K233698 · Ge Healthcare Japan Corporation · JAK · Apr 11, 2024 · Radiology
Device Facts
Record ID
K233698
Device Name
True Enhance DL
Applicant
Ge Healthcare Japan Corporation
Product Code
JAK · Radiology
Decision Date
Apr 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K233698 · Apr 11, 2024
True Enhance DL
Ge Healthcare Japan Corporation
Retrospectively collected clinical images
Retrospective clinical cases were used in a reader study to assess image quality and contrast enhancement performance of the deep learning algorithm compared to native 120 kVp images.
Reader study of sample clinical data; Retrospective reader study
Representative clinical cases with disease/pathology
Native 120 kVp image series
Comparative assessment of image quality; contrast enhancement rating
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Contrast enhancement of single energy CT images
Deep learning-based image processing method
—
—
—
—
Reader study of retrospectively collected representative clinical cases with disease/pathology.
4 (board certified radiologists)
Indications for Use
True Enhance DL is a deep learning-based image processing method trained to estimate monochromatic, 50 keV GSI images. The algorithm is intended to improve the contrast of 120 kVp, single energy images of the body. This device is intended to provide non-quantitative, adjunct information and should not be interpreted without the original 120 kVp image. True Enhance DL may be used for patients of all ages.
Device Story
True Enhance DL is a deep learning-based post-processing software for 120 kVp single energy CT images acquired on GE Revolution Ascend systems. It transforms 120 kVp input data into images estimating 50 keV monochromatic GSI (Gemstone Spectral Imaging) characteristics to improve contrast. Users select from four models based on contrast phase: CT Angiography, Arterial, Portal/Venous, or Delayed. The device operates within the existing CT acquisition/processing software environment. Radiologists view the enhanced output alongside original 120 kVp images to assist in diagnostic assessment. It provides non-quantitative, adjunctive information; it does not replace hardware-based GSI or dual-energy acquisitions. The device benefits patients by providing improved contrast visualization from standard single-energy CT scans.
Clinical Evidence
Clinical evidence consisted of a reader study using retrospective clinical cases. Four board-certified radiologists performed a comparative assessment of image quality and contrast enhancement between True Enhance DL output and native 120 kVp images. Additional evaluation included head-to-head material comparison of clinical and anthropomorphic phantom cases against 50 keV and 120 kVp reference images. Results validated that the software provides improved contrast compared to original 120 kVp images.
Technological Characteristics
Deep learning-based image processing software integrated into the Revolution Ascend CT platform. Utilizes existing hardware and reconstruction software (FBP, ASiR-V). Four specific deep learning models for contrast phases. Operates on 120 kVp single energy input. Developed under ISO 13485 and 21 CFR 820 design controls.
Indications for Use
Indicated for patients of all ages requiring contrast enhancement of 120 kVp single energy body CT images. Provides non-quantitative, adjunct information; must be used with original 120 kVp images.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Deep Learning Image Reconstruction, DLIR (K230807)
Submission Summary (Full Text)
{0}------------------------------------------------
April 11, 2024
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an eagle and text around it, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
GE Healthcare Japan Corporation % Laura Turner Regulatory Affairs Manager, Premium CT 7-127, Asahigaoka, 4-chome HINO-SHI, TOKYO 191-8503 JAPAN
Re: K233698
Trade/Device Name: True Enhance DL Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: November 17, 2023 Received: November 17, 2023
Dear Laura Turner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233698
Device Name True Enhance DL
Indications for Use (Describe)
True Enhance DL is a deep learning-based image processing method trained to estimate monochromatic, 50 keV GSI images. The algorithm is intended to improve the contrast of 120 kVp, single energy images of the body.
This device is intended to provide non-quantitative, adjunct information and should not be original 120 kVp image.
