipulation allows the *Attrax Putty* to be molded into specific shapes, mixed with autograft, or contoured into a bone defect or the graft aperture of a spinal interbody implant, as desired by the clinician. The subject implants are offered sterile and/or non-sterile with the accessory surgical instruments packaged as non-sterile to be sterilized by the end user.
The purpose of this submission is to expand the indications for use of the *NuVasive Thoracolumbar Interbody Systems* to include its use with *NuVasive Attrax Putty*.
## E. Indications for Use
### CoRoent Thoracolumbar System:
The *NuVasive CoRoent Thoracolumbar System* is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The *NuVasive CoRoent Thoracolumbar System* (*XL platform*) implants are intended for use in interbody fusions in the thoracic spine, from T1 to T12, and at the thoracolumbar junction (T12-L1), and the *CoRoent Thoracolumbar System* (*XL and L platforms*) implants are intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with
{11}
NUVASIVE
K203714
Page 3 of 7
degeneration of the disc confirmed by history and radiographic studies. The NuVasive CoRoent Thoracolumbar System (XL and L platforms) can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
## CoRoent XL Interfixated System:
The NuVasive CoRoent XL Interfixated System implants are indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive CoRoent XL Interfixated System implants are intended for use in interbody fusions in the lumbar spine, from L2 to L5, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive CoRoent XL Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
## Brigade System (Brigade Standalone and Brigade Hyperlordotic):
The Brigade System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Standalone System (lordotic angles of 8° and 12°) is a standalone system. The Brigade Hyperlordotic System (lordotic angles of 15° to 30°) must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.
The Brigade System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade System platform implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the Brigade System platform must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
## Brigade Lateral System:
The Brigade Lateral System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Lateral System must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA
{12}
NUVASIVE
K203714
Page 4 of 7
for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous bone graft, allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.
The Brigade Lateral System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade Lateral System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
## BASE Titanium Interfixated System:
The BASE Interfixated Titanium System is indicated for spinal fusion procedures in skeletally mature patients. The BASE Interfixated Titanium System 10° - 20° lordotic cages may be used as a standalone system. The BASE Interfixated Titanium System 25° - 30° lordotic cages must be used with supplemental internal spinal fixation systems (i.e., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.
The BASE Interfixated Titanium System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The BASE Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the BASE Interfixated Titanium System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
## Coalesce Thoracolumbar Interbody Fusion System:
The Coalesce Thoracolumbar Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The Coalesce Thoracolumbar Interbody Fusion System is intended for use in interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two
{13}
NUVASIVE
K203714
Page 5 of 7
adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Coalesce Thoracolumbar Interbody Fusion System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
## Cohere-Thoracolumbar Interbody Systems:
The NuVasive Cohere Thoracolumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. When used with or without the Cohere XLIF internal fixation, the system is indicated for use with supplemental spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive Cohere Thoracolumbar Interbody System is intended for use in interbody fusions in the thoracic spine, from T1 to T12, at the thoracolumbar junction (T12-L1), and in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Cohere Thoracolumbar Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
## Modulus-XLIF Interbody System:
The NuVasive Modulus XLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. When used with or without Modulus XLIF internal fixation, the system is intended for use with supplemental spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive Modulus XLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus XLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
## Modulus-TLIF Interbody System:
{14}
NUVASIVE
K203714
Page 6 of 7
The NuVasive Modulus TLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive Modulus TLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus TLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
## Modulus ALIF System:
The NuVasive Modulus ALIF System is indicated for spinal fusion procedures in skeletally mature patients. The Modulus ALIF System 10°-20° lordotic cages may be used as a standalone system. The Modulus ALIF System 25°-30° lordotic cages must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The System is designed for use with autogenous bone graft, allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive Modulus ALIF System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Modulus ALIF System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the Modulus ALIF System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
## NuVasive Attrax Putty:
Attrax Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., posterolateral spine, intervertebral disc space, and pelvis). These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bone structure. NuVasive Attrax Putty resorbs and is replaced with bone during the healing process. NuVasive Attrax Putty may be used in combination with autogenous bone in the posterolateral spine and pelvis. When used in intervertebral body
{15}
NUVASIVE
K203714
Page 7 of 7
fusion procedures, NuVasive Attrax Putty must be used on its own with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
## F. Technological Characteristics
As was established in this submission, the subject NuVasive Thoracolumbar Interbody Systems and NuVasive Attrax Putty are substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject devices were shown to be substantially equivalent and have the same technological characteristics to their predicate devices through comparison in areas including design, intended use, material composition, function, and range of sizes. These devices do not contain software or electrical equipment.
## G. Performance Data
The purpose of this submission is to expand the indications for use of the NuVasive Thoracolumbar Interbody Systems to include its use with NuVasive Attrax Putty. A clinical outcomes registry study and a clinical literature analysis of patients treated with the subject devices were performed. The registry data provided clinical and radiographic data from 75 patients treated with the NuVasive Modulus XLIF Device and Attrax Putty and was supplemented by a literature comparison of NuVasive Attrax Putty used with another interbody fusion device in the NuVasive Thoracolumbar Interbody System, the NuVasive CoRoent (PEEK) device. Comparisons of intervertebral body devices with Attrax Putty versus autograft or allograft, and versus posterolateral fusion supported the performance in intervertebral body fusion based on fusion and adverse event rates. Additional clinical and engineering rationale was provided to address technological differences between the intervertebral body fusion devices included in the submission to support use of NuVasive Attrax Putty with all the fusion devices in the NuVasive Thoracolumbar Interbody System. Bacterial endotoxin testing (BET) is performed for all implant systems per ANSI/AAMI ST72:2011/(R)2016. Based on the clinical data, it was determined that the NuVasive Thoracolumbar Interbody Systems filled with NuVasive Attrax Putty have a safety and effectiveness profile similar to the predicate devices.
## H. Conclusions
Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject NuVasive Thoracolumbar Interbody Systems and NuVasive Attrax Putty have been shown to be substantially equivalent to legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.