True Enhance DL may be used for patients of all ages.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the GE Healthcare logo. The logo consists of a purple circular emblem on the left and the text "GE HealthCare" on the right. The emblem features a stylized, intertwined "GE" design within the circle, also in purple. The text is in a sans-serif font, with "GE" slightly larger than "HealthCare", both in the same purple color.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVNESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.92:
| Date: | April 10, 2024 |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare Japan Corporation<br>7-127, Asahigaoka, 4-chome<br>Hino-shi, Tokyo, 191-8503, Japan |
| Primary Contact: | Laura Turner<br>Regulatory Affairs Manager<br>Phone: 1-262-200-1044<br>Email: laura.turner@ge.com |
| Secondary Contacts: | John Jaeckle<br>Chief Regulatory Affairs Engineer and Strategist<br>Phone: 1-262-424-9547<br>Email: john.jaeckle@med.ge.com<br>Veronica Aguayo<br>Regulatory Affairs Leader<br>Phone: 1-414-217-4331<br>Email: Veronica.Aguayo@ge.com |
| Subject Device Name: | True Enhance DL |
| Device Classification | Class II |
| Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed tomography x-ray system / JAK |
# Predicate Device Information
| Device Name: | Revolution Ascend |
|-------------------------------------|-------------------------------------------------------|
| Manufacturer: | GE Medical Systems, LLC |
| 510(k) Number: | K213938 cleared on February 4, 2022 |
| Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed tomography x-ray system / JA |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the GE Healthcare logo. The logo consists of a purple circular emblem with a stylized "GE" monogram inside. To the right of the emblem, the words "GE HealthCare" are written in purple sans-serif font. The logo is clean and modern, reflecting the company's focus on healthcare technology and innovation.
# Device Description
True Enhance DL is a deep learning-based image processing method for contrast enhanced images of the body obtained using the Revolution Ascend Family (K213938), which consists of multiple commercial configurations: Revolution Ascend Elite, Revolution Ascend Plus, and Revolution Ascend Select. True Enhance DL is intended to post-process single energy, 120 kVp images to output nonquantitative, adjunctive information with better contrast than single energy input data.
True Enhance DL brings four deep leaning models that the user can choose depending on different contrast enhancement phases. These four models are CT Angiography, Arterial, Portal/Venous, and Delayed True Enhance DL.
True Enhance DL is not intended to replace hardware based Monochromatic Imaging by Gemstone Spectral Imaging (GSI) technology or replicate GSI dual energy acquisitions. The device was trained to estimate monochromatic, 50 keV GSI images, and only enhances images from 120 kVp acquisitions on non-GSI Revolution Ascend systems.
#### Intended Use
True Enhance DL is a deep learning-based image processing method intended for contrast enhancement of single energy images.
#### Indications for Use
True Enhance DL is a deep learning-based image processing method trained to estimate monochromatic, 50 keV GSI images. The algorithm is intended to improve the contrast of 120 kVp, single energy images of the body.
This device is intended to provide non-quantitative, adjunct information and should not be interpreted without the original 120 kVp image.
True Enhance DL may be used for patients of all ages.
# Technological Characteristics
Since the True Enhance DL option is deployed within Revolution Ascend acquisition and processing software, it utilizes the same hardware and software platform to reconstruct images that achieve contrast enhancement from single kVp images. Use of True Enhance DL does not require that the acquisition parameters used for the 120 kVp input images to be altered. The table below summarizes the substantive feature / technological differences between the predicate device and the proposed device:
| Specification | Predicate Device | Proposed Device |
|---------------|--------------------------------|-----------------------------------------|
| | Revolution Ascend<br>(K213938) | True Enhance DL<br>on Revolution Ascend |
{5}------------------------------------------------
# GE HealthCare 510(k) Premarket Notification Submission True Enhance DL
Image /page/5/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem with a stylized "GE" monogram on the left, and the text "GE HealthCare" on the right. The emblem and text are both in a purple color.
| Specification | Predicate Device | Proposed Device |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| | Revolution Ascend | True Enhance DL |
| | (K213938) | on Revolution Ascend |
| Clinical Workflow | The Revolution Ascend User<br>Interface allows for<br>modifications of image<br>reconstruction settings (FBP,<br>ASIR-V, and other parameters)<br>based on users desired image<br>display and workflow set up. | Same |
| Supported kVps | 80, 100, 120, 140 | 120 |
| Supported Recon Kernels | - Standard | - Standard |
| | - Soft | - Soft |
| | - Detail | - Detail |
| | - Bone, Bone plus | |
| | - Chest | |
| | - Lung | |
| | - Ultra | |
| | - Edge, Edge plus | |
| Image Reconstruction | - Filtered back Projection<br>- ASiR-V<br>- Deep Learning Image<br>Reconstruction, DLIR<br>(K230807) | - Filtered back Projection<br>- ASiR-V |
The different technological characteristics do not raise different or new question of safety and effectiveness, and that the device is as safe and effective the legally marketed predicate.
# Determination of Substantial Equivalence
True Enhance DL has successfully completed the design control testing per GE's quality system. No additional hazards were identified, and no unexpected test results were observed. True Enhance DL was designed under the Quality System Regulations of 21CFR 820 and ISO 13485. GEHC believes that the successfully completed design verification & validation, as well as the extensive bench and clinical testing is sufficient to demonstrate the device's substantial equivalence to the predicate.
The following quality assurance measures have been applied to the development of the system:
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the GE Healthcare logo. The logo consists of a purple circular emblem with intertwined letters on the left, followed by the text "GE HealthCare" in a matching purple color. The text is written in a clean, sans-serif font, and the overall design is modern and corporate.
- Requirement Definition
- Risk Analysis and Control
- Technical Design Reviews
- . Formal Design Reviews
- . Software Development Lifecycle
- Code Review O
- O Software Unit Implementation
- Software Integrations and Integration Testing O
- System Testing
- Safety Testing (Verification) o
- o Image Performance Testing (Verification)
- Simulating Use Testing (Validation) o
- Software Release
GE believes True Enhance DL on Revolution Ascend is of comparable type and substantially equivalent to our currently marketed system Revolution Ascend (K213938). The testing and results did not raise different questions of safety and effectiveness than associated with predicate device.
# Summary of Non-Clinical Testing
The verification and validation testing have been successfully completed as required by design control procedures under GE Healthcare's quality system. This includes risk management, software verification and validation testing as well as image quality performance evaluation using well established metrics and methods. IQ evaluation include:
- . General IQ Performance testing of metrics identified in IEC 61223-3-5 Ed. 2 to study the overall performance in a standardized and referenceable manner.
- . IQ performance test comparison using standard IQ, QA, and anthropomorphic phantoms for representative conditions to characterize image quality of True Enhance DL on Revolution Ascend, and predicate Revolution Ascend
All the testing and results did not raise new or different questions of safety and effectiveness than associated with predicate device.
# Clinical Testing
The clinical testing was carried out in the form of a reader study of sample clinical data. The reader study used retrospectively collected representative clinical cases with disease/pathology, that were processed using True Enhance DL software. The images generated from the subject device were read by four board certified radiologists. The four radiologists each provided a comparative assessment of image quality related to diagnostic use. Additionally, the readers were asked to rate the contrast enhancement in the True Enhance DL series vs the native image series.
Additional representative clinical cases and anthropomorphic phantom cases from a GSI system generating both single energy 120 kVp and 50 keV monochromatic images were evaluated for CT
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for GE Healthcare. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is a stylized, intertwined "GE" in a purple color. The text is also in purple and is written in a clean, sans-serif font.
number in various anatomical regions to study image characteristics for different materials of the device output compared to 50 keV and 120 kVp reference images.
The result of this reader study and head-to-head material comparison validated that True Enhance DL software provides additional benefit by improving contrast in the True Enhance output when compared to the original 120 kVp single energy images.
#### Substantial Equivalence
The changes associated with True Enhance DL do not create a new Intended Use and represent technological characteristics that produce images that have demonstrated adequate image quality and contrast improvement relative to the 120 kVp input.
GE's quality system's design verification, and risk management processes did not identify any new questions of safety or effectiveness, hazards, unexpected results, or adverse effects stemming from the changes to the predicate.
Based on development under GE Healthcare's quality system, the successful system and software verification and validation testing, the additional engineering and the clinical testing demonstrates that True Enhance DL is substantially equivalent to and as effective for its Intended Use, as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.